Locking Bone Plates and Screws are intended for adult patients with age above 21 as indicated for fixation of fractures, including ulna, radius, humerus, femur and tibia.
Device Story
System consists of metallic locking bone plates, locking screws, and associated surgical instruments. Plates and screws provide internal fixation for bone fractures. Screws are available in self-tapping and self-drilling configurations. Devices are provided non-sterile; intended for sterilization via hospital autoclave prior to use. Used by surgeons in clinical/surgical settings to stabilize bone fractures, facilitating healing through rigid fixation.
Clinical Evidence
Bench testing only. Compliance with ASTM F382-99 (bone plates) and ASTM F543-07 (bone screws) demonstrated mechanical performance equivalent to predicates. Sterilization validation performed per ANSI/AAMI/ISO 17665-1.
Technological Characteristics
Materials: Titanium (ASTM F67-06) for plates; Ti6Al4V ELI (ASTM F136) for screws. Mechanical principle: Locking fixation. Sterilization: Moist heat (autoclave) to SAL 10^-6. Plate specs: 2-24 holes, 26-312mm length. Screw diameters: 3.5, 4.5, 5.0mm.
Indications for Use
Indicated for adult patients >21 years old requiring internal fixation of fractures of the ulna, radius, humerus, femur, and tibia.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Section III 510(k) Summary Premarket Notification 510(k) Submission
Project #:M0042013Aa
#### Section Ill 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.
K130340 The assigned 510(k) Number:
- Date of Submission: 01/27/2013 1.
- 2. Sponsor
Weigao Orthopaedic Device Co., Ltd. No 26 Xiangjiang Road, Tourist Resorts, Weihai, Shandong, 264203, China
Establishment Registration Number: 3006639944
Contact Person: Han Wang Position: Quality & Technique Manager Tel: +86-631-5788966 Fax: +86-631-5660958 Email: wanghan@wegortho.com
#### 3, Submission Correspondent
Ms. Diana Hong & Mr. Lee Fu Mid-Link Consulting Co., Ltd P.O. Box 237-023. Shanghai, 200237, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net
4. Proposed Device Identification
Proposed Device Name: Locking Bone Plates and Screws
Classification Name: Plate, Fixation, Bone Common Name: Bone Plates Class: Class II Product Code: HRS Regulation Number: 21 CFR 888.3030
{1}------------------------------------------------
## K130340 page 2 of 4
#### Premarket Notification 510(k) Submission
Section III 510(k) Summary
Project #:M0042013Aa
Review Panel: Orthopedic
Classification Name: Screw. Fixation. Bone Common Name: Bone Screws Class: Class II Product Code: HWC Regulation Number: 21 CFR 888.3040 Review Panel: Orthopedic
Intended Use Statement:
Locking Bone Plates and Screws are intended for adult patients with age above 21 as indicated for . fixation of fractures, including ulna, radius, humerus, femur and tibia.
- ను Predicate Device Identification
510(k) Number: K101400 Product Name: Locking Compression Plate Manufacturer: Changzhou Orthmed Medical Instrument Co., Ltd
510(k) Number: K100721 Product Name: Locking Bone Screw Manufacturer: Changzhou Orthmed Medical Instrument Co., Ltd
#### 6. Device Description
The proposed products, Locking Bone Plates and Screws, contain (1) locking plats with various specifications, (2) locking screws with various specifications and (3) Various specific instruments.
The bone plates are used for internal fixation of bones, screws are used for fix the plates on the bones, and instruments are used for completing the surgery.
Locking Screws are available in two kinds, which are self-dapping and self-drilling. Both of them share the same dimensions and materials.
These devices are provided un-sterilized, but shall be sterilized via autoclave method to achieve Sterility Assurance Level of 10th by hospital prior to use.
### 7 Non-Clinical Test Conclusion
Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
III-2
. •
{2}------------------------------------------------
ASTM F382-99 (Reapproved 2008), Standard Specification and Test Method for Metallic Bone Plates.
ASTM F543-07, Standard Specification and Test Methods for Metallic Medical Bone Screws. ANSI/AAMI/ISO17665-1: 2006 Sterilization of health care products -- Moist heat -- Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices.
- Substantially Equivalent Comparison and Conclusion જે.
| ITEM | Proposed Device<br>Locking Bone Plates and Screws | Predicate Device |
|----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| | | Locking Compression Plate<br>K101400<br>Locking Bone Screw K100721 |
| Product Code | Plate: HRS<br>Screw: HWC | Same |
| Regulation No. | Plate: 21 CFR 888.3030<br>Screw: 21 CFR 888.3040 | Same |
| Class | Class II | Same |
| Regulation No. | Plate: 21 CFR 888.3030 | Same |
| Intended Use | Locking Bone Plates and Screws are<br>intended for adult patients with age above<br>21 as indicated for fixation of fractures,<br>including ulna, radius, humerus, femur and<br>tibia. | Similar |
| Sterilization | Method: Autoclave<br>SAL: 10⁻⁶ | Same |
- Table III-1 General and Safety Comparison
{3}------------------------------------------------
### Premarket Notification 510(k) Submission
Section III 510(k) Summary
| ITEM | Proposed Device<br>Locking Bone Plates and Screws | Predicate Device<br>Locking Compression Plate<br>K101400<br>Locking Bone Screw K100721 |
|-----------------------------------------------|---------------------------------------------------------|----------------------------------------------------------------------------------------|
| Physical Specifications - Locking Bone Plates | | |
| Number of Holes | 2-24 | Similar |
| Length | 26-312 mm | Similar |
| Thickness | 3.3 / 3.6 / 4.2 mm | Similar |
| Width | 11 / 12.5 / 13.5 mm | Similar |
| Material | Titanium conforms to ASTM F67-06 | Similar |
| Physical Specifications - Locking Bone Screws | | |
| Diameters | 3.5. 4.5, 5.0 | Similar |
| Lengths | 10~120mm | Same |
| Material | Ti6Al4V ELI conforms to ASTM<br>F136 | Same |
| Mechanical Performance | | |
| Test items for bone<br>plates | Static four point bending<br>Dynamic four point bending | Same |
| Test standard for bone<br>plates | ASTM F 382-99 | Same |
| Test items for bone<br>screws | Torsional properties<br>Driving torque<br>Pull-out test | Same |
| Test standard for bone<br>screws | ASTM F543-07 | Same |
Table III-2 Specifications and Performance Comparison
. Difference in intended use, and physical specifications between the proposed and predicate device are discussed in the 510(k) submission documents, it is concluded that these differences will not affect the effectiveness and safety of the proposed device.
The proposed device. Locking Bone Plates and Screws, is determined to be Substantially Equivalent (SE) to the predicate device, Locking Compression Plate (K101400), and Locking Bone Screw (K100721), in respect of safety and effectiveness.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Weigao Orthopaedic Device Co., Ltd. % Mid-Link Consulting Co., Ltd Ms. Diana Hong General Manager P.O. Box 237-023, 200237 Shanghai China
Letter dated: April 5, 2013
Re: K130340 Trade/Device Name: Locking Bone Plates and Screws Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: February 7. 2013 Received: February 19, 2013
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
### Page 2 - Ms. Diana Hong
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Eyaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Section II Indications for Use
510(k) Number: K130340 Device Name: Locking Bone Plates and Screws
Indications for Use:
Locking Bone Plates and Screws are intended for adult patients with age above 21 as indicated for fixation of fractures, including ulna, radius, humerus, femur and tibia.
ZPRESCRIPTION USE (Part 21 CFR 801 Subpart D)
OR
OOVER-THE-COUNTER USE (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
# Elizabeth பூர்ลีที่ -S
Division of Orthopedic Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.