K133622 · Wipro GE Healthcare · DPS · Jan 23, 2014 · Cardiovascular
Device Facts
Record ID
K133622
Device Name
MAC 2000 ECG ANALYSIS SYSTEM
Applicant
Wipro GE Healthcare
Product Code
DPS · Cardiovascular
Decision Date
Jan 23, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2340
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The MACTM 2000 ECG analysis system is a portable device intended to be used by or under the direct supervision of a licensed healthcare practitioner using surface electrodes to acquire, analvze, display, and record information for adult and pediations in a hospital, medical professional's facility, clinics, physician`s office or outreach centers. NOTE: Pediatric populations are defined as patients between the ages of 0 and 15 years. The MACTM 2000 ECG analysis system provides the following modes of operation: - Resting ECG mode . - . Arrhythmia mode - Exercise mode for exercise stress testing (optional) . - RR analysis mode for RR interval analysis (optional) . The basic system prints 6 or 12 leads of ECG and is upgradeable to provide software options such as 12-lead ECG measurement and interpretive analysis. Arrhythmia detection is provided for the convenience of automatic documentation. Transmission and reception of ECG data to and from a central ECG cardiovascular information system is optional.
Device Story
Portable ECG analysis system; acquires, displays, and prints multi-lead ECG data via surface electrodes. Features integrated 7-inch color display, alphanumeric keyboard, and thermal writer for real-time/continuous waveforms and reports. Includes 'Hook up advisor' for lead quality monitoring. Optional features: exercise stress testing, RR interval analysis, barcode reader for patient data, and connectivity to central cardiovascular information systems via WiFi/network for data exchange and PDF export. Used by clinicians in hospitals, clinics, and outreach centers. Provides automated arrhythmia detection and interpretive analysis to assist clinical decision-making. Benefits include rapid diagnostic data acquisition and streamlined electronic record management.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical, mechanical, safety, and system design verification protocols. Quality assurance measures included risk analysis, requirements reviews, and design reviews.
Technological Characteristics
Portable ECG system with 7-inch integrated display and thermal writer. Connectivity via WiFi and network interface for central system integration. Supports 6 or 12-lead acquisition. Software-based interpretive analysis and arrhythmia detection. Complies with voluntary electrical and safety standards. Sterilization: Not applicable.
Indications for Use
Indicated for adult and pediatric (0-15 years) patients. Used by or under supervision of licensed healthcare practitioners in hospitals, clinics, physician offices, or outreach centers to acquire, analyze, display, and record ECG data.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
ﺮ
્રિ
### Premarket Notification 510(k) Summary As required by section 807.92
MACTM 2000 ECG Analysis System
# GENERAL COMPANY INFORMATION as required by 807.92(a)(1)
| Date: | [4th Nov 2013] |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Wipro GE Healthcare Private Ltd.<br>4, kadugodi industrial area<br>bangalore, INDIA 560067 |
| Primary Contact Person: | Sanjukta DebRoy<br>GE Healthcare.<br>Phone : +91 9632211022<br>Fax :+9180 28411645 |
| Secondary Contact Person: | Douglas Kentz<br>GE Healthcare<br>Phone : 1-414 581-8987<br>Fax : 1-414 362-2585 |
| Device:<br>Trade Name: | MACTM 2000 ECG Analysis System |
| Common/Usual Name: | Electrocardiograph |
| Classification Names: | Electrocardiograph,<br>Programmable Diagnostic Computer,<br>Transmitter receiver Electrocardiograph, Telephone |
| Product Code: | DPS,<br>DQK,<br>DXH |
| Regulation No: | 21 CFR 870.2340,<br>21 CFR 870.1425,<br>21 CFR 870.2920 |
| Predicate Device(s): | MAC 1600 ECG Analysis System<br>MAC 5500 ECG Analysis System<br>MAC 800 ECG Analysis System |
| Regulatory Class | II |
| Prescription Status | Prescription Device |
| Device Classification Panel | Cardiovascular |
| Device Description: | The MACTM 2000 ECG analysis system can print and display multiple<br>leads of ECG data. The device is capable of acquiring 2 additional ECG<br>leads beyond what is needed for classical 12 lead ECG acquisition. |
{1}------------------------------------------------
#### page 2 of 4 K133622
The MACTM 2000 ECG analysis system provides ECG measurement in both resting and stress mode. The device also supports arrhythmia as a mode of operation.
The Hook up advisor feature in MAC™ 2000 ECG analysis system advises the users of poor lead quality based on noise measurement.
The MACT14 2000 ECG Analysis System delivers multiple leads of ECG on full-size reports and includes an alphanumeric keyboard for patient demographics and other data entry, an integrated 7' color display, and an integrated thermal writer. The thermal writer will print real time continuous waveform, alphanumeric data and non real time reports.
The device has an optional feature of transmission and reception of ECG data, downloading of orders and patient demographics to and from a central ECG cardiovascular information system and supports exporting of ECG records in PDF. It also has an optional internal memory and removable storage to store resting ECG records.
The patient information can be entered with the help of optional bar code reader.
Clinical Trials Data Guard and audit trail options are also available to support electronic record requirements.
MAC™ 2000 ECG Analysis system can be used as a portable unit also.
Indication for Use:
The MACTM 2000 ECG analysis system is a portable device intended to be used by or under the direct supervision of a licensed healthcare practitioner using surface electrodes to acquire, analvze, display, and record information for adult and pediations in a hospital, medical professional's facility, clinics, physician`s office or outreach centers.
NOTE: Pediatric populations are defined as patients between the ages of 0 and 15 years.
The MACTM 2000 ECG analysis system provides the following modes of operation:
- Resting ECG mode .
- . Arrhythmia mode
- Exercise mode for exercise stress testing (optional) .
- RR analysis mode for RR interval analysis (optional) .
The basic system prints 6 or 12 leads of ECG and is upgradeable to provide software options such as 12-lead ECG measurement and interpretive analysis. Arrhythmia detection is provided for the convenience of automatic documentation.
Transmission and reception of ECG data to and from a central ECG cardiovascular information system is optional.
{2}------------------------------------------------
MAC™ 2000 ECG Analysis System uses the same fundamental technology as Technology: its predicate device MAC1600 ECG Analysis System in acquiring, analyzing, recording, displaying and printing ECG data for both adult and pediatric populations.
> The basic system prints 6 or 12 leads of ECG and provides optional transmission and reception of ECG data to and from a central ECG cardiovascular information system. The system can be upgraded with software options such as stress testing, interpretive analysis, storage and communication options which is similar to predicate device.
> The intended use of the predicate device differs from the proposed device on verbiage as it has been modified to increase clarity and also on the offering of RR analysis mode of operation which is not available for sale in United States.
> MACTM 2000 ECG Analysis System is similar to MAC 5500 Resting ECG Analysis Systems K073625 in the technology of downloading orders and patient demographics from a central ECG cardiovascular information system.
> MACTM2000 ECG Analysis System makes use of the similar technology of exporting ECG reports in PDF and WiFi communication from MAC800 Resting ECG Analysis System K090212.
> For more on the predicate device comparison refer to section 12 of this 510K submission.
Performance Standards
The MACTM 2000 ECG Analysis System complies with the voluntary standards as detailed in Section 9 of this submission.
Sterilization
Not Applicable
ي
{3}------------------------------------------------
#### Determination of Substantial Equivalence:
### Summary of Non-Clinical Tests:
Verification and Testing activities establish the performance, functionality. usability, safety, and reliability characteristics of MAC™ 2000 ECG analysis system.
MACTM 2000 ECG analysis system comply with voluntary standards as detailed in Section 09, 15, 16, 17 and 18 of this premarket submission.
The following quality assurance measures were applied to the development of the system:
- Risk Analysis ◆
- Requirements Reviews .
- . Design Reviews
Summary of Simulated Use Setting:
The Design verification of MAC"131 2000 ECG analysis system has been divided into several protocols that include electrical, mechanical, safety testing, reliability, and system design verification.
The performance testing included testing on unit level, system level, as well as usability and safety parameters.
The results of the Design verification testing protocols have been documented in Section 18 of this 510(k) application.
The results demonstrate that MAC"N 2000 ECG analysis system meets all design requirements and performance claims.
The subject of this premarket submission, MAC™ 2000 ECG analysis system , did not require clinical studies to support substantial equivalence.
#### Conclusion:
GE Healthcare considers the MAC"M 2000 ECG analysis system to be as safe and as effective as the predicate device, and the performance to be substantially equivalent to the predicate device.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Food and Drug Administration
### January 23, 2014
Wipro GE Healthcare Douglas Kentz 8200 West Tower Ave Milwaukee, WI 53223 US
Re: K133622
> Trade/Device Name: MAC 2000 ECG Analysis System Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS, DQK, DXH Dated: November 20, 2013 Received: November 22, 2013
Dear Douglas Kentz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Douglas Kentz
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Owen P., Faris -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# K133622
.
#### Indications for Use
510(k) Number (if known):
Device Name: MAC™ 2000 ECG analysis system
Indications for Use:
The MAC™ 2000 ECG analysis system is a portable device intended to be used by or under the direct supervision of a licensed healthcare practitioner using surface electrodes to acquire, analyze, display, and record information for adult and pediatric populations in a hospital, medical professional's facility, clinics, physician's office or outreach centers.
NOTE: Pediatric populations are defined as patients between the ages of 0 and 15 years.
The MACTM 2000 ECG analysis system provides the following modes of operation:
- . Resting ECG mode
- . Arrhythmia mode
- . Exercise mode for exercise stress testing (optional)
- . RR analysis mode for RR interval analysis (optional)
The basic system prints 6 or 12 leads of ECG and is upgradeable to provide software options such as 12-lead ECG measurement and interpretive analysis.
Arrhythmia detection is provided for the convenience of automatic documentation. Transmission and reception of ECG data to and from a central ECG cardiovascular information system is optional.
Prescription Use_ AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Digitally signed by
Owen P. Faris -S
Date: 2014.01.23
15:35:48-05'00'
Vol. 1 Section 4 Page 2 of 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.