PHASEAL, SYSTEM FOR SEALED HANDLING OF CHEMOTHERAPEUTIC AGENTS
K972527 · Quality System Consulting · FPA · Sep 18, 1997 · General Hospital
Device Facts
Record ID
K972527
Device Name
PHASEAL, SYSTEM FOR SEALED HANDLING OF CHEMOTHERAPEUTIC AGENTS
Applicant
Quality System Consulting
Product Code
FPA · General Hospital
Decision Date
Sep 18, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
PhaSeal® closed system for the preparation and administration of parenteral drugs
Device Story
PhaSeal is a closed system for handling parenteral, particularly cytotoxic, drugs. System comprises four components: Protector 20 (vial adapter), Injector Luer (syringe-to-component transfer), Connector Luer-Lock (syringe-to-IV line connection), and Infusion Set (IV administration). Principle of operation: patented double-membrane technique using elastomeric seals to prevent leakage during connection/disconnection; built-in pressure equalization chamber in Protector 20 prevents aerosol leakage by managing pressure/vacuum during fluid transfer. Used in clinical settings by healthcare professionals for drug preparation, administration, and waste handling. System ensures sealed transfer of diluent, drug, or air between vials, syringes, and IV lines, reducing provider exposure to hazardous substances.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
System components utilize elastomeric membranes for sealed connections. Pressure equalization is achieved via an integrated expansion chamber. Components include Luer fittings for compatibility with standard syringes and IV lines. Non-vented infusion set design. Single-use, sterile components.
Indications for Use
Indicated for the preparation and administration of parenteral drugs, particularly cytotoxic drugs, in a closed system to minimize exposure to hazardous drug aerosols and spills during reconstitution, administration, and disposal.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Becton Dickinson SafetyGlide Shielding IM Injection Needle (K951254)
Sherwood Davis & Geck Monoject Luer Adapter (K940961)
Codan Medlon IV Administration Set Catalog #443
Submission Summary (Full Text)
{0}------------------------------------------------
APP VI
K972527
DATA Sept. 18, 1997
## SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## Carmel Pharma AB PhaSeal® closed system for the preparation and administration of parenteral drugs
### Carmel Pharma Contact Person
Britt Novén Manager, Regulatory Affairs Carmel Pharma AB Box 5352 S-402 28 Göteborg, Sweden Phone: +46 31-703 26 90 +46 31-703 41 20 Fax:
### Device Name
PhaSeal® closed system for the preparation and administration of parenteral drugs.
- Protector 20 Injector Luer Connector Luer Lock Infusion Set
### Predicate Devices
- Bristol-Myres Squibb/Mead Johnson Oncology Products (Survival Technology), CytoGuard® . Drug Reconstitution Device, K883614
- Becton Dickinson, SafetyGlide™ Shielding IM Injection Needle, K951254 ●
- Sherwod Davis & Geck, Monoject® Luer Adapter, K940961 .
- Codan Medlon. IV Administration Set Catalog #443, preammendment device ●
### Product Description, Function, Safety and Efficacy
The PhaSeal™ closed system for preparation and administration of parenteral drugs has four component devices that are dedicated to each other to create the system. These single use devices are designed to promote safe handling of medications, particularly cytotoxic drugs. Leakage of drug into the environment is effectively avoided during all three phases of drug handling when the PhaSeal system is used: the preparation of the drug, the administration of the drug to the patient, and waste handling.
All drug transferring utilizes a patented double membrane technique. Each element is sealed off with an elastomeric membrane cover. The membranes are joined together and transfer is made via a specially cut injection cannula. When the elements of the system are separated after transfer, the membranes act as tight seals that prevent leakage.
PhaSeal utilizes a built-in, patented pressure equalization technique. Air passes from the vial into a special expansion chamber. Neither excess pressure nor vacuum can occur during drug preparation. This effectively prevents aerosol leakage.
- PhaSeal Protector 20 Drug Vial Transfer Adapter ●
The Protector 20 is fitted to the drug vial and is used as a docking station between the drug vial for the parenteral drug and Injector Luer. In addition the Protector 20 equilibrates the pressure
{1}------------------------------------------------
difference which occurs when fluid or air is added or removed from the drug vial. Liquid transfer takes place through tightly fitting elastomeric membranes to minimize exposure to potentially hazardous drug aerosols and spills that can occur during the reconstitution, administration and disposal processes.
### PhaSeal Injector Luer - Drug Transfer Needle Device ●
The Injector Luer has an encapsulated cannula that is permanently locked onto a syringe using a Luer fitting. Sealed transfer of diluent, drug or air, between the single-use syringe and the various components in the system can be made via the Injector Luer in both the preparation and administration phases.
### PhaSeal Connector Luer-Lock - Luer Lock Device .
The Connector Luer Lock ensures a sealed connection between the single-use syringe and Injector Luer and the patient's IV line. With the help of the Connector Luer Lock, injections can be made without drug spillage.
#### PhaSeal Infusion Set - Intravascular Administration Set .
The Infusion Set is a non-vented infusion device that has a built-in connector to be used as a way of making additions of parenteral drugs to infusion fluids in a closed system. The Infusion Set may be used to administer the infusion fluid.
### Comparison of Predicate Devices/Equivalence
#### PhaSeal Protector 20 .
Is substantially equivalent to the CytoGuard Drug Reconstitution Device in that they both serve as a docking station for drug vials and act to minimize or eliminate the aerolization of drug being transferred from the drug vial into a single-use syringe.
#### ● PhaSeal Injector Luer
Is substantially equivalent to the Becton Dickinson SafetyGlide Shielding IM Injection Needle in that they both transfer fluid and drug between drug vials and infusate. In addition they both have protected needles which provide protection from accidental needle sticks.
#### . PhaScal Connector Luer Lock
Is substantially equivalent to the Sherwod Davis & Geck, Monoject® Luer Adapter in that they both provide a means of connecting various components of a drug delivery system via a luer fitting.
#### . PhaSeal Infusion Set
Is substantially equivalent to the Codan Medlon, IV Administration Set - Catalog #443 in that they both use similar components and provide a pathway to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. (Neither device provides the needle or catheter.)
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows a black and white logo for the Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol of three stylized human profiles facing right. The profiles are stacked on top of each other, creating a sense of depth and unity. The overall design is simple and professional, conveying a sense of authority and trustworthiness.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Fred Schlador Consultant Quality System Consulting C/O Carmel Pharma AB 1425 Cressa Court 92009 Carlsbad, California
SEP 1 8 1997
Re : K972527 PhaSeal™ System For Sealed Handling Of Trade Name: Chemotherapeutic AG Regulatory Class: II Product Code: LHI Dated: June 30, 1997 Received: July 7, 1997
Dear Mr. Schlador:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਜੋ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note:
{3}------------------------------------------------
Page 2 - Mr. Schlador
this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to Other qeneral information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
27. Indications for Use Statement:
510(k) Number (if known): K472507
# Device Name: PhaSear® closed system for the preparation and administration of parenteral drugs
Indications for Use:
)
පිහිටා බව පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි ප
)
## PhaSeal Protector 20 - Drug Vial Transfer Adapter
The Protector 20 is fitted to the drug vial and is used as a docking station between the drug vial for the parenteral drug and Injector Luer. In addition the Protector 20 equilibrates the pressure difference which occurs when fluid or air is added or removed from the drug vial. Liquid transfer takes place through tightly fitting elastomeric membranes to minimize exposure to potentially hazardous drug aerosols and spills that can occur during the reconstitution, administration and disposal processes.
### PhaSeal Iniector Luer - Drug Transfer Needle Device
The Injector Luer has an encapsulated cannula that is permanently locked onto a syringe using a Luer fitting. Sealed transfer of diluent, drug or air, between the single-use syringe and the various components in the system can be made via the Injector Luer in both the preparation and administration phases.
## PhaSeal Connector Luer-Lock - Luer Lock Device
The Connector Luer Lock ensures a sealed connection between the single-use syringe and Injector Luer and the patient's IV line. With the help of the Connector Luer Lock, injections can be made without drug spillage.
## PhaSeal Infusion Set - Intravascular Administration Set
The Infusion Set is a non-vented infusion device that has a built-in connector to be used as a way of making additions of parenteral drugs to infusion fluids in a closed system. The Infusion Set may be used to administer the infusion fluid.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation
Patricia Casceneti
Prescription Use (Per 21 CFR 801.109)
Over the Counter Use √
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.