The Tempus Pro is a portable vital signs monitor intended to be used by clinicians and medically qualified personnel for the attended or unattended monitoring of single or multiple vital signs in clinical and pre-hospital care applications. The device is indicated for 3 & 5 Lead ECG monitoring and 12 Lead ECG recording with interpretation, impedance pneumography, non-invasive blood pressure (NIBP), end-tidal CO2 (ETCO2), respiration rate, pulse oximetry (SpO2), contact temperature, invasive pressure and extended pulse oximetry capability including; carboxy haemoglobin (SpCO), metheglobin (SpMet), total haemoglobin (SpHb) and total oxygen content (SpOC) measurements. The monitor is intended to be used as a stand-alone monitor or as a telemedicine system (transmitting patient data to other medical professionals located elsewhere). The device is indicated for adults, pediatrics and neonates.
Device Story
Tempus Pro is a portable multi-parameter vital signs monitor for pre-hospital and clinical use by trained professionals. Inputs include ECG leads, pulse oximetry sensors, NIBP cuffs, capnometry, temperature probes, invasive pressure transducers, integrated camera, GPS receiver, and headset. Device processes physiological signals and transmits data via wired/wireless connections to remote i2i software systems for telemedicine. Output includes real-time vital sign displays, 12-lead ECG interpretation, images, video, and voice. Healthcare providers use output for patient assessment and clinical decision-making. Benefits include remote patient monitoring and data transmission to off-site specialists.
Clinical Evidence
Bench testing only. No clinical data presented. Testing included safety (IEC 60601-1), defibrillation/electrosurgical protection, environmental (RTCA DO-160, MIL810, EN1789), ingress protection (IEC 60529), EMC (IEC 60601-1-2), invasive pressure (IEC 60601-2-34), pulse oximetry (IEC 9919), and wireless co-existence. Software validated per IEC 62304 and FDA guidance. Comparative bench testing demonstrated performance equivalence to predicates.
Technological Characteristics
Portable vital signs monitor. Sensing: ECG, pulse oximetry (Masimo SET Rainbow), NIBP, capnometry, temperature, impedance pneumography, invasive pressure. Connectivity: Wired/wireless data transmission to remote i2i system. Software: Includes mymisys32.dll for ECG interpretation. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-34, IEC 9919, IEC 60529, RTCA DO-160, MIL810, EN1789, EN13718-1, EN60068.
Indications for Use
Indicated for adults, pediatrics, and neonates requiring monitoring of vital signs including 3/5-lead ECG, 12-lead ECG with interpretation, impedance pneumography, NIBP, ETCO2, respiration rate, pulse oximetry (SpO2, SpCO, SpMet, SpHb, SpOC), contact temperature, and invasive pressure in clinical or pre-hospital settings.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
K133988 Page 1 of 4
### RDT Abbreviated 510(k) Submission - Tempus Pro (EF)
### 5. 510(k) Summary of Safety and Effectiveness
MAY 1 6 2014
#### 5.1 Submitter's Information
i
The submitter of this abbreviated pre-market notification is:
| Name: | Remote Diagnostic Technologies Limited |
|------------------------|--------------------------------------------------------------------------------|
| Address: | The Old Coach House, Farleigh Wallop, Basingstoke, RG25<br>2HT, United Kingdom |
| Company Phone No: | +44 (0) 1256 362 400 |
| Company Fax No: | +44 (0) 1256 362 415 |
| Contact Person: | Dr Leigh Cornock (Director of Systems Eng. & Test) |
| Date summary prepared: | 10th April, 2014 |
#### 5.2 Device Identification
| Device Trade Name: | Tempus Pro |
|----------------------------|----------------------------------------------------------|
| Common Name: | Patient Monitor (without arrhythmia detection or alarms) |
| Class: | II |
| Classification Panel: | 74 |
| Primary Product Code: | MWI |
| Primary Regulation Number: | 870.2300 |
| Secondary FDA product code | Regulation Number |
|------------------------------------------------------------------------|-------------------|
| Cardiac Monitor (DRT) | 870.2300 |
| Electrocardiograph (DPS) | 870.2340 |
| Impedance Plethysmograph (DSB) | 870.2770 |
| Non Invasive Blood Pressure Measurement System (DXN) | 870.2770 |
| Carbon Dioxide Gas Analyzer (CCK) | 868.1400 |
| Breathing frequency monitor (MNR) | 868.2375 |
| Oximeter (DQA) | 870.2700 |
| Clinical electronic thermometer (FLL) | 880.2910 |
| Computer, Blood-pressure (DSK) | 870.1110 |
| Transmitters and Receivers, Physiological Signal, Radiofrequency (DRG) | 870.2910 |
#### 5.3 Device Description
The Tempus Pro is a multi-parameter vital signs monitor designed for use in pre-hospital care and remote clinical locations by trained healthcare professionals. It provides 3&5 lead ECG monitoring, 12 lead ECG recording, pulse oximetry, non-invasive blood pressure, sidestream capnometry, contact temperature, impedance respiration, invasive pressure and user configurable alarms.
In addition, it provides the ability to transmit all vital signs data via wired or wireless connections to a software system (called i2i) expected to be based in a facility far from the user e.g. a response centre facility. In addition to sending all vital signs, the system can also send pictures or video via an integrated camera, geographic position by an integrated GP\$ receiver and voice via a wired or wireless headset.
{1}------------------------------------------------
### RDT Abbreviated 510(k) Submission - Tempus Pro (EF)
#### 5.4 Indications for Use
The following indications for the Tempus Pro remain unchanged by addition of the extended features for ECG recording, pulse oximetry and invasive blood pressure monitoring to the Tempus Pro, with the exception of removal of "(SpO2)" as additional pulse oximetry measurements are being added.
"The Tempus Pro is a portable vital signs monitor intended to be used by clinicians and medically qualified personnel for the attended or unattended monitoring of single or multiple vital signs in clinical and pre-hospital care applications. The device is indicated for 3 & 5 Lead ECG monitoring and 12 Lead ECG recording with interpretation, impedance pneumography, non-invasive blood pressure (NIBP), end-tidal CO2 (ETCO2), respiration rate, pulse oximetry (SpO2), contact temperature, invasive pressure and extended pulse oximetry capability including; carboxy haemoglobin (SpCO), metheglobin (SpMet), total haemoglobin (SpHb) and total oxygen content (SpOC) measurements.
The monitor is intended to be used as a stand-alone monitor or as a telemedicine system (transmitting patient data to other medical professionals located elsewhere).
The device is indicated for adults, pediatrics and neonates."
#### 5.5 Comparison with Cleared Device
The intended use and indications for use, plus the fundamental technology used in the Tempus Pro device, remain essentially unchanged by the addition of optional extended features to the ECG recording, pulse oximetry, and invasive blood pressure functions, which are specifically:
- I Extended Pulse Oximetry capability to include optional carboxy haemoglobin (SpCO). metheglobin (SpMet) total haemoglobin (SpHb) and total oxygen content (SpOC) measurements
- . Optional 12-Lead ECG Recording Interpretation
- . 2 optional additional Invasive Blood Pressure Channels
#### 5.6 Substantial Equivalence
In adding some of the optional features, it has been necessary to cite other predicates for specific functionality provided:
- . Pulse Oximetry - new Masimo SET Rainbow module and associated sensors/patient cable, incorporating original SET oxygen saturation technology as used in cleared Tempus Pro plus 4 new measurement capabilities, is substantially equivalent to the same module incorporated into the Masimo Radical-7 device (K110028)
- I Invasive Blood Pressure - additional external IBP module which uses same Medlab IBP module as already incorporated internally into the cleared Tempus Pro device (K130773)
- ECG Recording Interpretation enabling the interpretation output of the ■ mymisys32.dll, which is already an integral part of the cleared Tempus Pro, but the interpretation output is not currently displayed to the user. The dll is part of the CardioView software package from QRS Diagnostic (K083749).
{2}------------------------------------------------
### RDT Abbreviated 510(k) Submission - Tempus Pro (EF)
Hence, the new pulse oximetry module with sensors and cables, the additional invasive blood pressure, and the ECG recording 12 lead interpretation are the same as those used by the original equipment manufacturers in their cleared devices.
The 510(k) numbers for the predicate devices to which we are claiming substantial equivalence are K110028, K130773, and K083749.
#### 5.7 Summary of Non-Clinical Testing
i
ﺮ ﺍ
All of the above modifications have been made under well-established design control procedures, which ensure that appropriate risk management processes have been carried out to determine their impact and ensure that appropriate verification/ validation testing is performed.
The non-clinical testing carried out in relation to addition of the extended features described above is summarized in the following table:
| Area | Testing Performed |
|-----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Safety | The device has been tested to IEC60601-1. |
| Defibrillation and<br>electrosurgical<br>protection | The device has been tested for operation with a defibrillator<br>and operation with an electro-surgical unit according to<br>IEC60601-1 (and relevant particular standards). |
| Environmental | The device has been tested to a range of environmental<br>(temperature, altitude, humidity, vibration, shock) tests<br>according to RTCA DO-160, MIL810, EN1789, EN13718-1,<br>EN60068. |
| Ingress Protection | The device has been tested to IEC 60529 for solid and water<br>ingress. |
| EMC | The device has been tested to IEC 60601-1-2 for emissions<br>and immunity and RTCA DO-160 for radiated emissions.<br>(including immunity at 20 V/m) |
| Invasive pressure | The device has been tested to IEC 60601-2-34. |
| Pulse oximeter | The device has been tested to IEC 9919. |
| Comparative testing to<br>predicates | Comparative testing has been performed to demonstrate that<br>the performance of the device is equivalent to the predicates. |
| Software | The requirements of the FDA document Guidance for the<br>Content of Premarket Submissions for Software in Pre-Market<br>Submissions has been applied. In addition, the requirements<br>of IEC 62304 have been addressed. |
| Bench testing | All parameters of the device have been tested to confirm they<br>operate to specification across their stated performance range<br>and across their stated temperature range. |
| Bench testing | The product has been bench tested to confirm that all data is<br>transmitted reliably and accurately. |
{3}------------------------------------------------
## RDT Abbreviated 510(k) Submission - Tempus Pro (EF)
| Area | Testing Performed |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Wireless co-existence testing | The thermometer has been tested to confirm it operates reliably in the presence of other wireless fields as per the FDA Guidance for Radio-Frequency Wireless Technology in Medical Devices. |
In each case the results of this testing confirmed that acceptance criteria defined by the relevant standard, or other appropriate reference document had been met.
With respect to usability, no additional user validation was considered necessary, as the Tempus Pro (EF) is almost identical to the predicate Tempus Pro in terms of physical and interface features, including size and weight, user interface data layouts, button styles, menus, and layout of connectors, resulting the physical interaction with the device being unchanged.
#### 5.8 Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this Premarket Notification, Remote Diagnostic Technologies Limited conclude that this extended version of the Tempus Pro Patient Monitor is safe and effective, and substantially equivalent to the unmodified version of this device and other cleared devices used as predicates.
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Image /page/4/Picture/3 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized symbol to the right and text arranged in a circle to the left. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The symbol is a stylized representation of a human figure.
May 16, 2014
Remote Diagnostic Technologies Ltd. · Dr. Leigh Cornock Director of Systems Engineering and Test The Old Coach House, The Avenue Farleigh Wallop, Basingstoke, Hampshire, UK
K133988 Re:
> Trade/Device Name: Tempus pro extended features Regulation Number: 21 CFR 870.2300 Regulation Name: Monitor, Physiological, Patient (Without Arrhythmia Detection or Alarms) Regulatory Class: Class II Product Codes: MWI, DRT, DPS, DSB, DXN, CCK, MNR, DQA, FLL, DSK, DRG Dated: April 14, 2014 Received: April 23, 2014
Dear Dr. Leigh Cornock,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
### Page 2 - Leigh Cornock
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours
forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use Statement
K133988 Page 1 of 1
510(k) Number (if known):
Device Name: Tempus Pro (EF)
Indications for Use:
The Tempus Pro is a portable vital signs monitor intended to be used by clinicians and medically qualified personnel for the attended or unattended monitoring of single or multiple vital signs in clinical and pre-hospital care applications. The device is indicated for 3 & 5 Lead ECG monitoring and 12 Lead ECG recording with interpretation, impedance pneumography, non-invasive blood pressure (NIBP), endtidal CO2 (ETCO2), respiration rate, pulse oximetry (SpO2), contact temperature, invasive pressure and extended pulse oximetry capability including; carboxy haemoglobin (SpCO), metheglobin (SpMet), total haemoglobin (SpHb) and total oxygen content (SpOC) measurements.
The monitor is intended to be used as a stand-alone monitor or as a telemedicine system (transmitting patient data to other medical professionals located elsewhere).
The device is indicated for adults, pediatrics and neonates.
Prescription Use _V__ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Date:
2014.05.16
12:28:44 -04'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.