DINAMAP PRO 1000 MONITOR WITH SUPERSTAT, MODEL 1000
K022834 · Ge Medical Systems Information Technologies · MHX · May 13, 2003 · Cardiovascular
Device Facts
Record ID
K022834
Device Name
DINAMAP PRO 1000 MONITOR WITH SUPERSTAT, MODEL 1000
Applicant
Ge Medical Systems Information Technologies
Product Code
MHX · Cardiovascular
Decision Date
May 13, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The DINAMAP® Pro 1000 with SuperStat Monitor is intended to monitor the following adult, pediatric and neonate patient vital signs: Non-Invasive Blood Presure: · Able to measure oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure) · Will optimize performance in the presence of artifact due to vibration and patient motion. •Non-invasive blood pressure determinations acquired concurrently with the ECG parameter will report a determination in the presence of an irregular heart rhythm in the adult/pediatric mode. Heart/pulse rate, respiration rate, ECG Oxygen saturation (SpO2) by noninvasive pulse oximetry and Predictive temperature by resistive thermometry. The Pro 1000 with SuperStat Monitor also detects alarm limit conditions and is capable of recording up to two waveforms. Using the monitor a clinician can view, record and recall clinical data derived from each parameter.
Device Story
Monitor for adult, pediatric, and neonatal vital signs; inputs include oscillometric NIBP, ECG, SpO2, and resistive temperature sensors. Device processes signals to derive systolic, diastolic, and mean arterial pressure, heart/pulse rate, and respiration rate. Features SuperStat algorithm to optimize NIBP performance during motion/vibration and irregular heart rhythms. Used in hospitals and outpatient facilities by clinicians; portable via internal battery for intra-hospital transport. Outputs displayed on-screen; device records up to two waveforms and detects alarm limit conditions. Clinicians use data for patient assessment and clinical decision-making; networking capabilities allow central station monitoring.
Clinical Evidence
Clinical studies conducted to demonstrate performance of the SuperStat NIBP algorithm according to ANSI/AAMI SP10 standard (American National Standard for Electronic or Automated Sphygmomanometers).
Technological Characteristics
Software-driven electronic monitor; parameters include oscillometric NIBP (SuperStat algorithm), ECG, SpO2, and resistive thermometry. Connectivity via VHF, 900 MHz, or hardwire to central stations. Portable with internal battery. Compliant with ANSI/AAMI SP10 for NIBP.
Indications for Use
Indicated for monitoring vital signs (NIBP, heart/pulse rate, respiration, ECG, SpO2, temperature) in adult, pediatric, and neonatal patients in hospital, outpatient surgery, and healthcare facilities.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The letters and the border are in black, creating a strong contrast. The logo is simple, recognizable, and represents the General Electric brand.
K022834
MAY 1 3 2003
General Electric Company 4502 Woodland Corporate Blvd., Tampa, FL 33614 813 887-2000
April 7, 2003
# SUMMARY OF SAFETY AND EFFECTIVENESS
#### DINAMAP® Pro 1000 Monitor with SuperStat
#### A. Submitter
GE Medical Systems Information Technologies. 4502 Woodland Corporate Boulevard Tampa, FL 33614
#### B. Company Contact
Primary:
Melissa Robinson, Regulatory Affairs Specialist Secondary: Thomas English, Global QA/RA
### C. Device Name
Pro 1000 Monitor with SuperStat Trade Name: Common Name: Physiological Monitor, Patient Monitor Classification/Device Product Code: System, Measurement, Blood Pressure, Noninvasive-870.1130-DXN Computer, Blood Pressure-870.1110-DSK Alarm, Blood Pressure-870.1100-DSJ Oximeter-870.2700-DQA Oximeter, Ear-870.2710-DPZ Thermometer, Clinical Electronic-880.2910-FLL Monitor, Cardiac (including cardiotachometer & rate alarm)-870.2300-DRT Electrocardiograph-870.2340-DPS Adapter, Lead Switching, Electrocardiograph-870.2350-DRW Arrhythmia Detection and Alarm-870.1025-DSI Monitor, Breathing Frequency-868.2375-BZO Recorder, Paper Chart-870.2810-DSF
# D. Predicate/Legally Marketed Devices
DINAMAP® Pro 1000 V2 Monitor, K012915 GE Medical Systems Information Technologies
{1}------------------------------------------------
#### E. Device Description
The DINAMAP® Pro 1000 Monitor w/ SuperStat is intended to monitor a single patient's vital signs in the hospital, outpatient surgery and healthcare practitioner facilities. The patient populations include adult, pediatric, and neonatal. The device's networking capabilities include connection to a central station via VHF, 900 MHz or hardwire communication: host communications for use with other devices. In addition, the DINAMAP Pro 1000 Monitor may be operated from internal batteries making the device portable and suitable for intra-hospital transport.
#### F. Intended Use
The DINAMAP® Pro 1000 with SuperStat Monitor is intended to monitor the following adult, pediatric and neonate patient vital signs:
Non-Invasive Blood Pressure:
· Able to measure oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure)
· Will optimize performance in the presence of artifact due to vibration and patient motion.
•Non-invasive blood pressure determinations acquired concurrently with the ECG parameter will report a determination in the presence of an irregular heart rhythm in the adult/pediatric mode.
Heart/pulse rate, respiration rate, ECG
Oxygen saturation (SpO2) by noninvasive pulse oximetry and
Predictive temperature by resistive thermometry.
The Pro 1000 with SuperStat Monitor also detects alarm limit conditions and is capable of recording up to two waveforms. Using the monitor a clinician can view, record and recall clinical data derived from each parameter.
# G. Technological Characteristics
The DINAMAP® Pro 1000 Monitor with SuperStat has the same technological characteristics as the predicate device, the DINAMAP® Pro 1000 V2 Monitor. There are no new technological characteristics. The Pro 1000 Monitor with SuperStat and the Pro 1000 V2 Monitor are both software-driven electronic devices that include the same monitoring parameters.
# H. Parameter Technology
The DINAMAP PRO 1000 Monitor with SuperStat has the following parameter technologies:
·SuperStat NIBP algorithm tested according to the ANSI/AAMI SP10 standard.
- · Temperature-Same as the PRO 1000 V2 Monitor
- ·SpO2-Same as the PRO 1000 V2 Monitor
- •ECG-Same as the PRO 1000 V2 Monitor
{2}------------------------------------------------
•Respiration-Same as the PRO 1000 V2 Monitor
# I. Testing
Clinical studies were conducted to demonstrate performance (safety and effectiveness) of the DINAMAP PRO 1000 with SuperStat Monitor to the ANSI/AAMI SP10 Standard: American National Standard for Electronic or Automated Sphygmomanometers.
#### J. Conclusion
The DINAMAP® Pro 1000 with SuperStat Monitor is substantially equivalent to the currently marketed DINAMAP® Pro 1000 V2 Monitor, K012915 cleared 3/21/02.
{3}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 1 3 2003
GE Medical Systems Information Technologies c/o Ms. Melissa Robinson Regulatory Affairs Specialist 4502 Woodland Corporate Blvd. Tampa, FL 33614
Re: K022834
Trade Name: DINAMAP® Pro 1000 Monitor with SuperStat Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm Regulatory Class: Class III (three) Product Code: MHX Dated: March 17, 2003 Received: March 19, 2003
Dear Ms. Robinson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
Page 2 – Ms. Melissa Robinson
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Delettu
am D. Zucker rman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
April 7, 2003
# 510(K) Number (if known):
Device Name: DINAMAP® Pro 1000 Monitor with SuperStat
Indications for Use:
The DINAMAP® Pro 1000 with SuperStat Monitor is intended to monitor the following adult, pediatric and neonate patient vital signs:
Non-Invasive Blood Presure:
- · Able to measure oscillometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure)
- · Will optimize performance in the presence of artifact due to vibration and patient motion.
- •Non-invasive blood pressure determinations acquired concurrently with the ECG parameter will report a determination in the presence of an irregular heart rhythm in the adult/pediatric mode.
Heart/pulse rate, respiration rate, ECG
Oxygen saturation (SpO2) by noninvasive pulse oximetry and
Predictive temperature by resistive thermometry.
The Pro 1000 with SuperStat Monitor also detects alarm limit conditions and is capable of recording up to two waveforms. Using the monitor a clinician can view, record and recall clinical data derived from each parameter.
(Please Do Not Write Below This Line-Continue On Another Page If Needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
Or
Over-The Counter Use
DeleTu
**510(k) Number**
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.