The Tempus Pro Patient Monitor is intended to be used in remote or pre-hospital care situations by trained healthcare professionals e.g. nurse, EMT, paramedic, physician, military medic etc. The device is intended to be used primarily as a standalone monitor for traditional monitoring applications. It is expected that its real-time telemedicine capabilities will be used in a minority of applications. When its telemedicine features are used it is intended that this will be for the purpose of obtaining support in the diagnosis and treatment decisions for the patient e.g. where the patient is in a remote country and the user's organisation needs to make an extraction or repatriation decision.
Device Story
Tempus Pro is a portable multi-parameter vital signs monitor for pre-hospital and remote clinical use by trained professionals. Inputs include ECG (3/5/12-lead), impedance pneumography, NIBP, ETCO2, SpO2, contact temperature, invasive pressure, integrated camera (still/moving images), GPS, and voice (headset). Device functions as a standalone monitor or as a telemedicine system, transmitting data via wired Ethernet or WiFi to the i2i software system at a remote response center. Healthcare providers at the remote site receive real-time voice, vital signs, and visual data to support diagnosis and treatment decisions, such as extraction or repatriation. Benefits include remote clinical support for patients in isolated locations.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), environmental stress testing (RTCA DO-160, MIL810), and performance verification of physiological parameters against AAMI and IEC standards. Wireless data transmission reliability and co-existence were verified via bench testing.
Indicated for adults, paediatrics, and neonates. Used by clinicians/medically qualified personnel for attended or unattended monitoring of vital signs in clinical and pre-hospital settings. Parameters: 3/5-lead ECG, 12-lead ECG, impedance pneumography, NIBP, ETCO2, respiration rate, SpO2, contact temperature, and invasive pressure. Not an apnoea monitor; not for use in MRI environments.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
Welch Allyn Propaq 206 Encore Patient Monitor (K012451)
Submission Summary (Full Text)
{0}------------------------------------------------
Remote Diagnostic Technologies Ltd - Tempus Pro™ Patient Monitor 510k Summary of Safety and Effectiveness
JUN 0 5 2013
# 510(k) Summary of Safety and Effectiveness
### Submitter Information
Name:
Address:
Remote Diagnostic Technologies Ltd
The Old Coach House The Avenue Farleigh Wallop Basingsloke RG25 2HT United Kingdom
Phone Number:
+44 1256 362 415
Chris Hannan
January 2013
+44 1256 362 400
Contact Name:
Fax Number:
Date Prepared
Device Name
Portable Patient Monitor Common Name:
Tempus Pro7M Patient Monitor Proprietary Name:
Classification Name:
Monitor, physiological, patient (without arrhythmia detection or alarms)
#### Device Description
The Tempus Pro is a multi-parameter vital signs monitor designed for use in prehospital care and remote clinical locations by trained healthcare professionals. It provides 3 & 5 Lead ECG monitoring and 12 Lead ECG recording, impedance pneumography, non-invasive blood pressure (NIBP), end-tidal CO2 (ETCO2) and respiration rate, pulse oximely (SpO2), contact temperature and invasive pressure ..
In addition, it provides the ability to transmit all vital signs data via wired Ethernet or wireless WiFi connections to a software system (called i2i) expected to be based in a facility far from the user e.g. a response centre facility. In addition to sending all vital signs, the system can also capture and transmit other data including still or moving pictures via an integrated camera, geographic position by an integrated GPS receiver and voice via a wired or wireless headset.
{1}------------------------------------------------
## Remote Diagnostic Technologies Ltd - Tempus Pro™ Patient Monitor 510k Summary of Safety and Effectiveness
It is expected that the ability to transmit data in real-time will be performed in remote locations typically using satellite or terrestrial communications systems.
The Tempus Pro Is used in conjunction with 121 software, which provides a system for receiving real-time voice and medical data. The system enables users to receive voice, vital signs data and other medical data, still and moving video pictures from the Tempus Pro devices.
The i2i system can be used by commercial response centre service providers or by individuals or organisations wishing to provide their own internal service
I2i also supports a full patient records database.
#### Intended Use
The Tempus Pro Patient Monitor is intended to be used in remote or pre-hospital care situations by frained healthcare professionals e.g. nurse, EMT, paramedic, physician, military medic etc.
The device is intended to be used primarily as a standalone monitor for traditional monitoring applications. It is expected that its real-time telemedicine capabilities will be used in a minority of applications. When its telemedicine features are used it is intended that this will be for the purpose of obtaining support in the diagnosis and treatment decisions for the patient e.g. where the patient is in a remote country and the user's organisation needs to make an extraction or repatriation decision.
#### Indications for Use
The Tempus Pro is a portable vital signs monitor intended to be used by clinicians and medically qualified personnel for the attended or unattended monitoring of single or multiple vital signs in clinical and pre-hospital care applications. The device is indicated for 3 & 5 Lead ECG monitoring and 12 Lead ECG recording, impedance pneumography, non-invasive blood pressure (NIBP), end-tidal CO2 (ETCO2) and respiration rate, pulse oximetry (SpO2), contact temperature and invasive pressure.
The monitor is intended to be used as a stand-alone monitor or as a telemedicine system (transmitting patient data to other medical professionals located elsewhere).
The device is indicated for adults, paediatrics and neonates.
#### Contraindications
The Tempus Pro does not replace a physician's care. The device is not an apnoea monitor.
The Tempus Pro is not intended to be used in strong magnetic or electro-magnetic fields which are generated for medical purposes e.g. MRI.
{2}------------------------------------------------
## Remote Diagnostic Technologies Ltd - Tempus Pro™ Patient Monitor 510k Summary of Safety and Effectiveness
#### Predicate Devices
The Tempus Pro M Patient Monitor is predicated on our existing model, the Tempus IC Professional Patient Monitor, which was the subject of a previous submission (K101264). Additional predicates include: the Welch Allyn Propag 206 Encore Patient Monitor (K012451) to demonstrate substantial equivalence (SE) for the overall application and medical parameter modules.
The Tempus Pro Patient Monitor has the same intended use as the Tempus IC Professional Patient Monitor, with the following additions; the Tempus Pro Patient Monitor is also intended for monitoring neonates, and provides medical monitoring parameter functions for invasive pressure and impedance respiration.
#### Tesling
The Tempus IC Professional uses currently available (OEM) technology found in many legally marketed devices.
| Area | Testing Performed |
|-----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Safety | The device has been tested to IEC60601-1. |
| Defibrillation and<br>electrosurgical<br>protection | The device has been tested for operation with a<br>defibrillator and operation with an electro-surgical unit<br>according to IEC60601-1 (and relevant particular<br>standards). |
| Environmental | The device has been tested to a range of environmental<br>(temperature, altitude, humidity, vibration, shock) tests<br>according to RTCA DO-160, MIL810, EN1789,<br>EN13718-1, EN60068. |
| Ingress Protection | The device has been tested to IEC60529 for solid and<br>water ingress. |
| ECG monitoring | Testing to AAMI EC11 & EC13 and IEC60601-2-25 &<br>IEC60601-1-27 has been performed. |
| EMC | The device has been tested to IEC60601-1-2 for<br>emissions and immunity and RTCA DO-160 for radiated<br>emissions. |
| Alarms | The alarm functions of the product have been tested to<br>IEC60601-1-8. |
| Invasive pressure | The device has been tested to IEC60601-2-34. |
| Contact temperature | The device has been tested to EN12470-4 and<br>IEC80601-2-56 |
{3}------------------------------------------------
## K130773
## Remote Diagnostic Technologies Ltd - Tempus Pro™ Patient Monitor 510k Summary of Safety and Effectiveness
| Area | Testing Performed |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparative testing to<br>predicates | Comparative testing has been performed to demonstrate<br>that the performance of the device is equivalent to the<br>predicates. |
| Software | The requirements of the FDA document Guidance for the<br>Content of Premarket Submissions for Software in Pre-<br>Market Submissions has been applied. In addition, the<br>requirements of IEC62304 and IEC60601-1-4 have been<br>addressed. |
| Bench testing | All parameters of the device have been tested to confirm<br>they operate to specification across their stated<br>performance range and across their stated temperature<br>range. |
| Bench testing | The product has been bench tested to confirm that all<br>data is transmitted reliably and accurately. |
| Wireless range | The device has been tested to confirm it operates<br>reliably at its maximum stated range. |
| Wireless<br>co-existence<br>testing | The thermometer has been tested to confirm it operates<br>reliably in the presence of other wireless fields as per the<br>FDA Guidance for Radio-Frequency Wireless Technology in<br>Medical Devices. |
### Evidence of Conformity to Essential Principles
The device has been shown to conform to the essential principles for safety and performance defined in guidance prepared by the Global Harmonization Task Force Study Group1 (GHTF/SG1/N14R9:2005), with supporting evidence prepared in the summary technical documentation (STED) format recommended in final version of GHTF guidance (SG1/N011: 2008).
Specifically, this evidence includes performance testing, software validation, electrical safety, electromagnetic compatibility etc..
The design of this device utilises currently available (OEM) technology found in many legally marketed devices. In terms of measurement performance, the Tempus Pro 14 is effectively identical to the devices that incorporate the same OEM technology.
#### Conclusion
On the basis of these results and the above referenced testing, if is our determination that the device is safe, effective and performs as well as, or better than, the legally marketed predicate device(s).
{4}------------------------------------------------
Remote Diagnostic Technologies Ltd - Tempus Pro™ Patient Monitor 510k ic Technologies Lid - Tempus Fro - Patient Monitor STOX
Summary of Safety and Effectiveness
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
·.
.
:
:
Respectfully
.
: :
signature
Chris Hannan Programme & Regulatory Affairs Director
the submit and the submit of the subject of the subject of
. . . . .
. '
..............................................................................................................................................................................
・
.
1999 - 1999 - 1999
:: : :
.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged in a circular fashion around the symbol. The logo is black and white.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 5, 2013
Remote Diagnostic Technology Limited c/o Mr. Mark Job Regulatory Technology Services LLC 1394 25th Street NW Buffalo, MN 55313
Re: K130773
Trade/Device Name: Tempus Pro Regulatory Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including Cardiotachometer and Rate Alarm) Regulatory Class: II (two) Product Code: 74 MWI Dated: May 20, 2013 Received: May 21, 2013
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. Mark Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## K130773
# Statement of Indications for Use
510(k) Number (if known): Not known
Tempus Pro™ Patient Monitor
Indications for Use
Device Name:
The Tempus Pro is a portable vital signs monitor intended to be used by clinicians and medically qualified personnel for the attended monitoring of single or multiple vital signs in clinical and pre-hospital care applications. The device is indicated for 3 & 5 Lead ECG monitoring and 12 Lead ECG recording, impedance pneumography, noninvasive blood pressure (NIBP), end-tidal CO2 (ETCO2) and respiration rate, pulse oximetry (SpO2), contact temperature and invasive pressure.
The monitor is intended to be used as a stand-alone monitor or as a telemedicine system (transmitting patient data to other medical professionals located elsewhere).
The device is indicated for adults, paediatrics and neonates.
Prescription Use: YES (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use: NO (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
* -- Concurrence of EDRH, Office of Device Evaluation (ODE)-
Date: 2013.06.05 16:49:33-04'00' for Bram Zuckerman
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.