K090212 · Ge Medical Syst. Information Technologies · DPS · Feb 19, 2009 · Cardiovascular
Device Facts
Record ID
K090212
Device Name
MAC 800 RESTING ECG ANALYSIS SYSTEM
Applicant
Ge Medical Syst. Information Technologies
Product Code
DPS · Cardiovascular
Decision Date
Feb 19, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2340
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The MAC 800 is a portable ECG acquisition, analysis and recording system. The MAC 800 is intended to acquire, analyze, display and record information from adult and pediatric populations. Pediatric population is defined as patients between the ages of 0 and 15 years. The MAC 800 is intended to be used by trained operators in a hospital or medical professional's facility environment to record ECG signals from surface electrodes. The basic system shall provide 2 modes of operation: (1) Resting ECG mode and (2) Arrhythmia mode. The basic systems shall print 3, 6-leads of ECG. The device shall be upgradeable to provide software options such as 12-lead ECG measurement and interpretive analysis. Transmission and reception of ECG data to and from a central ECG cardiovascular information system shall be optional. The arrhythmia detection portion of the MAC 800 is provided to the customer for the convenience of automatic documentation. The MAC 800 is used under the direct supervision of a licensed healthcare practitioner.
Device Story
Portable ECG acquisition, analysis, and recording system; inputs ECG signals via surface electrodes. Features integrated 7" display, thermal writer, and keypad for data entry. Operates in Resting ECG or Arrhythmia modes; provides real-time waveform printing and optional 12-lead measurement/interpretation. Used in hospitals/clinics by trained operators; supports optional data transmission to central cardiovascular information systems. Output includes printed reports, PDF files on SD cards, or digital records. Assists clinicians in diagnostic documentation; arrhythmia detection provided for convenience only, not as a clinical alarm system.
Clinical Evidence
No clinical data; bench testing only. System verification and validation included module-level testing, integration testing, and performance/safety testing to confirm substantial equivalence.
Technological Characteristics
Portable ECG system with integrated thermal writer, 7" color display, and keypad. Supports 3, 6, and optional 12-lead ECG acquisition. Connectivity includes optional SD card storage, barcode/magnetic card scanners, and network transmission to central information systems. Software includes optional measurement and interpretive analysis modules. Operates on AC/battery power.
Indications for Use
Indicated for adult and pediatric (0-15 years) patients for ECG acquisition, analysis, display, and recording. Used by trained operators in hospitals or medical facilities under licensed practitioner supervision. Contraindicated for intra-cardiac use, as a vital signs monitor, or during patient transport.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
090212
# GE Healthcare 510(K) Premarket Notification Submission MAC 800 Resting ECG Analysis System
## 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | January 26, 2009 | | |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|------|
| Submitter: | Larry Lepley<br>Regulatory Affairs - Diagnostic Cardiology<br>9900 Innovation Drive<br>Wauwatosa, WI 53226 | | |
| Primary Contact Person: | Larry Lepley<br>Regulatory Affairs - Diagnostic Cardiology<br>9900 Innovation Drive<br>Wauwatosa, WI 53226<br>T: (414) 721-2593<br>F: (414) 721-3899 | | |
| Secondary Contact Person: | Patricia Taige<br>Regulatory Affairs - Diagnostic Cardiology<br>9900 Innovation Drive<br>Wauwatosa, WI 53226<br>T: (414) 721-3222<br>F: (414) 721-3899 | | |
| Device: Trade Name: | MAC 800 Resting ECG Analysis System | | |
| Common/Usual Name: | Electrocardiograph | | |
| Classification Names: | 21 CFR | Classification Name | Code |
| Product Code: | 870.2340 | Electrocardiograph | DPS |
| | 870.1425 | Programmable Diagnostic Computer | DQK |
| | 870.2920 | Transmitters and Receivers,<br>Electrocardiographs, Telephone | DXH |
| Predicate Device(s): | K081437, MAC 1600 ECG Analysis System | | |
| Device Description: | The MAC 800 ECG acquisition, analysis and recording system<br>can print and display multiple leads of ECG data. The MAC 800<br>will provide, in resting ECG mode, ECG quality information<br>using the hookup advisor. The hookup advisor advises users of<br>poor lead quality based on noise measurement. It can be upgraded<br>to provide options such as ECG measurement and interpretation<br>with 12SL. Transmission and reception of ECG data to and from<br>a central ECG cardiovascular information system is also optional.<br>Multiple QT correction formulas including Bazett, Framingham<br>and Fridericia will be available as a user selectable option.<br>Clinical Trials Data Guard and audit trail options are also | | |
FR Page 1 OF 3
{1}------------------------------------------------
## GE Healthcare 510(K) Premarket Notification Submission MAC 800 Resting ECG Analysis System
The MAC 800 delivers multiple leads of ECG on full-size reports and includes an SMS/text message telephone keypad for patient demographics and other data entry with T9 input method, an integrated 7" color display, and an integrated thermal writer. The thermal writer will print real time continuous waveform, alphanumeric data and non real time reports. The device can print the resting ECG report via the external laser printer. The device will have optional internal memory and removable storage to store resting ECG records. The device also can export the resting ECG record to PDF file on SD card as an optional function. An optional barcode reader and magnetic card scanner to enter patient information is available. The MAC 800 can be used as a portable unit.
The MAC 800 is a portable ECG acquisition, analysis and recording Intended Use: system.
> The MAC 800 is intended to acquire, analyze, display and record information from adult and pediatric populations. Pediatric population is defined as patients between the ages of 0 and 15 years.
> The MAC 800 is intended to be used by trained operators in a hospital or medical professional's facility environment to record ECG signals from surface electrodes.
The basic system shall provide 2 modes of operation: (1) Resting ECG mode and (2) Arrhythmia mode.
The basic systems shall print 3, 6-leads of ECG. The device shall be upgradeable to provide software options such as 12-lead ECG measurement and interpretive analysis.
Transmission and reception of ECG data to and from a central ECG cardiovascular information system shall be optional.
The arrhythmia detection portion of the MAC 800 is provided to the customer for the convenience of automatic documentation.
The MAC 800 is used under the direct supervision of a licensed healthcare practitioner.
Contraindications:
The MAC 800 is not designed to provide alarms for arrhythmia detection.
The device is not suitable for intra cardiac application.
It is not intended for use:
- · As a vital signs physiological monitor; or
- For use during patient transport.
E Page 2 of 3
{2}------------------------------------------------
# GE Healthcare
## 510(K) Premarket Notification Submission MAC 800 Resting ECG Analysis System
## Technology:
The MAC 800 Resting ECG Analysis System employs the same functional scientific technology as its predicate devices.
#### Determination of Summary of Non-Clinical Tests:
Substantial Equivalence: The MAC 800 Resting ECG Analysis System and its applications comply with voluntary standards as detailed in Section 9, 11 and 17 of this premarket submission. The following quality assurance
measures were applied to the development of the system:
- Risk Analysis .
- Requirements Reviews .
- Design Reviews
- Testing on unit level (Module verification)
- Integration testing (System verification)
- Final acceptance testing (Validation)
- Performance testing (Verification)
- Safety testing (Verification)
#### Summary of Clinical Tests:
The subject of this premarket submission, MAC 800 Resting ECG Analysis System, did not require clinical studies to support substantial equivalence.
Conclusion:
GE Healthcare considers the MAC 800 Resting ECG Analysis System to be as safe, as effective, and performance is substantially equivalent to the predicate device.
傍
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features an eagle-like symbol with three stylized lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the emblem.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# FEB 1 9 2009
GE Medical Systems Information Technologies c/o Mr. Larry Lepley Regulatory Affairs - Diagnostic Cardiology 9900 Innovation Drive Wauwatosa, WI 53226
Re: K090212 -
Trade Name: MAC 800 Resting ECG Analysis System Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS and DQK Dated: January 26, 2009 Received: January 28, 2009
#### Dear Mr. Lepley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controll provisions plathe Age melude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
### Page 2 - Mr. Larry Lepley
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notifications" (21CFR Part 807.97) For questions regarding posimarket surveillance, please contact CDRH s Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
87 Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for I levices and Radiological Health
Enclosure
{5}------------------------------------------------
# GE Healthcare
510(K) Premarket Notification Submission MAC 800 Resting ECG Analysis System
510(k) Number (if known)
510(k) Number (if known): `K090312
Device Name:
MAC 800 Resting ECG Analysis System
Indications for Use:
The MAC 800 is a portable ECG acquisition, analysis and recording system. The MAC 800 is intended to acquire, analyze, display and record information from adult and pediatric populations. Pediatric population is defined as patients between the ages of 0 and 15 years.
The MAC 800 is intended to be used by trained operators in a hospital or medical professional's facility environment to record ECG signals from surface electrodes.
The basic system shall provide 2 modes of operation: {1} Resting ECG mode and (2) Arrhythmia mode.
The basic systems shall print 3, 6-leads of ECG. The device shall be upgradeable to provide software options such as 12-lead ECG measurement and interpretive analysis.
Transmission and reception of ECG data to and from a central ECG cardiovascular information system shall be optional.
The arrhythmia detection portion of the MAC 800 is provided to the customer for the convenience of automatic documentation.
The MAC 800 is used under the direct supervision of a licensed healthcare practitioner.
#### Contraindications:
The MAC 800 is not designed to provide alarms for arrhythmia detection. The device is not suitable for intra cardiac application. It is not intended for use:
· As a vital signs physiological monitor; or
For use during patient transport.
Prescription Use X AND/OR .. Over-The-Counter Use (Port 21 CFR 801 Submart D) (Part 21:CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) for Zuckerman
2/20/09
Division of Cardiovascular Devices
510(k) Number K090212
ાર
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.