The YM6000 monitor is intended to be used to monitor electrocardiography, heart rate, pulse rate, noninvasive blood pressure (systolic, diastolic and mean arterial pressures), functional arterial oxygen saturation, invasive blood pressure, respiration, capnography (EtCO2 and InCO2) and temperature for adult, pediatric and neonate patients in all areas of a hospital and hospital-type facilities. Monitor users should be skilled at the level of a technician, doctor, nurse or medical specialist. - Note: Hospital use typically includes such areas as general care floors, operating rooms, special procedure areas, intensive and critical care area, within the hospital. Hospital-type facilities include physician office-based facilities, sleep labs, skilled nursing facilities, surgicenters, and sub-acutecenters. - Note: The medically skilled and trained user can be any clinicians like doctor and nurse who know how to take and interpret a patient's vital signs. These clinicians must take the direct responsibility for patient life. It can include a care-giver or medically trained interpreter who is authorized under the appropriate clinical facility procedure to support the patient care. Any inappropriate setting, especially the limit alarm notification setting can lead the hazardous situation that injures the patient or harms the patient or threats the patient life. The special attention to assign the trained user who can adjust the patient setting of the patient monitor should be essentially paid.
Device Story
YM6000 is a compact, AC/battery-powered patient monitor for hospital/clinical use. It acquires physiological signals via ECG, NIBP, SpO2, IBP, capnography, and temperature modules. Inputs are processed to display real-time vital signs and status on a TFT-LCD screen. Clinicians (doctors, nurses, technicians) use the device to track patient status and set alarm limits. Output data supports clinical decision-making by providing continuous monitoring of patient stability. Benefits include early detection of physiological changes and timely intervention. The device uses established modules (Mediana ECG, Omron NIBP, Nellcor SpO2, Philips Respironics capnography) to ensure accurate measurement.
Clinical Evidence
Bench testing only. The device underwent system verification and validation (Plan #PAA00-00005) using production-equivalent units. Performance was validated against established standards for each module: AAMI SP-10 and IEC60601-2-30 (NIBP), ISO9919 (SpO2), and IEC60601-2-34 (IBP). No clinical trial data was required or provided.
Technological Characteristics
Compact monitor (341x305x172mm, 5.5kg) with TFT-LCD display. Powered by AC (100-240VAC) or internal rechargeable lead-acid battery. Modules: Mediana MDE-1 (ECG), Omron M3200 (NIBP), Nellcor NELL-3 (SpO2), Philips Respironics Capnostat 5/LoFlo C5 (Capnography), Mediana MDT-1 (Temperature). Connectivity includes physiological signal transmission. Complies with AAMI SP-10, IEC60601-2-30, ISO9919, and IEC60601-2-34.
Indications for Use
Indicated for adult, pediatric, and neonate patients requiring monitoring of ECG, HR, pulse rate, NIBP, SpO2, IBP, respiration, capnography (EtCO2/InCO2), and temperature in hospital and hospital-type settings.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
Omron Model HBP-2070 Vital Signs Monitor (K082812)
Spacelabs Medical Inc. mCare 300 Vital Signs Monitors, Model 91220 (K062095)
Omron Healthcare Inc. Colin Press Mate Model BP-S510 (K063690)
Mediana Co., Ltd. Lucon M-Series Patient Monitor, Model M20 and M30 (K100217)
{0}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
### Submitter
Mediana Co.,Ltd.
Wonju Medical Industry Park, 1650-1, Donghwa-ri, Munmak-eup, Wonju-si, Gangwon-do, Korea Tel: (82) 2 542 3375 (82) 33 742 5400 Fax: (82) 2 542 7447 (82) 33 742 5483 Company Contact: Amy M.H.KIM, Manager of Regulatory Affairs Rachel J.E.OH, Specialist of Regulatory Affairs
Date Summary Prepared: May 5th, 2011
#### Device Name
| Trade Name: | YM6000, patient monitor |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Patient Monitor |
| Classification Name: | Monitor, Physiological, Patient (with arrhythmia dete ction or alarms)<br>(21CFR870.1205) also contains<br>Monitor, Cardiac (including cardiotachometer and rate ala rm)<br>(21CFR870.2300)<br>System, measurement, blood-pressure, non-invasive<br>(21CFR870.1130),<br>Electrocardiograph (21CFR870.2340)<br>Oximeter (21CFR870.2700)<br>Analyzer, gas, carbon-dioxide, gaseous-phase (21CFR868.1400)<br>Computer, Blood-Pressure (21CFR870.1110)<br>Transmitters and receivers, physiological signal, radiofrequency<br>(21CFR870.2910) |
| Classification: | Class II |
| Product code: | MHX |
#### Predicate Devices (Legally Marketed Devices)
The predicate devices for Patient monitors, Model YM6000 are:
- . Omron Healthcare Inc. Omron Model HBP-2070 Vital Signs Monitor cleared by FDA through 510(k) No. K082812,
- Spacelabs Medical Inc. mCare 300 Vital Signs Monitors, Model 91220 . • cleared by FDA through 510(k) No. K062095,
- Omron Healthcare Inc. Colin Press Mate Model BP-S510 t cleared by FDA through 510(k) No. K063690,
- Mediana Co., Ltd. Lucon M-Series Patient Monitor, Model M20 and M30 cleared by FDA through 510(k) No. K100217
{1}------------------------------------------------
### Device Description
The YM6000 patient monitor is to monitor electrocardiography (ECG), Arrhythmia, ST segment, heart rate (HR), pulse rate, noninyasive blood pressure (systolic, diastolic and mean arterial pressures); functional arterial oxyge n saturation, invasive b lood pressure, respiration, capnography (EtCO2 and InCO2) and temperature for adult, pediatric and neonate in patients in general hospital and alternate ca re facilities by medically trained personnel. This monitor is available for sale only upon the order of a physician or licensed health care professional.
The YM6000 patient monitor is a lightweight and compact device (341 × 305 × 172 (mm) (W × H × D) and 5.5 kg) powered by AC mains (100-240VAC, 50-60Hz) and also powered by internal battery. The monitor provides patient data and monitoring status on TFT-LCD displays.
#### Intended Use
The YM6000 monitor is intended to be used to mo nitor electrocardiography, heart rate, p ulse rate, noninvasive blood pressure (systolic, diastolic and mean arterial pressures), functional arterial oxygen satu ration, invasive blood pressure, respiration, capnography (EtCO2 and InCO2) and temperature for adult, pedi atric and neonate patients in all areas of a hospital and hospital-type facilities. Monitor users should be skilled at the level of a technici an, doctor, nurse or medical specialist.
- Note: Hospital use typically includes such areas as general care floors, operating rooms, special procedure areas, inten sive and critical care area, within the hospital. Ho spital-type facilities include physician office-based facilities, sleep lab s, skilled nu rsing facilities, surgicenters, and sub-acutecenters.
- Note: The medically skilled and trained user can be any clinicians like doctor and nurse who know how to take and interpret a patient's vital signs. These clinicians must take the direct responsibility for patient life. It can in clude a care-qiver or medically trained interpreter who is authorized under the appropriate clinical facility procedure to support the patient care. Any inappropriate setting, especially the limit alarm notification setting can lead the hazardous situation that injures the patient or harms the patient or threats the patient life. The sp ecial attention to assign the trained user who can adjust the patient setting of the patient monitor should be essentially paid.
### Summary of Technical Characteristics of the Device Compared to the Predicate Devices (Legally Marketed Devices)
The Mediana Patient Monitor, Model YM6000 is substantially equivalent to the Colin Press Mate Model BP-S510, Lucon M-series Patient Monitor, Model M20 and M30, Omron M odel HBP-2070 Vital Signs Monitor, and mCare 300 Vital Signs Monitor, Model 91220.
- . The Model YM600 uses the Mediana ECG module, MDE-1 manufactured by Mediana Co., Ltd. which is identical to the module used in the Model M20, M30 and Model HBP-2070. So the Electrocardiogram (ECG) and heart rate specifications for the Model YM600, Model M20, M30 and Model HBP-2070 are identical,
- The Model YM6000 uses the Omron NIBP module M3200 which is identical to the module used in the Model BP-S510, Model M20, M30 and Model HBP-2070. The non-invasive blood pressure (NIBP) and pulse rate specifications for the Model YM6000, Model BP-S510, Model M20, M30 and Model HBP-2070 are identical. The circuitry, algorithm and cuffs in
{2}------------------------------------------------
both models are the same manufactured by Omron Healthcare Inc. Also, all devices comply with the AAMI performance standard SP-10 & IEC60601-2-30 and have same measurement technology,
- The Model YM6000 uses the Nelloor SpO2 module, NELL-3, which is identical to the module used in the Model BP-S510, Model M20, M30 and Model HBP-2070. The SpO2 specifications for the Model YM6000, Model M20, M30 and Model HBP-2070 are identical. The circuitry, algorithm and sensors are the same as SpO2 module used in above monitors, is anufactured by Nellcor Inc. Also, all devices comply with the performance standard 1509919,
- The respiration can be obtained from either the Electrocardiogram (ECG) channel or the Capnography (EtCO2) channel. The algorithm of Impedance measurement in the Model M20, M30 and Model HBP-2070 is identical to that in Model YM6000 and the respiration specification for the Model YM6000, Lucon M-series (Model M20, M30) and Model HBP-2070 are identical because the respiration measurement (Thoracic Impedance) for Model YM6000, Lucon M-series (Model M20, M30) and Model HBP-2070 is from ECG module that are the same module designed and manufactured by Mediana Co., Ltd.,
- And also Lucon M-series (Model M20, M30) and Model 91220 have identical performance of respiration measurement (Airway Measurement) with Model YM6000. Because respiration measurement (Airway Measurement) is from Capnography module and these devices use the same Capnography module which manufactured by Philips Respironics Inc.,
- The Model YM6000 uses the thermometry MDT-1 module which is identical to used in the Lucon M-series (Model M20, M30) and Model HBP-2070. The temperature measurement method for the Model YM6000, Lucon M-series (Model M20, M30) and Model HBP-2070 are equivalent. Also all of device uses the YSI probes/probe covers,
- The pulse rate derived from the non-invasive blood pressure (NIBP) channel is from M3200 NIBP module in the Model YM6000. This M3200 module is identical to that in the Model BP-S510, Lucon M-series (Model M20, M30) and Model HBP-2070. And the pulse rate derived from the Pulse Oximetry (SpO2) channel is from NELL-3 Spo2 module in the Model YM6000. This NELL-3 module is identical to that in the Model BP-S510, Lucon M-series (Model M20, M30) and Model HBP-2070,
- The Model YM6000 uses the Respironics Capnography module, Capnostat 5 and LoFio C5 which have already been registered in FDA by Philips Respironics Inc. (#K042601 is for Capnostat 5, #K053174 is for LoFlo C5). Both modules are identical to the modules used in the Lucon M-series (Model M20, M30) and Model 91220. All devices have identical technology and operation of theory in Capnography module,
- The Model YM6000 uses the same measurement technology with the Model BP-S510 for Invasive Blood Pressure. The IBP technology is most traditional and typical method among measuring the patient parameter. The Invasive blood pressurement performance of the Model YM6000 is similar to the Model BP-S510 by using same measurement technology. And both devices comply with the IBP performance standard, IEC60601-2-34.
- The YM6000 monitor as well as the predicate devices has internal power source, a rechargeable lead acid battery and AC power.
{3}------------------------------------------------
The Mediana patient monitor, Model YM6000 substantially have been tested in accordance with the system V & V plan (#P AA00-00005) and summary included with the submission using production equivalent units prior to release to market.
4/4
A risk a nalysis identifying potential hazards and documenting mitigation of the hazards has been developed and applied as part of Mediana design control procedure. Mediana's quality system confirms to 21 CFR820, ISO13485 and CMDCAS ISO13485 certified by DNV (Det Norske Veritas).
## Conclusions
As stated above, the Mediana patient monitors, Model YM6000 is safe and effective and comply with the appropriate medical device standards and are substantially equivalent to the earl ier identified predicate devices.
- End of Section -
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its wing and body. The eagle is enclosed in a circle, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference of the circle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room ~WO66-G609 Silver Spring, MD 20993-0002
NOV - 9 2011
Mediana Co., Ltd. c/o Ms. Charlie Mack President International Regulatory Consultants 7325 Joyce Way Echo, OR 97826
Re: K112190
> Trade/Device Name: YM6000, Patient Monitor Regulatory Number: 21 CFR 870.1025 Regulation Name: Patient Physiological Monitor (with arrhythmia detection or alarms)
> Pogulatory Class - H ( Regulatory Class: II (two) Product Code: 74 MHX Dated: October 16, 2011 Received: October 24, 2011
Dear Ms. Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act (Act (Act that the provise approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Charlie Mack
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## INDICATIONS FOR USE
Applicant:
Mediana Co.,Ltd. Wonju Medical Industry Park, 1650-1, Donghwa-ri, Munmak-eup, Wonju-si, Gangwon-do. Korea Tel: (82) 2 542 3375 (82) 33 742 5400 Fax: (82) 2 542 7447 (82) 33 742 5483
510(k) Number:
Device Name: YM6000, Patient Monitor
#### Indications for Use:
The YM6000 monitor is intended to be used to monitor electrocardiography, heart rate, pulse rate, noninvasive blood pressure (systolic, diastolic and mean arterial pressures), functional arterial oxygen saturation, invasive blood pressure, respiration, capnography (EtCO2 and InCO2) and temperature for adult, pediatric a nd neonate patients in all areas of a hospital and hospital-type facilities. Monitor users should be skilled at the level of a technician, doctor, nurse or medical specialist.
- Note: Hospital use typically includ es such areas as gen eral care floors, opera ting rooms, special proced ure areas, intensive and critical care area, within the hospital. Hospital-type facilities include physician office-based facilities, sleep labs, skilled nursing facilities, surgicenters, and sub-acutecenters.
- Note: The medically skilled and trained user can be any clinicians like doctor and nurse who know how to take a nd interpret a patient's vital signs. T hese clinicians must take the direct respon sibility for pati ent life. It can include a care - -giver or medically trained interp reter who is authorize d under the appropria te clinical facility procedure to support the patient care. Any i nappropriate setting. especially the limit alarm or alarm notification setting can lead the hazardous situation that injures the patient or harms the patient or threats the patient life. The special attention to assign the trained user who can adjust the patient setting of the patient monitor should be essentially paid.
(Please do not write below this line - continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | X | OR Over-The-Counter | |
|------------------------------------|---------|---------------------|--|
| (Per 21CFR801.109) | | | |
| (Division Sign-Off) | | | |
| Division of Cardiovascular Devices | | | |
| 510(k) Number | K112190 | | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.