The ERMIS Sterilization Container System Is intended to allow sterilization of the enclosed medical device and also maintain sterility during transport and storage for 30 days. The system consists of different models, such as 1/1 Size, ¾ Size, ½ size, mini and dental containers which may vary in size, perforations and color. All models are available with perforated lids and either perforated or non-perforated bottoms. To enable proper organization of sterilized goods, different wire- or sleeve baskets exactly adapted to the specific container dimensions are available. All models are intended to be used with Ermis single use paper filters. The containers are reusable devices designed to be used with the following sterilization cycle parameters: Pre Vacuum cycle: 4 minutes 132°C (270°F) Drying Time minimum 20 minutes 3 minutes 135°C (273°F) Drying Time minimum 16 minutes Loading: Metal surgical instruments (scissors, clamps, forceps) and textiles
Device Story
Reusable sterilization container system; constructed of anodized aluminum alloy, stainless steel, and silicone. Designed to house metal surgical instruments and textiles during steam sterilization cycles (pre-vacuum). Features include perforated/non-perforated bottoms, perforated lids with filter systems, and optional safety lids for transport. Uses disposable paper filters to maintain microbial barrier for 30 days post-sterilization. Used in clinical environments to organize, sterilize, and store surgical goods. Healthcare providers load instruments into baskets within the container; the system is then processed in a steam sterilizer. The container protects contents from contamination during storage and transport, ensuring sterility until the point of use.
Clinical Evidence
Bench testing only. Validation performed for sterilant penetration/efficacy, maintenance of sterility, microbial barrier properties, and use life. Compliance with ISO 10993-1 for biocompatibility of indirect contact components confirmed.
Indicated for sterilization and maintenance of sterility for metal surgical instruments and textiles in pre-vacuum steam sterilizers. Suitable for use in clinical settings.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
SklarLite Rigid Sterilization Container Systems (K091720)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Ermis Medizintechnik. The logo consists of a stylized geometric shape on the left, followed by the word "Ermis" in a bold, sans-serif font. Below "Ermis" is the word "MEDIZINTECHNIK" in a smaller font size. The logo is set against a dark background, with the text and geometric shape appearing in white.
# K 123234
JUL 5 2013
# SECTION 05
# 510(k) Summary
DATE OF APPLICATION: 2013-06-21
APPLICANT:
Ermis Medizintechnik Dornierstr. 49 78532 Tuttlingen Germany Tel.: +49 7461 964857-0 Fax: +49 7461 964857-99
Establishment Registration Number: 3005071827
E-Mail: info@ermis-medizintechnik.de www.ermis-medizintechnik.de
# CONTACT PERSON: Mr. Mehmet Ermis
Managing Director Tel.: +49 7461 964857-0 Fax: +49 7461 964857-99
001_510K_Summary_v1.5.doc
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows a logo for Ermis Medizintechnik. The logo consists of a stylized graphic to the left of the word "Ermis" in a bold, sans-serif font. Below "Ermis" is the word "MEDIZINTECHNIK" in a smaller font size. The logo is black and white.
#### Device Name 1.
Ermis Medizintechnik Sterilization Container System consisting of:
- 1/1 Size, ¼ Size, ½ Size, Mini, Dental containers and Accessories
- · Sterilization Container Systems Trade Names:
Sterilization Container Common Name:
Sterilization Wrap Containers, Trays, Cassettes & Other Accessories Classification Name:
#### Classification Product Code / Subsequent Code 2.
| Device | Medical<br>Specialty | Review Panel | Product Code | Device Class | Regulation Number |
|------------------------------------------------------------------------------|----------------------|------------------|--------------|--------------|-------------------|
| Sterilization Wrap<br>Containers, Trays,<br>Cassettes & Other<br>Accessories | Part 880 | General Hospital | KCT | 2 | 880.6850 |
#### 3. Predicate Device
Ermis Medizintechnik Sterilization Container Systems are substantially equivalent to the following predicate devices, most recently cleared by the FDA:
| Predicate Device | 510(k) Number | 510(k) Holder |
|----------------------------------------------------|---------------|-------------------|
| Bahadir Sterilization Trays | K112190 | Bahadir USA Corp. |
| SklarLite Rigid Sterilization Container<br>Systems | K091720 | Sklar Instruments |
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for Ermis Medical Technology. The logo consists of a stylized building icon on the left, followed by the word "Ermis" in a bold, sans-serif font. Below "Ermis" are the words "MEDICAL TECHNOLOGY" in a smaller font. The background of the logo is black, and the text and icon are white.
#### Description of the Device 4.
Ermis Medizintechnik Sterilization Container Systems are reusable devices intended to allow sterilization of enclosed medical devices and to maintain sterility of the enclosed devices until used.
- Interchangeable components within the different sizes (Full size, ¼ size, ½ size, Wide Body, Den-. tal, Mini, Flat).
- Containers are available with perforated and non-perforated bottom and perforated lids. .
- Safety lids are available for 1/1 size, 3/4 size and 1/2 size. Safety lids protect filters during storage . and/or transportation. The Lids are not intended to be used during sterilization.
- The lids are equipped with a silicone based gasket. .
- All containers incorporate a filter system to be used with Ermis disposable paper filters without . chemical indicator.
- To ensure effective sterilization and organization of sterilized goods, wire and screen baskets are . available.
### Available sizes:
1/1 size:
580 x 280 x 100 mm . 580 x 280 x 135 mm
580 x 280 x 150 mm
580 x 280 x 200 mm
580 x 280 x 260 mm
3/4 size:
- 465 x 280 x 100 mm .
465 x 280 x 135 mm
- 465 x 280 x 150 mm
1/2 size:
- 285 x 280 x 100 mm .
- 285 x 280 x 135 mm
- 285 x 280 x 150 mm
- 285 x 280 x 200 mm
- 285 x 280 x 260 mm
Mini Container:
- 300 x 140 x 40 mm ◆
- 300 x 140 x 70 mm
300 x 140 x 100 mm .
Dental Container.
| | | | 300 x 190 x 40 mm | |
|--|--|--|-------------------|--|
| | | | | |
| | | | 300 x 190 x 65 mm | |
| | | | | |
- 300 x 190 x 80 mm
- 300 x 190 x 100 mm
- 300 x 190 x 130 mm
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Ermis Medizintechnik. The logo consists of a geometric shape on the left, followed by the word "Ermis" in a bold, sans-serif font. Below "Ermis" is the word "MEDIZINTECHNIK" in a smaller font. The logo is black and white.
#### 5. Indications for Use
The ERMIS Sterilization Container System is intended to allow sterilization of the enclosed medical device and also maintain sterility during transport and storage for 30 days. The system consists of different models, such as 1/1 Size, ¾ Size, ½ size, mini and dental containers which may vary in size, perforations and color. All models are available with perforated lids and either perforated or non-perforated bottoms. To enable proper organization of sterilized goods, different wire- or sleeve baskets exactly adapted to the specific container dimensions are available. All models are intended to be used with Ermis single use paper filters.
The containers are reusable devices designed to be used with the following sterilization cycle parameters:
Pre Vacuum cycle: 4 minutes 132°C (270°F) Drying Time minimum 20 minutes
3 minutes 135°C (273°F) Drying Time minimum 16 minutes
Loading: Metal surgical instruments (scissors, clamps, forceps) and textiles
### Maximum recommended loading:
| | | Maximum Recommended<br>Load in kg (lbs.) incl. Container weight | |
|----------------------------------|-------------|-----------------------------------------------------------------|------------|
| Model | Dimension | Instruments | Textiles |
| | 580x280x100 | 3.8 (8.3) | 3 (6.6) |
| 1/1 (Full-)<br>Size<br>Container | 580x280x135 | 5.2 (11.4) | 4.1 (9.0) |
| | 580x280x150 | 5.8 (12.7) | 4.6 (10.1) |
| | 580x280x200 | 7.7 (16.9) | 6.1 (13.4) |
| | 580x280x260 | 10 (22.0) | 8 (17.6) |
| | 465x280x100 | 3.1 (6.8) | 2.5 (5.5) |
| ¾ Size<br>Container | 465x280x135 | 4.2 (9.2) | 3.3 (7.2) |
| | 465x280x150 | 4.6 (10.1) | 3.7 (8.1) |
| | 285x280x100 | 1.9 (4.1) | 1.5 (3.3) |
| | 285x280x135 | 2.6 (5.7) | 2 (4.4) |
| ½ Size<br>Container | 285x280x150 | 2.8 (6.1) | 2.2 (4.8) |
| | 285x280x200 | 3.8 (8.3) | 3 (6.6) |
| | 285x280x260 | 4.9 (10.8) | 3.9 (8.5) |
| Mini<br>Container | 300x140x40 | 0.4 (0.8) | 0.3 (0.6) |
| | 300x140x70 | 0.7 (1.5) | 0.5 (1.1) |
| | 300x140x100 | 1 (2.2) | 0.8 (1.7) |
| Dental<br>Container | 310x190x40 | 0.6 (1.3) | 0.5 (1.1) |
| | 310x190x65 | 0.9 (1.9) | 0.7 (1.5) |
| | 310x190x80 | 1.2 (2.6) | 0.9 (1.9) |
| | 310x190x100 | 1.5 (3.3) | 1.2 (2.6) |
| | 310x190x130 | 1.8 (3.9) | 1.4 (3.0) |
QD1 510K Summary v1 5 don
Page 4 of 6
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a logo for "Ermis Medizintechnik". The logo consists of a geometric shape on the left, followed by the word "Ermis" in a bold, sans-serif font. Below "Ermis" is the word "MEDIZINTECHNIK" in a smaller font. The logo is set against a black background.
#### Technological Characteristics 6.
The Sterilization Containers are made of anodized aluminum alloy. The containers are constructed to be used in any conventional steam sterilizer and can be stacked safely on top of one another during sterilization without slipping out of place.
The ERMIS sterilization container systems allow a systematic organization of the entire instrument sterilization. Storage, transportation and disposal can be organized with the containers.
The filter retainer is locked by pressing down in center of the filter. The opening of the filter retainer succeeds by pressing the plate.
The performance of the subject container was validated for sterilant penetration/efficacy, maintenance of sterility, microbial barrier properties and use life. The container met acceptance criteria for all claimed tests.
| | Ermis System | Bahadir System | Sklar System |
|---------------------------------------------------------------------------------------------------|-----------------------------------------------------------|--------------------------------------------------|-----------------------------------------------------------|
| PROPERTIES | | | |
| Indicated for use containing in-<br>struments to be sterilized in vacu-<br>um steam sterilizers | Yes | Yes | Yes |
| Reusable | Yes | Yes | Yes |
| Closed system | Yes | Yes | Yes |
| Sealed | Yes | Yes | Yes |
| DESIGN | | | |
| Incorporates Filter System to per-<br>mit entry of sterilant | Yes | Yes | Yes |
| Incorporates filter system to pre-<br>vent microbial migration during<br>transport | Yes | Yes | Yes |
| Incorporates security lid to prevent<br>damage / contamination of filter<br>unit during transport | Yes | No | Yes |
| MATERIALS | | | |
| Container | Anodized Aluminium<br>Alloy, Stainless<br>Steel, Silicone | Aluminium Alloy,<br>Stainless Steel,<br>Silicone | Anodized Aluminium<br>Alloy, Stainless<br>Steel, Silicone |
| PERFORMANCE | | | |
| Load | Up to 22 lbs. | Up to 24.5 Ibs. | Up to 25 lbs. for all<br>container sizes |
| Lumened / cannulated instruments | No | Yes | Yes |
| Stainless steel medical devices | Yes | Yes | Yes |
#### Characteristics compared to the Predicate Device 6.1.
001_510K_Summary_v1.5.dac
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a logo for "Ermis Medizintechnik". The logo consists of a stylized graphic on the left, followed by the word "Ermis" in a bold, sans-serif font. Below "Ermis" is the word "MEDIZINTECHNIK" in a smaller font. The logo is black and white.
## 7. Testing
Testing in order to proof safety and effectiveness of Ermis Medizintechnik Sterilization Container Systems has been performed according to the requirements set out in ISO 10993-1.
### 7.1. Performance Testing
In order to demonstrate compliance to Safety and effectiveness requirements that apply to the device, performance testing has been conducted under consideration of the requirements for sterilization performance, reusability, maintenance of sterility and sterilization compatibility.
### 7.2. Biocompatibility
The devices subject to this submission consist of the same materials as the predicate devices and do not come into direct contact with the patient. Evaluation of biological risks resulting from devices or components with indirect patient contact has been performed in accordance to the requirements set out in ISO 10993-1.
#### Substantial Equivalence Summary / Conclusion 8.
Based on available 510(k) information provided herein, Ermis Medizintechnik Sterilization Container Systems are considered substantial equivalent to the predicate devices in terms of indications for use, material, technology, design and performance specifications.
There are no differences between the devices which would raise new issues of safety or effectiveness.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
### July 5, 2013
Ermis Medizintechnik C/O Mr. Andre Weingerl Regulatory Affairs Consultant Medagent GmbH & Company KG Griesweg 47 Muehlheim Baden-Wuerttmberg GERMANY 78570
Re: K123234
Trade/Device Name: Sterilization Container System Consisting of: 1/1 Size, 1/2 Size. Mini, Dental containers and Accessories Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT
Dated: June 6, 2013 Received: June 10, 2013
### Dear Mr. Weingerl:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
## Page 2 - Mr. Weingerl
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safetv/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Image /page/7/Picture/7 description: The image shows a signature and some text. The text reads "Sincerely yours, Tejashri Purohit-Sheth, M.D., Clinical Deputy Director, DAGRID, FOR". The signature is illegible. The text is likely part of a letter or document.
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Image /page/8/Picture/17 description: The image shows a logo for Ermis Medizintechnik. The logo features a stylized geometric shape on the left, followed by the word "Ermis" in a bold, sans-serif font. Below "Ermis" is the word "MEDIZINTECHNIK" in a smaller font. The logo appears to be in black and white.
# Indications for Use Statement
510(k) Number: K 123234
Device Name:
Ermis Medizintechnik Sterilization Container System consisting of: 1/1 Size, % Size, 1/2 Size, Mini, Dental containers and Accessories
Indications for Use:
The ERMIS Sterilization Container System Is intended to allow sterilization of the enclosed medical device and also maintain sterility during transport and storage for 30 days. The system consists of different models, such as 1/1 Size, ½ Size, ½ size, mini and dental containers which may vary in size, perforations and color. All models are available with perforated lids and either perforated or non-perforated bottoms. To enable proper organization of sterilized goods, different wire- or sleeve baskets exactly adapted to the specific container dimensions are available. All models are intended to be used with Ermis single use paper filters.
The containers are reusable devices designed to be used with the following sterilization cycle parameters:
Pre Vacuum cycle: 4 minutes 132°C (270°F) Drying Time minimum 20 minutes
3 minutes 135°C (273°F) Drying Time minimum 16 minutes
Loading: Metal surgical instruments (scissors, clamps, forceps) and textiles
Prescription Use (Parl 21 CFR 801 Subpart D) ···· AND/OR
Over-The-Counter Use" __ X (21 SFR 801 Subparl C)
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sreekanth Gutala -S
2013.07.02 14:27:40 -04'00
FDA
| Division Sign-Off) |
|--------------------|
|--------------------|
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
| 510(k) Number: | K123234 |
|----------------|---------|
|----------------|---------|
{9}------------------------------------------------
and the program and the comments of the comments of the comments of the comments of the comments of the first of the first of the first of the first of the first of the first
Image /page/9/Picture/1 description: The image shows a logo with the text "Ermis MEDIZINTECHNIK". The logo also includes a stylized graphic to the left of the text. The text "Ermis" is in a large, bold font, while "MEDIZINTECHNIK" is in a smaller font below it. There is a registered trademark symbol to the right of the word "Ermis".
Device overview Ermis Sterilization Container System
・
| | | Maximum Recommended<br>Load in kg (lbs.) incl. contain-<br>er weight | | Available configurations / Ac- |
|----------------------------------|-------------|----------------------------------------------------------------------|------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model | Dimension | Instruments | Textiles | cessories |
| 1/1 (Full-)<br>Size<br>Container | 580x280x100 | 3.8 (8.3) | 3 (6.6) | |
| | 580x280x135 | 5.2 (11.4) | 4.1 (9.0) | |
| | 580x280x150 | 5.8 (12.7) | 4.6 (10.1) | |
| | 580x280x200 | 7.7 (16.9) | 6.1 (13.4) | |
| | 580x280x260 | 10 (22.0) | 8 (17.6) | |
| ¾ Size<br>Container | 465x280x100 | 3.1 (6.8) | 2.5 (5.5) | Perforated Lid;<br>Perforated or non-perforated<br>bottom;<br>Optional safety-Lid for storage /<br>transportation;<br>Accessories: Stainless steel<br>wire- or sleeve baskets;<br>Available colours: Natural alu-<br>minium, anodized gold, red,<br>blue, green, black |
| | 465x280x135 | 4.2 (9.2) | 3.3 (7.2) | |
| | 465x280x150 | 4.6 (10.1) | 3.7 (8.1) | |
| ½ Size<br>Container | 285x280x100 | 1.9 (4.1) | 1.5 (3.3) | |
| | 285x280x135 | 2.6 (5.7) | 2 (4.4) | |
| | 285x280x150 | 2.8 (6.1) | 2.2 (4.8) | |
| | 285x280x200 | 3.8 (8.3) | 3 (6.6) | |
| | 285x280x260 | 4.9 (10.8) | 3.9 (8.5) | |
| Mini<br>Container | 300x140x40 | 0.4 (0.8) | 0.3 (0.6) | |
| | 300x140x70 | 0.7 (1.5) | 0.5 (1.1) | |
| | 300x140x100 | 1 (2.2) | 0.8 (1.7) | |
| Dental<br>Container | 310x190x40 | 0.6 (1.3) | 0.5 (1.1) | |
| | 310x190x65 | 0.9 (1.9) | 0.7 (1.5) | |
| | 310x190x80 | 1.2 (2.6) | 0.9 (1.9) | |
| | 310x190x100 | 1.5 (3.3) | 1.2 (2.6) | |
| | 310x190x130 | 1.8 (3.9) | 1.4 (3.0) | |
·
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.