The SklarLite™ Rigid Sterilization Container System is intended for use in hospitals and health care facilities to contain other medical devices that are to be sterilized. Containers allow sterilization of the enclosed medical devices, including surfaces and lumens, using high vacuum steam sterilizers at 270°F for 4 minutes with 30 minutes (minimum) dry time. The containers have been validated for sterilization of up to two instruments with lumens no smaller than 2.4 mm I.D. in diameter and no longer than 380 mm in length for the Full-Size (580mm x 280mm) and Mid-Size (465mm x 280mm) containers, and no smaller than 1.0 mm I.D. in diameter and no longer than 250 mm in length for the Half-Size (285mm x 280mm) containers have been validated for a combined maximum load of 25lbs. with up to two internal stacked baskets and up to two silicon mats (one mat per basket) for the Full-Size, Mid-Size and Half-Size containers. Both solid and perforated bottom containers, maximum depth 150mm/each, can be stacked a maximum of 3 containers high during sterilization. Sterilized devices may be stored and transported within the container. The container is intended to maintain sterility of the contents for up to six (6) months.
Device Story
SklarLite™ Rigid Sterilization Container System; reusable, stackable, sealed containers for medical device sterilization. Components: container bottoms (perforated/non-perforated), lids, baskets, labels, disposable filters, security seals. Used in hospitals/healthcare facilities; operated by sterile processing staff. Input: medical devices placed inside containers. Process: high vacuum steam sterilization (270°F); sterilant enters via perforated lids/bottoms; filters prevent microbial migration. Output: sterilized, protected instruments ready for storage/transport. Benefit: maintains sterility for up to 6 months; protects instruments during handling. Healthcare providers use containers to organize, sterilize, and store surgical instrumentation.
Clinical Evidence
Bench testing only. Validation included pre-vacuum steam sterilization cycles at 270°F. Testing confirmed sterility maintenance for 6 months and successful sterilization of inoculated stainless steel medical devices and lumens (2.4mm ID x 380mm and 1.0mm ID x 250mm). Conformance to AAMI/ANSI ST77:2006 standard demonstrated.
Technological Characteristics
Materials: Anodized aluminum alloy, stainless steel, silicon. Design: Rigid, stackable, closed container system with filter-based sterilant entry and microbial barrier. Dimensions: Various sizes (Full, Mid, Half). Sterilization: High vacuum steam. Connectivity: None. Software: None.
Indications for Use
Indicated for use in hospitals and health care facilities to contain medical devices during high vacuum steam sterilization (270°F, 4 min exposure, 30 min dry time). Validated for loads up to 25 lbs, including specific lumen constraints (2.4mm ID/380mm length for Full/Mid-size; 1.0mm ID/250mm length for Half-size).
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
Miltex Rigid Sterilization Container System (K050570)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: This image shows the logo and contact information for Sklar Instruments. The logo features a stylized image to the left of the company name, "Sklar," written in a bold, serif font. Below the logo, it says "INSTRUMENTS" in smaller, capitalized letters. The address is listed as 889 South Matlack Street, West Chester, PA 19382, followed by the phone number 610-430-3200, toll-free number 800-221-2166, fax number 610-429-0500, and website www.sklarcorp.com.
Exhibit#6 FDA 510(k) Pre-Market Notification SklarLite™ Rigid Sterilization Container System
APR - 9 2010
Exhibit 1
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: K091720.
### 1. Submitter's Identification:
Sklar Instruments 889 South Matlack Street West_Chester, PA_19382 800-221-2166 Tel:
Date Summary Prepared: April 2, 2010
- Name of the Device: SklarLite" Rigid Sterilization Container System 2. Class II, Product Code KCT
- Common or Usual Name: Closed Sterilization Container 3.
- 4. Predicate Device Information : Miltex Rigid Sterilization Container System, Miltex, Incorporated, K050570
### ર. Device Description :
Rigid Sterilization Container System consists of a family of rigid, re-The SklarLite" usable, stackable, sealed containers that provide an effective sterilization packaging method for medical devices. Container bottoms and lids, within a given size, are interchangeable. Containers can be fitted with safety lids, protecting filters during storage and/or transport. The system consists of the following components:
- Container bottoms (both perforated and non-perforated versions) .
- Container lids (both perforated and non-perforated_versions)_ .
- Container baskets .
- Container labels .
- Disposable filters .
The container system is designed for sterilant penetration through perforations in the lid and container bottom models that are perforated. Containers are offered in six colors. Filter media and sterilization indicator cards should be used in conjunction with the containers. Containers are protected by disposable plastic security seals.
{1}------------------------------------------------
#### 6. Intended Use:
The SklarLite™ Rigid Sterilization Container System is intended for use in hospitals and health care facilities to contain other medical devices that are to be sterilized. Containers allow sterilization of the enclosed medical devices, including surfaces and lumens, using high vacuum steam sterilizers at 270°F for 4 minutes with 30 minutes (minimum) dry time.
The containers have been validated for sterilization of up to two instruments with lumens no smaller than 2.4 mm I.D. in diameter and no longer than 380 mm in length for the Full-Size (580mm x 280mm) and Mid-Size (465mm x 280mm) containers, and no smaller than 1.0 mm I.D. in diameter and no longer than 250 mm in length for the Half-Size (285mm x 280mm) containers have been validated for a combined maximum load of 25lbs. with up to two internal stacked baskets and up to two silicon mats (one mat per basket) for the Full-Size, Mid-Size and Half-Size containers. Both solid and perforated bottom containers, maximum depth 150mm/each, can be stacked a maximum of 3 containers high during sterilization.
Sterilized devices may be stored and transported within the container. The container is intended to maintain sterility of the contents for up to six (6) months.
### Technological Characteristics as Compared to the Predicate Device: 8.
| | Sklar System | Miltex System |
|---------------------------------------------------------------------------------------------|-----------------------------------------------------------|--------------------------------------------------|
| PROPERTIES | | |
| Indicated for use containing<br>instruments to be sterilized in vacuum<br>steam sterilizers | Yes | Yes |
| Intended to be re-used | Yes | Yes |
| Closed System | Yes | Yes |
| Sealed | Yes | Yes |
| DESIGN | | |
| Incorporates a filter system to permit<br>entry of sterilant agents | Yes | Yes |
| Incorporates a filter system to prevent<br>microbial migration during transport | Yes | Yes |
| MATERIALS: | | |
| Container | Anodized Aluminum<br>-Alloy-stainless-steel-<br>& silicon | Aluminum alloy,<br>stainless-steel &-<br>silicon |
| Color | Silver, yellow, red,<br>blue, green and black | Silver |
{2}------------------------------------------------
## 9. Performance Data as Compared to the Predicate Device:
| | Sklar System | Miltex System |
|----------------------------------------------|------------------------------------------------|----------------------------------------|
| PERFORMANCE DATA | | |
| Conformance to appropriate AAMI<br>standards | Conforms to<br>AAMI/ANSI<br>ST77:2006 standard | Conforms to<br>AAMI/ANSI ST77<br>Draft |
| VALIDATION TESTING | | |
|-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| Pre-vacuum steam | Yes | Yes |
| Load | Up to 25 lbs. for all<br>container sizes. | Up to 16 lbs. (small)<br>Up to 20 lbs. (med.)<br>Up to 25 lbs. (large) |
| TEST ORGANISMS | | |
| Inoculated Lumens | (2) 2.4mm I.D. x<br>380mm, metal for<br>Full-and-Mid-Size<br>container models.<br>(2) 1.0mm I.D. x<br>250mm, metal for<br>Half-Size container<br>models. | 3mm I.D. x 400mm,<br>metal (large<br>containers) and 3mm<br>I.D. x 200mm, metal<br>(medium and small<br>containers) |
| Inoculated Stainless Steel Medical<br>Devices | Yes | Yes |
## 10. Conclusions:
The studies conducted on the SklarLite™ Rigid Sterilization Container System demonstrate that the device is substantially equivalent to the Miltex Rigid Sterilization Containers (K050570).
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines representing its wings. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA", which is arranged in a circular fashion around the left side of the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Sklar Corporation C/O Ms. Natalya Valerio Official Correspondent MDI Consultants, Incorporated 55 Northern Boulevard, Suite 200 Great Neck, New York 11021
APR - 9 2010
Re: K091720
Trade/Device Name: SklarLite™ Rigid Sterilization Container System Regulation Number: 21CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT Dated: April 2, 2010 Received: April 5, 2010
Dear Ms. Valerio:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 – Ms. Valerio
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Anthony D. Anton
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital,
Infection Control and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
Page _________ of ____________________________________________________________________________________________________________________________________________________________
510(k) Number (if known):
Device Name:_SklarLite™ Rigid Sterilization Container System
# Indications For Use:
The SklarLite™ Rigid Sterilization Container System is intended for use in hospitals and health care facilities to contain other medical devices that are to be sterilized. Containers allow sterilization of the enclosed medical devices, including surfaces and lumens, using high vacuum steam sterilizers at 270°F for 4 minutes with 30 minutes (minimum) dry time
The containers have been validated for sterilization of up to two instruments with lumens no smaller than 2.4 mm I.D. in diameter and no longer than 380 mm in length for the full-Size in (580mm x 280mm) and Mid-Size (465mm x 280mm) containers, and no smaller than 102c in diameter and no longer than 250 mm in length for the Half-Size (285mm x 280mm) container. Containers have been validated for a combined maximum load of 25lbs. with us to two internal stacked baskets and up to two silicon mats (one mat per basket) for the Full-Size, Mid-Size and Half-Size containers. Both solid and perforated bottom containers, maximum deeth 1500mmean, can be stacked a maximum of 3 containers high during sterilization.
Sterilized devices may be stored and transported within the container. The container is intended to maintain sterility of the contents for up to six (6) months.
Prescription Use (Per 21 CFR 801 Subpart D)
OR
Over-The Counter Use X (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Anesthesiology. General Hospital Infection Control, Dental Devices
510(k) Number: k091720
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Sklar Instruments. The logo consists of a stylized graphic to the left of the word "Sklar" in a bold, serif typeface. Below the word "Sklar" is the word "INSTRUMENTS" in a smaller, sans-serif typeface. The logo is black and white.
.
-
:
# Indications for Use
Page ___ 2____ of ____________________________________________________________________________________________________________________________________________________________
The sterilization containers which are the subject of this premarket notification are as follows:
| MODEL | Perforations | Dimensions | Load Limit |
|-------------------------------------------|---------------------------------------------|-----------------|-----------------|
| Full Size Containers<br>Basic/Safe Models | Perforated Lid and<br>Non Perforated Bottom | 580 x 280 x 100 | 25 lbs. |
| | | 580 x 280 x 135 | 25 lbs |
| | | 580 x 280 x 150 | 25 lbs |
| | | 580 x 280 x 200 | 25 lbs |
| | | 580 x 280 x 260 | 25 lbs |
| | Perforated Lid and<br>Perforated Bottom | 580 x 280 x 100 | 25 lbs |
| | | 580 x 280 x 135 | 25 lbs |
| | | 580 x 280 x 150 | 25 lbs |
| | | 580 x 280 x 200 | 25 lbs |
| | | 580 x 280 x 260 | 25 lbs |
| Mid Size Containers<br>Basic/Safe Models | Perforated Lid and<br>Non Perforated Bottom | 465 x 280 x 100 | 25 lbs |
| | | 465 x 280 x 135 | 25 lbs |
| | | 465 x 280 x 150 | 25 lbs |
| | Perforated Lid and<br>Perforated Bottom | 465 x 280 x 100 | 25 lbs |
| | | 465 x 280 x 135 | 25 lbs |
| | | 465 x 280 x 150 | 25 lbs |
| Half Size Containers<br>Basic/Safe Models | Perforated Lid and<br>Non Perforated Bottom | 285 x 280 x 100 | 25 lbs |
| | | | 285 x 280 x 135 |
| | | | 285 x 280 x 150 |
| | | 285 x 280 x 200 | 25 lbs |
| | | 285 x 280 x 260 | 25 lbs |
| | Perforated Lid and<br>Perforated Bottom | 285 x 280 x 100 | 25 lbs |
| | | | 285 x 280 x 135 |
| | | | 285 x 280 x 150 |
| | | | 285 x 280 x 200 |
| | | | 285 x 280 x 260 |
| | | 600 x 400 x 180 | 25 lbs |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.