Diagnostic ECG for 12 Channel Resting ECG Patient Population: Male/Female, Pediatric to Adult Environment of Use: Hospital, Clinic and Home Use
Device Story
The EKGCard™ System is a diagnostic electrocardiograph that captures cardiac electrical signals via 12-lead ECG cables and electrodes. The device requires a Type II PC Card slot and operates on a Windows CE Pocket PC 2002 platform. It functions as a data acquisition system, recording voltage versus time to produce a graphical ECG record. Used in hospitals, clinics, and home environments, it is operated by healthcare professionals. The system provides clinicians with visual ECG data to assist in diagnosing cardiac abnormalities, monitoring responses to drug therapy, and tracking changes in heart function. It lacks automated interpretation, printing, or transtelephonic transmission capabilities.
Clinical Evidence
Bench testing only. The device was tested for compliance with safety and performance standards including IEC 601-1, IEC 60601-1, IEC 601-2-25, ANSI/AAMI EC11, ANSI/AAMI EC53, ANSI/AAMI ES1, and electromagnetic compatibility standards (EN 60601-1-2, EN 61000 series).
Technological Characteristics
12-channel diagnostic electrocardiograph; Type II PC Card form factor; Windows CE Pocket PC 2002 platform. Features internal isolation and defibrillator protection. Uses standard snap or tab electrodes and banana plug connectors. Complies with IEC 60601-1, IEC 60601-2-25, and ANSI/AAMI EC11/EC53 standards. Non-sterile, prescription-only device.
Indications for Use
Indicated for 12-channel resting ECG diagnostics in pediatric and adult male and female patients. Prescription use only in hospital, clinic, or home settings.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
K030535
Image /page/0/Picture/1 description: The image shows a logo with the letters QRS in a bold, sans-serif font. The letters are arranged horizontally, with the Q on the left, the R in the middle, and the S on the right. A thin, black oval surrounds the letters, adding a sense of motion or orbit to the design. The logo is simple, yet eye-catching, and would be easily recognizable.
B.
JUN 1 1 2003
510(k) SUMMARY of SAFETY and EFFECTIVENESS
## A. General Information
| 1. | Submitter's Name: | QRS Diagnostic, LLC |
|----|----------------------|---------------------------------------------|
| 2. | Address: | 14755 27th Avenue No.<br>Plymouth, MN 55447 |
| 3. | Telephone: | 763-559-8492, Ext. 952 |
| 4. | Contact Person: | Brad Sorenson |
| 5. | Date Prepared: | December 17, 2002 |
| 6. | Registration Number: | 2133542 |
| | <b>Device</b> | |
| 1. | Name: | EKGCard™ System |
| 2. | Trade Name: | EKGCard™ System |
| 3. | Common Name: | Diagnostic Electrocardiograph |
| 4. | Classification Name: | Electrocardiograph (ECG or EKG) |
| 5. | Product Code: | DPS |
| 6. | Class: | II |
| 7. | Regulation Number: | 870.2340 |
14755 27th Avenue N. • Plymouth, MN 55447 (763) 559-8492 • FAX (763) 559-2961 www.QRSdiagnostic.com
{1}------------------------------------------------
#### C. Identification of Legally Marketed Devices
| 1. Name: | Biolog 3000 and ECG Interface |
|------------------|-----------------------------------|
| 2. K Number: | K974351 and K974220 |
| 3. Date Cleared: | April 1, 1998 and August 25, 1998 |
#### D. Description of the Device
The EKGCard™ System is a diagnostic electrocardiograph (ECG or EKG) for 12 channel resting ECG's. The patient population is for both male and female, pediatric and adult. The system has been tested to the following standards:
- IEC 601-1, 60601-1; 1991 1
- IEC 601-2-25; 1993
- . ANSI/AAMI EC11 - 1991
- . ANSI/AAMI EC53 - 1995
- 트 ANSI/AAMI ES1 - 1993
- 처 CUL 601-1
- UL 2601-1
- 1 EN60601-1-2
- 트 EN 25011
- . EN 61000-4-2
- 트 EN 61000-4-3
- 트 EN 61000-4-4
- 트 EN-61000-4-6
- EN-61000-4-8
- 180 10993
- . 21CFR Part 898
- . 21CFR Part 801
- 1 93/42/EEC
- . EN 980
- 트 21CFR Part820
The EKGCard™ System is an electrocardiograph that detects signals associated with cardiac activity and produces an ECG; a graphical record of the voltage versus time. ECG's are routinely used to diagnose cardiac abnormalities, determine a patient's response to drug therapy, and reveal trends of changes in heart function.
The System requires a Type II PC Card Slot and Windows CE Pocket PC 2002 and is a Prescription Device.
2/5
{2}------------------------------------------------
#### Intended Use Statement E.
Diagnostic ECG for 12 Channel Resting ECG Patient Population: Male/Female, Pediatric to Adult Environment of Use: Hospital, Clinic and Home Use
#### F. Components/ Part Numbers
- 6000-4332 User's Manual ●
- 7000-1000 WLD ● EKGCard
- ECG Cable 2010-3722 (IEC) ●
- ECG Cable 2010-3721 (AAMI) ●
- Snap Adapters 5000-1861 ●
- Tab Adapters 2000-1858 .
- 2000-1859 Snap Electrodes .
- Tab Electrodes 5000-1858 ●
#### G. Table of Comparisons
The following summary tables of comparisons compare the new device (EKGCard System) to the predicate devices: Biolog 3000 and ECG Interface.
| # | Area | New Device:<br>EKGCard™ System | Predicate Device:<br>Biolog 3000 or<br>ECG Interface | Same | Different |
|----|-----------------------------|------------------------------------|------------------------------------------------------|------|-----------|
| 1 | Indications for<br>Use | ECG and Connection to<br>Pocket PC | ECG and Connection<br>to Pocket PC | X | |
| 2 | Patient Population | Male/Female<br>Pediatric to Adult | Male/Female<br>Pediatric to Adult | X | |
| 3 | Environment | Hospital, Clinic, Home<br>Use | Hospital, Clinic,<br>Home Use | X | |
| 4 | Number of<br>Electrodes | 12 Lead ECG | 12 Lead ECG | X | |
| 5 | Batteries | No | No | X | |
| 6 | Internal Isolation | Yes | Yes | X | |
| 7 | Defibrillator<br>Protection | Yes | Yes | X | |
| 8 | Banana Plugs | Yes | Yes | X | |
| 9 | Types of<br>Electrodes | Snap or Tab | Snap or Tab | X | |
| 10 | CMRR | Yes | Yes | X | |
| 11 | Heart Rate | Yes | Yes | X | |
3/5
{3}------------------------------------------------
| 12 | Standards<br>AAMI, EC11 | Yes | Yes | X | |
|----|--------------------------------|-------------|-------------|---|---|
| 13 | Standards<br>60601-2-25 | Yes | Yes | X | |
| 14 | Interpretation | No | Yes | | X |
| 15 | Transtelephonic | No | Yes | | X |
| 16 | Cable Length<br>3 Feet, 5 Feet | Yes | Yes | X | |
| 17 | Type BF | Yes | Yes | X | |
| 18 | Filters | 50 or 60 Hz | 50 or 60 Hz | X | |
| 19 | Connection Status | Yes | Yes | X | |
| 20 | Electrode<br>Labeling | IEC or AAMI | IEC or AAMI | X | |
| 21 | Print EKG's | No | Yes | | X |
| 22 | Supplied<br>Non-Sterile | Yes | Yes | X | |
| 23 | Prescription<br>Device | Yes | Yes | X | |
| 24 | Safety Standards | Yes | Yes | X | |
| 25 | EMC Standards | Yes | Yes | X | |
| 26 | Operating<br>Conditions | Yes | Yes | X | |
#### H. Discussion of Similarities and Differences
The EKGCard and Biolog 3000 or ECG Interface have the following similarities:
- . Indications for Use
- Patient Population ●
- Environment Number of Electrodes .
- Batteries .
- . Internal Isolation
- Defibrillator Protection .
- Banana Plugs ●
- Types of Electrodes . CMRR
- Heart Rate .
- Standards EC11 ●
- IEC 60601-2-25 .
{4}------------------------------------------------
- Cable Length ●
- . Type BF
- Filters ●
- Connection Status ●
- Electrode Labeling ●
- . Non-Sterile
- Prescription Device ●
- Safety ●
- EMC ●
- Operating Conditions .
The differences, with comments, are the following:
- Interpretation The EKGCard does not have interpretation, whereas the . Biolog 3000 does or can.
- . Transtelephonic – The EKGCard does not have the capability to transmit ECG's trans-telephonically.
- Print ECG The EKGCard does not yet have the ability to print EKG's. .
The above differences do not raise any new types of safety or effectiveness questions.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the eagle's body is formed by three human profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 1 1 2003
ORS Diagnostics, LLC c/o Mr. Alan Barker British Standards Institution Maylands Avenue Hemel Hemostead Hertfordshire, HP2 4SQ United Kingdom
Re: K030535
Trade Name: EKGCard™ Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II (two) Product Code: DPS Dated: Undated Received: May 27, 2003
Dear Mr. Barker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 -- Mr. Alan Barker
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Bram D. Zuckerman, M.D.
Director
Division of Cardiovascular Devices
Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# K030535
510(k) Number: To be determined
Device Name: EKGCard™ System
## Indications for Use:
- Diagnostic ECG for 12 Channel Resting ECG ●
- Patient Population: Malc/Female/Pediatric to Adult -
- Environment of Use: Hospital, Clinic, and Home Use -
- Prescription Device by a Physician -
### PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Division Sign-Off)
Division of Cardiovascular Devices
**510(k) Number** K030533
OVER-THE-COUNTER USE (optional Form 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.