K130411 · Cas Medical Systems, Inc. · MWI · May 29, 2013 · Cardiovascular
Device Facts
Record ID
K130411
Device Name
740 SELECT
Applicant
Cas Medical Systems, Inc.
Product Code
MWI · Cardiovascular
Decision Date
May 29, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The 740 SELECT is indicated for use as a bedside, portable device for use by health care professionals, clinicians and medically qualified personnel for spot checking, continuous monitoring and recording of adult, pediatric and neonatal patients. The Monitor features world class technology to facilitate the monitoring of non-invasive blood pressure, pulse rate, functional arterial oxygen saturation (Sp02) and body temperature in a variety of clinical environments.
Device Story
740 SELECT is a portable, multi-parameter patient monitor for spot-checking or continuous monitoring of NIBP, pulse rate, SpO2, and predictive body temperature. Inputs include oscillometric NIBP cuff pressure, red/infrared light absorption for SpO2, and temperature probe signals. Device processes these inputs to calculate vital signs, displaying them on a color touchscreen. Operated by clinicians in various clinical environments; supports mobile stands, wall mounts, or tabletop use. Features include alarm management, trend recall, and patient data entry via screen keyboard. Connectivity includes serial data output for EMR integration and nurse call relay. Benefits include enhanced bedside workflow, automated alarm management, and flexible monitoring acuity. Powered by AC or internal Li-Ion battery.
Clinical Evidence
Clinical validation reports were cited for NIBP, pulse oximeter, and temperature functions. No specific study metrics (e.g., sensitivity, specificity) were provided in the summary document.
Technological Characteristics
Portable monitor with color touchscreen. Parameters: Oscillometric NIBP (CASMED MAXNIBP), SpO2 (Masimo Rainbow SET or Nellcor NELL-1), and predictive temperature. Connectivity: Serial data output, nurse call relay. Power: AC or internal Li-Ion battery. Software: Embedded system for data processing, alarm management, and user interface.
Indications for Use
Indicated for adult, pediatric, and neonatal patients requiring spot-check or continuous monitoring of blood pressure, pulse rate, SpO2, and body temperature in clinical settings.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
February 15, 2013
740 SELECT
## K130411
#### 6. 510(k) Summary
MAY 2 9 2013
Image /page/0/Picture/6 description: The image shows the word "CASMED" in all caps. The letters are large and bold, and they appear to be slightly distressed or textured. A small registered trademark symbol is located to the right of the letter "D".
:
. . . .
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
| Submitter: | CAS Medical Systems, Inc. |
|--------------|-----------------------------------------------------------------------------------------------------------------------------|
| Address: | 44 East Industrial Rd. Branford CT. 06405 USA |
| Contact: | Ron Jeffrey - Director, Regulatory Affairs<br>Phone - (203) 488-6056<br>Fax - (203) 488-9438<br>Email - rjeffrey@casmed.com |
| Prepared: | February 15, 2013 |
| Trade Name: | 740 SELECTTM |
| Common Name: | Monitor, Physiological, Patient |
Classification Name: Cardiovascular Monitoring Device 870.2300 (MWI)
{1}------------------------------------------------
February 15, 2013
740 SELECT
K130411 page 2 of 4
#### EQUIVALENCE (Predicate Device)
The 740 SELECT is equivalent to the following device:
* CAS 740 Series Monitor (K033048)
#### DESCRIPTION
The 740 SELECT Series Monitor is a portable and rugged non-invasive multi-parameter device used for spot checking, continuous monitoring, and recording of blood pressure, pulse rate, functional oxygen saturation (%Sp02), and predictive body temperature in a variety of clinical environments. The Monitor includes features that are optional or configurable.
The 740 SELECT Series Monitor is compact and lightweight in design, allowing it to be easily carried and or configured for use with a mobile stand, wall mount or on a table top.
A Lithium lon (Li-lon) rechargeable battery is provided for automatic switching to back-up power if an untended loss of AC power should occur as well as allow the monitor to be used for clinical applications when portable monitoring is required.
The 740 SELECT Series Monitor is equipped with a color touchscreen which is the primary interface for the user to control and configure the Monitor (for patient type, alarms, settings and various clinical workflows). A dedicated message area provides patient alarm and technical messages. Date and time as well as the Monitor ID are displayed. A battery gauge provides approximate battery capacity when the monitor is operating on battery power.
The Monitor supports both point-of-care and variable acuity monitoring. Spot vital signs measurements from multiple patients can be saved individually. The monitor can be used for continuous patient monitoring with full alarm support.when increased vigilance in necessary based on patient acuity.
The Stand-By Mode provides enhanced bedside patient workflow and alarm management permitting sensor attachment, automatic alarm suspension and switch to active monitoring.
The color coded Trends Display allows the user to recall all trended physiologic patient. measurement records. Trends can be recalled for saved snapshots as well as timed intervals.
The Monitor provides a screen keyboard for entry of patient data (e.g., age, weight, height, gender, date of birth, patient name, and patient identification number).
A serial data output is provided for connectivity to hospital EMR systems and external serial devices. An internal isolated relay switch is provided for nurse call systems. The monitor supports an optional externally connected thermal strip chart recorder which is directly powered from the serial port.
{2}------------------------------------------------
740 SELECT
#### page 3 of 4 130411
The MAXNIBP™ non-invasive blood pressure (NIBP) parameter automatically inflates an occluding cuff and, using the oscillometric measurement technique, determines systolic, diastolic and mean arterial pressure as well as pulse rate. Measurement results along with user prompts and error messages are displayed in the dedicated message area on the display screen. The frequency of NIBP determination can be selected by the user in varied times between one and ninety minutes. The AUTO, MANUAL and STAT modes are available to cover'a variety of clinical uses:
The pulse oximeter parameter (%SpO2) determines arterial hemoglobin oxygen saturation by measuring the absorption of red and infrared light passing through the tissue. Changes in absorption caused by pulsations of blood in the vascular bed are used to determine arterial saturation and pulse rate. The oximeter requires no routine calibration or maintenance. Monitors with the SpO2 option are provided with the ability to display the plethysmograph waveform (in spot and or continuous monitoring mode). A signal strength bar provides relative quality of the SpO2 value based on the measurement site where the sensor is placed. An audio pulše tone "beep" can be generated each time the SpO2 module detects a pulse.
A pulse rate is displayed and appropriately color coded to indicate the source parameter (SpO2 or NIBP) of the displayed.
The temperature parameter has the capability of taking temperature in either normal (predictive) or monitor mode. In the normal mode, the thermometer's nicroprocessor "predicts" body temperature in about four (4) seconds for oral temperatures, about ten (10) seconds for axillary temperatures and about fifteen (15) seconds for rectal temperatures. The default setting used by the monitor for Tcmperature determinations is the Predictive (normal) mode. Upper and lower Temperature alarm limits can be enabled. The last temperature measurement time is displayed for reference.
Alarm conditions are displayed as text messages in a dedicated area on the bottom of the display screen. A recurring series of alert tones are provided to indicate: an alarm limit has been violated, alert user to a specific monitor operation (e.g., battery low), or a condition exists that affects patient safety, Visual cues (flashing background) are displayed within a parameter cell to indicate an alert is currently active for that parameter. A solid background indicates an alert is currently active for that parameter but the audio tone has been silenced.
#### The 740 SELECT Intended Use
The 740 SELECT is indicated for use as a bedside, portable device for use by health care professionals, clinicians and medically qualified personnel for spot checking, continuous monitoring and recording of adult, pediatric and neonatal patients. The Monitor features world class technology to facilitate the monitoring of non-invasive blood pressure, pulse rate, functional arterial oxygen saturation (Sp02) and body temperature in a variety of clinical environments.
{3}------------------------------------------------
### 740 SELECT Monitor Technologies Compared to Predicate Devices
The 740 SELECT compares substantially to the cited predicate device in that it uses fundamentally the same technologies for the common parameters. The non-invasive blood pressure component is the CASMED MAXNIBP an oscillometric instrument having neither a stethoscope nor a microphone and does not need the quiet environment necessary to detect the auscultatory sounds. Oscillometric instruments work on the principle that when the artery opens during a portion of the pressure cycle, an oscillation is superimposed on the pressure inside the cuff due to a tiny enlargement of the circumference of the limb caused by the surge of blood under the cuff. The amplitude of the oscillometric signals change over the course of the deflation of the cuff. Oscillometric devices look for oscillation amplitude of certain percentages of the maximum amplitude at mean arterial pressure (MAP), defining one percentage as the systolic point and another as the diastolic point. Alternatively, a combination of the amplitude, the slope of the increase or decrease, and some other complex factors are used to find these points.
The 740 SELECT pulse oximeter can include either of two industry leading components specific to the customer's requirement. Masimo.Rainbow SET® or Nellcor's NELL I technology. The pulse oximeter parameter (%SpO2) determines arterial oxyhemoglobin saturation by measuring the absorption of red and infrared light passing through the tissue. Changes in absorption caused by pulsations of blood in the vascular bed are used to determine arterial saturation and pulse rate.
The temperature parameter has the capability of taking temperature in either normal (predictive) or monitor mode. In the normal mode, the thermometer's microprocessor "predicts" body temperature in about four (4) seconds for Oral temperatures, about ten (10) seconds for Axillary temperatures and in about fifteen (15) seconds for Rectal temperatures.
#### Non-Clinical Performance Testing to Demonstrate Substantial Equivalence
The 740 SELECT has successfully undergone extensive performance, safety, electromagnetic, and environmental testing to ensure it has been found to be substantially equivalent to the primary and secondary predicate devices. In addition to the above laboratory tests, CAS has conducted a full program of individual hardware, software and systems verification and validation studies of the monitor and accessories.
#### Clinical Testing to Show Substantial Equivalence
The 740 SELECT has successfully undergone extensive clinical validation for the indicated use. The premarket notification cites clinical validation reports for the NIBP, pulse oximeter and the temperature function. Commen 、
#### Conclusions Drawn from Clinical and Non-Clinical Testing
Clinical evaluation, safety testing, software and systems validation demonstrate the 740 SELECT is substantially equivalent to the predicate device.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
DEPARTMENT OF HEALTH & HUMAN SERVICES-USA
Re:
May 29, 2013
Cas Medical Systems, Inc. Ron Jeffrey 44 East Industrial Rd. Branford, CT 06405 US
> K130411 Trade/Device Name: 740 select Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiovascular Monitoring Device Regulatory Class: Class II Product Code: MWI Dated: February 15, 2013 Received: February 28, 2013
Dear Ron Jeffrey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, Misbranding by reference to premarket notification (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htmfor the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Owen PD Earis -S
for Bram D. Zuckerman, M.D. Director Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
## Indications for Use
510(k) Nümber: K130411
Device Name: 740 SELECT
Indications for Use:
The 740 SELECT is indicated for use as a bedside, portable device for use by health care professionals, clinicians and medically qualified personnel for spot checking, continuous monitoring and recording of adult, pediatric and neonatal patients. The Monitor features world class technology to facilitate the monitoring of non-invasive blood pressure, pulse rate, functional arterial oxygen saturation (Sp02) and body temperature in a variety of clinical environments.
Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Owen P. Faris -S
2013.05.29 09:25:50
-04'00'
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.