The LifeShirt Real-Time is intended for use during daily activities of living and sleep, for the purpose of recording physiological data for later analysis by a physician. Respiration, ECG, pulse oximetry, blood pressure, temperature, and body position data may be collected. The system is intended to provide analysis of breathing patterns as an aid in classifying apneas as well as displaying heart rate changes from electrocardiographic waveforms in the wake and sleeping states as well as activities of daily living. The LifeShirt Real-Time is indicated for reuse by multiple adult patients in applications that may include pharmaceutical studies in which respiratory information is a useful indicator, or the general healthcare market where patients may be monitored at home and the data provided to their physicians as an aid to diagnosis and treatment.
Device Story
LifeShirt Real-Time is a wearable diagnostic system for recording physiological data (respiration, ECG, pulse oximetry, blood pressure, temperature, body position) during daily activities and sleep. System includes a garment, recorder/transmitter, wireless flash card, and analysis software. Data is recorded on-body and transmitted wirelessly to a networked computer for physician review. Device aids in apnea classification and heart rate monitoring. Used in home or clinical environments; operated by patients or healthcare providers. Physician reviews output to assist in diagnosis and treatment decisions. Wireless transmission capability allows remote data access.
Clinical Evidence
Bench testing only. Testing verified accurate data recording and transmission to the PC system and electromagnetic compatibility.
Technological Characteristics
Components: garment, recorder/transmitter, wireless flash card, VivoMonitor software, VivoLogic analysis software. Optional accessories: pulse oximeter, blood pressure cuff, temperature sensor. Connectivity: wireless transmission to networked computer. Power: battery-operated recorder. Data processing: digital signal recording and analysis.
Indications for Use
Indicated for adult patients in home or clinical settings, including pharmaceutical studies, for monitoring respiration, ECG, pulse oximetry, blood pressure, temperature, and body position during daily activities and sleep to aid in apnea classification and heart rate analysis.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
Predicate Devices
Reusable Multiple-Patient LifeShirt with VivoLogic Analysis Software
GMP Companies, Inc., LifeSync® Wireless ECG System
Life Sensing Instrument Company, Teletrens Model TM10 Multi-Parameter Portable Patient Transmitter
{0}------------------------------------------------
<043604
pi/3
### 510(k) SUMMARY
## VivoMetrics, Inc. LifeShirt™ Real-Time
## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
VivoMetrics, Inc. 121 North Fir Street, Suite E Ventura, California 93001 Phone: 805-667 -2225 Facsimile: 805-667 -6846 Contact Person: Keith Gilroy Date Prepared: December 28, 2004
#### Name of Device and Name/Address of Sponsor
LifeShirt Real-Time
VivoMetrics, Inc. 121 North Fir Street, Suite E Ventura, California 93001
#### Common or Usual Name
Programmable Diagnostic Computer
#### Classification Name
Programmable Diagnostic Computer
#### Predicate Devices
Reusable Multiple-Patient LifeShirt with VivoLogic Analysis Software
GMP Companies, Inc., LifeSync® Wireless ECG System
Life Sensing Instrument Company, Teletrens Model TM10 Multi-Parameter Portable Patient Transmitter
Mortara Instrument Inc., Ambulatory X-12 Telemetry Module
{1}------------------------------------------------
# Purpose of the Special 510(k) notice.
The purpose of this Special 510(k) is to modify the LifeShirt. The only difference between the devices is that the new device is indicated for use with an optional temperature probe, which is supplied, and uses a wireless data card in its recorder to permit the wireless transmission of its recorded data to a networked computer.
## Intended Use
The LifeShirt Real-Time is intended for use during daily activities of living and sleep, for the purpose of recording physiological data for later analysis by a physician. Respiration, ECG, pulse oximetry, blood pressure, temperature, and body position data may be collected. The system is intended to provide analysis of breathing patterns as an aid in classifying apneas as well as displaying heart rate changes from electrocardiographic waveforms in the wake and sleeping states as well as activities of daily living. The LifeShirt Real-Time is indicated for reuse by multiple adult patients in applications that may include pharmaceutical studies in which respiratory information is a useful indicator, or the general healthcare market where patients may be monitored at home and the data provided to their physicians as an aid to diagnosis and treatment.
## Technological Characteristics
The LifeShirt Real-Time consists of the following components: the LifeShirt garment, LifeShirt Recorder/Transmitter, a wireless flash card, a data Flash Card, VivoMonitor™ real-time software and VivoLogic® analysis and reporting software. A cleared pulse oximeter, blood pressure cuff, and a temperature sensor are optional accessories provided for use with the device.
## Performance Data
Performance testing included verification that data recorded by the LifeShirt recorder is accurately transmitted and received by the PC computer system. In addition, testing was performed to demonstrate electromagnetic compatibility.
## Substantial Equivalence
LifeShirt Real-Time has the same intended use and similar indications, principles of operation, and technological characteristics as its predicate devices.
{2}------------------------------------------------
The minor differences in the LifeShirt Real-Time and the cleared LifeShirt which are the optional temperature probe and wireless communication do not raise any new questions of safety or effectiveness as the predicate devices have similar features. Performance data demonstrates that the LifeShirt Real-Time is as safe and effective as the predicate Multiple Patient Use LifeShirt. Thus, the LifeShirt Real-Time is substantially equivalent to its predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features the department's name arranged in a circular pattern around a stylized eagle emblem. The eagle is depicted with three stripes forming its body and wing, and the text is written in a sans-serif font.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 2 9 2005
Vivometrics, Inc. c/o Mr Howard M. Holstein Hogan & Hartson, L.L.P. 555 Thirteenth Street, NW Washington, DC 20004-1109
#### Re: K043604
Trade Name: LifeShirt Rel-Time with Vivo Monitor Software Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II (two) Product Code: DQK Dated: April 15, 2005 Received: April 15, 2005
#### Dear Mr. Holstein:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave teviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 10 use stated in the encreate) to the enactment date of the Medical Device Amendments, or to conninered pror to they 2011/09/2017 accordance with the provisions of the Federal Food, Drug, de vices that have been require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r our may, mereleve, mains of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is elaborned (over as only of the major regulations affecting your device can inay be subject to such additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 -- Mr Howard M. Holstein
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r rease be actived a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must of any I cueral butther and step and the mot limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Of It in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. product in the low you to begin marketing your device as described in your Section 510(k) I mb lotter will and will your he FDA finding of substantial equivalence of your device to a legally prematics notineteem. The a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you a the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general micromational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Bhimmafor
Bran D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for Use Statement
510(k) Number (if known):___
Device Name: LifeShirt Real-Time with Vivo Monitor Software
Indications for Use:
The LifeShirt Real-Time is intended for use during daily activities of living and sleep, for the purpose of recording physiological data for later analysis by a oloop, ior are piration, ECG, pulse oximetry, blood pressure, temperature, and body physiousn. feespirations of the system is intended to provide analysis of possibing patterns as an aid in classifying apneas as well as displaying heart rate broaming patterns as ardiographic waveforms in the wake and sleeping states as onaligos from see of daily living. The LifeShirt Real-Time is indicated for reuse by well about patients in applications that may include pharmaceutical studies in which respiratory information is a useful indicator, or the general healthcare which toopirates may be monitored at home and the data provided to their physicians as an aid to diagnosis and treatment.
Prescription Use _V (Per 21 C.F.R. 801.109) AND/OR
Over-The-Counter Use (Per 21 C.F.R. 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Blsimmons
(Division Sign-Off) Division of Cardiovascular Devices 510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.