K140430 · Cas Medical Systems, Inc. · MWI · Jun 13, 2014 · Cardiovascular
Device Facts
Record ID
K140430
Device Name
740 SELECT
Applicant
Cas Medical Systems, Inc.
Product Code
MWI · Cardiovascular
Decision Date
Jun 13, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The 740 SELECT series of monitors is indicated for use as a portable, multi-parameter, variable acuity device for use by health care professionals, clinicians and medically qualified personnel, in a variety of clinical environments and hospital departments, for non-invasive spot checking, and/or continuous monitoring and recording of: . Blood pressure and pulse rate of adult, pediatric and neonatal patients; . Functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult, pediatric and neonatal patients; . Additionally the Masimo rainbow® Pulse CO-Oximetry™ and Masimo SET technology provides noninvasive monitoring of carboxyhemoglobin saturation (SpCO®) and/or respiration (RRa), Other information displayed on the 740 SELECT with the Masimo Rainbow SET option includes: Signal IQ Waveform, Low Signal IQ (Low SIQ), Perfusion Index (PI), and/or Pleth Variability Index (PVI®) indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused: . Carbon Dioxide concentration of the expired and inspired breath and respiration rate and FICO2 (fractional concentration of inspired CO2) of adult, pediatric, infant or neonatal patient and included, with the Oridion option, an Integrated Pulmonary Index (IPI); . Intermittent predictive body temperature (oral, axillary, rectal) of adult, pediatric and neonatal patients: . Infrared (over the temporal artery) measurement of body temperature of people of all ages.
Device Story
740 SELECT is a portable, multi-parameter monitor for spot-checking or continuous patient monitoring. Inputs include NIBP (oscillometric cuff), SpO2/SpCO/RRa (optical sensors), CO2 (sidestream/mainstream gas analyzers), and temperature (predictive probe or infrared scanner). Device processes signals to display vital signs, waveforms, and indices (e.g., PVI, IPI, Signal IQ). Used by clinicians in hospital departments; output displayed on front panel LED/digital display. Provides real-time physiological data to assist clinical decision-making; enables monitoring of well/poorly perfused patients under motion/no-motion conditions. Benefits include rapid, non-invasive assessment of patient status across diverse acuity levels.
Clinical Evidence
Clinical validation performed for NIBP, pulse oximetry, CO2, and temperature functions. Evidence includes OEM clinical validation reports for integrated modules and internal system verification/validation studies. No specific clinical trial metrics (e.g., sensitivity/specificity) provided in summary.
Technological Characteristics
Portable multi-parameter monitor. Parameters: NIBP (oscillometric), SpO2/SpCO/RRa (optical), CO2 (infrared gas analysis), Temperature (predictive/infrared). Connectivity: External attachment for CO2 modules. Display: LED digital display. Software: Microprocessor-based control for predictive temperature and NIBP algorithms. No specific materials or sterilization standards cited.
Indications for Use
Indicated for adult, pediatric, and neonatal patients (all ages for temporal temperature) requiring non-invasive spot checking or continuous monitoring of blood pressure, pulse rate, SpO2, SpCO, respiration rate, CO2, and body temperature in clinical/hospital settings by healthcare professionals.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the word "CASMED" in bold, black letters. The letters are large and fill most of the frame. A small registered trademark symbol is located to the right of the letter "D".
JUN 1 3 2014
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
| Submitter: | CAS Medical Systems, Inc. |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Address: | 44 East Industrial Rd. Branford CT. 06405 USA |
| Contact: | Ron Jeffrey - Director, Regulatory Affairs<br>Phone - (203) 488-6056<br>Fax - (203) 488-9438<br>Email - rjeffrey@casmed.com |
| Prepared: | February 14, 2014 |
| Trade Name: | 740 SELECT |
| Common Name: | Monitor, Physiological, Patient |
| Classification Name: | Cardiovascular Monitoring Device 870.2300 (MWI) |
Substantially Equivalent Device(s):
The 740 SELECT is equivalent to the following devices:
- CASMED 740 SELECT (Phase 1) Monitor (K130411);
- Masimo Corporation Radical 7 Pulse Co-Oximeter (K120657);
- Covidien (Nellcor) Bedside Respiratory Patient Monitoring System (K 130320); .
- Exergen Corporation TemporalScanner Thermometer (K011291; ●
- PHASEIN AB (Masimo) Infrared Sidestream Gas Analyzer (ISA) (K103604); .
- PHASEIN AB (Masimo) Infrared Mainstream Gas Analyzer (IRMA) (K 123403);
- Oridion (Covidien) Capnostream20 (K094012)
រ
#### DESCRIPTION
The 740 SELECT is a rugged portable non-invasive multi-parameter device used for spot checking or continuous monitoring of blood pressure, pulse rate, functional oxygen saturation (%Sp02), body temperature and CO2 in various modes of operation.
The non-invasive blood pressure (NIBP) parameter automatically inflates an occluding cuff and, using the oscillometric step-deflation measurement technique, determines systolic, diastolic and mean arterial pressure as well as pulse rate. Measurement results along with user prompts and error messages are displayed on the front panel. The frequency of NIBP determination can be selected by the user in
{1}------------------------------------------------
varied times between one and ninety minutes. The auto and manual operating modes cover a variety of clinical uses.
The pulse oximeter parameter (%SpO2) determines arterial oxyhemoglobin saturation by measuring the absorption of red and infrared light passing through the tissue. Changes in absorption caused by pulsations of blood in the vascular bed are used to determine arterial saturation and pulse rate. The oximeter requires no routine calibration or maintenance. Oxygen saturation and heart rate are displayed on a light emitting diode (LED) digital display. On each detected pulse, the perfusion LED indicates patient perfusion signals. This bar graph gives the user a pulse-by-pulse visual indication of waveform signal quality. An audio "beep" can be enabled that is generated each time the SpO2 module detects a pulse. In addition to %SpO2 the SELECT (equipped with the Masimo rainbow® Pulse CO-Oximetry™ and Masimo SET ® oximetry) can monitor Pulse Rate (PR), Carboxyhemoglobin (SpC0®), Acoustic Respiration Rate (RRa) and Pleth Variability Index (PVI®). Alternately, the customer can choose a Nellcor Pulse Oximeter at this time of ordering.
The predictive temperature parameter has the capability of taking temperature in either normal (predictive) or monitor mode. In the normal mode, the thermometer's microprocessor "predicts" body temperature in about four (4) seconds for oral temperatures, about ten (10) seconds for axillary temperatures and about fifteen (15) seconds for rectal temperatures. The default setting used by the monitor for Temperature determinations is the Predictive (normal) mode. Alternately, the customer can choose an Exergen TemporalScanner (infrared) Thermometer at the time of ordering. The Exergen TemporalScanner Thermometer has the capability of taking Temporal Artery (TA) Temperature in about 3 seconds. The Exergen Temp scanner has the ability to measure the patient's Arterial temperature in °F or °C.
The 740 SELECT has (2) optional choices for the parameter of CO2 monitoring. The customer may order either the Masimo - Phasein™ Infrared Sidestream Gas Analyzer (ISA™) or the Infrared Mainstream Gas Analyzer (IRMATM). Alternately, the customer can select the option of the Oridion Sidestream MicroPod™ EtCO2. Each of these CO2 options is externally attached to the SELECT.
### The 740 SELECT Intended Use
う
The 740 SELECT series of monitors is indicated for use as a portable, multi-parameter, variable acuity device for use by health care professionals, clinicians and medically qualified personnel, in a variety of clinical environments and hospital departments, for non-invasive spot checking, and/or continuous monitoring and recording of:
- . Blood pressure and pulse rate of adult, pediatric and neonatal patients;
- . Functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult, pediatric and neonatal patients;
- . Additionally the Masimo rainbow® Pulse CO-Oximetry™ and Masimo SET technology provides noninvasive monitoring of carboxyhemoglobin saturation (SpCO®) and/or respiration (RRa), Other information displayed on the 740 SELECT with the Masimo Rainbow SET option includes: Signal IQ Waveform, Low Signal IQ (Low SIQ), Perfusion Index (PI), and/or Pleth Variability Index (PVI®) indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused:
- . Carbon Dioxide concentration of the expired and inspired breath and respiration rate and FICO2 (fractional concentration of inspired CO2) of adult, pediatric, infant or neonatal patient and included, with the Oridion option, an Integrated Pulmonary Index (IPI);
{2}------------------------------------------------
- . Intermittent predictive body temperature (oral, axillary, rectal) of adult, pediatric and neonatal patients:
- . Infrared (over the temporal artery) measurement of body temperature of people of all ages.
#### 740 SELECT Monitor Technologies Compared to Predicate Devices
The 740 SELECT compares substantially to the cited predicate device(s) in that it uses fundamentally the same technologies, adopted from the parameters manufacturer, without alteration. The non-invasive blood pressure component is the CASMED MAXIQ™ an oscillometric instrument having neither a stethoscope nor a microphone and does not need the quiet environment necessary to detect the auscultatory sounds. They work on the principle that when the artery opens during a portion of the pressure cycle, an oscillation is superimposed on the pressure inside the cuff due to a tiny enlargement of the circumference of the limb caused by the surge of blood under the cuff. The amplitude of the oscillometric signals change over the course of the deflation of the cuff. Oscillometric devices look for oscillation amplitude of certain percentages of the maximum amplitude at mean arterial pressure (MAP), defining one percentage as the systolic point and another as the diastolic point. Alternatively, a combination of the amplitude, the slope of the increase or decrease, and some other complex factors are used to find these points.
The 740 SELECT pulse oximeter can include either of two industry leading components specific to the customer's requirement. Masimo rainbow Pulse CO-Oximetry™ and Masimo SET® or Covidien (Nellcor's) OxiMax technology. The pulse oximeter parameter (%SpO2) determines arterial oxyhemoglobin saturation by measuring the absorption of red and infrared light passing through the tissue. Changes in absorption caused by pulsations of blood in the vascular bed are used to determine arterial saturation and pulse rate. Additional leading edge functions, as offered by the pulse oximeters manufacturer, are made available to the customer with the 740 SELECT.
The FILAC™ 3000 temperature parameter has the capability of taking temperature in either normal (predictive) or monitor mode. In the normal mode, the thermometer's microprocessor "predicts" body temperature in about four (4) seconds for Oral temperatures, about ten (10) seconds for Axillary temperatures and in about fifteen (15) seconds for Rectal temperatures. Exergen's TemporalScanner is a hand held device that measures the skin temperature over the temporal artery. The Exergen TemporalScanner Thermometer has the capability of taking Temporal Artery (TA) Temperature in about 3 seconds. The Exergen Temp scanner has the ability to measure the patient's Arterial temperature in ºF or °C.
For the CO2 parameter, the 740 SELECT allows for the external attachment of a choice of two manufacturers solutions. Offered are the Masimo -Phaseln™ (Masimo) Sidestream Gas Analyzer (ISA) or the Mainstream Gas Analyzer (IRMA). The other choice is the Oridion Micro Pod™ capnography module.
#### Non-Clinical Performance Testing to Demonstrate Substantial Equivalence
The 740 SELECT has successfully undergone performance, safety, electromagnetic, and environmental testing to ensure it has been found to be substantially equivalent to the primary and secondary predicate devices. In addition to the above laboratory tests, CAS has conducted a full program of individual hardware, software and systems verification and validation studies of the monitor and accessories.
{3}------------------------------------------------
### Clinical Testing to Show Substantial Equivalence
The 740 SELECT has successfully undergone clinical validation for the indicated use. The premarket notification cites clinical validation reports for the NIBP, and the OEM's clinical work for the pulse oximeters, CO2 and the temperature functions.
### Conclusions Drawn from Clinical and Non-Clinical Testing
Clinical evaluation, safety testing, software and systems validation demonstrate the 740 SELECT is substantially equivalent to the predicate device(s).
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake winding around it, overlaid on a stylized human figure. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 13, 2014
CAS Medical Systems, Inc. Ron Jeffrey Director, Regulatory Affairs 44 East Industrial Rd. Branford, Connecticut 06405
K140430 Re:
> Trade/Device Name: 740 Select Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (With Cardiotachometer and Rate Alarm) Regulatory Class: Class II Product Code: MWI Dated: May 5, 2014 Received: May 6, 2014
Dear Mr. Jeffrey,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{5}------------------------------------------------
Page 2 - Ron Jeffrey
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Mitchell Stein
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health ... ..
Enclosure
{6}------------------------------------------------
#### K140430
## Indications for Use
510(k) Number:
#### Device Name: 740 SELECT
The 740 SELECT series of monitors is indicated for use as a portable, multi-parameter, variable aculty device for use by health care professionals, clinicians and medically qualified personnel, in a variety of clinical environments and hospital departments, for non-invasive spot checking, and/or continuous monitoring and recording of:
- Blood pressure and pulse rate of adult, pediatric and neonatal patients; .
- Functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult, pediatric and . neonatal patients;
- Additionally the Masimo Rainbow SET technology provides noninvasive monitoring of . carboxyhemoglobin saturation (SpCO) and/or respiration (RRa). Other information displayed on the 740 SELECT with the Masimo Rainbow SET option includes: Signal IQ Waveform, Low Signal IQ (Low SIQ), Perfusion Index (PI), and/or Pleth Variability Index (PVI) indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused;
- Carbon Dioxide concentration of the expired and inspired breath and respiration rate of adult, . pediatric, infant or neonatal patient and included, with the Oridion option, an Integrated Pulmonary Index (IPI):
- Intermittent predictive body temperature (oral, axillary, rectal) of adult, pediatric and neonatal ● patients;
- Infrared (over the temporal artery) measurement of body temperature of people of all ages. .
| Prescription Use ✔ | AND/OR | Over-the-Counter Use _ |
|-----------------------------|--------|------------------------|
| (Part 21 CFR 801 Subpart D) | | (21 CFR 807 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Date:
2014.06.13
09:46:52 -04'00'
Page 1 of 1
0.41001
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.