HICAT Air is a software application for: - Aiding diagnosis in the ear-nose-throat region - Aiding treatment planning in the ear-nose-throat region - Aiding comparisons of different treatment options - Aiding treatment planning for oral appliances
Device Story
HICAT Air is a software application for visualization and segmentation of medical imaging data (CT, CBCT) and optical impressions. Used by medical professionals in clinical settings to aid diagnosis and treatment planning in the ENT region. Inputs include DICOM radiological data and STL/proprietary optical surface data. The software performs 3D rendering, airway segmentation, and geometric analysis. Outputs include visualized anatomical structures and planning data, which can be exported to CAD/Rapid Prototyping systems for manufacturing therapeutic devices like oral appliances. It assists clinicians in evaluating and comparing treatment options, potentially improving treatment planning accuracy and patient outcomes.
Clinical Evidence
Bench testing only. Verification and validation activities included design reviews, unit/code/integration/system tests, and usability testing. Bench testing specifically verified airway segmentation performance and geometric dimension calculation accuracy.
Technological Characteristics
Pure software device. Operates on Windows 7/8/8.1 PCs. Uses C# and C++ programming languages. Implements ray-casting algorithms for 3D volume rendering and Watershed (graph-cut) algorithms for segmentation. Connectivity via Ethernet. Complies with DICOM (NEMA PS 3.1-3.20), ISO 14971, IEC 62304, and IEC 62366 standards.
Indications for Use
Indicated for use by qualified medical professionals for diagnosis, treatment planning, and comparison of treatment options in the ear-nose-throat (ENT) region, including planning for oral appliances.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized caduceus or a series of interconnected human profiles. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 16, 2015
HICAT GmbH % Mr. Frederik Kunze Head of Quality Management and Regulatory Affairs Brunnenallee 6 Bonn, North Rhine - Westphalia 53177 GERMANY
Re: K152661 Trade/Device Name: HICAT Air Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: September 22, 2015 Received: September 23, 2015
Dear Mr. Kunze:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael O'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br>Indications for Use | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| 510(k) Number (if known) | |
| K152661 | |
| Device Name | |
Device Proprietary Name: HICAT Air
Indications for Use (Describe)
HICAT Air is a software application for:
- · Aiding diagnosis in the ear-nose-throat region
- Aiding treatment planning in the ear-nose-throat region
- · Aiding comparisons of different treatment options
- · Aiding treatment planning for oral appliances
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
| FORM FDA 3881 (8/14) | |
|----------------------|--|
|----------------------|--|
{3}------------------------------------------------
# 510(k) Summary for HICAT Air
#### Content and format as required by section 21 CFR 807.92
( http://www.fda.gov/ MedicalDevices/ DeviceRegulationandGuidance/ HowtoMarket YourDevice/PremarketSubmissions/PremarketNotification510k/ucm142651.htm\
#### 1. SUBMITTER/510(K) HOLDER
HICAT GmbH Brunnenallee 6 53177 Bonn Germany
| Establishment<br>Registration Number: | HICAT GmbH is not registered yet |
|---------------------------------------|----------------------------------|
| Telephone: | +49 (228) 854697-38 |
| Facsimile: | +49 (228) 854697-99 |
| Primary Contact: | Mr. Frederik Kunze |
| Secondary Contact: | Dr. Manfred Breuer |
| Date Prepared: | September 14th, 2015 |
#### 2. DEVICE NAME AND DEVICE CLASSIFICATION
| Proprietary Name: | HICAT Air |
|-------------------------|-----------------------------------------------------------------------------|
| Common/Usual Name: | Radiological Visualization Software for Diagnosis and Treatment<br>Planning |
| Classification Name: | System, Image Processing, Radiological |
| Regulation Description: | Picture archiving and communications system |
| Product Code: | LLZ |
| Regulation Number: | 21 CFR 892.2050 |
| Classification Class: | Class II Product |
#### 3. PREDICATE DEVICES
- . Primary Predicate: SICAT Function (K133320)
- Reference predicate: Dolphin Imaging 11.5 (K110430) .
## 4. DEVICE DESCRIPTION
HICAT Air is a pure software device.
HICAT Air is a software application for the visualization and segmentation of imaging information of the ear-nose-throat (ENT) region.
{4}------------------------------------------------
The imaging data originates from medical scanners such as CT or Cone Beam – CT (CBCT) scanners.
This information can be complemented by the imaging information from optical impression systems. The additional information about the exact geometry of the tooth surfaces can be visualized together with the radiological data.
HICAT Air is also used as a software system to aid qualified medical professionals with the diagnosis, and followed by the evaluation, comparison and planning of ENT treatment options.
The medical professionals' input information and planning data may be exported from HICAT Air to be used as input data for CAD or Rapid Prototyping Systems for the manufacturing of therapeutic devices such as oral devices.
## 5. Indications for Use
HICAT Air is a software application for:
- Aiding diagnosis in the ear-nose-throat region
- Aiding treatment planning in the ear-nose-throat region
- Aiding comparisons of different treatment options
- Aiding treatment planning for oral appliances
## 6. Device Comparison Table
The following table shows a summary of the indications for use, technological characteristics, design and function of HICAT Air and the predicate devices.
| HICAT Air<br>Proposed | Reference predicate<br>Dolphin Imaging 11.5<br>(K110430) | Primary predicate<br>SICAT Function<br>(K133320) |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common / Usual Name | | |
| Radiological Visualization<br>Software for Diagnosis and<br>Treatment Planning | Not available | Radiological Visualization<br>Software for Diagnosis and<br>Dental Treatment Planning |
| Device Description | | |
| HICAT Air is a pure software<br>device. HICAT Air is a<br>software application for the<br>visualization and<br>segmentation of imaging<br>information of the ear-nose-<br>throat (ENT) region. The<br>imaging data originates from<br>medical scanners such as CT<br>or Cone Beam – CT (CBCT)<br>scanners. | Dolphin Imaging is a pure<br>software device. Dolphin<br>Imaging is a software<br>application for the<br>visualization of imaging<br>information of the oral-<br>maxillofacial region. The<br>imaging data originates<br>from medical scanners<br>such as CT or DVT<br>scanners. | SICAT Function is a pure<br>software device. SICAT<br>Function is a software<br>application for the<br>visualization and<br>segmentation of imaging<br>information of the oral-<br>maxillofacial region. The<br>imaging data originates from<br>medical scanners such as CT<br>or Cone Beam - CT scanners. |
| This information can be<br>complemented by the<br>imaging information from<br>optical impression systems.<br>The additional information<br>about the exact geometry of | Dolphin Imaging software<br>is designed for capturing,<br>storing and presenting<br>patient images and<br>assisting in treatment<br>planning and case | This information can be<br>complemented by the<br>imaging information from<br>optical impression systems<br>and jaw tracking devices. The<br>additional information about |
{5}------------------------------------------------
| HICAT Air<br>Proposed | Reference predicate<br>Dolphin Imaging 11.5<br>(K110430) | Primary predicate<br>SICAT Function<br>(K133320) |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| the tooth surfaces can be<br>visualized together with the<br>radiological data.<br>HICAT Air is also used as a<br>software system to aid<br>qualified medical<br>professionals with the<br>diagnosis, and followed by<br>the evaluation, comparison<br>and planning of ENT<br>treatment options.<br>The medical professionals'<br>input information and<br>planning data may be<br>exported from HICAT Air to<br>be used as input data for<br>CAD or Rapid Prototyping<br>Systems for the<br>manufacturing of therapeutic<br>devices such as oral devices. | diagnosis. Results<br>produced by the software's<br>diagnostic and treatment<br>planning tools are<br>dependent on the<br>interpretation of trained<br>and licensed practitioners.<br>Acquisition of 2D image<br>data can be performed<br>from multiple types of<br>devices, including digital<br>radiography devices such<br>as x-ray machines, generic<br>imaging devices that<br>support the TWAIN<br>interface standard, such as<br>flatbed scanners and<br>standard image file formats<br>(Jpeg, Bitmap, etc.) stored<br>on a computer, camera<br>memory card or similar.<br>The following features are<br>available:<br>Import from a variety of 3D<br>dataset formats (DICOM)<br>High-quality, fast 3D<br>rendering<br>Three-dimensional airway<br>segmentation and analysis<br>in the ear-nose throat-<br>region<br>Airway analysis:<br>Analyze the airway by<br>drawing a border around<br>your selected portion of the<br>volumetric scan; the<br>program will automatically<br>fill in and display all the<br>airway space within that<br>border, then report back<br>telling you the volume of<br>airway space in cubic<br>millimeters. It will also<br>locate, display and<br>measure, in square<br>millimeters, the most<br>constricted spot of that<br>airway.<br>Cross sections in Multiple | the exact geometry of the<br>tooth surfaces and the<br>mandibular movement can be<br>visualized together with the<br>radiological data.<br>SICAT Function is also used<br>as a software system to aid<br>qualified dental professionals<br>with the evaluation and<br>planning of dental treatment<br>options.<br>The dental professionals'<br>treatment planning<br>information may be exported<br>from SICAT Function to be<br>used as input data for CAD or<br>Rapid Prototyping Systems<br>for the manufacturing of<br>therapeutic devices such as<br>oral appliances. |
{6}------------------------------------------------
| | HICAT Air<br>Proposed | Reference predicate<br>Dolphin Imaging 11.5<br>(K110430) | Primary predicate<br>SICAT Function<br>(K133320) |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | Planar View (MPV)<br>Precise volume orientation | |
| | | Volume stitching:<br>Combine two separate<br>volumes into one<br>2D facial photo wrap onto<br>volume | |
| | | Volume-to-volume<br>superimposition | |
| | | 3D nerve marking | |
| | | TMJ analysis | |
| | | Generate cephalometric<br>and panoramic radiographs | |
| | | 3D-2D measurements:<br>distance, angle and area | |
| | | Create animated movies<br>with automated scripts | |
| | | Design image layouts and<br>reports | |
| | | Images export to other<br>applications | |
| | | 3D implant treatment<br>planning and simulation | |
| | | Distinguish biological<br>structures via their<br>radiolucency with a<br>Hounsfield Unit measuring<br>tool | |
| Indications for Use | | | |
| Indications for<br>Use | HICAT Air is a software<br>application for:<br>• Aiding diagnosis in the<br>ear-nose-throat region.<br>• Aiding treatment<br>planning in the ear-<br>nose-throat region.<br>• Aiding comparisons of<br>different treatment<br>options.<br>• Aiding treatment<br>planning for oral<br>appliances | Intended Use Statement of<br>510(k) Summary K110430:<br>"Dolphin Imaging software<br>is designed for use by<br>specialized dental practices<br>for capturing, storing and<br>presenting patient images<br>and assisting in treatment<br>planning and case<br>diagnosis. Results<br>produced by the software's<br>diagnostic and treatment<br>planning tools are | Indications for Use<br>Statement of 510(k)<br>Summary K133320:<br>SICAT Function is a software<br>application for the<br>visualization and<br>segmentation of imaging<br>information of the oral-<br>maxillofacial region. The<br>imaging data originates from<br>medical scanners such as CT<br>or CBCT scanners. It is also<br>used as a software system to |
| | | dependent on the<br>interpretation of trained | aid qualified dental<br>professionals with the |
{7}------------------------------------------------
| | HICAT Air<br>Proposed | Reference predicate<br>Dolphin Imaging 11.5<br>(K110430) | Primary predicate<br>SICAT Function<br>(K133320) |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | and licensed practitioners."<br>Product information from Appendix B-2 indicates among others: Aiding diagnosis in the ENT-region Aiding treatment planning in the ENT-region Aiding comparisons of different treatment options | evaluation of dental treatment options. The dental professionals' planning data may be exported from SICAT Function and used as input data for CAD or Rapid Prototyping Systems. |
| Contraindications | none | unknown | none |
| | Medical Data Viewing | | |
| Data types<br>visualized | 3D volume data, optical<br>impressions | 3D volume data, 2D<br>radiological x-ray data,<br>images and photographs | 3D volume data, optical<br>impressions, jaw motion<br>tracking data |
| Imaging data<br>visualization<br>region | ENT region including the<br>oral-maxillofacial region | ENT region including the<br>oral-maxillofacial region | oral-maxillofacial region |
| | Viewing Modes | | |
| Orthogonal slices | Axial, coronal, sagittal<br>Features and algorithms are<br>identical to SICAT Function. | Axial, coronal, sagittal | Axial, coronal, sagittal |
| Special Dental<br>View:<br>Panoramic View | Panoramic view (ray sum)<br>based on a panoramic curve.<br>Panoramic curve can be<br>manually adjusted.<br>(available in the optical<br>impression registration<br>window)<br>Features and Algorithms are<br>identical to SICAT Function. | Yes | Panoramic view (ray sum)<br>based on a panoramic curve.<br>Panoramic curve can be<br>manually adjusted. |
| Special dental<br>view:<br>Cephalometric<br>View | No | Yes | No |
| Special dental<br>views:<br>TSA and LSA | Transversal slice (TSA) and<br>longitudinal slice (LSA), both<br>with respect to panoramic<br>curve.<br>(available in the optical<br>impression registration<br>window)<br>Features and Algorithms<br>identical to SICAT Function | Single transversal slices are<br>generated. | Transversal slice (TSA) and<br>longitudinal slice (LSA), both<br>with respect to panoramic<br>curve. |
| 3D volume<br>rendering | Yes, volume rendering using<br>a standard and well<br>established ray-casting<br>algorithm with integration<br>along rays, and volume<br>rendering utilizing standard | Yes | Yes |
{8}------------------------------------------------
| | HICAT Air<br>Proposed | Reference predicate<br>Dolphin Imaging 11.5<br>(K110430) | Primary predicate<br>SICAT Function<br>(K133320) |
|---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| 3D surface<br>rendering of<br>volume data | shader based slicing<br>algorithms available on<br>typical consumer graphics<br>hardware.<br>Yes, using a standard and<br>well established front-to-<br>back ray-casting algorithm<br>with an iso-surface<br>threshold. | Yes | No |
| View<br>manipulating<br>tools | Zoom, Pan, Change of<br>orientation<br>Features and Algorithms<br>identical to SICAT Function | Yes | Zoom, Pan, Change of<br>orientation |
| Color<br>manipulating<br>tools | Brightness, Contrast<br>Features and Algorithms<br>identical to SICAT Function | Yes | Brightness, Contrast |
| Scrolling through<br>slices | For all slices<br>Features and Algorithms<br>identical to SICAT Function | Yes | For all slices (including<br>panoramic slicing window) |
| 3D volume<br>clipping | Yes<br>Features and Algorithms<br>identical to SICAT Function | Yes | Yes |
| Length<br>measurement | Yes<br>Features and Algorithms<br>identical to SICAT Function | Yes | Yes |
| Area<br>measurement | Yes | Yes | No |
| Angle<br>measurement | Yes<br>Features and Algorithms<br>identical to SICAT Function | Yes | Yes |
| Overall Length<br>Measurement<br>Accuracy | 100 µm<br>Algorithms identical to SICAT<br>Function | unknown | 100 μm |
| Overall Angular<br>Measurement<br>Accuracy | 1 degree<br>Algorithms identical to SICAT<br>Function | unknown | 1 degree |
| Segmentation of Anatomical Structures | | | |
| Segmentation of<br>anatomical<br>structures | Yes, using a segmentation<br>wizard.<br>Algorithm: Water Shed (a<br>type of graph-cut algorithm,<br>which is identical to SICAT<br>Function ) | Yes<br>Segmentation algorithm is<br>unknown. | Yes, using a segmentation<br>wizard.<br>Algorithm: Water Shed (a<br>type of graph-cut algorithm) |
| Airway<br>Segmentation | Airway Segmentation of the<br>Airway using the<br>segmentation wizard. | Yes<br>Segmentation algorithm is<br>unknown. | No |
| Segmentation of<br>mandible and<br>fossa | No | No | Yes, using a segmentation<br>wizard.<br>Algorithm: Water Shed (a<br>type of graph-cut algorithm) |
{9}------------------------------------------------
| | HICAT Air<br>Proposed | Reference predicate<br>Dolphin Imaging 11.5<br>(K110430) | Primary predicate<br>SICAT Function<br>(K133320) |
|------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Mesh texturing of<br>the segmented<br>airway | The Airway is texturized by a<br>standard algorithm. | Yes | No |
| | Optical Impression Data Visualization | | |
| Optical<br>impression<br>(maxilla,<br>mandible, or a<br>part thereof) | Optical impression (maxilla,<br>mandible, or a part thereof) | No | Optical impression (maxilla,<br>mandible, or a part thereof) |
| | Polygonal mesh in 3D view<br>and panorama view, cut<br>through polygonal mesh<br>(contour) in slice view<br>Features and Algorithms<br>identical to SICAT Function. | | Polygonal mesh in 3D view<br>and panorama view, cut<br>through polygonal mesh<br>(contour) in slice view |
| Optical<br>impression name | A list view in a dedicated<br>toolbar contains the names<br>of all loaded optical<br>impressions.<br>Features and Algorithms<br>identical to SICAT Function. | No | A list view in a dedicated<br>toolbar contains the names<br>of all loaded optical<br>impressions. |
| Show/Hide an<br>optical impression | Yes<br>Visibility applies to all views<br>simultaneously.<br>Algorithms identical to SICAT<br>Function. | No | Yes<br>Visibility applies to all views<br>simultaneously. |
| | Optical Impression Data Manipulation | | |
| Registration of<br>surface data<br>(from optical<br>impressions) to<br>volume data | Semi-automatic, using user<br>specified reference points.<br>Features and Algorithms<br>identical to SICAT Function. | No | Semi-automatic, using user<br>specified reference points. |
| Fine tuning the<br>accuracy of the<br>registration | Restart of the registration<br>with a different set of<br>reference points.<br>Features and Algorithms<br>identical to SICAT Function. | No | Restart of the registration<br>with a different set of<br>reference points. |
| Delete | Yes<br>Features and Algorithms<br>identical to SICAT Function. | No | Yes |
| | I/O | | |
| Volume data<br>import | DICOM | DICOM | DICOM |
| Optical surface<br>data / optical<br>impression import | Standard STL format and<br>proprietary SSI or SIXD<br>container format.<br>Features and Algorithms<br>identical to SICAT Function. | No | Standard STL format and<br>proprietary SSI or SIXD<br>container format. |
| Data<br>Import/Export<br>Data Compression | A "study" consisting of<br>planning data (including<br>segmentation information),<br>volume data, optical surface<br>data may be exported and<br>imported in a SICAT<br>proprietary format.<br>Lossless ZIP compression | No | A "study" consisting of<br>planning data (including<br>segmentation information),<br>volume data, jaw motion<br>data and optical surface data<br>may be exported and<br>imported in a SICAT<br>proprietary format. |
{10}------------------------------------------------
| | HICAT Air<br>Proposed | Reference predicate<br>Dolphin Imaging 11.5<br>(K110430) | Primary predicate<br>SICAT Function<br>(K133320) |
|------------------------------------------------------------------------------------------------------|---------------------------------------------------------|------------------------------------------------------------------|--------------------------------------------------|
| | Features and Algorithms<br>identical to SICAT Function. | | Lossless ZIP compression |
| Export of data for<br>CAD and rapid<br>proto-typing | Yes | No | Yes |
| | | Misc. functions | |
| Soft tissue<br>simulation with<br>photo mapping | No | Yes | No |
| Cross sections in<br>Multiple Planar<br>View (MPV) | No | Yes | No |
| Volume stitching:<br>Combine two<br>separate volumes<br>into one | No | Yes | No |
| 2D facial photo<br>wrap onto volume | No | Yes | No |
| Volume-to-<br>volume<br>superimposition | No | Yes | No |
| 3D nerve marking | No | Yes | No |
| Generate<br>panoramic<br>radiographs | Yes | Yes | Yes |
| Generate<br>cephalometric<br>radiographs | No | Yes | No |
| Create animated<br>movies with<br>automated scripts | No | Yes | No |
| Design and print<br>handout reports | Yes | Yes | No |
| Image export to<br>other applications | Yes | Yes | Yes |
| Main Functions | | | |
| Three-<br>dimensional<br>airway<br>segmentation and<br>analysis in the<br>ear-nose throat-<br>region | Yes<br>Image: 3D airway rendering | Yes<br>Image: Display of airway volume and most constricted area | No |
| Airway<br>Visualization | 3D surface rendering | 3D surface rendering | No |
| Airway Analysis:<br>Calculation of | Yes | Yes | No |
{11}------------------------------------------------
| | HICAT Air<br>Proposed | Reference predicate<br>Dolphin Imaging 11.5<br>(K110430) | Primary predicate<br>SICAT Function<br>(K133320) |
|-----------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| airway volume | Yes | Yes | No |
| Airway Analysis:<br>Calculation of the<br>airway cross<br>section<br>dimensions | Yes | Yes | No |
| Airway Analysis:<br>Calculation of the<br>position and size<br>of the area with<br>minimum cross<br>section in airway | Yes | Yes | No |
| Simulation of<br>orthodontic<br>procedures,<br>osteotomies and<br>distractions | No | Yes | No |
| Cephalometric<br>analysis | No | Yes | No |
| TMJ analysis | No | Yes | Yes |
| 3D implant<br>treatment<br>planning and<br>simulation | No | Yes | No |
| | | Programming Language | |
| | C# and C++ | Unknown | C# and C++ |
| | | System Requirements | |
| PC Hardware<br>Requirements<br>(Minimum) | | Minimum Imaging<br>Requirements (Rev 05/15) | |
| | CPU: 2,0 GHz Dualcore | CPU: Intel® Core 2 | CPU: 1 GHz |
| | RAM: 4 GB | RAM: 2 GB (4 GB for<br>Dolphin 3D) | RAM: 1 GB |
| | HDD: 5 GB free space | HDD: 20 GB free space | HDD: 5 GB free space |
| | Video card: Dedicated,<br>DirectX 10 or higher<br>512 MB video memory<br>Shader Model 3<br>"INTZ"-Surfaces<br>Current driver | Video card:…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.