MODIFICATION TO: MAC 5500 RESTING ECG ANALYSIS SYSTEM
K073625 · Ge Healthcare Systems Information Technologies · DPS · Jan 25, 2008 · Cardiovascular
Device Facts
Record ID
K073625
Device Name
MODIFICATION TO: MAC 5500 RESTING ECG ANALYSIS SYSTEM
Applicant
Ge Healthcare Systems Information Technologies
Product Code
DPS · Cardiovascular
Decision Date
Jan 25, 2008
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2340
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The MAC 5500 ECG Analysis System is intended to acquire, analyze, display, and record electrocardiographic information adult and pediatric populations. Basic systems deliver 3, 6, 12, or 15 lead ECG's, interpretive analysis, vector loops, and can be upgraded to provide software analysis options such as high resolution signal averaging of QRS and P wave portions of the electrocardiogram. Transmission and reception of ECG data to and from a central ECG cardiovascular information system is optional. The MAC 5500 is intended to be used under the direct supervision of a licensed healthcare practitioner, by trained operators in a hospital or medical professional's facility.
Device Story
MAC 5500 ECG Analysis System acquires, analyzes, displays, and records surface ECG signals via CAM14 patient data acquisition module. System provides 3, 6, 12, or 15 lead ECGs, interpretive analysis, and vector loops; optional upgrades include high-resolution signal averaging of QRS and P wave portions. Device features alphanumeric keyboard, VGA display, integrated thermal writer, and removable storage. Used in hospitals or medical facilities by trained operators under physician supervision. Mobile unit can be trolley-mounted or used as desktop. Optional connectivity allows transmission/reception of ECG data to/from central cardiovascular information systems. Output assists clinicians in cardiac assessment.
Clinical Evidence
No clinical data provided. Substantial equivalence based on functional technology and performance measurements compared to predicate device.
Technological Characteristics
System components: processing unit and CAM14 patient acquisition module. Features: 3, 6, 12, 15 lead ECG acquisition, interpretive analysis, vector loops, high-resolution signal averaging. Connectivity: optional network link to central cardiovascular information system. Power: rechargeable battery or AC. Form factor: mobile trolley or desktop. Software: interpretive analysis and signal averaging algorithms.
Indications for Use
Indicated for acquisition, analysis, display, and recording of electrocardiographic information in adult and pediatric populations. Intended for use by trained operators under direct supervision of licensed healthcare practitioners in hospitals or medical facilities.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for GE Healthcare. The logo consists of the GE monogram on the left and the words "GE Healthcare" on the right. The GE monogram is a circular design with the letters "GE" intertwined.
December 21, 2007
JAN 25 2008
| 2. | 510(k) Summary | | | |
|----|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|-----|
| | Date: | 12/17/2007 | | |
| | Submitter: | GE Medical Systems Information Technologies 9900 Innovation Drive<br>Wauwatosa, WI 53226 USA | | |
| | Primary Contact: | Margaret Mucha<br>Regulatory Affairs Leader<br>GE Medical Systems Information Technologies<br>Phone: 414-416-2716<br>Fax: 414-721-3863 | | |
| | Secondary Contact: | Matthias Buerger<br>Global Director QA/RA,<br>Diagnostic Cardiology<br>GE Medical Systems Information Technologies<br>Phone: 414-416-2317<br>Fax: 414-721-3863 | | |
| | Device: | Trade Name: MAC 5500 ECG Analysis System | | |
| | Common/Usual Name: | Electrocardiograph | | |
| | Classification Names1: | 21 CFR 870.2340 | Electrocardiograph | DPS |
| | | 21 CFR 870.1425 | Programmable Diagnostic Computer | DQK |
| | | 21 CFR 870.2920 | Transmitters and Receivers,<br>Electrocardiograph, Telephone | DHX |
| | Predicate Device: | K042177 MAC 5500 ECG Analysis System | | |
| | Device Description: | The MAC 5500 ECG Analysis System is designed to acquire, analyze,<br>display, and record ECG signals from surface ECG electrodes. The<br>device consists of two basic components: the processing unit and the<br>patient acquisition module. Models provide rechargeable battery<br>operation and/or optional trolley for transporting the equipment. | | |
The Classification Names/Numbers and Code assigned during the clearance of the MAC 5500 ECC Analysis System for K042777 was MHX - Detector Arrhythmia Detector and Alarm (including ST-segment measurement and alarm) (870.1025). GE Medical Systems Information Technologies formally requested on 12/14/07 that the MAC 5500 ECG be reclassified as DPS – Electrocardiograph (870.2340) as this code more accurately reflects the intended use of this device as this device is not intended to detect an arrhythmia and alarm or produce an audible signal when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation ocurs, which is the Identification statement contained in Code MHX (Section 870.1025). The intended use more closely matches the Identification statement for Section 870.2340 Electrocardiograph, which states that an Electrocarding a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart. This submission is reflective of this requested code change to DPS for the MAC 5500.
{1}------------------------------------------------
Page 20 of 727
Image /page/1/Picture/1 description: The image shows the GE Healthcare logo. The logo consists of the GE monogram in a circle on the left, followed by the words "GE Healthcare" in a simple, sans-serif font. The text is aligned horizontally with the monogram.
December 21, 2007
Device Description: (continued) The MAC 5500 can deliver 3, 6, 12, or 15 lead ECG's, interpretive analysis, vector loops, and can be upgraded to provide software analysis options such as high resolution signal averaging of QRS and P wave portions of the electrocardiogram. Transmission and reception of ECG data to and from a central ECG cardiovascular information system is optional.
The MAC 5500 system acquires ECG data using the CAM14 patient data acquisition module. By placing the data acquisition device closer to the patient, signal fidelity is improved and noise is reduced. MAC 5500 delivers 12 or 15 lead ECG's on full-size reports. The MAC 5500 incorporates a alphanumeric keyboard for patient demographics and other data entry, a full size VGA graphics and waveform display, an integrated thermal writer and removable data storage.
Additionally, the MAC 5500 utilizes battery power for customer convenience and can transmit and receive ECGs to and from a central ECG cardiovascular information system via optional communication links. The system is intended as a mobile device but the main unit can be separated from the trolley and used as a desktop unit.
## Indications for Use
The MAC 5500 ECG Analysis System is intended to acquire, analyze, display, and record electrocardiographic information adult and pediatric populations. Basic systems deliver 3, 6, 12, or 15 lead ECG's, interpretive analysis, vector loops, and can be upgraded to provide software analysis options such as high resolution signal averaging of QRS and P wave portions of the electrocardiogram. Transmission and reception of ECG data to and from a central ECG cardiovascular information system is optional.
The MAC 5500 is intended to be used under the direct supervision of a licensed healthcare practitioner, by trained operators in a hospital or medical professional's facility.
Technology The MAC 5500 ECG Analysis System employs the same functional technology as the predicate devices.
Conclusion: The results of these measurements demonstrated that the MAC 5500 ECG Analysis System is as safe, as effective, and performs as well as the predicate devices.
{2}------------------------------------------------
Public Health Service
Image /page/2/Picture/2 description: The image is a black and white seal for the Department of Health & Human Services, USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, which is a common symbol of the United States.
JAN 25 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
GE Medical Systems Information Technologies c/o Ms. Margaret Mucha Regulatory Affairs Leader 9900 Innovation Drive Wauwatosa, WI 53226
Re: K073625
Trade/Device Name: MAC 5500 Resting ECG Analysis System Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II (two) Product Code: DPS, DQK, DHX Dated: December 21, 2007 Received: December 26, 2007
Dear Ms. Mucha:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 – Ms. Margaret Mucha
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Bfsimima fr
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the GE Healthcare logo. The logo consists of the GE monogram in a circle on the left, followed by the words "GE Healthcare" in a simple, sans-serif font. The text is aligned horizontally with the monogram.
December 21, 2007
3. Intended Use
510(k) Number (if known):
Page 1 of 1
Device Name:
MAC 5500 ECG Analysis System
Intended Use:
The MAC 5500 ECG Analysis System is intended to acquire, analyze, display, and record electrocardiographic information adult and pediatric populations. Basic systems deliver 3, 6, 12, or 15 lead ECG's, interpretive analysis, vector loops, and can be upgraded to provide software analysis options such as high resolution signal averaging of QRS and P wave portions of the electrocardiogram. Transmission and reception of ECG data to and from a central ECG cardiovascular information system is optional.
The MAC 5500 is intended to be used under the direct supervision of a licensed healthcare practitioner, by trained operators in a hospital or medical professional's facility.
Prescription Use X (Per 21 CFR 801.109 Subpart D) OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Brammer
Division of Cardiovascular Devices
510(k) Number K073625
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.