The MobileECG System is intended to provide an analysis of rhythm and of detailed morphology of complex cardiac complexes for the resting 12 lead ECG in all situations, whether in a hospital or primary care setting. It is adjunct to the diagnosis of all commonly recognized ECG abnormalities such as myocardial infarction (MI), including acute MI, ventricular hypertrophy, abnormal ST-T changes and common abnormalities of rhythm. The MobileECG System is intended for use in adults and children of any age from birth upwards.
Device Story
MobileECG System consists of Model 950-12LR HHC wireless module and CardioVu interpretation software. Module captures 12-lead electrical heart activity; transmits data wirelessly to laptop, netbook, tablet, or desktop computer (up to 100 feet). Wireless transmission reduces cable-related artifacts. CardioVu software receives, displays, and stores ECG data in real-time; applies ECG algorithms to provide advisory interpretation. Used in hospitals or primary care settings by physicians or qualified medical professionals. Output serves as advisory tool for clinical decision-making; not intended as sole diagnostic means. Benefits include patient comfort and reduced motion artifacts by eliminating tethered cables.
Clinical Evidence
Bench testing only. No clinical data provided. Equivalence established through comparative non-clinical testing against standards including ANS/AAMI EC11, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4, ANSI/AAMI EC53, and FCC Part 15.247.
Technological Characteristics
System comprises Model 950-12LR HHC wireless module and CardioVu software. Wireless communication via dedicated secure link. Standards: ANS/AAMI EC11 (ECG performance), IEC 60601-1-1 (medical electrical systems), IEC 60601-1-2 (EMC), IEC 60601-1-4 (programmable electrical medical systems), ANSI/AAMI EC53 (ECG cables/leads), FCC Part 15.247 (radio frequency).
Indications for Use
Indicated for adults and pediatric patients (birth upwards) requiring resting 12-lead ECG analysis for rhythm and morphology, including detection of myocardial infarction, ventricular hypertrophy, ST-T changes, and rhythm abnormalities. Intended for use in hospital or primary care settings as an adjunct to clinical diagnosis.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
2011
510(k) K103427
# SUMMARY OF SAFETY AND EFFECTIVENE
## 510(k) SUMMARY of Safety and Effectiveness
## 1. General Information
| Submitter's Name:<br>Address: | Memtec Corporation<br>68 Stiles Road Unit D<br>Salem, NH 03079 |
|-------------------------------|----------------------------------------------------------------|
| Telephone: | 603 893-8080 Ext. 204 |
| Contact Person: | Dennis Garboski |
| Trade Name: | MobileECG System |
| Common Name: | Electrocardiograph (per 21 CFR 870.2340) |
| Class: | II |
#### 2. Predicated Devices
The legally marketed predicated devices to which equivalence is being claimed is:
| Micromedical, Inc. | Biolog 3000 | K974351 |
|--------------------|-----------------|---------|
| Datrix, Inc. | Cardio WiFi and | K053083 |
| Datrix, Inc. | CardioServer | K052883 |
| QRS Diagnostic | EKGCARD SYSTEM | K030535 |
| QRS Diagnostic | Cardioview 32 | K083749 |
#### 3. Description of Device
·
The MobileECG System is an electrocardiograph consisting of Model 950-12LR HHC module and CardioVu interpretation software.
The Model 950-12LR HHC module is a small, light weight, single battery, wireless module designed to transmit 12Lead electrical heart activity of a patient to a computer. This module will transmit a patient's ECG wirelessly to a laptop, netbook, tablet, or desktop computer containing the CardioVu interpretation software from up to 100 feet away.
Elimination of the cable reduces a source of artifact resulting in high-quality ECG tracings. Wireless dedicated and secure communication links the Model 950-12LR HHC module to CardioVu interpretation software.
The CardioVu software receives, displays, and stores the patients ECG data wirelessly on a laptop, netbook, tablet, or desktop computer in "real time". The software also has the data using ECG algorithms for review by a physician or other qualified medical professional as an advisory basis only in conjunction with the physician's knowledge of ECG. The wireless feature allows the patient not to be tethered to the computer or a bulky box thus reducing motion artifacts caused by movement of the connecting cables and providing comfort to the patient.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the text "K103427 P2/2" in a handwritten style. The text appears to be a code or identifier, possibly related to a document or item. The handwriting is clear and legible, with distinct characters.
#### 4. Indications For Use
The MobileEG System is intended to provide an interpretation of the resting 12 lead ECG in all situations, whether in a hospital or primary care setting. It is capable of diagnosing all commonly recognized ECG abnormalities such as myocardial infarction (MI), including acute MI, ventricular hypertrophy, abnormal ST-T changes and common abnormalities of rhythm.
The MobileECG System is intended for use in adults and children of any age from birth upwards.
The MobileECG System's interpretation software is not intended as sole means of diagnosis and is offered to physicians and clinicians on an advisory basis only in conjunction with the physician's knowledge of ECG.
#### 5. Non-clinical Tests Used in Determination of Substantial Equivalence
Non-clinical tests were performed to compare the MobileECG System to the predicated devices.
The following applicable standards were used to compare the MobileECG System to the predicated devices: ANS/AAMI EC11, IEC 60601-1-2, IEC 60601-1-1, IEC 60601-1-4, 21 CFR 898, ANSI/AAMI EC53, AND FCC Part 15.247.
### 6. Conclusions From Non-clinical Testing
After comparing predicated devices to Memtec's MobileECG System, results show that with the intended use, the Model 950-12LR HHC module with CardioVu interpretation software is equivalent in safety and effectiveness. Therefore Memtec supports a claim of substantial equivalence for the MobileECG System.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. Inside the circle is an image of an eagle with its wings spread, with three lines representing the feathers. The eagle is facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Memtec Corporation c/o Mr. Dennis Garboski President 68 Stiles Road Unit D Salem, NH 03079
MAR 2 4 2511
K103427 Re:
> Trade/Device Name: MobileECG Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II (two) Product Code: DPS Dated: March 14, 2011 Received: March 16, 2011
Dear Mr. Garboski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 –Mr. Dennis Garboski
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
17 Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
Indications for Use
510(k) Number(if known) KIO 3427
Device Name: MobileECG System
Indications for Use
The MobileECG System is intended to provide an analysis of rhythm and of detailed morphology of complex cardiac complexes for the resting 12 lead ECG in all situations, whether in a hospital or primary care setting. It is adjunct to the diagnosis of all commonly recognized ECG abnormalities such as myocardial infarction (MI), including acute MI, ventricular hypertrophy, abnormal ST-T changes and common abnormalities of rhythm.
The MobileECG System is intended for use in adults and children of any age from birth upwards.
Prescription Use
Over-The-Counter Use_
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(Please Do Not Write Below This Line)
Concurrence of CDRH, Office of Device Evaluation (ODE)
A.D.O. for B.Zuckerman
(Division Sign-Off) ular Device 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.