K051187 · Asahi Kasei Medical Co., Ltd. · KDI · Jun 8, 2005 · Gastroenterology, Urology
Device Facts
Record ID
K051187
Device Name
REXEED SERIES DIALYZERS
Applicant
Asahi Kasei Medical Co., Ltd.
Product Code
KDI · Gastroenterology, Urology
Decision Date
Jun 8, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The line of the REXEED Series Dialyzers is a family of high permeability hollow fiber dialyzers intended for the treatment of patient with acute or chronic renal failure. REXEED-U/S are intended for single use and REXEED-UR/R for reuse. This is the same intended use as the existing APS cleared under K001250 and K041726, with the exception of using mixture of hydrogen peroxide and peroxyacetic acid (e.g., Renaline) as reprocessing agent.
Device Story
REXEED Series Dialyzers are high-permeability hollow fiber dialyzers for acute or chronic renal failure treatment. Input: blood and dialysate fluids. Operation: blood flows through crimped polysulfone hollow fibers; solutes/water removed via diffusion/convection. Output: filtered blood returned to patient. Used in clinical settings by physicians/trained staff. Models REXEED-U/S are single-use; REXEED-UR/R are reusable up to 15 cycles. Reprocessing requires hydrogen peroxide/peroxyacetic acid mixtures and volumetric ultrafiltration controllers. Benefits: effective waste removal and fluid management for renal patients.
Clinical Evidence
Bench testing and in vitro studies were conducted to verify design modifications, including fiber crimping and reprocessing agent compatibility. Clinical studies were performed to confirm performance under reuse conditions. All test results met acceptance criteria.
Indicated for hemodialysis treatment of patients with acute or chronic renal failure. Must be used by a physician familiar with hemodialysis. REXEED-UR/R models are indicated for reuse up to 15 cycles on the same patient when reprocessed using hydrogen peroxide and peroxyacetic acid mixtures (e.g., Renalin) in systems with volumetric ultrafiltration controllers, provided residual Total Cell Volume (TCV) is at least 80% of original.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
{0}------------------------------------------------
K051187 Page 101
CONFIDENTIAL
# 14.510 (K) SUMMARY
Asahi REXEED Series Dialyzers
| Manufacturer: | Asahi KASEI Medical Company, Ltd.<br>9-1, Kanda Mitoshirocho<br>Chiyoda-ku, Tokyo 101-8482<br>Japan |
|----------------------------------------|-----------------------------------------------------------------------------------------------------|
| Date summary was prepared: | April, 30, 2005 |
| Name(s) of the device: | Asahi REXEED Series Dialyzers |
| Identification of predicate device(s): | Asahi APS Series Dialyzers<br>Fresenius Optiflux<br>Gambro Polyflux |
### Description of the device:
The line of Asahi REXEED Series Dialyzers (hereinafter called REXEED) is a family of high permeability hollow fiber dialyzers intended for the treatment of patients with acute or chronic renal failure. REXEED-U/S are designed for single use and REXEED-UR/R are designed for reuse. The REXEED dialyers are constructed of reusable hollow fiber (polysulfone) membranes, housed within a plastic casing of styrene butadiene block copolymer and are gamma sterilized prior to shipment.
Modifications made in this 510(k) including the followings. (1) The shape modification of polysulfone hollow fibers to the crimped was implemented. The REXEED dialyzers with crimped hollow fibers is added to the existing APS Series Dialyzers (hereafter called APS) cleared under K001250 and; (2) the REXEED dialyzers membrane surface area ranges from 1.5m² to 2.5m², while that of APS ranges from 0.8 m² to 2.1 m² and; (3) mixture of hydrogen peroxide and peroxyacetic acid (e.g., Renalin®) is used as reprocessing agent) and; (4) accordingly, change in the Instruction For Use (IFU) in regard to reprocessing agent.
### Intended Use:
The line of the REXEED Series Dialyzers is a family of high permeability hollow fiber dialyzers intended for the treatment of patient with acute or chronic renal failure. REXEED-U/S are intended for single use and REXEED-UR/R for reuse. This is the same intended use as the existing APS cleared under K001250 and K041726, with the exception of using mixture of hydrogen peroxide and peroxyacetic acid (e.g., Renaline) as reprocessing agent.
{1}------------------------------------------------
Ko 511/87 Page 2 of 2
Special 510(k) Premarket Notification
CONFIDENTIAL
#### Evaluation of Design Modifications
As the basis for Asahi KASEI Medical's device evaluation studies and overall process for managing medical device risk, the company has performed a risk analysis using procedures based on ISO 14971 (2000) "Medical Devices - Application of Risk Management to Medical Devices". The risk analysis method used to assess the impact of the modification was Failure Modes and Effects Analysis (FMEA). Design verification tests based on the result of risk analysis and design input were performed to verify those modifications. All test results meet the acceptance criteria, and proved that those modifications to be appropriate.
#### Conclusion:
Asahi KASEI Medical made three modifications to the original APS cleared under K001250 and K041726. All design verification tests based on the result of risk analysis and design input proved that the REXEED dialyzers are substantially equivalent in intended use, design, principle of operation, technology, materials, specifications, and performance to the existing APS cleared under K001250 and K041726, Fresenius Optiflux F200A cleared under K003498, and Gambro Polyflux 24S and Polyflux 24R cleared under K010667, with the exception of using mixture of hydrogen peroxide and peroxyacetic acid (e.g., Renalin®).
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three wing-like shapes extending from its body. The words "U.S. Department of Health & Human Services" are arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 9200 Coffeet and Ante Administration Rockville 2006 orgate Boulevard Rockville MD 20850
APR 2 1 2009
Asahi Kasei Medical Co., Ltd. % David L. West, Ph.D. Vice President, Medical Device Development Quintiles Consulting 1801 Rockville Pike, Suite 300 ROCKVILLE MD 20852
Re: K051187
Trade/Device Name: REXEED Series Dialyzers, Models 15U, 15UR, 18U, 18UR, 21U, 21UR, 25U, 25UR, 15S, 15R, 18S, 18R, 21S, 21R, 25S, AND 25R Regulation Number: 21 CFR 876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: KDI and MSF Dated: May 9, 2005 Received: May 9, 2005
Dear Dr. West:
This letter corrects our substantially equivalent letter of June 8, 2005.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
|----------------|----------------------------------|----------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | (240) 276-0115 |
| 21 CFR 892.xxx | (Radiology) | (240) 276-0120 |
| Other | | (240) 276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of device adverse events under the MDR regulation (21CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please fo to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act form the Division of Small Manufacturers, International and Consumer Assistance at is toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.htm.
Sincerely yours,
Janine M. Morris
Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Ko51187
## CONFIDENTIAL
# 16.INDICATIONS FOR USE STATEMENT
# INDICATIONS FOR USE STATEMENT
| 510(k) Number: | |
|----------------------|-------------------------------|
| Device Name: | Asahi REXEED Series Dialyzers |
| Indications for Use: | |
- REXEED Series Dialyzers is intended for use for hemodialysis treatment of patients A. who have chronic or acute renal failure.
- who have chrome of acute renar land." B. REAEED Series Diary2015 mast 50 abd familiar with the conditions of the patient.
- physician familiar with nemodiaryins and under single or initial use एं REAEED Series Dialyzers nave beers under reprocessing and reuse conditions for alle KEAED-ORNY Borros Diary Loresults from these evaluations, REXEED-UR/R up to 15 use cycles. Dased on the reserve on the same patient. If reprocessing Series Dialyzels may be reprocessor ter read that the reuse be done under the conditions as existed in the in vitro and confirmatory clinical studies as conditions as "Existed" in "the "in "the "ht" "ht" "hat REXEED-UR/R Series Dialyzers recommisinced inimediately corrent reprocessed with agents and/or processes other than these, and the performance of the dialyzers under other conditions are not known and cannot be recommended. Accordingly:
- (1) The reprocessed dialyzer may be used only if the residual Total Cell Volume The reprocessed that 20% of the original TCV and if such dialyzer otherwise meets (TCV) is at least 60% of these instruction for use and the instruction of the the acceptance officilized. Furthermore, the policies, instructions, and criteria reprocessing system annized. Furthernore, mag dialyzer performance, residual of the ind/or dialyzer leakage or damage) should be followed.
- (2) The reprocessing agent may mixture of hydrogen peroxide and peroxyacetic The reprocessing agent may intinues of the Renatron® Dialyzer Reprocessing System (RS 8300), manufactured by Renal Systems, Inc.
- (3) The instructions provided by the manufacturer of the chosen reprocessing agent must be followed in reprocessing the dialyzer.
- (4) The reprocessed dialyzer may be used only on dialysis systems equipped with volumetric ultrafiltration controllers.
Concurrence of CDRH, Office of Device Evaluation (ODE)
ש ロ
Prescription Use (per 21 CFR 801.109) Over-the Counter Use
Division of Reproductive, Abdominal,
and Radiological Devices
K051187
510(k) Number
REXEED Series Dialyzers Asahi KASEI Medical Company., Ltd.
37 of 113
Page 37 of 38
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.