Software as a Medical Device, Pediatric, 3rd-Party Reviewed
Indications for Use
The CardioServer ECG Management System software is intended to be marketed to medical professionals and for point-of-care use. The software is designed to provide a database used through out the medical community to store, display, edit and print high resolution ECG data received from devices such as electrocardiographs. The CardioServer ECG Management System software allows medical professionals responsible for the diagnosis and treatment of patients (adult and pediatric) with heart disease to: review and edit specific patient ECG data including intervals such as QT measurements and algorithm generated preliminary interpretative statements. ECG records are all associated by patient ID and other demographic data. Secure access to the database is provided.
Device Story
CardioServer ECG Management System is a software-only database for storage, display, and printing of high-resolution ECG data. Input: ECG data transferred from Datrix Cardio WiFi electrocardiograph. Operation: Software utilizes University of Glasgow ECG algorithm to provide automated interpretive analysis and serial comparison of ECGs. Output: Display of 1, 3, or 12 leads, full disclosure, user-selected strips, and physician-editable interpretations. Used in medical community/point-of-care settings by medical professionals. Healthcare providers use output to review patient ECG data, perform interval measurements via on-screen calipers, and support diagnosis/treatment of heart disease. Benefits include centralized data management, secure access, and automated preliminary interpretation to assist clinical decision-making.
Clinical Evidence
Bench testing only. Substantial equivalence demonstrated through performance, functionality, and reliability testing of the software in accordance with FDA Guidance Document: General Principles of Software Validation.
Technological Characteristics
Software-only system; Windows 2000/2003 Server environment. Connectivity: Wireless LAN, LAN, USB2. Data processing: University of Glasgow ECG Algorithm for interpretation and serial comparison. Features: 1/3/12-lead display, full disclosure, on-screen calipers, secure access, audit logging, and EHR integration via industry-standard formats.
Indications for Use
Indicated for adult and pediatric patients with heart disease. Used by medical professionals for point-of-care review, editing, and management of ECG data.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
### Appendix G
#### 510(k) Summary of Safety and Effectiveness for Datrix, Inc CardioServer ECG Management System
#### 1. Date Summary Prepared:
9-26-2005
#### 2. SUBMITTER'S NAME AND ADDRESS:
Linda Gluckman Datrix, Inc 340 State Place Escondido, CA 92029
#### 3. Contact Person:
Linda Gluckman Datrix QA Manager
#### 4. Device Name:
| Proprietary (trade) Name: | Datrix CardioServer ECG Management System |
|---------------------------|--------------------------------------------|
| Common Name: | ECG Management System |
| Classification Name: | Programmable Diagnostic Computer/ 870.1425 |
| Product Code: | DQK |
| Class: | II |
#### 5. PREDICATE DEVICE: The legally marketed device/s to which equivalence is being claimed is: Quinton Pyramis ECG Management System 3303 Monte Villa Parkway Bothell, WA 98021 K032038
{1}------------------------------------------------
## Appendix G
#### 6. Device Description:
CardioServer ECG Management System is a software only ECG management database that stores, displays, and prints high resolution ECG data transferred from a Datrix Cardio WiFi electrocardiograph device. All ECG records are associated by patient ID, and a final record including physician interpretation can be created.
The software system analyzes data using an ECG analysis algorithm developed under direction of Dr. Peter MacFarlane, University of Glasgow (note: the same algorithm is contained in the predicate device to which equivalency is being claimed). The ECG display is able to show 1, 3, or 12 leads at once, full disclosure, user-selected strips, and interpretations editable bv physician. Hardware requirements: are Windows 2000 or 2003 Server operating system; Pentium IV, 2GHz (minimum); 512 MB RAM (minimum); 10/100 Ethernet (minimum); RAID 5 storage; 1024x768 monitor; and standard back-up technology.
#### 7. Intended Use:
The CardioServer ECG Management System software is intended to be marketed to medical professionals and for point-of-care use. The software is designed to work with ECG Management databases located through out the medical community that commonly store, retrieve, display, edit, and print high-resolution records of ECG data received from devices such as the Datrix Cardio WiFi electrocardiograph.
The CardioServer ECG Management System software allows medical professionals responsible for the diagnosis and treatment of patients (adult and pediatric) with heart disease to do the following:
- 1. Review specific patient ECG data which is available through on-screen display, printer, email, fax, or Hospital Information System (HIS) results which report by interface or other customized data bases.
- 2. Analyze ECG data using calipers for measurement of various intervals.
The CardioServer system does not modify the original ECG waveform information, but it does provide an automated ECG interpretive analysis of the data.
The ECG data display can show 1, 3, or 12 leads at once, full disclosure, user-selected strips, interpretations (editable by physician), and lists studies using various filters (study type, patient, etc.), Resulting ECG records are
{2}------------------------------------------------
## Appendix G
all associated by patient ID and include the physician interpretation.
#### 8. Non-clinical Tests Used in Determination of Safety:
The substantial equivalence of the Datrix ECG Management System is Demonstrated by the following non-clinical testing:
- Testing to the FDA Guidance Document: General Principles of Software . Validation; Final Guidance for Industry and FDA Staff.
- Testing for the performance, functionality, and reliability characteristics of . the software device followed established test procedures in a quality system.
#### 9. Conclusions from Non-clinical Testing:
Datrix's CardioServer ECG Management system meets the FDA Guidance Document: General Principles of Software Validation; Final Guidance for Industry and FDA Staff with acceptable results, demonstrating substantial equivalence.
### 10. Technological Comparison with Predicate:
#### Both systems:
- Receive ECGs via wireless LAN, LAN, USB2, and other standard . transmission modes;
- Store, archive, and display ECGs; .
- Export ECGs in industry standard formats to electronic health . records;
- Print ECGs; ●
- Utilize Windows standards; .
- Are delivered as software only system; ●
- Provide secure access to patient health information; .
- Log all interaction with patient health information; .
- Provide clinical access to ECG data across the users network; ●
- Utilize the University of Glasgow ECG Algorithm for data . interpretation;
- Utilize the University of Glasgow ECG Algorithm for serial . comparison of ECGs;
- Allow user editing of ECG interpretation; .
- Provide ECG measurements: .
- Contain calipers for user edit of ECG measurements; .
- Generate management reports such as departmental ECG volume. .
{3}------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------4
Datrix CServer 510(k) Summary Appendix G
{4}------------------------------------------------
Image /page/4/Picture/11 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes around a staff. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV - 1 2005
Datrix, Inc. c/o Mr. Mark Job Regulatory Technology Services, LLC 1394 25th Street NW Buffalo, MN 55313
Re: K052883
Trade Name: Datrix CardioServer ECG Management System Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II (two) Product Code: DQK Dated: October 10, 2005 Received: October 13, 2005
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, F12A may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Mark Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Blummatfor
Brand D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Appendix E
#### Indications for Use Statement
510(k) Number (if known): Kos2883
Device Name: Datrix ECG Management System
Datrix CardioServer Model:
Indications for Use: The CardioServer ECG Management System software is intended to be marketed to medical professionals and for point-of-care use. The software is designed to provide a database used through out the medical community to store, display, edit and print high resolution ECG data received from devices such as electrocardiographs.
The CardioServer ECG Management System software allows medical professionals responsible for the diagnosis and treatment of patients (adult and pediatric) with heart disease to: review and edit specific patient ECG data including intervals such as QT measurements and algorithm generated preliminary interpretative statements. ECG records are all associated by patient ID and other demographic data. Secure access to the database is provided.
Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
B. Bummuma
/
sion of Cardlovascular Devices k) Number 252000
1/1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.