PHASEAL, CLOSED SYSTEM FORHANDLING OF PARENTERAL DRUGS, ADDITIONAL ADMINISTRATION DEVICES. C80 INFUSION ADAPTER, A10 PRO
K980381 · Carmel Pharma Ab. · LHI · Mar 3, 1998 · General Hospital
Device Facts
Record ID
K980381
Device Name
PHASEAL, CLOSED SYSTEM FORHANDLING OF PARENTERAL DRUGS, ADDITIONAL ADMINISTRATION DEVICES. C80 INFUSION ADAPTER, A10 PRO
Applicant
Carmel Pharma Ab.
Product Code
LHI · General Hospital
Decision Date
Mar 3, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
PhaSeal Infusion Adapter - Intravascular Administration Set The Infusion Adapter serves as the connecting part between the IV bag and an external IV line. (Example IV regulators.) The Infusion Adapter has a built in Connector which makes it possible to admix drugs into the infusion solution using the sealed double membrane technique. PhaSeal Protection Cap - Special accessories The Protection Cap is intended to be used as a mechanical cover for the membrane in the bayonet fitting of PhaSeal devices The Protection Cap mates with the other PhaSeal components equipped with the bayonet fitting. One end of the Protection Cap has a male bayonet fitting and in the other a female bayonet fitting. PhaSeal Secondary Set - Intravascular Administration Set The Secondary Set is a non-vented infusion set used when drug is handled as an admixture and is administered via Intravenous infusion. The Secondary Set has a built in Connector which makes it possible to admix drugs into the infusion solution using the sealed PhaSeal technique. PhaSeal Extension Set - Intravascular Administration Set The Extension Set serves as the port for bolus injection with PhaSeal if there is no Luer Lock fitting, for Connector Luer Lock in the patients IV line. The Extension Set has a built in Connector which makes it possible inject drugs into the IV line of the patient using the sealed double membrane technique.
Device Story
PhaSeal is a closed system for parenteral drug preparation and administration; designed to prevent leakage of hazardous medications, particularly cytotoxic drugs. System utilizes patented double membrane technique; each component sealed with elastomeric membrane cover. During drug transfer, membranes join; specially cut injection cannula facilitates flow. Upon separation, membranes act as tight seals. Components include Infusion Adapter (connects IV bag to external line), Protection Cap (mechanical cover for bayonet fittings), Secondary Set (non-vented infusion set for admixtures), and Extension Set (port for bolus injection). Used in clinical settings by healthcare providers to ensure safe medication handling; protects clinicians from exposure to hazardous drugs; maintains sterility and containment during drug transfer and waste handling.
Clinical Evidence
No clinical data provided; substantial equivalence is based on design, material, and functional comparison to legally marketed predicate devices.
Technological Characteristics
System utilizes elastomeric membrane covers and injection cannulas for double-membrane sealing. Components feature bayonet fittings. Materials include unspecified elastomers and plastics consistent with IV administration sets. Single-use, non-vented design. No electronic or software components.
Indications for Use
Indicated for use in the preparation and administration of parenteral drugs, including cytotoxic medications, to patients requiring intravenous therapy. Used by clinicians to facilitate safe drug handling and prevent environmental leakage.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
Codan Medlon, IV Administration Set (preamendment device)
Codan Medlon Inc. (BC 103) Special Accessories Luer Lock Plug
Codan Medloc Inc. Vented Set (BC 589)
Submission Summary (Full Text)
{0}------------------------------------------------
K98038
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## Carmel Pharma AB
MAR - 3 1998
## PhaSeal® closed system for the preparation and administration of parenteral drugs
### Carmel Pharma Contact Person
Britt Novén Manager, Regulatory Affairs Carmel Pharma AB Box 5352 S-402 28 Göteborg, Sweden +46 31-703 04 00 Phone: Fax: +46 31-703 04 00 (Visitors: Flöjelbergsgatan 16B, Mölndal)
### Device Name
PhaSeal® closed system for the preparation and administration of parenteral drugs.
Infusion Adapter Protection Cap Secondary Set Extension Set
### Predicate Devices
- Codan Medlon, IV Administration Set -- Catalog Rev 3/1/93 B&P, preamendment device .
- Carmel Pharma ab, PhaSeal, a closed system for parenteral drugs, K972527 .
### Product Description, Function, Safety and Efficacy
PhaSeal is a closed system for preparation and administration of parenteral drugs where the component devices are dedicated to each other to create the system. These single use devices are designed to promote safe handling of medications, particularly cytotoxic drugs. Leakage of drug into the environment is effectively avoided during all three phases of drug handling when the PhaSeal system is used: the preparation of the administration of the drug to the patient, and waste handling.
All drug transferring utilizes a patented double membrane technique. Each component devices is sealed off with an elastomeric membrane cover. The membranes are joined together and transfer is made via a specially cut injection cannula. When the component devices of the system are separated after transfer, the membranes act as tight seals that prevent leakage.
- · PhaSeal Infusion Adapter Intravascular Administration Set
The Infusion Adapter serves as a the connecting part between the IV bag and an external IV line. (Example IV regulators.) The Infusion Adapter has a built in Connector which makes it possible to admix drugs into the infusion solution using the sealed double membrane technique.
{1}------------------------------------------------
1780581
- · PhaSeal Protection Cap Special accessories
The Protection Cap is intended to be used as a mechanical cover for the membrane in the bayonet fitting of PhaSeal devices The Protection Cap mates with the other PhaSeal components equipped with the bayonet fitting. One end of the Protection Cap has a male bayonet fitting and in the other a female bayonet fitting.
- · PhaSeal Secondary Set Intravascular Administration Set
The Secondary Set is a non-vented infusion set used when drug is handled as an admixture and is administered via Intravenous infusion. The Secondary Set has a built in Connector which makes it possible to admix drugs into the infusion solution using the sealed PhaSeal technique.
- · PhaSeal Extension Set Intravascular Administration Set
The Extension Set serves as the port for bolus injection with PhaSeal if there is no Luer Lock fitting, for Connector Luer Lock in the patients IV line. The Extension Set has a built in Connector which makes it possible inject drugs into the IV line of the patient using the sealed double membrane technique.
## Comparison of Predicate Devices/Equivalence
#### ● PhaSeal Infusion Adapter
Is substantially equivalent to Codan Medlon Inc. Vented Set (C 302 and C430) and Carmel Pharma ab, Infusion Set in that they all use similar components and provide a pathway to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. (Neither device provides the needle or catheter.)
#### . PhaSeal Protection Cap
Is substantially equivalent the Codan Medlon Inc. (BC 103) Special Accessories Luer Lock Plug in both being a protection and mechanical cover for a connections to IV systems. Protection Cap is substantially equivalent with Carmel Pharma ab Connector in design, the bayonet fitting and material.
#### PhaSeal Secondary Set .
Is substantially equivalent to Codan Medlon Inc. Vented Set (C 302) and Carmel Pharma ab, Infusion Set in that they all use similar components and provide a pathway to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. (Neither device provides the needle or catheter.)
#### ● PhaSeal Extension Set
Is substantially equivalent to Codan Medloc Inc. Vented Set (BC 589) and Carmel Pharma ab, Infusion Set in that they all use similar components and provide a pathway to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. (Neither device provides the needle or catheter.)
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with three heads, representing health, services, and people. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA", which is arranged in a circular fashion around the figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR - 3 1998
Mr. Britt Novén Manager, Regulatory Affairs Carmel Pharma AB Box 5352 S-402 28 Göteborg, Sweden
K980381 Re : PhaSeal® closed system for the preparation Trade Name: and administration of parenteral drugs Regulatory Class: II Product Code: LHI January 30, 1998 Dated: February 2, 1998 Received:
Dear Mr. Novén:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does
{3}------------------------------------------------
Page 2 - Mr. Novén
not affect any obligation you might have under sections 531
through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to Other general information on your responsibilities under the Act may be wor obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fde.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
A. Timothy Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radioloqical Health
Enclosure
{4}------------------------------------------------
## 27. Indications for Use Statement:
510(k) Number (if known):
Device Name: PhaSeal® closed system for the preparation and administration of parenteral drugs
Indications for Use:
### PhaSeal Infusion Adapter - Intravascular Administration Set
The Infusion Adapter serves as the connecting part between the IV bag and an external IV line. (Example IV regulators.) The Infusion Adapter has a built in Connector which makes it possible to admix drugs into the infusion solution using the sealed double membrane technique.
## PhaSeal Protection Cap - Special accessories
The Protection Cap is intended to be used as a mechanical cover for the membrane in the bayonet fitting of PhaSeal devices The Protection Cap mates with the other PhaSeal components equipped with the bayonet fitting. One end of the Protection Cap has a male bayonet fitting and in the other a female bayonet fitting.
## PhaSeal Secondary Set - Intravascular Administration Set
The Secondary Set is a non-vented infusion set used when drug is handled as an admixture and is administered via Intravenous infusion. The Secondary Set has a built in Connector which makes it possible to admix drugs into the infusion solution using the sealed PhaSeal technique.
## PhaSeal Extension Set - Intravascular Administration Set
The Extension Set serves as the port for bolus injection with PhaSeal if there is no Luer Lock fitting, for Connector Luer Lock in the patients IV line. The Extension Set has a built in Connector which makes it possible inject drugs into the IV line of the patient using the sealed double membrane technique.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation
Patricia Crescenti
(Division Sign-Off)
Division of Dental, Infection Control,
and General Hospital Devices
510(k) Number X98038
Over the Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.