MOTIONLOC SCREW FOR NCB POLYAXIAL LOCKING PLATE SYSTEM

K123918 · Zimmer, Inc. · HRS · Feb 14, 2013 · Orthopedic

Device Facts

Record IDK123918
Device NameMOTIONLOC SCREW FOR NCB POLYAXIAL LOCKING PLATE SYSTEM
ApplicantZimmer, Inc.
Product CodeHRS · Orthopedic
Decision DateFeb 14, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

MotionLoc Screws, when used with the NCB Proximal Humeral or NCB Proximal Tibial plates, are indicated for temporary internal fixation and stabilization of fractures and osteotomies of long bones, including: - Comminuted fractures - Fractures in osteopenic bone - Nonunions/Malunions MotionLoc Screws, when used with the NCB Distal Femoral plates, are indicated for temporary internal fixation and stabilization of fractures and osteotomies of long bones, including: - Comminuted fractures - Supracondylar fractures - Fractures in osteopenic bone - Nonunions/Malunions MotionLoc Screws, when used with the NCB Straight Narrow Shaft plates, are indicated for temporary internal fixation and stabilization of humeral and tibial shaft fractures and osteotomies, including: - Periprosthetic fractures - Comminuted fractures - Fractures in osteopenic bone - Nonunions/Malunions MotionLoc Screws, when used with the NCB Periprosthetic Proximal Femur, Distal Femur or Curved Femur Shaft plates, are indicated for temporary internal fixation and stabilization of fractures and osteotomies of long bones, including: - Periprosthetic fractures - Comminuted fractures - Supracondylar fractures - Fractures in osteopenic bone - Nonunions/Malunions

Device Story

MotionLoc Screw is a titanium alloy bone screw used with NCB Polyaxial Locking Plate System for internal fixation. Features standard NCB front thread, reduced-diameter mid-section, collar, and locking head. Provides unicortical fixation in far cortex; locked into plate without rigid fixation in near cortex. Enables polyaxial placement (30° cone) via locking caps. Non-contact bridging concept reduces periosteal blood supply impairment; spacers (1, 2, or 3mm) control plate-to-bone distance. Used by surgeons in clinical settings for fracture stabilization. Reduces required shaft screws from four to three. Benefits include reduced construct stiffness and suitability for osteopenic bone.

Clinical Evidence

No clinical data provided. Safety and effectiveness supported by non-clinical bench testing (axial fatigue, torsional fatigue) and literature references regarding use in osteopenic bone. Biocompatibility testing performed per ISO 10993-1.

Technological Characteristics

Titanium alloy bone screw; polyaxial locking mechanism (30° cone); unicortical fixation; non-contact bridging design. Sterilized via gamma radiation (min 20kGy). Shelf life 10 years. Biocompatibility per ISO 10993-1.

Indications for Use

Indicated for temporary internal fixation and stabilization of long bone fractures and osteotomies, including comminuted, supracondylar, periprosthetic, nonunion/malunion fractures, and fractures in osteopenic bone.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for Zimmer Biomet. The logo consists of a stylized letter "Z" enclosed in a circle, positioned above the word "zimmer" in a lowercase, sans-serif font. The "Z" is black, and the word "zimmer" is also black. P.O. Box 708 Warsaw, IN 46581-0708 574 267-6131 K123918 510(k) Summary page 1 of 4 **FEB 14 2013** Sponsor: Zimmer, Inc. P.O. Box 708 Warsaw, IN 46581-0708 Contact Person: Stephen H. McKelvey Senior Project Manager, Trauma Regulatory Affairs Telephone: (574) 372-4944 Fax: (574) 371-8760 Date: Trade Name: Zimmer . MotionLoc™ Screw for NCB® Polyaxial Locking Plate System Common Name: Classification Names and References: Classification Panel: Predicate Devices: Purpose and Device Description: Zimmer . MotionLoc Screw for NCB Polyaxial Locking Plate System Bone Screw December 18, 2012 Plate, Fixation, Bone (21 CFR 888.3030, Product Code HRS) Orthopedics/87 The predicate devices are: - Zimmer MotionLoc Screw For NCB Polyaxial Locking Plate . System, K101696, cleared September 10, 2010 - . Zimmer NCB Periprosthetic Femur Polyaxial Locking Plate System (K100111, cleared April 12, 2010) - Zimmer NCB Straight Narrow Shaft Plates (K113718, cleared . January 27, 2012) - Synthes 4.5mm Locking Compression Plate (LCP) with . Extended Indications (K082807, cleared January 13, 2009) This submission extends the indications of the existing MotionLoc Screw for NCB Polyaxial Locking Plate System by adding to the list of compatible plates to include the NCB Periprosthetic Proximal Femur, Distal Femur, Curved Femur Shaft and Straight Narrow Shaft plates and by allowing the MotionLoc screw to be used for fractures in osteopenic bone. In addition, this submission decreases the number of required MotionLoc screws in the shaft of the bone from four to three. {1}------------------------------------------------ The MotionLoc Screw is a member of the NCB Screw family and is used as an alternative for standard NCB Screws in applications where a surgeon desires reduced stiffness in a construct. The MotionLoc Screw has a standard NCB Screw front thread section, a mid-section with a reduced core-diameter, a collar region, and a standard NCB Screw head for engagement in NCB locking plates. The MotionLoc Screws provide unicortical fixation in the far cortex of a diaphysis and are locked into the plate, without being rigidly fixed in the near cortex underlying the plate. The NCB technology allows for polyaxial screw placement (30° cone) of the MotionLoc Screws with screw locking achieved using previously cleared Locking Caps (K042695, cleared 10/29/2004) that are threaded into the plate holes. In the locked mode the NCB plate acts as an internal fixator without contact between the plate and the bone surface thus reducing the risk of periosteal blood supply impairment. This Non-Contact Bridging concept can be specifically controlled through the use of 1, 2, or 3mm spacers (also previously cleared in K042695), which are threaded into the plate holes prior to plate insertion. Plates, screws, spacers and locking caps are made of titanium alloy. Intended Use: MotionLoc Screws, when used with the NCB Proximal Humeral or NCB Proximal Tibial plates, are indicated for temporary internal fixation and stabilization of fractures and osteotomies of long bones, including: - · Comminuted fractures - · Fractures in osteopenic bone - · Nonunions/Malunions MotionLoc Screws, when used with the NCB Distal Femoral plates, are indicated for temporary internal fixation and stabilization of fractures and osteotomies of long bones, including: - · Comminuted fractures - · Supracondylar fractures - · Fractures in osteopenic bone - · Nonunions/Malunions MotionLoc Screws, when used with the NCB Straight Narrow Shaft plates, are indicated for temporary internal fixation and stabilization of humeral and tibial shaft fractures and osteotomies, including: {2}------------------------------------------------ - · Periprosthetic fractures - · Comminuted fractures - · Fractures in osteopenic bone - · Nonunions/Malunions MotionLoc Screws, when used with the NCB Periprosthetic Proximal Femur, Distal Femur or Curved Femur Shaft plates, are indicated for temporary internal fixation and stabilization of fractures and osteotomies of long bones, including: - · Periprosthetic fractures - · Comminuted fractures - · Supracondylar fractures - · Fractures in osteopenic bone - Nonunions/M - alunions . The physical attributes (dimensions, materials, specifications manufacturing processes, etc.) have not changed from the predicate MotionLoc Screws. They still have a standard NCB Screw front thread section, a mid-section with a reduced corediameter, a collar region, and a standard NCB Screw head for engagement in NCB locking plates. As before, these MotionLoc Screws provide unicortical fixation in the far cortex of a diaphysis and are locked into the plate, without being rigidly fixed in the near cortex underlying the plate. The NCB technology allows for polyaxial screw placement (30° cone) of the MotionLoc Screws with screw locking achieved using previously cleared Locking Caps (K042695, cleared 10/29/2004) that are threaded into the plate holes. In the locked mode the NCB plate acts as an internal fixator without contact between the plate and the bone surface thus reducing the risk of periosteal blood supply impairment. This Non-Contact Bridging concept can be specifically controlled through the use of 1, 2, or 3mm spacers (also previously cleared in K042695), which are threaded into the plate holes prior to plate insertion. Plates, screws, spacers and locking caps are made of titanium alloy. The proposed device is similar in intended use to the predicate devices. Comparison to Predicate Devices: {3}------------------------------------------------ Performance Data (Nonclinical and/or Clinical): ## Non-Clinical Performance and Conclusions: Sterilization Validation - Sterilization has not changed with the new indications or reducing the number of required MotionLoc screws in the shaft of the bone from four to three. These devices have been validated to demonstrate that at a minimum gamma dose of 20kGy these screws can be terminally sterilized to a SAL of 10° or better. Shelf Life - Shelf life has not changed with the new indications or reducing the number of required MotionLoc screws in the shaft of the bone from four to three. Accelerated aging showed that these sterile screws have a shelf life of 10 years. Sterile Packaging - Packaging has not changed. The sterile screws can withstand normal distribution and storage conditions and maintain the sterile barrier properties throughout the specified product shelf life. Biocompatibility - Biocompatibility testing on the screw implant materials was conducted per ISO 10993-1 and Good Laboratory Practices (21 CFR § 58). All testing passed. The new indications or reducing the number of required MotionLoc screws in the shaft of the bone from four to three does not change the fundamental scientific technology of any of the devices. Each sterile device uses the same operating principle and incorporates the same basic labeling. Testing/Literature - The results of non-clinical (lab) performance testing demonstrate that the devices are safe and effective and substantially equivalent to the predicate devices when used with the newly indicated plate and for use with three screws. Testing/analysis performed included axial fatigue strength and torsional fatigue. In addition, data from the included literature references indicates that the devices are safe and effective for the newly indicated bone quality (osteopenic). ### Clinical Performance and Conclusions: Clinical data and conclusions were not needed for this device. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol that resembles three abstract human figures. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 14. 2013 Zimmer. Incorporated % Stephen H. McKelvey, MA, RAC Senior Project Manager, Trauma Regulatory Affairs P.O. Box 708 Warsaw, Indiana 46581-0708 Re: K123918 Trade/Device Name: Zimmer® MotionLoc™ Screw for NCB® Polyaxial Locking Plate System Regulation Number: 21 CFR 888.3030 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HRS Dated: December 18, 2012 Received: December 19, 2012 Dear Mr. McKelvey: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {5}------------------------------------------------ ## Page 2 - Stephen H. McKelvey, MA, RAC CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Image /page/5/Picture/7 description: The image shows the name "Erin Keith" in a stylized font. The word "Erin" is in a bold, sans-serif font, while "Keith" is in a more decorative font with geometric patterns filling the letters. The overall design is simple but eye-catching due to the contrast in font styles. Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # 510(k) Number (if known): K123918 ### Device Name: Zimmer® MotionLoc™ Screw For NCB® Polyaxial Locking Plate System #### Indications for Use: MotionLoc Screws, when used with the NCB Proximal Humeral or NCB Proximal Tibial plates, are indicated for temporary internal fixation and stabilization of fractures and osteotomies of long bones, including: - · Comminuted fractures - · Fractures in osteopenic bone - · Nonunions/Malunions MotionLoc Screws, when used with the NCB Distal Femoral plates, are indicated for temporary internal fixation and stabilization of fractures and osteotomies of long bones, including: - · Comminuted fractures - · Supracondylar fractures - · Fractures in osteopenic bone - · Nonunions/Malunions MotionLoc Screws, when used with the NCB Straight Narrow Shaft plates, are indicated for temporary internal fixation and stabilization of humeral and tibial shaft fractures and osteotomies, including: - · Periprosthetic fractures - · Comminuted fractures - · Fractures in osteopenic bone - · Nonunions/Malunions MotionLoc Screws, when used with the NCB Periprosthetic Proximal Femur, Distal Femur or Curved Femur Shaft plates, are indicated for temporary internal fixation and stabilization of fractures and osteotomies of long bones, including: - · Periprosthetic fractures - · Comminuted fractures - · Supracondylar fractures - · Fractures in osteopenic bone - Nonunions/Malunions Prescription Use X (Part 21 CFR 801 Subpart D) . AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (Please do not write below this line - Continue on another page if needed) Concurrence of CDRH, Office of Device Evaluation (ODE) Image /page/6/Picture/30 description: The image shows the name "Michael C. Owens" in a stylized font. The first name, "Michael," is written in a simple, bold font. The initials "C.O." are stylized with a concentric line pattern, and the last name, "Owens," is written in a simple, bold font. - Division of Orthopedic Devices
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...