The Equivital™ Vital Signs Physiological Monitor is an ambulatory multiparameter vital signs telemetry device intended for monitoring of adults (16 - 65years) in hospital care facilities, the home, workplace, and alternate care settings. The device consists of a chest belt hamess and a body wom electronics module (SEM) supported by the chest belt). The device collects and transmits ECG data and rate, respiration data and rate, skin temperature, body orientation and motion. The monitor is indicated for use as a general patient monitor, to provide physiological information as part of an occupational welfare monitoring system, and for general research and performance measurement purposes. Federal Law (US) restricts this device to sale by or on the order of a physician
Device Story
Equivital™ Vital Signs Physiological Monitor; ambulatory multiparameter telemetry system. Components: chest belt with skin electrodes/expansion sensor; body-worn electronics module (SEM). Inputs: ECG signals, thoracic movement (respiration), skin temperature, motion/orientation. Processing: digitizes signals; derives heart rate, breathing rate, activity level (none/low/high); detects R-waves via hardware processing. Outputs: physiological waveforms, vital sign rates, alerts for predefined boundaries. Connectivity: Bluetooth or low-power radio (40.68MHz). Usage: hospital, home, workplace, research; operated by clinicians or users. Data viewed via PC application or third-party integration. Benefits: continuous remote monitoring of patient physiology; facilitates occupational welfare and performance tracking.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified against standards: EN60601-1 (safety), IEC60601-1-2 (EMC), ISO 10993-1 (biocompatibility), ANSI/AAMI EC13 (cardiac monitors), and ASTM E 1112-00 (temperature). Additional functional and environmental performance testing conducted.
Technological Characteristics
Chest belt with skin electrodes and expansion sensor; body-worn electronics module (SEM). Energy: rechargeable or primary disposable batteries. Connectivity: Bluetooth or 40.68MHz low-power radio. Standards: EN60601-1, IEC60601-1-2, ISO 10993-1, ANSI/AAMI EC13, ASTM E 1112-00. Software: structured lifecycle per EN60601-1-4.
Indications for Use
Indicated for ambulatory monitoring of adults (16-65 years) in hospitals, homes, workplaces, and alternate care settings. Provides physiological data including ECG, respiration, skin temperature, body orientation, and motion for general patient monitoring, occupational welfare, and research/performance measurement.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "hidalgo" followed by four filled circles. Below the circles, there is handwritten text that reads "K061993" and "P 1/4". The text appears to be a label or identifier associated with the image.
## 510(k) Summary
#### Submitter
NOV - S 2006
: :
Hidalgo Ltd Stable Block at the Grange 20 Market St Cambridge CB4 5QG United Kingdom
Contact: J.Pisani - Operations Director Tel: +441954233430
#### Manufacturer
Hidalgo Ltd Stable Block at the Grange 20 Market St Cambridge CB4 5QG United Kingdom
Hidalgo propose to register as the manufacturer of the device in accordance with 21CFR Part 807.
#### Name of Device
Equivital™ Vital Signs Physiological Monitor Device Trade Name:
Device Common Name: Ambulatory Patient Monitor
### Device Classification and Product Code:
| 74 DRT | 870.2300 - Monitor, Cardiac (incl Cardiotachometer & rate alarm) |
|--------|------------------------------------------------------------------|
| 74DPS | 870.2340 - Electrocardiograph |
| 73BZQ | 868.2375 - Monitor, Breathing Frequency |
| 80FLL | 880.2910 - Thermometer, Electronic, Clinical |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "hidalgo" followed by five filled circles. Below the circles, there is handwritten text that reads "K061993 p2/4". The text appears to be a label or identifier for the image.
#### Predicate Devices
| GMP Wireless Medicine | LifeSync™ | - K030795 |
|-----------------------|------------------------------------------|-----------|
| Dymedix Inc | Re-usable Respiratory Effort Belt Sensor | - K040605 |
| Nexan Inc | NX-300 | - K003520 |
| Protocol Systems | Propaq Encore 200 | - K012451 |
| Respironics | Actiheart® | - K052489 |
| Vivometrics | Life Shirt™ Real Time | -K043604 |
The device has been compared to the following predicates:
#### Device Description
The Equivital™ Vital Signs Physiological Monitor comprises two components:
i) A chest belt containing skin electrodes and an expansion sensor
ii) A battery powered electronics module which connects directly to the chest belt and which acts to record, digitise and transmit the physiological information wirelessly to a receiving display device.
The device offers continuous monitoring of two views of the user's heart electrical activity (ECG) and respiratory breathing frequency inferred from thoracic cavity movement and uses this data to derive a Heart Rate and a Breathing Effort Rate.
The sensor also provides the following information:
- ECG and Respiration physiological waveforms. .
- an indication of the user's activity level (none, low or high) derived from a movement . detection sensor.
- . Body orientation.
- Chest skin surface temperature. .
- alternate secondary measurement of heart rate based on the detection of the user's R . wave using a separate hardware processing function.
- indications and alerts if physiology exceeds predefined boundaries. .
The device offers two variants for the wireless interface: a low power radio interface designed for military applications and a general purpose interface using Bluetooth™ technology.
The device offers two battery power options for the user, rechargeable or primary disposable cells.
The wireless data provided by the sensor may be viewed using a standalone PC based viewing application, or integrated into third party monitoring applications.
#### Intended Use Summary
The Equivital™ Vital Signs Physiological Monitor is an ambulatory multiparameter vital signs telemetry device intended for monitoring of adults (16 - 65years) in hospital care facilities, the home, workplace, and alternate care settings.
{2}------------------------------------------------
The device consists of a chest belt hamess and a body wom electronics module (SEM) supported by the chest belt).
The device collects and transmits ECG data and rate, respiration data and rate, skin temperature, body orientation and motion.
The monitor is indicated for use as a general patient monitor, to provide physiological information as part of an occupational welfare monitoring system, and for general research and performance measurement purposes.
hidalo
#### Technological Characteristics
Substantial equivalence has been measured by the technological review and comparison of performance data for the following.
| Function | Predicate 1 | Additional Predicates |
|----------------------------------------------------------------------------------------|-----------------------------------------------------------|------------------------------------------------------------------------------|
| ECG Heart Rate,<br>Breathing Rate<br>Temperature and<br>General Physical<br>Properties | Protocol Systems Propaq<br>Encore 200<br>K003520 | Nexan Inc NX-300<br>K003520 |
| Chest Expansion<br>Respiration | Dymedix<br>Re-usable Respiratory<br>Effort Belt - K040605 | Nexan Inc NX-300<br>K003520<br>Vivometrics LifeShirt<br>Real Time<br>K043604 |
| Activity/Motion<br>Detection | Respironics<br>Actiheart - K052489 | Vivometrics LifeShirt<br>Real Time<br>K043604 |
| Bluetooth telemetry<br>and ECG transmission | GMP Wireless Medicine -<br>LifeSyncTM K030795 | |
| Low Power 40.68MHz<br>PAN | Respironics<br>Actiheart - K052489 | |
The device technological characteristics compare to the following predicates:
#### Performance Measurement Summary
Performance measurement and review to the applicable sections of the following standards has been undertaken and successfully demonstrated as recommended by available guidrance from the agency:
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "hidalgo" followed by four solid black circles. Below the circles, there is handwritten text that reads "K061993" and "P 4/4". The text is in black ink and appears to be a label or identifier for the image.
- EN60601-1: Medical Electrical Equipment -- Part 1 General Requirements for Safety. .
- IEC60601-1-2: 2001 Medical Electrical Equipment Part 1-2: Collateral Standard: . Electromagnetic Compatibility.
- ISO 10993-1: Biological Evaluation of Medical Devices: Evaluation and Testing. .
- ANSI/AAMI EC13: 2002 Cardiac Monitors, heart rate meters, and alarms. The device was . not evaluated against the criteria for defibrillation and pacemaker immunity as it is not indicated for use in either of these cases.
- ASTM E 1 12-00 Standard Specification for intermittent determination of patient . temperature.
Additional voluntary testing has been undertaken to the following:
- . General functional test.
- Environmental Performance Test. .
The device software has been developed using a structured software lifecycle which meets the requirements of EN60601-1-4 Programmable Systems, which also applies to ongoing maintenance of the software.
#### Substantial Equivalence Conclusion
This pre-market notification has demonstrated the substantial equivalence of the Equivital™ Vital Signs Physiological Monitor to the predicates identified, by comparison to the describition material and performance testing of the device.
{4}------------------------------------------------
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV - 3 2006
Hidalgo Limited c/o Mr. Justin Pisani Operations Director 20 Market Street Cambridge, CB4 5QG United Kingdom
Re: K061993
Equivital Vital Signs Physiological Monitor Regulation Number: 21 CFR 870 1025 Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm) Regulatory Class: Class II Product Code: MHX Dated: September 29, 2006 Received: October 5, 2006
Dear Mr. Pisani:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA max publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Justin Pisani
Please be advised that FDA's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that vour device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
B.Bimmofor for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K061993
Device Name: Equivital™ Vital Signs Physiological Monitor
Indications For Use:
The Equivital™ Vital Signs Physiological Monitor is an ambulatory multiparameter vital signs The Equivilal™ Vital Signs Friyslological Rions and and States (16 - 65years) in hospital care facilities, the home, workplace and alternate care settings.
The device consists of a chest belt hamess and a body worn electronics module (SEM) supported by the chest belt.
The device collects and transmits ECG data and rate, respiration data and rate, skin temperature, body orientation and motion.
The monitor is indicated for use as a general patient monitor, to provide physiological The monitor is inalcated for ose as a general patient maniforing system, and for general research and performance measurement purposes.
Federal Law (US) restricts this device to sale by or on the order of a physician
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
B/Bummerman
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K011993
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.