The Equivital™ Vital Signs Physiological Monitor is an ambulatory multiparameter vital signs telemetry device intended for monitoring of adults (16 - 65 years) in hospital care facilities, the home, workplace, and alternate care settings. The device consists of a chest belt harness and a body worn electronics module (SEM) supported by the chest belt. The device collects and transmits ECG data and rate, respiration data and rate, skin temperature, body orientation and motion. The monitor is indicated for use as a general patient monitor, to provide physiological information as part of an occupational welfare monitoring system, and for general research and performance measurement purposes. Federal Law (US) restricts this device to sale by or on the order of a physician.
Device Story
Equivital EQ02 is an ambulatory multiparameter telemetry monitor consisting of a chest belt harness with integrated skin electrodes and expansion sensors, plus a body-worn electronics module (SEM). It continuously monitors ECG (two views), respiratory breathing frequency (via thoracic movement), skin temperature, body orientation, and activity level. The device digitizes and transmits physiological data wirelessly via low-power radio or Bluetooth to a receiving display device (standalone PC or third-party monitoring application). It provides secondary heart rate detection via R-wave processing and secondary skin temperature via infrared thermometer. The system alerts users if physiology exceeds predefined boundaries. Used in hospitals, homes, and workplaces; operated by clinicians or patients. Benefits include continuous, non-invasive physiological tracking for clinical, occupational, or research purposes.
Clinical Evidence
Bench testing only. Performance verified against standards: EN60601-1 (safety), IEC60601-1-2 (EMC), ISO 10993-1 (biocompatibility), ANSI/AAMI EC13 (cardiac monitors), ASTM E1965-98 (infrared thermometers), and IP67 (environmental). No clinical trial data presented.
Technological Characteristics
System includes chest belt harness and battery-powered electronics module. Sensors: skin electrodes, expansion sensor, movement sensor, infrared thermometer. Connectivity: Bluetooth and low-power 40.68MHz radio. Power: rechargeable or primary disposable cells. Software developed per EN60601-1-4. Materials: biocompatible per ISO 10993-1. Form factor: body-worn.
Indications for Use
Indicated for ambulatory monitoring of adults (16-65 years) in hospitals, homes, workplaces, and alternate care settings. Used as a general patient monitor for physiological data collection, occupational welfare monitoring, and research/performance measurement.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
# KII3054 hidalgo
FEB 1 7 2012
## 510(k) Summary Equivital Vital Signs Physiological Monitor EQ02
| Submitter: | Hidalgo Ltd |
|------------|------------------------------------------------------------------------------------------------------------|
| | Unit F<br>Buckingway Business Park<br>Anderson Road<br>Swavesey<br>Cambridge<br>CB24 4UQ<br>United Kingdom |
| Contact: | Dr Ekta Sood - Clinical Director<br>Tel: +441954233430 |
| Date: | 2011-10-10 |
#### Name of Device:
| Device Trade Name: | Equivital™ Vital Signs Physiological Monitor EQ0 |
|---------------------|--------------------------------------------------|
| Device Common Name: | Ambulatory Patient Monitor |
#### Device Classification and Product Code:
- Class Ilb EU:
Medical Device Directive: 93/42/EEC as amended by 2007/47/EC
- USA: Class II
- Panel: Cardiovascular
Product Code: MHX
Regulation Number: 870.1025
Classification Name: Monitor, Physiological, Patient
(With Arrhythmia Detection Or Alarms)
Image /page/0/Picture/16 description: The image shows the logo for Jaltek. The logo consists of the word "jaltek" in lowercase letters, with a stylized globe above it. The globe appears to have land masses and oceans depicted on its surface.
Hidalgo Ilmited - A Jaltek Group Company
Unit F, Buckingway Business Park, Anderson Road, Swavesey, Cambridge, CB24 4UQ Fax: +44 (0) 1954 233431 Web: www.hidalgo.co.uk Tel: +44 (0) 1954 233430 Registered in England under Company No. 3315718. VAT No. 688 647 072.
Page 1 of 4
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "hidalgo" followed by a series of circles. The first circle is white, while the remaining four circles are black. Above the word "hidalgo" is the text "K113054".
#### Predicate Devices
The device has been compared to the following predicate:
| Hidalgo | EQ01 | - K061993 |
|------------------------|------------------------------------------|-----------|
| Additional Predicates: | | |
| GMP Wireless Medicine | LifeSync™ | - K030795 |
| Dymedix Inc | Re-usable Respiratory Effort Belt Sensor | - K040605 |
| Nexan Inc | NX-300 | - K003520 |
| Protocol Systems | Propaq Encore 200 | - K012451 |
| Respironics | Actiheart® | - K052489 |
| Vivometrics | Life Shirt™ Real Time | -K043604 |
#### Device Description
The Equivital™ Vital Signs Physiological Monitor comprises two components:
i) A chest belt containing skin electrodes and an expansion sensor
ii) A battery powered electronics module which connects directly to the chest belt and which acts to record, digitise and transmit the physiological information wirelessly to a receiving display device.
The device offers continuous monitoring of two views of the user's heart electrical activity (ECG) and respiratory breathing frequency inferred from thoracic cavity movement and uses this data to derive a Heart Rate and a Breathing Effort Rate.
The sensor also provides the following information:
- ECG and Respiration physiological waveforms .
- An indication of the user's activity level derived from a movement detection sensor .
- Body orientation and indications for change in orientation .
- Skin surface temperature .
- Alternate secondary measurement of heart rate based on the detection of the user's R . wave using a separate hardware processing function
- . Indications and alerts if physiology exceeds predefined boundaries
- Alternate secondary skin temperature measurement using infrared thermometer .
The device offers two variants for the wireless interface: a low power radio interface and a general purpose interface using Bluetooth™ technology.
The device offers two battery power options for the user, rechargeable or primary disposable cells.
The wireless data provided by the sensor may be viewed using a standalone PC based viewing application, or integrated into third party monitoring applications.
{2}------------------------------------------------
# KII.3054 hidalao……………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………
#### intended Use Summary
The Equivital™ Vital Signs Physiological Monitor is an ambulatory multiparameter vital signs telemetry device intended for monitoring of adults (16 - 65 years) in hospital care facilities, the home, workplace, and alternate care settings.
The device consists of a chest belt harness and a body worn electronics module (SEM) supported by the chest belt.
The device collects and transmits ECG data and rate, respiration data and rate, skin temperature, body orientation and motion.
The monitor is indicated for use as a general patient monitor, to provide physiological information as part of an occupational welfare monitoring system, and for general research and performance measurement purposes.
### Technological Characteristics
The device does not contain tissues of animal origin as referred to in EC commission directive 2003/32/EC.
Substantial equivalence has been demonstrated by technological review and comparison fo the predicate Equivital EQ01 device and verified by the Design Control Process.
In addition substantial equivalence remains to the Equivital EQ01 predicate devices cited in that device's submission.
| Function | Predicate Device | Additional Predicates |
|----------------------------------------------------------------------------------------|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| ECG Heart Rate,<br>Breathing Rate<br>Temperature and<br>General Physical<br>Properties | Hidalgo Equivital EQ01 –<br>K061993 | Protocol Systems<br>Propaq Encore 200 –<br>K012451<br>Nexan Inc NX-300 –<br>K003520 |
| Chest Expansion<br>Respiration | Hidalgo Equivital EQ01 –<br>K061993 | Dymedix Re-usable<br>Respiratory Effort Belt –<br>K040605<br>Nexan Inc NX-300 –<br>K003520<br>Vivometrics LifeShirt<br>Real Time –<br>K043604 |
| Activity/Motion<br>Detection | Hidalgo Equivital EQ01 –<br>K061993 | Respironics Actiheart –<br>K052489<br>Vivometrics LifeShirt<br>Real Time – |
Technological Characteristics Compared:
Page 3 of 4
{3}------------------------------------------------
K113054
| Function | Predicate Device | Additional Predicates |
|---------------------------------------------|-------------------------------------|-------------------------------------------------|
| | | K043604 |
| Bluetooth telemetry<br>and ECG transmission | Hidalgo Equivital EQ01 -<br>K061993 | GMP Wireless<br>Medicine - LifeSync™<br>K030795 |
| Low Power 40.68MHz<br>PAN | Hidalgo Equivital EQ01 -<br>K061993 | Respironics Actiheart -<br>K052489 |
#### Performance Measurement Summary
Performance measurement and review to the applicable sections of the following standards has been undertaken and successfully demonstrated as recommended by available guidance from the agency:
- EN60601-1: Medical Electrical Equipment Part 1 General Requirements for Safety. .
- . IEC60601-1-2: 2001 Medical Electrical Equipment Part 1-2: Collateral Standard: Electromagnetic Compatibility.
- ISO 10993-1: Biological Evaluation of Medical Devices: Evaluation and Testing. ●
- ANSI/AAMI EC13: 2002 Cardiac Monitors, heart rate meters, and alarms. The device was . not evaluated against the criteria for defibrillation and pacemaker immunity as it is not indicated for use in either of these cases.
- ASTM E1965 98 (2009) Standard Specification for Infrared Thermometers for internittent . determination of patient temperature.
- . IP67: Environmental Performance Testing
The device software has been developed using a structured software lifecycle which meets the requirements of EN60601-1-4 Programmable Systems, which also applies to ongoing maintenance of the software.
#### Substantial Equivalence Conclusion
EQ02 has the same intended use and fundamental scientific technology as EQ01. The changes made do not raise any new questions of safety and effectiveness.
In addition the testing undertaken has demonstrated that the Equivital EQ02 device is as safe, as effective and performs as well as the legally marketed predicate Equivital EQ01 device.
To conclude substantial equivalence of the Equivital™ Vital Signs Physiological Monitor EQ02 has been demonstrated to the identified predicate Equivital™ Vital Signs Physiological Monitor EQ01.
Page 4 of 4
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
FEB 1 7 2012
Hidalgo Limited c/o Dr. Ekta Sood Unit F, Buckingway Business Park Anderson Road Swavesey Cambridge Cambridgeshire CB24 4UQ GB
Re: K113054 Trade/Device Name: Equivital Regulation Number: 21 CFR 870.1025 Regulation Name: Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms) Regulatory Class: Class II Product Code: MHX Dated: January 27, 2012 Received: January 30, 2012
Dear Dr. Sood:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{5}------------------------------------------------
#### Page 2 - Dr. Ekta Sood
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
R. D. Zaharov, M.P.
fo
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### Indications for Use
510(k) Number (if known):_K113054
Device Name: Equivital™ EQ02 Vital Signs Physiological Monitor
Indications For Use:
The Equivital™ Vital Signs Physiological Monitor is an ambulatory multiparameter vital signs telemetry device intended for monitoring of adults (16 - 65years) in hospital in the mital in the mater vital care facilities, the home, workplace for monitoning of addits (11)
facilities, the home, workplace and alternate care settings.
The device consists of a chest belt harness and a body worn electronics module (SEM) supported by the chest belt.
The device collects and transmits ECG data and rate, respiration data and rate, skin
temperature, body orientation and motion temperature, body orientation and motion.
The monitor is indicated for use as a general patient monitor, to provide physiological information as part of an occupational welfare monitor, to provide physiology
research and performance measuroment purpos research and performance measurement purposes.
Federal Law (US) restricts this device to sale by or on the order of a physician
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(Please do not write below this line-continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
ion Slanovascular Devices inn of Car
510(k) Number K113054
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.