K100423 · ETHICON, Inc. · MPN · May 4, 2010 · General, Plastic Surgery
Device Facts
Record ID
K100423
Device Name
DERMABOND NX ADHESIVE MODEL: DNX6, DNX12
Applicant
ETHICON, Inc.
Product Code
MPN · General, Plastic Surgery
Decision Date
May 4, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
DERMABOND® NX Topical Skin Adhesive is intended for topical application only, to hold closed easily approximated skin edges of wounds from surgical incisions, including incisions from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. DERMABOND® NX Topical Skin Adhesive may be used in conjunction with, but not in place of, deep dermal stitches.
Device Story
DERMABOND NX is a sterile, liquid topical skin adhesive (2-octyl cyanoacrylate) in a pen-style applicator with a crushable ampoule. Designed for single-use, it is applied to skin edges in a single layer; the syrup-like liquid polymerizes within minutes to form a flexible, microbial barrier. Used by clinicians in surgical or clinical settings to approximate skin edges. Higher viscosity formulation allows for one-pass application; modified compound improves heat dissipation during curing. Benefits include efficient wound closure and protection against microbial penetration.
Clinical Evidence
No clinical data. Substantial equivalence supported by biocompatibility testing (acute systemic toxicity, ocular/skin irritation, sensitization, intracutaneous reactivity, subcutaneous implantation, cytotoxicity) and bench testing (activation force, compression force, expressed mass, wound burst strength, tensile/shear testing, water resistance).
Technological Characteristics
Liquid 2-octyl cyanoacrylate monomer with viscosity increasing agent, stabilizer, and D&C Violet #2 colorant. Supplied in a pen-style applicator with a crushable ampoule. Sterilized via dry heat (ampoule) and 100% ethylene oxide (final package). Higher viscosity than predicate for single-pass application; includes heat dissipation agent.
Indications for Use
Indicated for topical closure of easily approximated skin edges of surgical incisions (including minimally invasive) and simple, cleansed, trauma-induced lacerations. May be used with deep dermal stitches.
Regulatory Classification
Identification
A tissue adhesive for the topical approximation of skin is a device intended for topical closure of surgical incisions, including laparoscopic incisions, and simple traumatic lacerations that have easily approximated skin edges. Tissue adhesives for the topical approximation of skin may be used in conjunction with, but not in place of, deep dermal stitches.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: “Tissue Adhesive for the Topical Approximation of Skin.” See § 878.1(e) of this chapter for the availability of this guidance document.(b)
*Tissue adhesive for non-topical use* —(1)*Identification.* A tissue adhesive for non-topical use, including adhesives intended for use in the embolization of brain arteriovenous malformation or for use in ophthalmic surgery, is a device used for adhesion of internal tissues and vessels.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 878.3 of this chapter.
{0}------------------------------------------------
Page 1 of 13
MAY - 4 2010
510(k) Summary as required by 21 Section 5.
510(k) Owner:
Ethicon, Inc. P.O. Box 151, Route 22 West Somerville, NJ 08876-0151 Telephone: (908) 218-3323 Fax: (908) 218-2595
Contact Person: Neelu Medhekar, WW Director, Regulatory Affairs, Ethicon Products
February 1, 2010 Date:
Trade Name: DERMABOND® NX Topical Skin Adhesive
Common Name: Topical Skin Adhesive
Classification Name: Tissue Adhesive for the Topical Approximation of Skin
Product Code: MPN
Predicate Device: DERMABOND® Topical Skin Adhesive (P960052): Reclassified from Class III PMA to Class II 510(k) - May 5, 2008 - Docket number 2006p-071.
#### Device Description:
DERMABOND® NX Topical Skin Adhesive is a sterile, liquid topical skin adhesive containing a monomeric (2-octyl cyanoacrylate) formulation with a viscosity increasing agent, stabilizer, and colorant (D&C Violet #2). It is provided as a single use applicator in a blister package. The pen style applicator is composed of a crushable ampoule contained within a plastic applicator. As it is applied to easily approximated skin edges in a single layer; the liquid is syrup-like in viscosity and polymerizes within minutes. Studies have shown that following application, DERMABOND NX Topical Skin Adhesive acts as a barrier to prevent microbial penetration.
CONFIDENTIAL
{1}------------------------------------------------
K100423
## Indications for Use:
DERMABOND® NX Topical Skin Adhesive is intended for topical application only, to hold closed easily approximated skin edges of wounds from surgical incisions, including incisions from minimally invasive surgery, and simple, thoroughly cleansed, traumainduced lacerations. DERMABOND® NX Topical Skin Adhesive may be used in conjunction with, but not in place of, deep dermal stitches.
These indications for use are identical to those of Dermabond (predicate device).
# Technological Characteristics:
The technological characteristics of DERMABOND NX Topical Skin Adhesive are equivalent in performance to the predicate device, DERMABOND Topical Skin Adhesive.
DERMABOND NX consists of a liguid topical skin adhesive formulation packaged within a dispensing applicator. The device is supplied in a sterile single-use package for use in wound closure procedures. DERMABOND NX is a higher viscosity formulation to allow precise application of the adhesive to the intended area and allow a single application of the adhesive to the wound area. The topical skin adhesive is designed to bond to the skin to provide flexible wound closure maintaining wound approximation and providing a microbial barrier.
The main difference between DERMABOND NX and the predicate device is the adjusted higher viscosity to allow a one pass application and a different compound to increase the heat dissipation, reducing heat generated during the adhesive curing process on the skin.
### Determination of Substantial Equivalence:
# 1. Biocompatibility
The biocompatibility of DERMABOND NX Topical Skin Adhesive device has been evaluated through the use of the recommended biocompatibility tests. Tests included acute systemic toxicity, ocular irritation, primary skin irritation, Kligman sensitization, intracutaneous reactivity, subcutaneous implantation and MEM (cytotoxicity). The results were substantially equivalent to those of the predicate device. These results are not substantially different when compared to the predicate device when the same procedures were used.
{2}------------------------------------------------
Ki00423
### 2. Performance Bench Testing:
Page 3 of 3
All testing for DERMABOND NX Topical Skin Adhesive has been completed according to Design Control requirements of 21CFR820.30. The testing plan and results were based on the FDA guidance document "Guidance for Industry and FDA Staff: Class II Special Controls Guidance Document: Tissue Adhesive for the Topical Approximation of Skin". Bench tests included device activation force, compression force, expressed mass, wound burst strength, flatwise tensile and overlap shear testing, skin tensile and T-peel testing and water resistance testing. In all cases, DERMABOND NX met specifications and demonstrated equivalence to the predicate device with the minor changes of better heat dissipation and increased viscosity.
### 3. Sterilization and Shelf-Life:
Sterilization of DERMABOND NX Topical Skin Adhesive is the same as the predicate device, DERMABOND family of products (P960052). The sterilization process consists of two parts: 1) The dry heat sterilization of formulation filled ampoules containing liquid topical skin adhesive. And 2) The 100% ethylene oxide (EO) sterilization of the device after final packaging. Both of these processes provide a minimum sterility assurance level (SAL) of 10 which are the same as those used for sterilization of the predicate DERMABOND products.
Both real-time and accelerated stability testing data has been collected in support of this submission. Two years equivalent of accelerated testing has been completed supplementing the real time data of currently 12 months. Real time data will continue to be collected and analyzed to support a two (2) year shelf life labeling claim.
Based on extensive biocompatibility and bench testing, Dermabond NX has been demonstrated to be substantially equivalent to its predicate device from a safety perspective and has also demonstrated higher viscosity and better heat dissipation.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the upper portion of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ethicon, Inc. % Ms. Neelu Medhekar Director, Regulatory Affairs P.O. Box 151, Route 22 West Somerville, New Jersey 08876-0151
MAY - 4 2010
Re: K100423
Trade/Device Name: DERMABOND® NX Topical Skin Adhesive Regulation Number: 21 CFR 878.4010 Regulation Name: Tissue adhesive Regulatory Class: Class II Product Code: MPN Dated: February 05, 2010 Received: February 16, 2010
Dear Ms. Medhekar;
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Ms. Neelu Medhekar
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Muli. A. Millerm
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### INDICATIONS FOR USE
510(k) Number (if known):
7
Device Name: DERMABOND® NX Topical Skin Adhesive
Indications For Use:
DERMABOND NX Topical Skin Adhesive is intended for topical application only. to hold closed easily approximated skin edges of wounds from surgical incisions, including incisions from minimally invasive surgery, and simple, thoroughly cleansed, trauma-induced lacerations. DERMABOND NX Topical Skin Adhesive may be used in conjunction with, but not in place of, deep dermal stitches.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use NO (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Krone for MKM
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K100423
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.