PONTIS SUTURES AND SUTURE ANCHORS WITH OPTIONAL CRIMPS
K101126 · Core Essence Orthopaedics, Inc. · MBI · Dec 2, 2010 · Orthopedic
Device Facts
Record ID
K101126
Device Name
PONTIS SUTURES AND SUTURE ANCHORS WITH OPTIONAL CRIMPS
Applicant
Core Essence Orthopaedics, Inc.
Product Code
MBI · Orthopedic
Decision Date
Dec 2, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
PONTiSTM Sutures with Optional Crimps are indicated to secure soft tissue to soft tissue reattachment in the hand: - Collateral Ligaments around the PIP, DIP and MCP Joints . - Flexor and Extensor Tendons . PONTiSTM Suture Anchors with Optional Crimps are indicated to secure soft tissue to bone reattachment in the hand: - Collateral Ligaments around the PIP, DIP and MCP Joints . - Flexor and Extensor Tendons .
Device Story
PONTiS™ system comprises 316L stainless steel multifilament sutures, optional stainless steel crimp collars, and 3.0mm self-tapping threaded stainless steel suture anchors. Used in hand surgery for ligament and tendon reattachment. Surgeon uses disposable driver to insert anchor into pre-drilled bone hole; suture strands secure soft tissue to bone or to other soft tissue. Crimp collars, applied via specialized crimping instrument, secure multiple suture strands simultaneously, replacing traditional knot-tying. Adjustable tendon holder facilitates tensioning. System provides mechanical fixation to support active motion post-repair. Benefits include low-profile fixation in small bones and secure multi-strand attachment.
Clinical Evidence
Bench testing only. Evaluated biocompatibility (cytotoxicity, sensitization, irritation per ISO 10993), suture tensile strength, and crimped loop pull-strength. Suture anchors compared to reNOVO™ in foam bone models; tested for insertion torque, pull-out failure force, and cyclic pull-out performance. Results showed PONTiS™ performance equivalent to knotted suture products and sufficient to support active motion in hand tendon repair.
Technological Characteristics
Materials: 316L stainless steel. Components: 3.0mm self-tapping threaded anchors, multifilament sutures (USP 4-0, 3-0), and crimp collars. Principle: Mechanical fixation via threaded anchor in bone or crimped collar for soft tissue. Connectivity: None. Sterilization: Sterile, single-use. Software: None.
Indications for Use
Indicated for soft tissue-to-soft tissue or soft tissue-to-bone reattachment in the hand, specifically collateral ligaments (PIP, DIP, MCP joints) and flexor/extensor tendons.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
## Subject 510(k) Number: K101126
### Sponsor
Core Essence Orthopaedics, Inc 575A Virginia Drive Fort Washington, PA 19034
FDA Establishment Registration Number 3004613836
#### Official Contact
Jeff Miller Vice President Operations Core Essence Orthopaedics, Inc. 575A Virginia Drive Fort Washington, PA 19034 Phone - (215) 310-9534 Fax - (215) 660-5015
## Proprietary Name
## 1. PONTiSTM Sutures and Suture Anchors with Optional Crimps
#### Common Name
- 1. Suture Anchor, Stainless Steel Suture
#### Regulatory Class & Device Product Codes, Suture Anchors
Class II 21 CFR § 888.3040 Smooth or threaded metallic bone fixation fastener Class II 21 CFR § 878.4495 Suture, Nonabsorbable, Steel, Monofilament and Multifilament, Sterile
Product Codes (Panel 87, Orthopedic):
#### PONTiS™ Suture Anchor with optional crimp
- MBI (fastener, fixation, nondegradable, soft tissue) .
Date Prepared: November 26, 2010
{1}------------------------------------------------
### PONTiS™ Sutures with optional crimp
· GAQ
The substantial equivalence of the PONTiS™ families of sutures and crimp collars is based on the equivalence in intended use, materials, operational principals, and indications to other metallic sutures or cables secured with a crimp, including the Core Essence PONTiS™ ferroFibre stainless steel suture (K081060), the Pioneer Surgical SDB Cerclage System (K992616), the Ortheon Medical Tenofix tendon repair devices (K023594).
The substantial equivalence of the PONTiS™ 3.0mm suture anchors are based on the equivalence in intended use, materials, operational principals, and indications to reNOVO™ and other suture anchors including the Core Essence Orthopaedics reNOVO™ (K071520) and the J&J DePuy Mitek Micro Anchor (K962511).
# Brief Description of the Device, Sutures with Crimp Collars
This premarket notification covers line extensions for the PONTiS™ product line. The ferroFibre™ stainless steel (SS) sutures were cleared for marketing in 2008, and are marketed in the PONTiS ™ product group. The PONTiS™ product family has been expanded to include stainless steel suture anchors geometrically similar to the reNOVO™ titanium product family (cleared in 2007), in combination with ferroFibre™ SS Sutures and optional suture crimps.
The PONTiS [1] implants incorporate non-absorbable ferroFibre [1] 316 L Stainless Steel Surgical sutures and are available in United States Pharmacopoeia (USP) sizes 4-0 and 3-0 in various lengths.
316L Stainless steel is used in a wide variety of medical devices including previously approved implants of this type. The ferroFibre™ Stainless Steel Sutures may be secured by either knot tying or crimping with a collar.
The ferroFibre™ multifilament sutures are supplied sterile, armed with cutting needles.
The implant components also include stainless steel collars specific to each suture size. The collars are used to secure multiple suture strands together simultaneously by crimping. For example, crimps are sized to accept between four (4) and eight (8) strands of each specific size of ferroFibre™ suture for crimping. A crimping instrument is supplied to crimp the collars to the suture strands. An adjustable tendon holder with suture tensioning slots is provided to facilitate suture cable tightening and tensioning. Excess suture is cut and removed.
{2}------------------------------------------------
#### Indications for Use
PONTiSTM Sutures with Optional Crimps are indicated to secure soft tissue to soft tissue reattachment in the hand:
- Collateral Ligaments around the PIP, DIP and MCP Joints .
- Flexor and Extensor Tendons .
Safety and Performance, PONTiSTM Anchors with ferroFibre™ Sutures and Crimp Collar Securing Method
A collection of tests was conducted to characterize biocompatibility, diameter and tensile strength of the sutures in accordance with:
- 0 · ISO 10993 standards
- 0 USP 32-NF 26 Monographs <861>, <871>, <881>
- Class II Special Control Guidance, Surgical Suture; Guidance for Industry and � FDA, June 3, 2003
In addition, a crimped suture loop pull-strength test was developed and used to verify that multiple sutures secured together using a crimped multi-strand collar had sufficient strength to secure a tendon repair. In the performance tests, the PONTiSTM Sutures secured with Crimp Collars had sufficient strength and exhibited equivalent results to knot-secured suture products currently marketed for hand tendon repair as confirmed by a comparison to knotted sutures and to the literature.
The Crimp-Collar-secured multi-strand suture tensile strengths were sufficient to support active motion without resistance following tendon reattachment in the hand.
All materials were subjected to biocompatibility reviews or tests including cytotoxicity, sensitization, and irritation. All tests were performed in accordance with Part-10993 of the International Standard Organization (ISO) Standard (Biological Evaluation of Medical Devices). The performance studies and biocompatibility reviews indicate that the PONTISTM Sutures, secured with crimped Collars are safe for their intended use.
#### Basis for Substantial Equivalence, Sutures with alternate Crimp Collar securing method:
Core Essence Orthopaedics PONTiS™ ferroFibre™ sutures secured with crimp collars are substantially equivalent to other sutures currently marketed in conjunction with suture anchors and present no substantial differences in design, material, intended use and function to previously approved products. Additionally, the subject device labeling is consistent both with FDA's guidance as well as current medical practice.
The substantial equivalence of the PONTiS™ families of sutures and crimp collars is based on the equivalence in intended use, materials, operational principals, and indications to other metallic sutures or cables secured with a crimp, including the Core Essence
{3}------------------------------------------------
PONTiS™ ferroFibre stainless steel suture (K081060), the Pioneer Surgical SDB Cerclage System (K992616), the Ortheon Medical Tenofix tendon repair devices (K023594).
# Brief Description of the Device, Suture Anchors
### PONTiS™ Suture Anchors
The stainless steel PONTiS™ Suture Anchors are 3.0mm in diameter, are selftapping, and fully threaded. The anchors are each available with 4-0 and 3-0 multistrand ferroFibre™ stainless steel sutures. A disposable anchor driver holds the excess suture and delivers the anchor directly into the bone. The strands are then used to secure the soft tissue to bone.
Prior to installation, a hole is created in the bone using k-wire, drill or punch. The anchor is threaded into this hole. The anchor has self-tapping flutes.
Unlike the previously cleared titanium reNOVO™ Suture Anchor, PONTiS™ Suture Anchors are fully threaded allowing thread fixation in both cancellous and cortical bone.
The shorter overall length of the 3.0mm anchors, when compared to the self-drilling titanium reNOVO™ anchors, allows use in smaller/thinner bones where a lower profile is needed to achieve results without subsequent tissue injury.
The PONTiS™ threaded suture anchors are stainless steel to assure compatibility with the stainless steel in the ferroFibre™ stainless steel multi-strand sutures and the stainless steel crimp collars.
The size of the nonabsorbable stainless steel ferroFibre™ sutures used in the PONTiS™ Anchors is USP 4-0 and 3-0.
A single use driver and hand piece holds the excess suture and delivers the preloaded anchor into the bone. The suture strands are used to reapproximate and secure the soft tissue to bone.
The PONTiS™ Suture Anchors are provided sterile for single use applications.
The sizes and materials are designed to address the indications cited.
{4}------------------------------------------------
#### Indications for Use
PONTiSTM Suture Anchors with Optional Crimps are indicated to secure soft tissue to bone reattachment in the hand:
- Collateral Ligaments around the PIP, DIP and MCP Joints .
- Flexor and Extensor Tendons .
#### Safety and Performance, Suture Anchors
The following safety and performance data has been provided to support substantial equivalence of the PONTiS™ Suture Anchors to reNOVO™ suture anchors.
Performance Testing: The reNOVO™ and the PONTiS™ Suture Anchors were compared in standardized foam bone model materials. The Anchor insertion torque, suture strength (knot pull), pull-out failure mechanism (pull-out or suture break strength) and pullout failure force were documented. In addition cyclic pull-out performance was characterized. The insertion torque was quantified in hard bone (bovine cortical bone) to assure the integrity of the anchor and anchor driver interface. The suture anchors were also functionally (vs. empirically) evaluated in animal and human cadaver bone by surgeons and engineers. The pullout strengths were sufficient to support active motion without resistance following tendon reattachment in the hand.
In all performance tests the PONTiS™ Sutures and Anchors secured by knots or crimps exhibited equivalent results to products currently marketed for the same indications as confirmed by a comparison to the literature documenting the comparative characteristics of suture anchors.
All materials were subjected to biocompatibility reviews or tests including cytotoxicity, sensitization, and irritation. The performance studies and biocompatibility reviews indicate that the PONTiS™ Sutures and Anchors are safe for their intended use.
#### Basis for Substantial Equivalence
The substantial equivalence of the PONTiS™ 3.0mm suture anchors are based on the equivalence in intended use, materials, operational principals, and indications to reNOVO™ and other suture anchors including the Core Essence Orthopaedics reNOVO™ (K071520) and the J&J DePuy Mitek Micro Anchor (K962511).
Relative to the hand, the PONTiS™ suture anchors have the same intended use (bone anchoring function), but different technological characteristics than the Mitek Micro-Anchors. (threads vs. barbs). The barbed predicate device from J&J Mitek incorporates the USP 4-0 suture sizes like the PONTiS™ suture anchors, and is indicated for soft tissue reattachment procedures in the hand.
The 3.0mm PONTiS™ suture anchors have threaded profiles similar to the 3.0mm reNOVO™ anchors. The PONTiS™ suture anchors are minor modifications of the
{5}------------------------------------------------
original titanium reNOVO™ anchors, and are indicated for the same procedures in the hand. The PONTiS™ suture anchors will be sold pre-assembled with stainless steel ferroFibre™ suture.
### Conclusion
The substantial equivalence of the PONTiS™ families of sutures and crimp collars is based on the equivalence in intended use, materials, operational principals, and indications to other metallic sutures or cables secured with a crimp, including the SBD Cerclage system and the Tenofix tendon repair system.
The Core Essence Orthopaedics PONTiS™ sutures used in the anchors are substantially equivalent to the previously cleared Core Essence ferroFibre™ sutures, and the Core Essence Orthopaedics PONTiS™ suture anchors are substantially equivalent to reNOVO™ and J&J Mitek Micro-Anchors currently marketed, and present no substantial differences in design, material, intended use and function.
# END OF 510K SUMMARY
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized abstract symbol resembling a bird or a human figure with outstretched arms, composed of three curved lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Core Essence Orthopaedics, Inc. % Mr. Jeff Miller Vice President Operations 575A Virginia Drive Fort Washington, Pennsylvania 19034
DEC - 2 2010
Re: K101126
Trade/Device Name: PONTis™ Sutures and Suture Anchors with Optional Crimps Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: MBI, GAQ Dated: November 26, 2010 Received: November 29, 2010
Dear Mr. Miller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{7}------------------------------------------------
Page 2 - Mr. Jeff Miller
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vo
Sincerely yours,
Mark N. Mallekson
Mark I Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
K101126
# Statement of Indications for
・・・
DEC = 2 2010
510 (K) NUMBER IF KNOWN: K10176 (K) MANUFACTURER: Core Essence Orthopaedics, Inc. DEVICE NAME: PONTiS™ Sutures and Suture Anchors with Optional Crimps
PONTiSTM Sutures with Optional Crimps are indicated to secure soft tissue to soft tissue reattachment in the hand:
. · .
- . Collateral Ligaments around the PIP, DIP and MCP Joints
- Flexor and Extensor Tendons .
PONTiSTM Suture Anchors with Optional Crimps are indicated to secure soft tissue to bone reattachment in the hand:
- . Collateral Ligaments around the PIP, DIP and MCP Joints
- Flexor and Extensor Tendons ●
Oputu for mxm
(Division Sign-Off) Division of Surgical, Orthopedic. and Restorative Devices
510(k) Number K101126
Prescription Use XX (Per 21 CFR 801 Subpart D) and/or
Over-the-Counter Use _ NO (Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE, IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.