DERMA+FLEX QS HIGH VISCOSITY TISSUE ADHESIVE MODEL: QS70404
K101276 · Chemence Medical Products, Inc. · MPN · Jul 28, 2010 · General, Plastic Surgery
Device Facts
Record ID
K101276
Device Name
DERMA+FLEX QS HIGH VISCOSITY TISSUE ADHESIVE MODEL: QS70404
Applicant
Chemence Medical Products, Inc.
Product Code
MPN · General, Plastic Surgery
Decision Date
Jul 28, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
derma+flex® QS™ High Viscosity Tissue Adhesive is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. derma+flex® QS™ High Viscosity Tissue Adhesive may be used in conjunction with, but not in place of, deep dermal sutures.
Device Story
derma+flex® QS™ is a sterile, liquid topical skin adhesive composed of 2-octyl cyanoacrylate and D & C Violet #2 colorant. Supplied in a single-use aluminum collapsible tube with two applicator tips (dauber with foam surface for uniform spreading; nozzle for detailed application). Upon application to skin, the syrup-like liquid polymerizes within minutes to hold closed skin edges. Used in clinical settings for surgical incisions and trauma-induced lacerations. Increased viscosity reduces risk of unintended migration from wound site. Provides a topical barrier to hold skin edges together, potentially reducing need for traditional sutures or staples in simple wounds; used adjunctively with deep dermal sutures for complex closures.
Clinical Evidence
Bench testing only. Performance and safety profile established in accordance with the Class II Special Controls Guidance Document: Tissue Adhesive for Topical Approximation of Skin.
Technological Characteristics
Monomeric 2-octyl cyanoacrylate formulation with D & C Violet #2 colorant. Syrup-like viscosity. Single-use aluminum collapsible tube. Applicators: dauber with foam surface and nozzle tip. Polymerization occurs upon contact with skin. Non-electronic, non-software device.
Indications for Use
Indicated for topical closure of easily approximated skin edges of surgical incisions (including minimally invasive punctures) and simple, cleansed, trauma-induced lacerations. May be used with deep dermal sutures. No specific age or gender restrictions noted.
Regulatory Classification
Identification
A tissue adhesive for the topical approximation of skin is a device intended for topical closure of surgical incisions, including laparoscopic incisions, and simple traumatic lacerations that have easily approximated skin edges. Tissue adhesives for the topical approximation of skin may be used in conjunction with, but not in place of, deep dermal stitches.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: “Tissue Adhesive for the Topical Approximation of Skin.” See § 878.1(e) of this chapter for the availability of this guidance document.(b)
*Tissue adhesive for non-topical use* —(1)*Identification.* A tissue adhesive for non-topical use, including adhesives intended for use in the embolization of brain arteriovenous malformation or for use in ophthalmic surgery, is a device used for adhesion of internal tissues and vessels.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 878.3 of this chapter.
Predicate Devices
High Viscosity Dermabond Topical Skin Adhesive
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Chemence Medical Products. The logo includes a medical symbol with a snake wrapped around a staff, followed by the company name in block letters. Above the logo, there is a handwritten code "K101276".
Page 1 of ②
# Section 5. Abbreviated 510(k) Summary
Submitted By: Chemence Medical Products, Inc. 185 Bluegrass Valley Pkwy. Alpharetta, GA 30005-2222 Phone: 678-879-5446 678-879-0251 Fax:
JUL 2 8 2010
Contact Person: Patricia D. Lehman Director of Regulatory Affairs Phone: 678-879-5446 Ext. 201 Email: plehman@chemence-us.com
Date of Summary: April 30, 2010
Device Trade Name: derma+flex QS High Viscosity Tissue Adhesive
Product Code: MPN
Tissue Adhesive (21CFR 878.4010)
High Viscosity Dermabond Topical Skin Adhesive
Common Name: Topical Skin Adhesive
Classification Name:
Predicate Device:
Device Description:
derma+flex® QS™ High Viscosity Tissue Adhesive is a sterile. liquid topical skin adhesive containing a monomeric (2-octy) cyanoacrylate) formulation and the colorant D & C Violet #2. It is provided in a single use aluminum collapsible tube packaged in a RXM 48gaPET-200LDPE Film/1059B uncoated Tyvek pouch also containing 2 Indothene HD Grade HD50MA180 applicator tips. The dauber applicator is comprised of a selfpuncturing cap and a foam surface, which allows spreading of the adhesive with uniformity. The nozzle applicator is also a self-puncturing cap with an elongation that enables detailed application of the adhesive. As applied to skin, the liquid is syrup-like in viscosity and polymerizes within minutes. derma+flex® QS™ High Viscosity Tissue Adhesive has a syrup-like viscosity. The increased viscosity derma+flex® QS™ is intended to reduce the risk of unintended placement of the adhesive during application due to migration of the liquid adhesive from the wound site.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Chemence Medical Products. The logo consists of a caduceus symbol on the left, followed by the word "CHEMENCE" in large, bold letters. Below that, the word "MEDICAL" is printed in a smaller font, and below that, the word "PRODUCTS" is printed in an even smaller font.
Page 2 of 2
## Indications for Use:
derma+flex® QS™ High Viscosity Tissue Adhesive is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. derma+flex® QS™ High Viscosity Tissue Adhesive may be used in conjunction with, but not in place of, deep dermal sutures.
#### Substantial Equivalence:
derma+flex® QS™ High Viscosity Tissue Adhesive is substantially equivalent to High Viscosity Dermabond Topical Skin Adhesive with regard to Indication for Use, formulation, technology, target population, intended application, mechanism of action and performance at achieving their intended use. Both devices present the same main chemical characteristics (2-Octyl Cyanoacrylate) and equivalent formulating materials with similar chemical characteristics. Devices shown to be equivalent in all performance and safety tests. derma+flex® QS™ High Viscosity Tissue Adhesive was evaluated in tests to establish a performance and safety profile in accordance with the Class II Specials Controls Guidance Document: Tissue Adhesive for Topical Approximation of Skin.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
JUL 2 8 2010
: 1.110
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Chemence Medical Products, Inc. % Ms. Patricia D. Lehman Director of Regulatory Affairs 185 Bluegrass Valley Parkway, Suite 100 Alpharetta, Georgia 30005-2222
Re: K101276
Trade/Device Name: derma+flex® QS™ High Viscosity Tissue Adhesive Regulation Number: 21 CFR 878.4010(a) Regulation Name: Tissue adhesive Regulatory Class: Class II Product Code: MPN Dated: May 04, 2010 Received: May 06, 2010
Dear Ms. Lehman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
Page 2 - Ms. Patricia D. Lehman
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
cerely yours,
For Peter S. Rummel
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K101276.
### Indications for Use
້ັມັບLİ
510(k) Number:
Device Name:
derma+flex® QS™ High Viscosity Tissue Adhesive
Indications for Use:
derma+flex® QS™ High Viscosity Tissue Adhesive is intended for topical application only to hold closed easily approximated skin edges of wounds from surgical incisions, including punctures from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations.
derma+flex® QS™ High Viscosity Tissue Adhesive may be used in conjunction with, but not in place of, deep dermal sutures
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Kuore for MXM
Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
**510(k) Number** K101276
4 - 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.