RENOVO SUTURE ANCHORS, MODELS 100-2001 THRU 100-5502X

K071520 · Core Essence Orthopaedics, LLC · MBI · Aug 10, 2007 · Orthopedic

Device Facts

Record IDK071520
Device NameRENOVO SUTURE ANCHORS, MODELS 100-2001 THRU 100-5502X
ApplicantCore Essence Orthopaedics, LLC
Product CodeMBI · Orthopedic
Decision DateAug 10, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

reNOVO Suture Anchors are intended to secure soft tissue to bone of: The Shoulder: Bankart Repair SLAP Lesion Repair Acromio-Clavicular Separation Roatator Cuff Repair Capsule Repair Biceps Tenodesis Deltoid Repair The Elbow: Ulnar or Radial Collateral Ligament Reconstruction Bicep Tendon Reconstruction Tennis Elbow Repair The Hand and Wrist: Scapholunate Ligament Reconstruction Ulnar / Radial Collateral Ligament Reconstruction Collateral Ligaments around the PIP, DIP and MCP Joints Flexor and Extensor Tendons The Ankle and Foot: Lateral Stabilization Medial Stabilization Achilles Tendon Repair / Reconstruction Hallux Valgus Reconstruction Mid and Rear Foot Reconstruction

Device Story

reNOVO Suture Anchor System consists of threaded titanium anchors (2.0mm–5.5mm) and nonabsorbable UHMW polyethylene sutures. Device includes single-use driver and handpiece for delivery. Surgeon inserts preloaded anchor into bone; suture strands reapproximate and secure soft tissue to bone. Used in orthopedic surgical settings for various ligament and tendon repairs. Provides mechanical fixation to facilitate tissue healing.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Materials: Titanium (ASTM F136/ISO 5832-3) anchor; UHMW polyethylene suture. Design: Threaded, self-drilling, self-tapping anchor with eyelet. Dimensions: 2.0mm to 5.5mm diameter. Sterilization: Provided sterile for single use.

Indications for Use

Indicated for soft tissue-to-bone fixation in shoulder, elbow, hand, wrist, ankle, and foot procedures, including ligament reconstructions, tendon repairs, and stabilization surgeries.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ page of ### 510(K) SUMMARY AUG 1 0 2007 #### Subject 510(k) Number K071520 ### Sponsor Core Essence Orthopaedics, LLC 301 Oxford Valley Road Suite 905B Yardley, PA 19067 ### FDA Establishment Registration Number 3004613836 ### Official Contact Shawn T. Huxel, CEO & President Core Essence Orthopaedics, LLC 301 Oxford Valley Road Suite 905B Yardley, PA 19067 Phone - (215) 310-9534 Fax - (609) 482-4957 Mobile - (908) 896-5893 ### Proprietary Name reNOVO™ Suture Anchor System ### Common Name Suture Anchor ### Classification Name and Reference Sec. 888.3040 Smooth or threaded metallic bone fixation fastener Regulatory Class Class II # Device Product Code/Subsequent Code (Panel 87) MBI/GAT Date Prepared 9 August, 2007 ### Brief Description of Device The reNOVO™ Suture Anchors range in size from 2.0mm to 5.5mm in diameter. Core Essence Orthopaedics, LLC Confidential {1}------------------------------------------------ Page 2 of 3 The anchor portion consists of a threaded titanium (ASTM F136/ ISO 5832-3) segment that provides a self drilling and self tapping thread. The anchor eyelet provides a means to attach the range of suture sizes utilized on the anchors. The size of sutures used in the reNOVO line range from size 3-0 thru size 2 (USP) nonabsorbable UHMW polyethylene UltraFibre™ sutures. A single use driver and handpiece hold the excess suture and deliver the preloaded anchor into the bone. The suture strands are used to reapproximate and secure the soft tissue to the bone. The reNOVO™ Suture Anchor will be provided sterile for single use applications. The sizes and materials are designed to address the indications cited. ### Indications for Use reNOVO Suture Anchors are intended to secure soft tissue to bone of: The Shoulder: Bankart Repair SLAP Lesion Repair Acromio-Clavicular Separation Roatator Cuff Repair Capsule Repair Biceps Tenodesis Deltoid Repair The Elbow: Ulnar or Radial Collateral Ligament Reconstruction Bicep Tendon Reconstruction Tennis Elbow Repair The Hand and Wrist: Scapholunate Ligament Reconstruction Ulnar / Radial Collateral Ligament Reconstruction Collateral Ligaments around the PIP, DIP and MCP Joints Flexor and Extensor Tendons The Ankle and Foot: Lateral Stabilization Medial Stabilization Achilles Tendon Repair / Reconstruction Hallux Valgus Reconstruction Mid and Rear Foot Reconstruction Core Essence Orthopaedics, LLC {2}------------------------------------------------ Page 3 of 3 # Basis for Substantial Equivalence The substantial equivalence of the reNOVO Suture Anchors is based on the equivalence in intended use, materials, operational principals, and indications to: | Mitek | Zimmer | Arthrotek | Arthrex | Smith &<br>Nephew | United<br>States<br>Surgical | |---------|---------|-----------|---------|-------------------|------------------------------| | K962511 | K962397 | K012503 | K971723 | K053344 | K040594 | | K962793 | K926384 | K973775 | K960516 | K972326 | | | K982420 | | K943806 | | | | In conclusion: Core Essence Orthopaedics reNOVO suture anchors are substantially equivalent to the currently marked devices and present no substantial differences in design, material, intended use and function to the products on the table above. ## END OF 510(K) SUMMARY {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, with three lines representing the feathers. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Core Essence Orthopaedics, LLC % Mr. Shawn T. Huxel CEO & President 301 Oxford Valley Road Suite 905B Yardley, PA 19067 AUG 1 0 2007 Re: K071520 > Trade/Device Name: reNOVOTM Suture Anchor System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI, HWC, JDR Dated: June 1, 2007 Received: June 4, 2007 Dear Mr. Huxel: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2 -- Mr. Shawn T. Huxel This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or 240-276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Barbara Buchin Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use # 510 (K) NUMBER IF KNOWN: K071520 MANUFACTURER: Core Essence Orthopaedics, LLC DEVICE NAME: reNOVO Suture Anchors reNOVO Suture Anchors are intended to secure soft tissue to bone of: The Shoulder: Bankart Repair SLAP Lesion Repair Acromio-Clavicular Separation Roatator Cuff Repair Capsule Repair Biceps Tenodesis Deltoid Repair The Elbow: Ulnar or Radial Collateral Ligament Reconstruction Bicep Tendon Reconstruction Tennis Elbow Repair The Hand and Wrist: Scapholunate Ligament Reconstruction Ulnar / Radial Collateral Ligament Reconstruction Collateral Ligaments around the PIP, DIP and MCP Joints Flexor and Extensor Tendons The Ankle and Foot: Lateral Stabilization Medial Stabilization Achilles Tendon Repair / Reconstruction Hallux Valgus Reconstruction Mid and Rear Foot Reconstruction Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use Y (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Juliane Avenue for mym Division of General, Restor and Neurological Devices Page 1 of / **510(k) Number** K071520
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...