K123900 · Philips Medical Systems · MHX · Apr 18, 2013 · Cardiovascular
Device Facts
Record ID
K123900
Device Name
SURESIGNS VM4, SURESIGNS VM6, SURESIGNS VM8
Applicant
Philips Medical Systems
Product Code
MHX · Cardiovascular
Decision Date
Apr 18, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The SureSigns VM4, VM6 and VM8 Patient Monitors are for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments. Additionally, the monitor is intended for use in transport situations within a healthcare facility
Device Story
SureSigns VM4, VM6, and VM8 are multi-parameter patient monitors used in healthcare facilities for bedside and transport monitoring. Devices accept inputs from sensors for ECG, respiration, NBP, SpO2, IBP, CO2, and temperature. Monitors process these signals to display real-time waveforms and numerical values, perform arrhythmia analysis, and trigger alarms based on user-defined limits. Healthcare professionals operate the devices to observe patient status and make clinical decisions. Modifications include addition of arrhythmia analysis to VM4, LED display backlighting, RoHS-compliant CO2 module, increased trend database (120 hours), and software hooks for future central station connectivity. Output is visualized on the monitor screen and can be printed; device provides continuous physiological surveillance to assist in patient management.
Clinical Evidence
Bench testing only. Verification and validation activities included system-level tests, performance tests, and safety testing based on hazard analysis. Pass/Fail criteria were based on specifications cleared for the predicate devices.
Technological Characteristics
Multi-parameter patient monitor. Components: LED display backlight (mercury-free), RoHS-compliant microMedi OEM CO2 module. Connectivity: Software hooks for future central station integration. Arrhythmia analysis software. Parameters: ECG, Respiration, NBP, SpO2, IBP, CO2, Temperature.
Indications for Use
Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients (adult, pediatric, neonatal). Parameters include ECG, respiration, NBP, SpO2, IBP, CO2, and temperature.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
## K123900 pg 1 of 3
#### 510K Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807-92(c).
- 1. The submitter of this pre-market notification is:
Mary Kruitwagen Philips Medical Systems 3000 Minuteman Road Andover, MA 01810 United States
Tel: 978-659-4932 Fax: 978-659-7323 Email: mary.kruitwagen@philips.com
This summary was prepared on December 17, 2012.
- 2. The names of the subject devices are the Philips SureSigns Series Patient Monitors, SureSigns VM4, VM6, and VM8 Patient Monitors
- 3. The trade names of the devices are the SureSigns VM4, SureSigns VM6, and SureSigns VM8 Patient Monitors.
- 4. The common usual name is multi-parameter patient monitor
- 5. The Classification names are as follows:
| Device Panel | Classification | ProCode | Description | Applicable<br>Subject<br>Devices |
|-----------------------------------------|----------------|---------|--------------------------------------------------------------------------|----------------------------------|
| Circulatory System<br>Devices | 870.1025, II | MHX | Monitor, Physiological, Patient (with<br>arrhythmia detection or alarms) | VM4, VM6,<br>VM8 |
| | 870.1110, II | DSJ | Alarm, Blood Pressure | VM4, VM6,<br>VM8 |
| | 870.1110, II | DSK | Computer, Blood Pressure | VM4, VM6,<br>VM8 |
| | 870.1130, II | DXN | System, Measurement, Blood<br>Pressure, Non-Invasive | VM4, VM6,<br>VM8 |
| | 870.2300, II | DRT | Monitor, Cardiac (incl.<br>Cardiotachometer & Rate Alarm | VM4, VM6,<br>VM8 |
| | 870.2700, II | DQA | Oximeter | VM4, VM6,<br>VM8 |
| | 870.2900, I | DSA | Cable, Transducer and Electrode,<br>incl. Patient connector | VM4, VM6,<br>VM8 |
| General Hospital<br>and Personal Use | 880.2910, II | FLL | Thermometer, Electronic, Clinical | VM4 |
| Anesthesiology &<br>Respiratory Therapy | 868.1400, II | CCK | Analyzer, Gas, | VM8. |
6. The modified devices are substantially equivalent to previously cleared Philips device, SureSigns VM4, VM6 and VM8 Patient Monitors marketed pursuant to K052707, K080495, K090483, and K101067. The predicate 510Ks for this submission are: K103652 and K094012 (May 12 2010) by Oridion Capnography Inc.
{1}------------------------------------------------
7. The modifications are as follows:
- 1. Add arrhythmia analysis to the VM4, using the same arrhythmia software algorithm as in the VM6 and VM8\
- Add standby mode into CO2 menu to preserve "time to calibration' for situations 2. where the monitor is in use, but not CO2 monitoring.
- 3. The LCD display backlight has changed to an LED backlight. The new backlight does not contain mercury.
- Replace the current Oridion CO2 module with a RoHS compliant Oridion CO2 module. 4. This module, also referred to as the microMedi OEM board, is included in the Oridion Capnostream 20. and is cleared under K094012.
- 5. Enhancement requests include:
- Display the word 'Filter' in the ECG trace pane when the filter bandwidth setting . is enabled. Filter will also be printed on strip recordings
- Add a choice to enable a SpO2 high/low and/or SpO2 Desaturation alarm delay . time. When enabled, the SpO2 alarm delay will result in the SpO2 alarm generation once the delay time has elapsed.
- The temperature label, previously called temp, is now called T1. ●
- Add a chime tone when a manually initiated NBP measurement is completed. .
- The patient demographic information may be edited through an additional access ● point of the title bar.
- Separate Heart Rate (HR) and Pulse (Pulse) into distinct viewing panes with . separate alarm limits. No other changes were made.
- . Increase the trend database to allow for 120 hours of data to be stored
- Add software hooks that will allow for connection to the Philips central station (trade 6. name yet to be determined) once the central stations is developed, cleared and released. The following information is sent to the central:
- Patient ID information ●
- Patient type (adult, pediatric, neonatal)
- Monitor name
- Serial number of the connected monitor
- Waveforms
- Measurements
- Alarm limits
- Alarms ●
- ECG lead setting .
- Pace detection setting .
- NBP interval mode setting
- Demo mode notification
- . Working mode notification (whether monitor is in monitor mode, spot check mode and whether CO2 is in standby, if applicable)
- . Audio off/audio paused
- Waveform information including: Arrhythmia off, pacer pulse detection off, cannot analyze ECG, ECG (wave) freeze, IBP zero, Filter on.
- The subject devices have the same intended Use and Indications for Use as the legally 8. marketed predicate device:
"The SureSigns VM4, VM6 and VM8 Patient Monitors are for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics, and neonates in
{2}------------------------------------------------
# K123900 pg 3 of 3
healthcare environments. Additionally, the monitor is intended for use in transport situations within a healthcare facility",
Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
Standard and optional parameters include:
- · ECG
· Respiration
• NBP
• SpOz
• IBP
· CO2
- · Temperature
- ல் The subject devices have the same fundamental technological characteristics as the legally marketed predicate devices. The subject devices use the same design as the predicate devices.
i
- 10. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the subject devices with respect to the predicates. Testing involved system level tests, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device, the specifications of the subject device and test results showed substantial equivalence. The results demonstrate that the Philips SureSigns VM4, VM6 and VM8 Patient Monitors meet all reliability requirements and performance claims and supports a determination of substantial equivalence.
#### Page 113 of 175
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized caduceus or a bird-like figure with flowing lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
April 18, 2013
Philips Medical Systems North America, Co. c/o: Mary Kruitwagen Regulatory Engineer 3000 Minuteman Road . Andover, MA 01810
Re: K123900
Trade Name: SureSigns VM4, SureSignsVM6, and SureSigns VM8 Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm) Regulatory Class: Class II Product Code: MHX, DSJ, DSK, DXN, DRT, DOA, DSA, FLL, CCK Dated: March 12, 2013 Received: March 19, 2013
Dear Ms. Mary Kruitwagen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{4}------------------------------------------------
### Page 2 - Ms. Mary Kruitwagen
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Owen P. Faris -S
for
Bram D. Zuckerman, M.D. Director Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K123900
## Indications for Use
510 (k) Number (if known):
Device Name: SureSigns VM4 (reference number: 863063) SureSigns VM6 (reference numbers: 863064, 863065) SureSigns VM8 (reference numbers: 863066, 863068)
#### Indications for Use:
Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
Standard and optional parameters include:
- · ECG
- · Respiration
- NBP
- SpO2
- IBP
- · CO2
- · Temperature
Prescription Use: YES AND/OR over-the-counter Use: _________________________________________________________________________________________________________________________________________________ NO (Part 21 CFFR 801 Subpart D) (21 CFR 807 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE)
Owen P. Faris -S
2013.04.18
15:59:41 -04'00'
Page of of of
Page 27 of 175
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.