K094012 · Oridion Capnography, Inc. · CCK · May 12, 2010 · Anesthesiology
Device Facts
Record ID
K094012
Device Name
CAPNOSTREAM20 WITH MICROMEDICO2 MODULE
Applicant
Oridion Capnography, Inc.
Product Code
CCK · Anesthesiology
Decision Date
May 12, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.1400
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Capnostream20 combined capnograph/pulse oximeter monitor is intended to provide professionally trained health care providers the continuous, non invasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and respiration rate, and for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. It is intended for use with neonatal, pediatric and adult patients in hospitals, hospital type facilities, intra hospital transport and home environments. The Capnostream20 monitor provides the clinician with an integrated pulmonary index (IPI). The IPI is based on four parameters provided by the monitor: end tidal carbon dioxide, respiration rate, oxygen saturation and pulse rate. The IPI is a single index of an adult or pediatric patient's ventilatory status displayed on a scale of 1 - 10, where 10 indicates optimal pulmonary status. IPI monitoring displays a single value that represents the patient's pulmonary parameters and alerts clinicians to changes in the patient's pulmonary status. The IPI is an adjunct to, and is not intended to replace, vital sign monitoring.
Device Story
Bedside monitor integrates CO2 capnography and pulse oximetry modules; processes physiological signals to display real-time numeric values and waveforms. Features Integrated Pulmonary Index (IPI) algorithm; calculates single index (1-10 scale) from end-tidal CO2, respiration rate, SpO2, and pulse rate to represent ventilatory status. Used by physicians, nurses, and trained providers in critical care, anesthesia, and neonatal settings. Provides visual/audible alarms when thresholds are crossed. IPI serves as clinical adjunct to standard vital sign monitoring; assists clinicians in identifying changes in patient pulmonary status. New microMediCO2 module offers enhanced processing/memory, reduced power, and improved flow regulation compared to predicate.
Clinical Evidence
Bench testing only. Verification and validation of the microMediCO2 module as a standalone unit and when integrated into the Capnostream20 monitor were successfully completed. Hazard analysis concluded residual risks are acceptable.
Technological Characteristics
Two-parameter bedside monitor; CO2 analyzer (21 CFR 868.1400) and pulse oximeter (21 CFR 870.2700). Connectivity via RS232 or USB. Standards: ISO 21647, ISO 9919, IEC/EN 60601-1, IEC/EN 60601-1-2, IEC 60601-1-8, UL 60601-1, ISO 14971. Non-sterile. Module features smaller lamp, new solenoid, and molded parts.
Indications for Use
Indicated for neonatal, pediatric, and adult patients requiring continuous, non-invasive monitoring of CO2 concentration (expired/inspired), respiration rate, functional arterial oxygen saturation (SpO2), and pulse rate in hospital, clinical, transport, and home settings.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
{0}------------------------------------------------
SPECIAL-510(k) Submission for CAPNOSTREAM20 two parameter bedside monitor with microMediCO2 CO2 module 510(k) Summary of Safety and Effectiveness 1094012
# 7. 510(k) Summary of Safety and Effectiveness
This summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92
### (This section is not confidential)
MAY 1 2 2010
: "!
### DATE THIS SUMMARY WAS PREPARED
December 15th, 2009
## SUBMITTER'S NAME AND ESTABLISHMENT ADDRESS:
Oridion Capnography Inc. 160 Gould Street Needham, MA 02494
## ESTABLISHMENT REGISTRATION NUMBER
3003941644
#### CONTACT PERSON:
Rachel Weissbrod, Director of Regulatory Affairs Oridion Medical 1987 Ltd. Har Hotzvim Science Park POB 45025 91450 Jerusalem, Israel Telephone: 857-919-2923 Fax: +972-2-586-6680 Email: rachel.weissbrod@oridion.com
## DEVICE INFORMATION
Trade Name: Capnostream20 Common Name: Two Parameter Bedside Monitor Classification Name: Capnograph/Pulse Oximeter Regulation Number: 868.1400, Carbon Dioxide Analyzer (Classification CCK) 870.2700 Pulse Oximeter (Classification DQA) Device Listing Number: D001115.
{1}------------------------------------------------
SPECIAL-510(k) Submission for CAPNOSTREAM20 two parameter bedside monitor with microMediCO2 CO2 module 510(k) Summary of Safety and Effectiveness
## PREDICATE DEVICE
Capnostream20 with the microMediCO2 CO2 board is substantially equivalent to the following commercially available device:
| Manufacturer | Device | 510(k)No. | Clearance Date |
|--------------------------|---------------|-----------|--------------------|
| Oridion 1987 Medical Ltd | Capnostream20 | K082268 | February 9th, 2009 |
## DEVICE DESCRIPTION
The Capnostream20 bedside monitor is a two parameter monitor consisting of a CO2 capnography module and a pulse oximetry module implemented in a host device. The host device displays parameters received from the respective modules and generates alarms when preset alarm thresholds are crossed. The device is classified as CCK Class II according to 21 CFR § 868.1400 - Carbon Dioxide Analyzer with DQA 21 CFR § 870.2700 Pulse Oximeter listed as an additional or alternate classification.
This device has two modules that are classified as follows:
- 21 CFR 868.1400, Carbon Dioxide Analyzer (Classification CCK) .
- 21 CFR 870.2700 Pulse Oximeter (Classification DQA) .
Each module is controlled by dedicated software that is an integral part of the respective module.
## INTENDED USE
The Capnostream20 combined capnograph/pulse oximeter monitor is intended to provide professionally trained health care providers the continuous, non invasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and respiration rate, and for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpQ₂) and pulse rate. It is intended for use with neonatal, pediatric and adult patients in hospital type facilities, intra-hospital transport and home environments.
The Capnostream20 monitor provides the clinician with an integrated pulmonary index (IPI). The IPI is based on four parameters provided by the monitor: end tidal carbon dioxide, respiration rate, oxygen saturation and pulse rate. The IPI is a single index of an adult or pediatric patient's ventilatory status displayed on a scale of 1 - 10, where 10 indicates optimal pulmonary status. IPI monitoring displays a single value that represents the
{2}------------------------------------------------
SPECIAL-510(k) Submission for CAPNOSTREAM20 two parameter bedside monitor with microMediCO2 CO2 module 510(k) Summary of Safety and Effectiveness o ratient's pulmonary parameters and alerts clinicians to changes in the patient's pulmonary status.
The IPI is an adjunct to, and is not intended to replace, vital sign monitoring
## COMPARISON TO PREDICATE DEVICE
The Capnostream20 with the micromediCO2 CO2 module is substantially equivalent to the predicate Capnostream20 with the minimediCO2 CO2 module.
The new device meets the safety and performance standards met by the predicate device.
The functional features and the intended use of the Capnostream20 with the new CO2 module are substantially equivalent to the predicate device. No significant changes were made to the monitor hardware or software specifications to support the integration of the new module. Slight mechanical changes were made to the monitor to accommodate the smaller dimensions of the new module. The micromediCO2 module is smaller in size than the miniMediCO2 module, in order to facilitate integration inside other OEM monitors, and provides enhanced processing and memory capacity, to support software applications for future development. The module provides the following inputs to the host monitor:
FiCO2, EtCO2 numeric, EtCO2 waveform, Respiratory Rate, IPI (Integrated Pulmonary Index), and continuous CO2 numeric and waveform.
The micromediCO2 includes a smaller lamp, a new solenoid, broader use of molded parts to improve production quality, increased processing power, reduced power consumption and improved flow regulation.
The module software provides full backward compatibility with the miniMedicO2 and supports RS232 or USB communication with the host monitor.
A hazard analysis was carried out on the module and on the Capnostream20 monitor with the new module. This hazard analysis concluded that any residual risks were judged as acceptable when weighed against the intended benefits of use of the system. Verification and validation of the new module as a standalone and when integrated in the monitor were successfully completed.
{3}------------------------------------------------
. '
| Attribute | Capnostream20 Bedside<br>Monitor with<br>microMediCO2CO2 module | Predicate Device- Capnostream20<br>Bedside Monitor K082268 |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>use | The indications for use are identical to<br>the indications for use in the predicate<br>device | The Capnostream20 combined<br>capnograph/pulse oximeter monitor is intended<br>to provide professionally trained health care<br>providers the continuous, non invasive<br>measurement and monitoring of carbon<br>dioxide concentration of the expired and<br>inspired breath and respiration rate, and for<br>the continuous non-invasive monitoring of<br>functional oxygen saturation of arterial<br>hemoglobin (SpO2 and pulse rate). It is<br>intended for use with neonatal, pediatric and<br>adult patients in hospitals, hospital type<br>facilities, intra hospital transport and home<br>environments. |
| Target<br>population | It is intended for use with neonatal,<br>pediatric, and adult patients. | It is intended for use with neonatal, pediatric,<br>and adult patients. |
| Design | Identical to the Capnostream20<br>described in K082268 with the<br>exception of the introduction of the<br>micromediCO2 CO2 module | See K082268 |
| Where Used | It is to be used by physicians, nurses<br>and other trained health care<br>providers in critical care patient<br>settings, such as anesthesiology,<br>intensive care medicine, neonatal<br>intensive care and other health care<br>areas. | It is to be used by physicians, nurses and<br>other trained health care providers in critical<br>care patient settings, such as anesthesiology,<br>intensive care medicine, neonatal intensive<br>care and other health care areas. |
| Performance<br>Standards | ISO 21647<br>ISO 9919 | ISO 21647<br>ISO 9919 |
| Safety<br>Standards | IEC/EN 60601-1<br>IEC/EN 60601-1-2<br>IEC 60601-1-8<br>UL 60601-1<br>ISO 14971 | IEC/EN 60601-1<br>IEC/EN 60601-1-2<br>IEC 60601-1-8<br>UL 60601-1<br>ISO 14971 |
| Biocompatibility | There are no issues of<br>biocompatibility for this device and no<br>biocompatibility testing was done. | There are no issues of biocompatibility for this<br>device and no biocompatibility testing was<br>done. |
| Sterility | This device does not require<br>sterilization and is shipped marked<br>non-sterile. | This device does not require sterilization and is<br>shipped marked non-sterile. |
{4}------------------------------------------------
SPECIAL-510(k) Submission for CAPNOSTREAM20 two parameter bedside monitor with microMediCO2 CO2 module 510(k) Summary of Safety and Effectiveness
## CONCLUSION
Capnostream20 with the micromediCO2 CO2 module does not raise any new potential safety risks and is equivalent in performance to the existing legally marketed device. Therefore, the device is substantially equivalent to the predicate device with respect to safety, effectiveness, and intended use.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three swooping lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002
# MAY 12 2010
Ms. Rachel Weissbrod Director of Regulatory Affairs Oridion Capnography, Incorporated 160 Gould Street Needham, Massachusetts 02494
Re: K094012
Trade/Device Name: Capnostream20 with Micromedico2 Module Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: II Product Code: CCK Dated: April 13, 2010 Received: April 15, 2010
Dear Ms. Weissbrod:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
#### Page 2- Ms. Rachel Weissbrod
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Nh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{7}------------------------------------------------
SPECIAL-510(k) Submission for CAPNOSTREAM20 two parameter bedside monitor with microMediCO2 CO2 module Statement of Indications for Use
# 6. Statement of Indications for Use
# TWO PARAMETER CAPNOSTREAM MONITOR (This document is not confidential) Indications for Use
December 15th, 2009
Ko94012 510(k) Number (if known)
Device Name: Capnostream 20
#### Indications for Use:
The Capnostream20 combined capnograph/pulse oximeter monitor is intended to provide professionally trained health care providers the continuous, non invasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and respiration rate, and for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. It is intended for use with neonatal, pediatric and adult patients in hospitals, hospital type facilities, intra hospital transport and home environments.
The Capnostream20 monitor provides the clinician with an integrated pulmonary index (IPI). The IPI is based on four parameters provided by the monitor: end tidal carbon dioxide, respiration rate, oxygen saturation and pulse rate. The IPI is a single index of an adult or pediatric patient's ventilatory status displayed on a scale of 1 - 10, where 10 indicates optimal pulmonary status. IPI monitoring displays a single value that represents the patient's pulmonary parameters and alerts clinicians to changes in the patient's pulmonary status.
The IPI is an adjunct to, and is not intended to replace, vital sign monitoring.
Prescription Use X (Per 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
L. Schull
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: k094012
Page 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.