K143002 · Changzhou Dingjian Medical Appliance Co., Ltd. · HRS · Jun 5, 2015 · Orthopedic
Device Facts
Record ID
K143002
Device Name
Metallic Locking Bone Plate and Screw System
Applicant
Changzhou Dingjian Medical Appliance Co., Ltd.
Product Code
HRS · Orthopedic
Decision Date
Jun 5, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Metallic Locking Bone Plate and Screws System are intents with age above 21 as indicated for fixation of fractures, including ulna, radius, humerus, femur and tibia.
Device Story
Metallic Locking Bone Plate and Screw System; internal fixation device for bone fractures. System comprises bone plates with taper-thread holes and locking screws with taper-thread heads. When screws are tightened into plates, system functions as angle-fixed device; maintains anatomical reduction to restore physiological function during bone healing. Used by surgeons in clinical settings for orthopedic fracture fixation. Provided non-sterile; requires steam sterilization by healthcare provider prior to use. Single-use only.
Clinical Evidence
Bench testing only. No clinical or animal studies performed. Mechanical performance verified via ASTM F382 (static/dynamic four-point bending) and ASTM F543 (torsional properties, driving torque, pull-out test).
Technological Characteristics
Materials: Titanium (ASTM F67) and Ti6Al4V ELI (ASTM F136). Design: Locking plate with taper-thread holes and locking screws. Mechanical standards: ASTM F382-99(2008) for plates, ASTM F543-07 for screws. Sterilization: Steam (autoclave) required by user. Non-implantable software: None.
Indications for Use
Indicated for adult patients age >21 requiring fixation of fractures of the ulna, radius, humerus, femur, and tibia.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, representing the department's focus on people and health.
June 5, 2015
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Changzhou Dingjian Medical Appliance Company, Ltd % Mr. Mike Gu OSMUNDA Medical Device Consulting Company, Ltd 7th Floor Jingui Business Buildng No. 982 Cogyuan Road, Baiyun District Guangzhou, 510420 Guangdong China
Re: K143002/S003 Trade/Device Name: Metallic Locking Bone Plate and Screw System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: May 25, 2015 Received: May 28, 2015
Dear Mr. Mike Gu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement on last page.
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K143002
Device Name Metallic Locking Bone Plate and Screw System
Indications for Use (Describe)
The Metallic Locking Bone Plate and Screws System are intents with age above 21 as indicated for fixation of fractures, including ulna, radius, humerus, femur and tibia.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for D&J medical. The logo consists of a stylized graphic of three concentric arcs with circles at the ends, colored in shades of green and teal. To the right of the graphic, the text "D&J medical" is written in green. Below the text, there are three Chinese characters in gray.
Premarket Notification Submission_ Metallic Locking Bone Plate and Screw System
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided: SUBMITTER l.
Changzhou Dingjian Medical Appliance Co., Ltd.
No. 19 Juxiang Road
Wujin Economic Development Zone
Changzhou, Jiangsu Province, 213149, China
Mike Gu Primary Contact Person:
Regulatory Affairs Manager
OSMUNDA Medical Device Consulting Co., Ltd
Tel: (+86) 20-6232 1333
Fax: (+86) 20-8633 0253
Secondary Contact Person: Li Qingsong
Changzhou Dingjian Medical Appliance Co., Ltd.
Tel: 0086-13584330156
Fax: 0086-519-89603573
| Date Prepared: | April 13, 2015 | |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| DEVICE<br>II. | | |
| Name of Device: | Metallic Locking Bone Plate and Screw System | |
| Common/Usual Name: | Bone Fixation Plate and Bone Fixation Screw | |
| Classification Names: | Plate, Fixation, Bone (21 CFR 888.3030) | |
| | Screw, Fixation, Bone (21 CFR 888.3040) | |
| Regulation Class: | ============================================================================================================================================================================== | |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for D&J medical. The logo consists of a teal and light green graphic on the left, followed by the text "D&J medical" in teal and light green. Below the text are four Chinese characters in gray.
market Notification Submission_ Metallic Locking Bone Plate and Screw System
Product Code: HRS and HWC
lll. PREDICATE DEVICE
Locking Bone Plates and Screws K130340;
Spacer K123918;
These predicates have not been subject to a design-related recall.
No reference devices were used in this submission.
#### IV. DEVICE DESCRIPTION
The proposed device, Metallic Locking Bone Plate and Screw System consists of bone plate and screw. The bone plates are used for internal fixation of bone fracture, including ulna, radius, humerus, femur and tibia. Implant is made of Ti6Al4V Eli and titanium which conforms to ASTM F 136 and ASTM F67.
The proposed devices are provided un-sterilized. They shall be sterilized prior to use by healthcare provider. The proposed devices shall never be reused.
#### V. INDICATIONS FOR USE
The Metallic Locking Bone Plate and Screw System are intended for adult patients with age above 21 as indicated for fixation of fractures, including ulna, radius, humerus, femur and tibia.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Metallic Locking Bone Plate and Screw System employs the same technology as its predicate devices K130340 and K123918.
The Metallic Locking Bone Plate and Screw System is a structure formed by locking plate (which has at least one taper-thread hole) and locking screw (which has a head of taper-thread). Wherein the metal locking plate is a fracture fixation device with threaded holes, When the screw is revolved tightened in the plate, the plate system becomes an angle-fixed device. Locking plate can have both locking and nonlocking screw holes fitting with various types of screws (also known as locking compression plate). In practice, positioning a plurality of screws in the locking plate maintains fracture anatomical reduction, which restores physiological function after bone healing.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains a logo for D&J medical. The logo consists of a teal and light green circular design on the left, followed by the text "D&J medical" in teal and light green. Below the text, there are three Chinese characters in gray. The logo appears to be for a medical practice or company.
| Specification | Predicate Device | Proposed Device |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| | Locking Bone Plates and Screws K<br>130340 | Metallic locking bone plate and<br>screw system |
| Manufacturer | Weigao Orthopaedic Device Co.,<br>Ltd. | Changzhou Dingjian Medical<br>Appliance Co., Ltd. |
| Class | II | II |
| Product Code | Plate: HRS; Screw: HWC | Plate: HRS; Screw: HWC |
| Regulation Number | Plate: 21CFR 888.3030 | Plate: 21CFR 888.3030 |
| | Screw: 21CFR 888.3040 | Screw: 21CFR 888.3040 |
| Intended Use | Locking bone plates and screws are<br>intended for adult patients with age<br>above 21 as indicated for fixation of<br>fracture. | Metallic locking bone plate and<br>screw systemis intended for adult<br>patients with age above 21 as<br>indicated for fixation of fracture. |
| Indications for Use | Bone fracture in ulna, radius,<br>humerus, femur and tibia | Bone fracture in ulna, radius,<br>humerus, femur and tibia |
| Patient Population | Adult patients of age above 21 | Adult patients of age above 21 |
| Prescription/OTC Use | Prescription use | Prescription use |
| Mechanical<br>Performance | FOR PLATES: ASTM F 382-99<br>FOR SCREWS: ASTM F543-07 | FOR PLATES: ASTM F 382-99(2008)<br>FOR SCREWS: ASTM F543-07 |
| Materials | PLATES: titanium which conforms<br>to ASTM F67<br>Screw: Titanium alloy (TiAl4V ELI)<br>which conforms to ASTM F136 | PLATES: titanium which conforms<br>to ASTM F67<br>Screw: Titanium alloy (TiAl4V ELI)<br>which conforms to ASTM F136 |
| Biocompatibility | PLATES: titanium which conforms<br>to ASTM F67<br>Screw: Titanium alloy (TiAl4V ELI)<br>which conforms to ASTM F136 | PLATES: titanium which conforms<br>to ASTM F67<br>Screw: Titanium alloy (TiAl4V ELI)<br>which conforms to ASTM F136 |
| Sterility | Provided as non-sterile, needs<br>autoclave prior to use | Provided as non-sterile, needs<br>steam sterilization prior to use |
| Test items for bone<br>plates | Static four point bending<br>Dynamic four point bending | Static four point bending<br>Dynamic four point bending |
| Test items for bone<br>screws | Torsional properties<br>Driving torque<br>Pull-out test | Torsional properties<br>Driving torque<br>Pull-out test |
510(k) Premarket Notification Submission_ Metallic Locking Bone Plate and Screw System
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for D&J medical. The logo consists of a stylized teal and light green graphic to the left, resembling three people holding hands in a circle. To the right of the graphic is the text "D&J medical" in teal and light green. Below the text are four gray Chinese characters.
emarket Notification Submission_ Metallic Locking Bone Plate and Screw System
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
# No clinical performance testing:
Bench tests were conducted to verify that proposed device meets all design specifications as was substantially equivalence to the predicate device. The following bench testing was performed to support a determination of substantial equivalence:
ASTM F382 standard specification and test method for metallic bone plates, including the following items:
- . Static four point bending
- . Dynamic four point bending
ASTM F543 standard specification and test methods for metallic medical bone screw including the following item:
- Torsional properties
- Driving torque
- Pullout test
# Animal and clinical study
The subject of this premarket submission, metallic locking bone plate system, does not require clinical studies to support substantial equivalence.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for D&J medical. The logo consists of a stylized graphic of three concentric arcs with circles at the ends, colored in shades of teal and light green. To the right of the graphic, the text "D&J medical" is displayed, with "D&J" in teal and "medical" in light green. Below the text, there are three Chinese characters in gray.
510(k) Premarket Notification Submission_ Metallic Locking Bone Plate and Screw System
#### VIII. CONCLUSIONS
The non-clinical data support the safety of the device and the performance testing report demonstrate that the metallic locking bone plate system should perform as intended in the specified use conditions. Changzhou Dingjian Medical Appliance Co., Ltd., Inc considers the metallic locking bone plate system does not raise any new issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.