SURESIGNS VM SERIES PATIENT MONITORS, SURESIGNS VM4, SURESIGNS VM6, AND SURESIGNS VM8
K103652 · Philips Medical Systesms · MHX · Jan 13, 2011 · Cardiovascular
Device Facts
Record ID
K103652
Device Name
SURESIGNS VM SERIES PATIENT MONITORS, SURESIGNS VM4, SURESIGNS VM6, AND SURESIGNS VM8
Applicant
Philips Medical Systesms
Product Code
MHX · Cardiovascular
Decision Date
Jan 13, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The SureSigns VM Series Patient Monitors are for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments. Additionally, the monitor is intended for use in transport situations within a healthcare facility.
Device Story
SureSigns VM4, VM6, and VM8 are multi-parameter patient monitors used in healthcare environments and transport. Devices accept inputs from sensors for ECG, respiration, NBP, SpO2, IBP, CO2, and temperature. Monitors process these signals to provide real-time physiological data, alarms, and recordings. Operated by healthcare professionals to monitor adults, pediatrics, and neonates. Output displayed on-screen for clinical review; enables continuous patient assessment and timely intervention. Modifications include updated alarm limit ranges for heart rate sourced from ECG, SpO2, or NBP to allow source-specific thresholds.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, including system-level performance tests and safety testing based on hazard analysis. Pass/Fail criteria were based on specifications cleared for the predicate devices.
Technological Characteristics
Multi-parameter patient monitor. Parameters: ECG, respiration, NBP, SpO2, IBP, CO2, temperature. Connectivity: Patient sensors/cables. Software-based alarm management. Class II device. Design is identical to predicate series.
Indications for Use
Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients, including adults, pediatrics, and neonates. Parameters include ECG, respiration, NBP, SpO2, IBP, CO2, and temperature.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
JUHAN 1. 3 2011
## 510K Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807-92(c).
- The submitter of this pre-market notification is: 1.
Larry Milana Philips Medical Systems 3000 Minuteman Road Andover, MA 01810 United States
Tel: 978-659-7860 Fax: 978-659-3819 Email: larry.milana@philips.com
This summary was prepared on December 10, 2010.
- The names of the subject devices are the Philips SureSigns Series Patient Monitors, 2. SureSigns VM4, VM6, and VM8 Patient Monitors
- 3. The trade names of the devices are the SureSigns VM4, SureSigns VM6, SureSigns VM8 Patient Monitors.
- 4. The common usual name is multi-parameter patient monitor
- 5. The Classification names are as follows:
| Device Panel | Classification | ProCode | Description | Applicable<br>Subject<br>Devices |
|-----------------------------------------|----------------|---------|-----------------------------------------------------------------------------|----------------------------------|
| Circulatory System<br>Devices | 870.1025, II | MHX | Monitor, Physiological,<br>Patient (with arrhythmia<br>detection or alarms) | VM4, VM6,<br>VM8 |
| | 870.1110, II | DSJ | Alarm, Blood Pressure | VM4, VM6,<br>VM8 |
| | 870.1110, II | DSK | Computer, Blood Pressure | VM4, VM6,<br>VM8 |
| | 870.1130, II | DXN | System, Measurement,<br>Blood Pressure, Non-<br>Invasive | VM4, VM6,<br>VM8 |
| | 870.2300, II | DRT | Monitor, Cardiac (incl.<br>Cardiotachometer & Rate<br>Alarm | VM4, VM6,<br>VM8 |
| | 870.2700, II | DQA | Oximeter | VM4, VM6,<br>VM8 |
| | 870.2900, I | DSA | Cable, Transducer and<br>Electrode, incl. Patient<br>connector | VM4, VM6,<br>VM8 |
| General Hospital and<br>Personal Use | 880.2910, II | FLL | Thermometer, Electronic,<br>Clinical | VM4 |
| Anesthesiology &<br>Respiratory Therapy | 868.1400, II | CCK | Analyzer, Gas, | VM8 |
December 10, 2010
{1}------------------------------------------------
- 6. The modified devices are substantially equivalent to previously cleared Philips device, SureSigns VM Series Patient Monitors marketed pursuant to K052707, K080495, K090483, and K101067.
- 7. The modifications are as follows:
- . The alarm limit range for the Heart Rate (HR) sourced from the ECG, the SP02, or the NBP measurement has been modified to allow for separate ranges for each source. This modification allows the device to offer a wider heart rate (pulse) measurement range for the ECG and SP02 sources while keeping the existing NBP range.
- . Modification of the current Indications for Use to include a list of the measurements the device can perform. The Indications for Use will now read as follows:
Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
Standard and optional parameters include:
·ECG ·Respiration ·NBP ·SpO2 ·IBP ·CO2 ·Temperature
- 8. The subject devices have the same Intended Use as the legally marketed predicate device :
"The SureSigns VM Series Patient Monitors are for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environments. Additionally, the monitor is intended for use in transport situations within a healthcare facility".
- ഗ് The subject devices have the same fundamental technological characteristics as the legally marketed predicate devices. The subject devices use the same design as the predicate devices. The change to heart rate does not change the fundamental characteristics.
- 10. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the subject devices with respect to the predicates. Testing involved system level tests, performance tests, and safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device, the specifications of the subject device and test results showed substantial equivalence. The results demonstrate that the Philips SureSigns VM4, VM6 and VM8 Patient Monitors meet all reliability requirements and performance claims and supports a determination of substantial equivalence.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JAN 1 3 2011
Philips Medical Systems c/o Mr. Lawrence Milana Regulatory Engineer 3000 Minuteman Road Andover, MA 01810
Re: K103652
Trade/Device Name: SureSigns VM4, SureSigns VM6, SureSigns VM8 Regulation Number: 21 CFR 870.1025 Regulation Name: Monitor, Physiological, Patient (with arrhythmia detection or alarm) Regulatory Class: Class II Product Code: MHX Dated: December 10, 2010 Received: December 14, 2010
Dear Mr. Milana:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
.
,
・・・・・・
:
{3}------------------------------------------------
Page 2 - Mr. Lawrence Milana
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
W.M.d.
3 Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
103652
## Indications for Use
JAN 1 3 2011
510 (k) Number (if known): K103652
Device Name: SureSigns VM4 (reference number: 863063) SureSigns VM6 (reference numbers: 863064, 863065) SureSigns VM8 (reference numbers: 863066, 863068)
## Indications for Use:
Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
Standard and optional parameters include:
·ECG
- ·Respiration
- •NBP
- ·SpO2
- •IBP
- ·CO2
- ·Temperature
Prescription Use: YES AND/OR over-the-counter Use: NO (Part 21 CFFFR 801 Subpart D) (21 CFR 807 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CORH, Office of Device Evaluation (ODE)
W.M.
ivision Sign-Off) Division of Cardiovascular Devices
Page 1_of 1
510(k) Number
December 10, 2010
Page 22 of 68
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.