HIGH FLUX HEMODIALYSIS MEMBRANE DIALYZER OR HIGH FLUX HOLLOW FIBER DIALYZER, MODELS APS-R, -M, -S, -E, AND -EX
Applicant
Asahi Medical Co., Ltd.
Product Code
KDI · Gastroenterology, Urology
Decision Date
Jul 23, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The line of APS Series Dialyzers is a family of high permeability hollow fiber dialyzers intended for the treatment of patient with acute or chronic renal failure. The APS Series Dialyzers are intended for both single use or for reuse with a maximum of 15 reprocessing reuse cycles per patient. This is the same intended use as the predicate device - APS Series Dialyzers cleared under K001250.
Device Story
Asahi APS Series Dialyzers are high-permeability hollow fiber dialyzers for hemodialysis treatment of acute or chronic renal failure. Device consists of polysulfone hollow fiber membranes within a styrene butadiene block polymer casing. Available in 'wet' (factory-primed) or 'dry' models. Used in clinical settings by physicians/trained staff. Modifications include revised pre-cleaning labeling, housing design to improve flow diffusion, and updated shipping carton materials. Device supports single use or reuse (up to 15 cycles) when reprocessed with specific agents and systems. Output is the filtered blood/dialysate exchange; healthcare providers monitor performance via standard dialysis system metrics. Benefits include effective solute removal and fluid management for renal patients.
Clinical Evidence
Confirmatory clinical studies and in vitro testing were conducted to validate performance under single use and reuse conditions (up to 15 cycles). Results demonstrated that modified dialyzers meet acceptance criteria for performance, residual blood, and leakage.
Indicated for hemodialysis treatment of patients with acute or chronic renal failure. Must be used by a physician familiar with hemodialysis and patient conditions. May be reused up to 15 cycles on the same patient if reprocessed using specified agents (4% formaldehyde or Renalin) and systems (Seratronics DRS4/DPS4 or Renatron RS 8300), provided residual Total Cell Volume is at least 80% and dialysis system uses volumetric ultrafiltration controllers.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
{0}------------------------------------------------
JUL 2 3 2004
Special 510(k) Premarket Notification
CONFIDENTIAL
ﮯ
## 510 (K) SUMMARY
#### APS Series Dialyzers
Manufacturer:
Asahi Medical Company, Ltd. 9-1, Kanda Mitoshirocho Chiyoda-ku, Tokyo 101-8482 Japan
Date summary was prepared:
June 24, 2004
Name(s) of the device:
Asahi APS Series Dialyzers
Identification of predicate device(s):
Asahi APS Series Dialyzers
Description of the device:
The line of Asahi Polysulfone (APS) Series Dialyzers is a family of high permeability hollow fiber dialyzers intended for the treatment of patients with acute or chronic renal failure. The Asahi APS Series Dialyzers are designed for both single use or for reuse with a maximum of 15 reprocessing reuse cycles per patient. They are constructed of reusable, hollow fiber (polysulfone) membranes, housed within a plastic casing of styrene butadiene block polymer and are gamma sterilized prior to shipment.
The Asahi APS Series Dialyzers are offered for sale in both a "wet" and a "dry" model. The wet and dry dialyzers are identical to each other except that the wet models filled at the factory with a fluid to facilitate priming by the user and the dry models are not filled. The use of a wet or dry dialyzer is a matter of user preference.
Modifications made to the Asahi APS Series Dialyzers subject of this 510(k) include (1) labeling revised to provide additional information concerning pre-cleaning; (2) modification of the dialyzer housing to diffuse the flow water used in pre-cleaning; and, (3) modification of the shipping carton partition materials and mold.
APS Series Dialyzers Asahi Medical Company., Ltd.
{1}------------------------------------------------
2
## Intended Use:
The line of APS Series Dialyzers is a family of high permeability hollow fiber dialyzers intended for the treatment of patient with acute or chronic renal failure. The APS Series Dialyzers are intended for both single use or for reuse with a maximum of 15 reprocessing reuse cycles per patient. This is the same intended use as the predicate device - APS Series Dialyzers cleared under K001250.
## Evaluation of Design Modifications
As the basis for Asahi Medical's device evaluation studies and overall process for managing medical device risk, the company has performed a risk analysis using procedures based on ISO 14971 (2000) Medical Devices -- Application of Risk Management to Medical Devices. The risk analysis method used to assess the impact of the modification was Failure Modes and Effects Analysis (FMEA). Design verification tests based on the result of risk analysis were performed to verify those modifications. All test results meet the acceptance criteria, and proved that those modifications to be appropriate.
### Conclusion:
Asahi Medical made three modifications to the original APS Series Dialyzers cleared under K001250. All design verification tests based on the result of risk analysis proved that modified APS Series Dialyzers are substantially equivalent in intended use, design, principle of operation/technology, materials, specifications and performance to the original APS Series Dialyzers.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a circular logo with a stylized bird in the center. The bird is composed of three curved lines that suggest feathers or wings. The bird is facing to the right. The text is arranged around the top half of the circle, following the curve of the circle. The text is in a sans-serif font and appears to be in all lowercase letters.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 2 3 2004
Asahi Medical Company, Ltd. c/o Mr. David L. West, Ph.D. Vice President Quintiles Consulting 1801 Rockville Pike ROCKVILLE MD 20852
Re: K041726
Trade/Device Name: High Flux Hemodialysis Membrane Dialyzer or High Flux Hollow Fiber Dialyzer, Models APS-R, -M, -S, -E, and -EX Regulation Number: 21 CFR §876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: 78 KDI Dated: June 25, 2004 Received: June 25, 2004
Dear Dr. West:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your bector. 310(x) presidentially equivalent (for the indications ferenced above and nave decemined by marketed predicate devices marketed in interstate for use stated in the enclosure) to regally to regally the Medical Device Amendments, or to commerce prior to May 20, 1776, the enactines with the provisions of the Federal Food, Drug, devices that have occarred in asse approval of a premarket approval application (PMA). and Cosmette Act (rec) market the device, subject to the general controls provisions of the Act. The 1 ou may, therefore, mance the act include requirements for annual registration, listing of general controls provisions of the fiel labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), IT your device is classified (see above) and are regulations affecting your device can be it may be subject to additional controls. Extraige angest 800 to 898. In addition, FDA may round in the Code of Casterning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Picase oe advised that I Dr s issuation of one device complies with other requirements of the Act that I DA has made a decemmanen administered by other Federal agencies. You must of any I ederal statutes and regulations including, but not limited to: registration and listing comply with an the Act s requirements, metaling, our manufacturing practice requirements as set (21 CFR Part 807), labemig (21 CFR Part 820); and if applicable, the electronic form in the quanty 35 stems (QB) registeriors 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter.
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Înternet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
## 510(k) Number: Device Name:
## Asahi APS Series Dialyzers
Indications for Use:
- APS Series Dialyzers are intended for use for hemodialysis treatment of patients A.. who have chronic renal failure or acute renal failure.
- APS Series Dialyzers must be used in accordance with the instructions for a B. physician familiar with hemodialysis and familiar with the conditions of the patient.
- APS Series Dialyzers have been tested in vitro and in confirmatory clinical studies एं under single or initial use and under reprocessing and reuse conditions for up to 15 reuse cycles. Based on the results from these evaluations, Asahi APS Series Dialyzers may be reprocessed for reuse on the same patient. If reprocessing and reuse is practiced, it is recommended that the reuse be done under the conditions as existed in the in vitro and confirmatory clinical studies as recommended immediately below. It is noted that the Asahi APS Series Dialyzers have not been tested for reuse when reprocessed with agents and/or processes other than these, and the performance of the dialyzers under other conditions are not known and cannot be recommended. Accordingly:
- The reprocessed dialyzer may be used only if the residual Total Cell Volume (1) (TCV) is at least 80% of the original TCV and if such dialyzer otherwise meets the acceptance criteria of these instruction for use and the instruction of the reprocessing system utilized. Furthermore, the policies, instructions, and criteria of the institution for reuse (e.g., concerning dialyzer performance, residual blood, and/or dialyzer leakage or damage) should be followed.
- The reprocessing agent may be either (1) 4% formaldehyde (also known as (2) formalin) in conjunction with the Seratronics Dialyzer Reprocessing Systems Dialyzer Reprocessing and Preparation (DRS4TM and DPS4TM), for manufactured by Seratronics, Inc., or (2) Renalin ® (peroxyacetic acid) in conjunction with the Renatron @ Dialyzer Reprocessing System (RS 8300), manufactured by Renal Systems, Inc.
- The instructions provided by the manufacturer of the chosen reprocessing agent (3) must be followed in reprocessing the dialyzer.
- The reprocessed dialyzer may be used only on dialysis systems equipped with (4) volumetric ultrafiltration controllers.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Concurence of CDRH, Office of Device Evaluation (ODE)
Image /page/4/Picture/14 description: The image shows a simple line drawing of a square with a line extending upwards from the top left corner. Another line extends from the top right corner, intersecting the first line to form a triangle shape above the square. The lines are black against a white background.
Prescription Use (per 21 CFR 801.109)
Image /page/4/Picture/16 description: The image shows a simple, thick-lined square. The square is oriented vertically, with its sides appearing to be of equal length. The lines forming the square are solid and black, contrasting with the white background. There are no other elements or details present in the image.
Over-the Counter Use
APS Series Dialyzers Asahi Medical Company., Ltd.
David b. Layman
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devic 510(k) Number.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.