K152476 · Chemence Medical Products, Inc. · MPN · Feb 9, 2016 · General, Plastic Surgery
Device Facts
Record ID
K152476
Device Name
Exofin High Viscosity Tissue Adhesive
Applicant
Chemence Medical Products, Inc.
Product Code
MPN · General, Plastic Surgery
Decision Date
Feb 9, 2016
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 878.4010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Exofin® High Viscosity Tissue Adhesive is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. Exofin® High Viscosity Tissue Adhesive may be used in conjunction with, but not in place of, deep dermal sutures.
Device Story
Exofin High Viscosity Tissue Adhesive is a sterile, liquid topical skin adhesive containing 2-octyl cyanoacrylate and D & C Violet #2 colorant. Provided in a single-use aluminum collapsible tube with an applicator featuring a self-puncturing cap and soft elastomeric brush. Used by clinicians to hold closed skin edges of surgical incisions or trauma-induced lacerations. Syrup-like viscosity prevents migration from wound site during application. Polymerizes within minutes to form a protective film. Acts as a microbial barrier while intact. Applied topically; intended for prescription use.
Clinical Evidence
Bench testing only. No clinical data provided. Performance evaluated via ASTM standards: wound closure strength (ASTM F2458-05), adhesive strength in tension (ASTM F2258-05), T-Peel adhesion strength (ASTM F2256-05), and lap-shear strength (ASTM F2258-05). Biocompatibility testing per ISO 10993-1 confirmed non-cytotoxic, non-irritating, and non-sensitizing properties. Microbial barrier properties demonstrated in vitro.
Technological Characteristics
Monomeric 2-octyl cyanoacrylate formulation with D & C Violet #2. Syrup-like viscosity. Applicator: self-puncturing cap with elastomeric brush. Sterilization: dry heat and ethylene oxide gas. Packaging: aluminum tube in RXM 48gaPET-200LDPE/Tyvek pouch. Testing standards: ASTM F2458-05, F2258-05, F2256-05; ISO 10993-1.
Indications for Use
Indicated for topical closure of easily approximated skin edges of surgical incisions from minimally invasive surgery and simple, cleansed, trauma-induced lacerations. May be used with deep dermal sutures. Not for use in place of deep dermal sutures.
Regulatory Classification
Identification
A tissue adhesive for the topical approximation of skin is a device intended for topical closure of surgical incisions, including laparoscopic incisions, and simple traumatic lacerations that have easily approximated skin edges. Tissue adhesives for the topical approximation of skin may be used in conjunction with, but not in place of, deep dermal stitches.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: “Tissue Adhesive for the Topical Approximation of Skin.” See § 878.1(e) of this chapter for the availability of this guidance document.(b)
*Tissue adhesive for non-topical use* —(1)*Identification.* A tissue adhesive for non-topical use, including adhesives intended for use in the embolization of brain arteriovenous malformation or for use in ophthalmic surgery, is a device used for adhesion of internal tissues and vessels.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 878.3 of this chapter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol is a stylized representation of human figures in profile, layered to create a sense of depth and connection. The figures are silhouetted in black, contrasting with the white background.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 9, 2016
Chemence Medical Products, Inc. Kenneth Broadley, Ph.D. Vice President Regulatory Affairs, Quality Assurance and NPD 200 Technology Drive Alpharetta, Georgia 30005
Re: K152476
Trade/Device Name: Exofin High Viscosity Tissue Adhesive Regulation Number: 21 CFR 878.4010 Regulation Name: Tissue adhesive Regulatory Class: Class II Product Code: MPN Dated: January 13, 2016 Received: January 14, 2016
Dear Dr. Broadley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# David Krause -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152476
Device Name Exofin® High Viscosity Tissue Adhesive
#### Indications for Use (Describe)
Exofin® High Viscosty Tissue Adhesive is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations.
Exofin® High Viscosity Tissue Adhesive may be used in conjunction with, but not in place of, deep dermal sutures.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Chemence Medical. The logo consists of a red diamond shape with a white "C" inside, above the word "CHEMENCE" in red and the word "MEDICAL" in blue. The "CHEMENCE" text also has a registered trademark symbol next to it.
# 510(k) Summary (In accordance with 21 CFR 807.92)
# Chemence Medical, Inc. Exofin® High Viscosity Tissue Adhesive
# 1. Submitter
| Submitted by: | Chemence Medical, Inc. |
|-----------------|-------------------------------------------|
| | 200 Technology Drive |
| | Alpharetta, GA 3005-3926 |
| | Phone: 844-633-4583 |
| | Fax: 678-820-3320 |
| Contact Person: | Dr. Kenneth N. Broadley |
| | Vice President |
| | Regulatory Affairs, Quality Assurance and |
| | New Product Development |
| | Chemence Medical, Inc. |
- Date of Summary: February 4th, 2016
# 2. Device
| Device Trade Name: | Exofin® High Viscosity Tissue Adhesive |
|-----------------------|----------------------------------------|
| Common or Usual Name: | Topical Skin Adhesive |
| Classification Name: | Tissue Adhesive (21 CFR 878.4010) |
| Regulatory Class: | Class II |
| Product Code: | MPN |
# 3. Predicate Devices
Legally marketed devices to which equivalence is claimed:
| Device Name: | Derma+Flex QS Tissue Adhesive |
|-------------------|------------------------------------|
| 510(k) Clearance: | K101276 |
| Device Name: | Dermabond NX Topical Skin Adhesive |
| 510(k) Clearance: | K100423 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Chemence Medical. The logo consists of a red and blue diamond shape with a stylized "C" in the center, where the left side of the "C" is red and the right side is blue. Below the diamond shape, the word "CHEMENCE" is written in red, and below that, the word "MEDICAL" is written in blue.
## 4. Device Description:
Exofin® High Viscosity Tissue Adhesive is a sterile, liquid topical skin adhesive containing a monomeric (2-octyl cyanoacrylate) formulation and the colorant D & C Violet #2. It is provided in a single use aluminum collapsible tube packaged in a RXM 48gaPET-200LDPE Film/1059B uncoated Tyvek pouch containing an applicator. The applicator is comprised of a self-puncturing cap and a soft elastomeric brush, which allows the adhesive to spread uniformly. As applied to skin, the liquid is syrup-like in viscosity and polymerizes within minutes. Exofin® High Viscosity Tissue Adhesive has a syrup-like viscosity. The increased viscosity is intended to reduce the risk of unintended placement of the adhesive during application due to migration of the liquid adhesive from the wound site. In vitro studies have shown that Exofin®High Viscosity Tissue Adhesive acts as a barrier to microbial penetration as long as the adhesive film remains intact. Clinical studies were not conducted to demonstrate microbial barrier properties and a correlation between microbial barrier properties and a reduction in infection have not been established.
### 5. Intended Use:
Exofin® High Viscosity Tissue Adhesive is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. Exofin® High Viscosity Tissue Adhesive may be used in conjunction with, but not in place of, deep dermal sutures.
## 6. Comparison of Technological Characteristics with the Predicate Device.
Exofin® High Viscosity Tissue Adhesive is very similar to Dermabond NX Topical Skin Adhesive and derma+flex QS with regard to intended use, mechanism of action and performance characteristics. All devices contain the same principle chemical ingredient, 2-octyl cyanoacrylate. Exofin® High Viscosity Tissue Adhesive has a different applicator from either of the two predicates and also has a larger volume of adhesive.
#### 7. Performance Data
Testing was performed in accordance with the FDA Class II Special Controls Guidance Document: Tissue Adhesive for the Topical Approximation of Skin.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Chemence Medical. The logo consists of a red and blue diamond shape with a white "C" in the center. Below the diamond shape is the word "CHEMENCE" in red, and below that is the word "MEDICAL" in blue. The Chemence logo is a registered trademark.
# Performance Testing
The following tests were performed on Exofin® High Viscosity Tissue Adhesive to demonstrate substantial equivalence:
- Wound closure strength (ASTM F2458-05)
- Adhesive strength in tension (ASTM F2258-05)
- T-Peel adhesion strength (ASTM F2256-05)
- Lap-Shear Strength (ASTM F2258-05)
- Heat of polymerization
- Degradation rate ●
- Viscosity ●
- Microbial barrier properties
### Biocompatibility Testing
The biological evaluation of Exofin® High Viscosity Tissue Adhesive was performed in accordance with ISO 10993-1, "Biological Testing of Medical Devices – Part 1: Evaluation and testing within a risk management process. The following test reports were provided in this submission:
- Cytotoxicity
- Intracutaneous irritation
- . Sensitization
Under the conditions of the studies, Exofin® High Viscosity Tissue Adhesive was considered to be noncytotoxic, non-irritating and non-sensitizing.
#### Sterilization and Shelf Life
Exofin® High Viscosity Tissue Adhesive is sterilized by dry heat and ethylene oxide gas. The shelf life of the device has been determined through both real time and accelerated aging studies.
## 8. Conclusion
Based on the testing carried out, Exofin® High Viscosity Tissue Adhesive is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.