K120384 · Bd Medical-Medical Surgical Systems · LHI · Sep 12, 2012 · General Hospital
Device Facts
Record ID
K120384
Device Name
PROTECTOR, INJECTOR, CONNECTOR
Applicant
Bd Medical-Medical Surgical Systems
Product Code
LHI · General Hospital
Decision Date
Sep 12, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The PhaSeal system is a closed system drug transfer device (CSTD) that mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols and spills. The PhaSeal Protector also prevents microbial ingress.
Device Story
PhaSeal is a sterile, single-use closed system drug transfer device (CSTD). It utilizes a double-membrane technique with self-sealing elastomeric membranes and a bayonet fitting to connect components. A single-lumen cannula perforates the membranes for liquid transfer; retraction of the cannula seals the membranes. The system mechanically prevents environmental contaminants from entering and drug vapors/aerosols from escaping. Used in clinical settings by healthcare professionals to transfer hazardous drugs, the device minimizes exposure risks to personnel and maintains the sterility of drug vial contents. It acts as a mechanical barrier, ensuring drug integrity and safety during parenteral drug preparation and administration.
Clinical Evidence
No clinical studies were submitted. Evidence consists of two peer-reviewed publications evaluating sterility maintenance. Study 1 (n=1328 syringes) showed a 1.8% failure rate over 168 hours, with 98.2% probability of sterility. Study 2 confirmed these results, showing a 99.7% probability of sterility at 168 hours. Microbial ingress testing performed per FDA guidance showed zero microbial ingress on PhaSeal Protector samples.
Technological Characteristics
Sterile, single-use CSTD. Features self-sealing elastomeric membranes and bayonet fittings. Employs a double-membrane, single-lumen cannula design for mechanical containment of hazardous drugs and prevention of microbial ingress. No electronic or software components.
Indications for Use
Indicated for use as a closed system drug transfer device (CSTD) to minimize healthcare worker exposure to hazardous drug vapors, aerosols, and spills, and to prevent microbial ingress into sterile drug solutions during transfer.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
K120384
## Section 5 - 510(k) Summary
Date Prepared:
Submitted Name: Submitter Address:
Submitter Telephone No:
Name of Device:
Predicate Device(s):
Description of Device:
Indications for Use:
Technological Characteristics of New Device vs. Predicate Device
Summary of Non-Clinical Test Submitted to Determine Substantial Equivalence:
21 August 2012
SEP 12 2012
Becton, Dickinson and Company 1 Becton Drive Franklin Lakes, NJ 07417 USA
201 847 5473
PhaSeal Closed System Drug Transfer Device
PhaSeal Protector: K972527, K001368, K090634 PhaSeal Injector: K972527, K980381, K001368, K023747, K060866, K092782, K110023 PhaSeal Connector: K972527, K001368, K092782
The PhaSeal® System is a sterile single-use closed system drug transfer device. The closed transfer of liquid takes place through a double membrane technique utilizing self-sealing elastomeric membranes, tightly fitted together through a bayonet fitting on all PhaSeal components. A single lumen cannula perforates the double membranes for the transfer of liquid. When the cannula is retracted the membranes seal off and transfer of environmental contaminants into the system and/or escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols and spills. The PhaSeal® Protector also prevents microbial ingress.
The PhaSeal system is a closed system drug transfer device (CSTD) that mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols and spills. The PhaSeal Protector also prevents microbial ingress.
The technological characteristics of the subject device are IDENTICAL to those of the predicate device(s)
Two peer-reviewed publications - see Appendix II are submitted to support the amended Indications for Use Statement. As well as Mircobial Ingress testing on the interface between the drug vial and PhaSeal Protector.
{1}------------------------------------------------
- 1. "Utility of the PhaSeal Closed System Drug Transfer Device," Derek M. McMichael, PharmD; Dawn Moore Jefferson, PharmD; E. Thomas Carey; Pharm D; Ryan A. Forrey, PharmD, MS; Susan M. Spivey, PharmD. DDS: John M. Mulvaney, MHA: James A. Jorgenson, RPh, MS; and R. Douglas Haughs, BS, Am J Pharm Benefits, 2011; 3(1): 9-16).
- 2. "Second Look Utilization of a Closed-System Transfer Device (PhaSeal)," E.Thomas Carey, Pharm D; Ryan A. Forrey, PharmD, MS; Douglas Haughs, BS; Dawn Moore Jefferson, PharmD; James A. Jorgenson, RPh, MS; Derek M. McMichael, PharmD; John M. Mulvaney, MHA; Susan M. Spivey, PharmD, DDS, Am J Pharm Benefits, 2011; 3(6): 11-18).
The objective of the first study, entitled "Utility of the PhaSeal Closed System Drug Transfer Device," was to assess the ability of the PhaSeal closed system drug transfer device to prevent contamination of parenteral drug products. The PhaSeal closed system drug transfer device was applied to vials containing sterile culture media. The vials were entered using the PhaSeal system and samples were removed at 24, 48, 72, 96 and 168 hours. Samples were tested by an independent microbiology laboratory for evidence of contamination. A total of 1328 syringes were produced at 4 different institutions. Visual, microscopic, and microbiologic subculture analyses were performed. A failure rate of 1.8% was observed, and supported the alternate hypothesis at the 99% confidence level that the PhaSeal system is capable of maintaining sterility in a controlled environment. Secondary analysis of the data was conducted based on time to failure. The analysis indicated that at the 16-hour mark there is a 98.2% probability that the vials will not be contaminated. Results of this study show that the PhaSeal closed system drug transfer device provides a mechanical barrier to the entry of contaminants into sterile solutions. The study also demonstrates that solutions are expected to remain sterile for up to 168 hours.
The objective of the second study, entitled "Second Look at the Utilization of Closed-System Transfer Devices (PhaSeal)," was to confirm and refine the results and conclusions of the first study. Aliquots of sterile culture medium were transferred from test vials of sterile culture medium to intravenous bags of sterile medium over a 7-day
{2}------------------------------------------------
time period utilizing the PhaSeal Closed System Transfer Device. The IV bag test samples were then held under controlled incubation for 14 days and monitored for evidence of contamination by an independent microbiology laboratory. The results indicate that at the 168-hour mark, the probability of failure was 0.3%. In other words, at 168 hours, the probability that the vial would remain sterile is 99.7%. Although the use of Closed System Transfer Devices has traditionally focused on reducing exposure of healthcare workers to hazardous substances, this study further demonstrates the ability of the PhaSeal System to maintain the sterility of drug vial contents.
Finally, Microbial Ingress testing on the interface between the drug vial and protector was performed in accordance with "Guidance for Industry and FDA Staff - Intravascular Administration Sets Premarket Notification Submissions [510(k)]". Based on the results of the study, microbial ingress was NOT observed on any of the samples with the PhaSeal Protector. These results demonstrate the subject device's self-sealing elastomeric membranes are in fact capable of preventing the transfer of microbes from outside to the inside of the sterile drug solutions.
Clinical studies are not submitted to support substantial equivalence.
Conclusions drawn from publication entitled "Utility of the PhaSeal Closed System Drug Transfer Device": The analysis indicated that at the 168-hout mark there is a 98.2% probability that the vials will not be contaminated. Results of this study show that the PhaSeal closed system drug transfer device provides a mechanical barrier to the entry of contaminants into sterile solutions. The study also demonstrates that solutions are expected to remain sterile for up to 168 hours.
Conclusions drawn from the publication entitled "Second Look at the Utilization of Closed-System Transfer Devices (PhaSeal)": The results indicate that at the 168-hour mark, the probability of failure was 0.3%. In other words, at 168 hours, the probability that the vial would remain sterile is 99.7%. Although the use of Closed System Transfer Devices has traditionally focused on reducing exposure of
Summary of Non-Clinical Test Submitted to Determine Substantial Equivalence
Conclusions Drawn from Non-Clinical Tests
{3}------------------------------------------------
healthcare workers to hazardous substances, this study further demonstrates the ability of the PhaSeal System to maintain the sterility of drug vial contents.
Conclusions drawn from Microbial Ingress studies performed in accordance with "Guidance for Industry and FDA Staff - Intravascular Administration Sets Premarket Notification Submissions [510(k)]": Based on the results of the study, microbial ingress was NOT observed on any of the samples with the PhaSeal Protector. These results demonstrate the subject device's self-sealing elastomeric membranes are in fact capable of preventing the transfer of microbes from outside to the inside of the sterile drug solutions.
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular fashion around the eagle.
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. John Roberts Regulatory Affairs Specialist Becton, Dickinson and Company - Medical Surgical Systems 1 Becton Drive Franklin Lakes, New Jersey 07417
SEP 12 2012
Re: K120384
Trade/Device Name: PhaSeal® - A Closed System Transfer Device Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: LHI Dated: August 23, 2012 Received: August 24, 2012
Dear Mr. Roberts:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
### Page 2- Mr. Roberts
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reportung (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference.to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
for Timothy
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known):
PhaSeal® – A Closed System Transfer Device Device Name:
Indications for Use:
The PhaSeal system is a closed system drug transfer device (CSTD) that mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug or vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols and spills. The PhaSeal Protector also prevents microbial ingress.
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | X | AND/OR | Over-The-Counter Use<br>(21 CFR 801 Subpart C) | |
|-------------------------------------------------|---|--------|------------------------------------------------|--|
|-------------------------------------------------|---|--------|------------------------------------------------|--|
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER
PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Page ___ of ___ | <img alt="Signature" src="signature.png"/><br>(Division Sign-Off)<br>Division of Anesthesiology, General Hospital<br>nfection Control, Dental Devices |
|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.