K143676 · GE Healthcare Finland Oy · MHX · Jul 31, 2015 · Cardiovascular
Device Facts
Record ID
K143676
Device Name
B40i Patient Monitor
Applicant
GE Healthcare Finland Oy
Product Code
MHX · Cardiovascular
Decision Date
Jul 31, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The B40i is a portable multi-parameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport. The B40i is intended for use under the direct supervision of a licensed health care practitioner. The B40i is not intended for use during MRI. The B40i can be a stand-alone monitor or interfaced to other devices via a network. The B40i monitors and displays: ECG (including ST segment, arrhythmia detection), invasive blood pressure, heart/pulse rate, oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), functional oxygen saturation (SpO2) and pulse rate via continuous monitoring (including monitoring conditions of clinical patient motion or low perfusion), temperature with a reusable or disposable electronic thermometer for continual monitoring Esophageal/Nasopharvngeal/Tympanic/Rectal/Bladder/Axillary/Skin/Airway/Room/Myo cardial/Core/Surface temperature, impedance respiration, respiration rate, airway gases (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiratory rate) and Entropy.
Device Story
B40i is a portable multi-parameter patient monitor; inputs include ECG, invasive/non-invasive blood pressure, SpO2, temperature, impedance respiration, and airway gases. Device processes signals to display real-time physiological data, generate alarms, and trend information. Used in hospital settings and intra-hospital transport by clinicians. Features 12.1-inch LCD, integrated keypad, and network interface for central station connectivity. Expands impedance respiration monitoring to neonates using algorithms from predicate Aware Transport (K042642). Supports optional modules for Entropy and respiratory gas analysis. Provides clinicians with continuous patient status updates to facilitate timely clinical decision-making and patient safety monitoring.
Clinical Evidence
No clinical data was performed for the B40i. Substantial equivalence is supported by bench testing, including module verification, system integration testing, and performance/safety testing against recognized standards (e.g., IEC 60601 series, AAMI/ANSI EC13, ISO 21647, ISO 9919).
Technological Characteristics
Multi-parameter monitor with 12.1-inch TFT LCD, LED backlight, and 18-key interface. Powered by Lithium-Ion battery. Connectivity via LAN. Materials include standard medical-grade components. Sensing principles: electrical potential (ECG), oscillometric (NIBP), bridge-type transducers (IBP), thermally sensitive resistors (Temp), non-dispersive infrared/paramagnetic (Gas), and impedance variation (Respiration). Software runs on Linux OS. Sterilization not applicable (external device).
Indications for Use
Indicated for monitoring and recording physiological parameters of adult, pediatric, and neonatal patients in hospital environments and during intra-hospital transport. Not for use during MRI. Requires direct supervision of a licensed health care practitioner.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized caduceus-like symbol with three human profiles facing right, representing health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol, indicating the department's name and national affiliation.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 31, 2015
GE Healthcare Finland Oy % Joel Kent Regulatory Affairs Manager GE Healthcare 86 Pilgrim Road Needham, Massachusetts 02492
Re: K143676
Trade/Device Name: B40i Patient Monitor Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm). Regulatory Class: Class II Product Code: MHX, BZO, CBO, CBR, CBS, CCK, CCL, DSK, DOA, DRT, DSB, DXN, FLL, GWQ, NHO, NHO, NHQ, NHP Dated: June 17, 2015 Received: June 23, 2015
Dear Joel Kent,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Bram D. Zuckerman -S
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### INDICATIONS FOR USE
510(k) Number (if known): K143676
Device Name: B40i
Indications for use:
The B40i is a portable multi-parameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport. The B40i is intended for use under the direct supervision of a licensed health care practitioner.
The B40i is not intended for use during MRI.
The B40i can be a stand-alone monitor or interfaced to other devices via a network. The B40i monitors and displays: ECG (including ST segment, arrhythmia detection), invasive blood pressure, heart/pulse rate, oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), functional oxygen saturation (SpO2) and pulse rate via continuous monitoring (including monitoring conditions of clinical patient motion or low perfusion), temperature with a reusable or disposable electronic thermometer for continual monitoring
Esophageal/Nasopharvngeal/Tympanic/Rectal/Bladder/Axillary/Skin/Airway/Room/Myo cardial/Core/Surface temperature, impedance respiration, respiration rate, airway gases (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiratory rate) and Entropy.
Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
## 510(k) Summary
In accordance with 21 CFR 807.92(c) the following summary of information is provided:
## Owner/Contact/Date (807.92(a)(1)):
| Date: | 19 December 2014 |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Healthcare Finland Oy<br>Kuortaneenkatu 2,<br>00510 Helsinki<br>Finland<br>Phone: +358-40-539 7764 |
| Primary Contact Person: | Joel Kent<br>Manager, Quality and Regulatory Affairs<br>GE Medical Systems Information Technologies, Inc.<br>Telephone: +1 617 851 0943<br>Fax at +1 781 433-1344<br>E-mail: joel.kent@med.ge.com |
| Secondary Contact Person: | Anssi Ruokonen<br>Regulatory Affairs Leader<br>GE Healthcare Finland Oy<br>Kuortaneenkatu 2,<br>00510 Helsinki<br>Finland<br>Phone: +358-10-394 3686<br>E-mail: anssi.ruokonen@med.ge.com |
Device names (807.92(a)(2)):
Trade Name: B40i Common/Usual Name: Multi-parameter patient monitor
{4}------------------------------------------------
| Classification Names: | 21 CFR 870.1025 Arrhythmia detector and alarm (including<br>ST-segment measurement and alarm) |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification Product<br>Code: | MHX |
| Subsequent Product Codes | BZQ CBQ CBR CBS CCK<br>CCL DQA DRT DSB DSK DXN<br>FLL GWQ NHO NHP NHQ |
| Predicate Device(s)<br>(807.92(a)(3): | K133576 Monitor B40<br>K071073 Patient Data Module |
| Device Description<br>(807.92(a)(4)): | The proposed monitor B40i is a multi-parameter patient monitor<br>that is developed based on the predicate Monitor B40 (K133576)<br>platform. The proposed monitor B40i provides support for<br>optional modules (E-Entropy module (K061907) and<br>CARESCAPE Respiratory modules (E-sCO and E-sCAiO)<br>(K123195). The proposed monitor B40i is also compatible with<br>CARESCAPE Respiratory modules (E-sCOV and E-sCAiOV)<br>(K123195) but with disabled spirometry function. The proposed<br>monitor B40i supports Airway Gas Option (N-CAiO). The<br>proposed monitor B40i expands the impedance respiration<br>parameter feature to cover the neonatal patient population<br>compared to the predicate Monitor B40 (K133576). This<br>parameter feature patient population extension to cover neonatal<br>patient population uses Patient Data Module as predicate<br>(K071073), which impedance respiration implementation<br>especially concerning the algorithm used is based on the predicate<br>Aware Transport (K042642). The proposed monitor B40i utilizes<br>12 inch LCD display panel and LED backlight with an integrated<br>keypad and a pre-configuration patient parameter measurement<br>module. The proposed monitor B40i interfaces with the optional<br>E-MiniC (K052582) and Thermal Recorder with an extension<br>rack. As with the predicate Monitor B40, the proposed monitor<br>B40i includes features and subsystems that are optional or<br>configurable. The proposed monitor B40i interfaces to a variety<br>of existing central station systems via a cabled network interface.<br>As with the predicate Monitor B40, the proposed monitor B40i<br>has a mounting plate on the bottom of the monitor. The monitor<br>can be mounted in a variety of ways (e.g. shelf, countertop, table,<br>wall, pole, or head/foot board) using existing mounting<br>accessories. |
{5}------------------------------------------------
| Intended Use (807.92(a)(5)): | The B40i is a portable multi-parameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport. |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The B40i is intended for use under the direct supervision of a licensed health care practitioner. |
| | The B40i is not intended for use during MRI. |
| | The B40i can be a stand-alone monitor or interfaced to other devices via a network. |
| | The B40i monitors and displays: |
| | ECG (including ST segment, arrhythmia detection), invasive blood pressure, heart/pulse rate, oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), functional oxygen saturation (SpO2) and pulse rate via continuous monitoring (including monitoring during conditions of clinical patient motion or low perfusion), temperature with a reusable or disposable electronic thermometer for continual monitoring. |
| | Esophageal/Nasopharyngeal/Tympanic/Rectal/Bladder/Axillary/Skin/Airway/Room/Myocardial/Core/Surface temperature, impedance respiration, respiration rate, airway gases (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiratory rate) and Entropy. |
| Technology (807.92(a)(6)): | The proposed monitor B40i is a modified system based on the predicate Monitor B40 (K133576). In addition to the labeling differences with the predicate Monitor B40 (K133576), the proposed monitor B40i also expands the impedance respiration parameter feature to cover the neonatal patient population compared to the predicate Monitor B40 (K133576). Patient Data Module (K071073) is the predicate, which impedance respiration implementation especially concerning the algorithm used is based on the predicate Aware Transport (K042642). The fundamental technology of the proposed monitor B40i is the same as the predicate devices. The proposed monitor B40i is as safe and effective as the predicate devices as summarized in the comparison table below. |
{6}------------------------------------------------
| Feature/Function | Current Device/System<br>the legally marketed<br>predicate Monitor B40V2.1<br>(K133576) | Proposed Device/System<br>Name<br>proposed Monitor B40i | Change<br>Explanation/Notes |
|---------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|
| Intended Use description to include: | | | |
| Claims/ features | The Monitor B40 is a portable<br>multi-parameter unit to be used<br>for monitoring and recording of,<br>and to generate alarms for,<br>multiple physiological<br>parameters of adult, pediatric,<br>and neonatal patients in a<br>hospital environment and during<br>intra-hospital transport.<br>The Monitor B40 is intended for<br>use under the direct supervision<br>of a licensed health care<br>practitioner.<br>The Monitor B40 is not intended<br>for use during MRI.<br>The Monitor B40 can be a<br>stand-alone monitor or<br>interfaced to other devices via a<br>network.<br>The Monitor B40 monitors and<br>displays: ECG (including ST<br>segment, arrhythmia detection),<br>invasive blood pressure,<br>heart/pulse rate, oscillometric<br>non-invasive blood pressure<br>(systolic, diastolic and mean<br>arterial pressure), functional<br>oxygen saturation (SpO2) and<br>pulse rate via continuous<br>monitoring (including monitoring<br>during conditions of clinical<br>patient motion or low perfusion),<br>temperature with a reusable or<br>disposable electronic<br>thermometer for continual<br>monitoring.<br>Esophageal/Nasopharyngeal/Ty<br>mpanic/Rectal/B ladder/Axi<br>llary/Skin/Airway/Room/Myocard<br>ial/Core/Surface temperature,<br>impedance respiration,<br>respiration rate, airway gases<br>(C02, 02, N20, anesthetic<br>agents, anesthetic agent<br>identification and respiratory<br>rate) and Entropy. | The B40i is a portable multi-<br>parameter unit to be used for<br>monitoring and recording of, and<br>to generate alarms for, multiple<br>physiological parameters of<br>adult, pediatric, and neonatal<br>patients in a hospital<br>environment and during intra-<br>hospital transport.<br>The B40i is intended for use<br>under the direct supervision of a<br>licensed health care practitioner.<br>The B40i is not intended for use<br>during MRI.<br>The B40i can be a stand-alone<br>monitor or interfaced to other<br>devices via a network.<br>The B40i monitors and displays:<br>ECG (including ST segment,<br>arrhythmia detection), invasive<br>blood pressure, heart/pulse rate,<br>oscillometric non-invasive blood<br>pressure (systolic, diastolic and<br>mean arterial pressure),<br>functional oxygen saturation<br>(SpO2) and pulse rate via<br>continuous monitoring (including<br>monitoring during conditions of<br>clinical patient motion or low<br>perfusion), temperature with a<br>reusable or disposable<br>electronic thermometer for<br>continual monitoring<br>Esophageal/Nasopharyngeal/Ty<br>mpanic/Rectal/Bladder/Axillary/S<br>kin/Airway/Room/Myocardial/Cor<br>e/Surface temperature,<br>impedance respiration,<br>respiration rate, airway gases<br>(CO2, O2, N2O, anesthetic<br>agents, anesthetic agent<br>identification and respiratory<br>rate) and Entropy. | Identical (only the<br>name of the monitor is<br>different) |
| Patient<br>Population | Adult, pediatric and neonate | Adult, pediatric and neonate | Identical |
| Environment of<br>Use | hospital environment and during<br>intra-hospital transport | hospital environment and during<br>intra-hospital transport | Identical |
| Hardware: | | | |
| Software Media | Embedded | Embedded | Identical |
| Battery Type | Lithium-Ion | Lithium-Ion | Identical |
| Battery run in time | >=2:15<br>typical monitor configuration:<br>- ECG, NIBP cycle time 5min,<br>SpO2, 2x INVP, 2x Temp<br>and continuous CO2 use<br>- Ambient temperature 25 °C<br>- Display brightness 70% | >=2:15<br>typical monitor configuration:<br>- ECG, NIBP cycle time 5min,<br>SpO2, 2x INVP, 2x Temp and<br>continuous CO2 use<br>- Ambient temperature 25 °C<br>- Display brightness 70% | Identical |
| Operating Systems Infrastructure Software: | | | |
| Operating System | Linux | Linux | Identical |
| Networking | | | |
| Networking<br>Interface | LAN | LAN | Identical |
| User Interface | | | |
| Front Key Pad | 18 hard keys with Trim Knob,<br>and 1 power On/Standby | 18 hard keys with Trim Knob,<br>and 1 power On/Standby | Identical |
| Display: | | | |
| Size | 12.1-inch | 12.1-inch | Identical |
| Types | TFT LCD | TFT LCD…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.