K121409 · Asahi Kasei Medical Co., Ltd. · KDI · Feb 6, 2013 · Gastroenterology, Urology
Device Facts
Record ID
K121409
Device Name
ASAHI REXEED-SX/LX SERIES DIALYZER
Applicant
Asahi Kasei Medical Co., Ltd.
Product Code
KDI · Gastroenterology, Urology
Decision Date
Feb 6, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
REXEED-SX/LX Dialyzers is intended for use for hemodialysis treatment of patients who have chronic or acute renal failure. REXEED SX/LX Dialyzers must be used in accordance with the instructions for a physician familiar with hemodialysis and familiar with the conditions of the patient. REXEED SX/LX Dialyzers have been tested in vitro under single use conditions.
Device Story
The Asahi REXEED-SX/LX Series Dialyzer is a single-use, high-permeability hollow fiber dialyzer. It functions as a blood-cleansing component in hemodialysis systems for patients with renal failure. The device consists of polysulfone hollow fiber membranes encased in a styrene-butadiene block copolymer housing. It is operated by trained clinicians in a clinical setting. During hemodialysis, blood flows through the hollow fibers while dialysate flows on the outside, allowing for the removal of waste products and excess fluid via diffusion and convection. The device is sterilized via electron beam irradiation. Modifications in this submission include an enhanced manufacturing process for material removal, updated labeling, and revised specification limits for fiber leakage and physical properties to improve safety.
Clinical Evidence
Bench testing only. Design verification and validation were performed to confirm that modifications met acceptance criteria. No clinical data was required or provided.
Technological Characteristics
High permeability hollow fiber dialyzer. Materials: Polysulfone (hollow fibers), styrene-butadiene block copolymer (casing). Sterilization: Electron beam irradiation. Single-use. No software or electronic components.
Indications for Use
Indicated for hemodialysis treatment of patients with chronic or acute renal failure.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
{0}------------------------------------------------
### 510(k) Summary of Safety and Effectiveness A.0
## A.1 Submitter Information
# FEB 0 6 2013
| Company Name and Address: | Contact Name: |
|-----------------------------------|-----------------------------------------|
| Asahi Kasei Medical Company, Ltd. | Patsy J.Trisler, JD, RAC |
| 1-105 Kanda Jinbocho, Chiyoda-ku, | Vice President, Regulatory and Clinical |
| Tokyo 101-8101, Japan | Qserve America, Qserve Group B.V. |
| | 154 Main Street, Suite #2 |
| | Charlestown, NH 03603 |
| | Telephone: 301-652-5344 |
| | Fax: 509-807-7204 |
| | Email: patsy.trisler@qservegroup.com |
#### Date Prepared: May 9, 2012 A.2
#### A.3 Name of Device
- 3.1 Trade Name: Asahi REXEED-SX/LX Series Dialyzer
- 3.2 Common Name: High Flux Hemodialysis Membrane Dialyzer or High Flux Hollow Fiber Dialyzer
#### 3.3 Classification Name, Class, Product Code and Panel
| Classification Name and Regulation<br>Number | Class | Product Code | Panel |
|-----------------------------------------------------------------------------------------------|-------|--------------|---------------------------------|
| High Permeability Hemodialysis<br>Systems, Title 21 Code of Federal<br>Regulations § 876.5860 | II | KDI | Gastroenterology<br>and Urology |
#### Substantial Equivalence Claimed to Predicate Device A.4
APS/REXEED, cleared for commercial distribution via 510(k) Premarket Notifications K001250 dated August 16, 2000; K041726 dated July 23, 2004; K051187 dated June 8, 2005; and K082515, dated October 3, 2008.
#### A.5 Device Description
The line of Asahi REXEED-SX/LX Series Dialyzer is a family of high permeability hollow fiber dialyzers intended for the treatment of patients with acute or chronic renal failure. REXEED-SX/LX Series Dialyzer is designed for single use. REXEED-SX/LX Series Dialyzer is constructed of hollow fiber (polysulfone) membranes, housed within a plastic casing of styrene-butadiene block copolymer and are subjected to electron beam irradiation prior to shipment.
{1}------------------------------------------------
K121409
Page 2 of 2
This Special 510(k) describes the following modifications:
- I Manufacturing Process Change: insertion of an additional step to enhance removal of manufacturing materials.
- . Labeling Change: changes to the Instructions for Use to emphasize importance of following instructions for use.
- I Device Modification: change in specification limit of leakage and related physical properties of hollow fibers for safer treatment and operation.
#### Intended Use and Indications for Use A.6
- A. REXEED-SX/LX Dialyzers is intended for use for hemodialysis treatment of patients who have chronic or acute renal failure.
- B. REXEED SX/LX Dialyzers must be used in accordance with the instructions for a physician familiar with hemodialysis and familiar with the conditions of the patient.
- C. REXEED SX/LX Dialyzers have been tested in vitro under single use conditions.
#### A.7 Evaluation of Design Modifications
As the basis for Asahi's device evaluation studies and overall process for managing medical device risk, the company has performed a risk analysis using procedures based on ISO 14971(2007) "Medical Devices - Application of Risk Management to Medical Devices". The risk analysis method used to assess the impact of the modification was Failure Modes and Effects Analysis (FMEA). Design validation or where appropriate, design verification, and process validation based on the result of risk analysis and design input were performed to verify those modifications. All test results meet the acceptance criteria, and proved that those modifications to be appropriate.
#### A.8 Conclusion:
Asahi made modifications to the REXEED-SX/LX Series Dialyzer cleared under K001250, K041726, K051187, and K082515, resulting in the modified REXEED-SX/LX Series Dialyzer. The information and data provided in this Special 510(k) Premarket Notification establish that the modified REXEED-SX/LX Series Dialyzer is substantially equivalent in intended use/indications for use, design, principle of operation, technology, materials, specifications, and performance to the predicate REXEED-SX/LX Series Dialyzer.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all caps and is smaller than the caduceus symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 6, 2013
Asahi Kasei Medical Co., Ltd. % Ms. Patsy J. Trisler, JD, RAC Vice President, Regulatory & Clinical Affairs Oserve Group B.V. 154 Main Street, Suite #2 CHARLESTOWN NH 03603
Re: K121409
Trade/Device Name: Asahi REXEED-SX/LX Series Dialyzer Regulation Number: 21 CFR§ 876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: KDI Dated: December 20, 2012 Received: December 27, 2012
Dear Ms. Trisler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{3}------------------------------------------------
Page 2 - Ms. Patsy J. Trisler, JD, RAC
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Benjamin/R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Statement of Indications for Use
## 510(k) Number (if known): K121409
Device Name: Asahi REXEED-SX/LX Series Dialyzer
### Indications for Use:
- A. REXEED-SX/LX Dialyzers is intended for use for hemodialysis treatment of patients who have chronic or acute renal failure.
- B. REXEED SX/LX Dialyzers must be used in accordance with the instructions for a physician familiar with hemodialysis and familiar with the conditions of the patient.
- C. REXEED SX/LX Dialyzers have been tested in vitro under single use conditions.
X AND/OR Over-the-Counter Use: Prescription Use: __ (21 CFR § 801 Subpart D) (21 CFR §07 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Benjamin R. Fjisher -S 2013.02.06 1/5:06:01 -05'00'
Division Sian-Off lon of Reproductive, Gastro-Renal, and logical Device 510(k) Numbe
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.