K023594 · Ortheon Medical, LLC · GAQ · May 15, 2003 · General, Plastic Surgery
Device Facts
Record ID
K023594
Device Name
TENO FIX
Applicant
Ortheon Medical, LLC
Product Code
GAQ · General, Plastic Surgery
Decision Date
May 15, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4495
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Ortheon Medical Teno Fix™ tendon repair system is indicated for the repair of severed or lacerated digital flexor tendons.
Device Story
Teno Fix™ is a sterile, implantable tendon repair system used by surgeons in clinical settings to repair severed or lacerated digital flexor tendons. The system comprises stainless steel anchors, stop beads, and a stainless steel core suture. It is supplied with single-use disposable installation instruments for anchor placement and a crimping instrument to secure the stop bead to the suture, allowing for precise control of the implanted device length. By reducing the gap between severed tendon ends, the device facilitates improved healing compared to traditional sutures. Clinical performance is assessed by the surgeon during the procedure and post-operative follow-up to ensure effective tendon repair.
Clinical Evidence
Evidence includes comparative bench testing, canine studies, and a human clinical study. Canine testing demonstrated biocompatibility, implant stability, and improved healing via reduced gap distance between severed tendon ends. Human clinical study results indicate the Teno Fix™ system performs as well as or better than predicate devices, confirming safety and effectiveness.
Technological Characteristics
Implantable stainless steel components (anchors, stop beads, suture). Materials conform to ASTM F899-95 (stainless steel for instruments) and ASTM F138-00 (wrought 18Cr-14Ni-2.5Mo stainless steel for implants). Supplied sterile via radiation sterilization (ANSI/AAMI/ISO 11137-1994). Includes manual installation and crimping instruments.
Indications for Use
Indicated for the repair of severed or lacerated digital flexor tendons in patients requiring surgical intervention.
Regulatory Classification
Identification
A stainless steel suture is a needled or unneedled nonabsorbable surgical suture composed of 316L stainless steel, in USP sizes 12-0 through 10, or a substantially equivalent stainless steel suture, intended for use in abdominal wound closure, intestinal anastomosis, hernia repair, and sternal closure.
Special Controls
*Classification.* Class II (special controls). The device, when it is a steel monofilament suture that is uncoated and does not incorporate barbs, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9. The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
510(k), Ortheon Medical Teno Fix™ Repair System
Page 60 of 61
# Summary of Safety and Effectiveness for the Ortheon Medical Teno Fix™ Tendon Repair Device
Submitted by
MAY 1 5 2003
Ortheon Medical, LLC 7151 University Blvd. Winter Park, Florida 32792 Phone: (407) 671-2944
## Identification of a Legally Marketed Predicate Device
The Teno Fix™ tendon repair system is substantially equivalent to the Surgical Stainless Steel Sutures manufactured and marketed by Ethicon Incorporated, a Johnson & Johnson Company pursuant to K931271, as well as to the Surgical Stainless Steel Sutures manufactured by Peters Pharmaceutical Laboratory and marketed in the USA by CardioThoracic Systems, Inc., under the Acier brand name pursuant to K991073. The Teno Fix™ tendon repair system is also substantially equivalent to Prolene sutures, which are also manufactured and marketed by Ethicon Incorporated, a Johnson & Johnson Company, pursuant to reclassification as Class II devices on May 31, 1991.
#### General Description
The Teno Fix™ Tendon Repair System is delivered sterile. It consists of the implantable stainless steel components, implantation accessories, and installation instruments. The implantable components of the device include the anchors, stop beads, and suture. The stainless steel suture is shipped pre-assembled with a crimped stop bead and tapered needle.
The Teno Fix™ Tendon Repair System is supplied with two preloaded, single-use disposable installation instruments for implantation of the Teno Fix™ anchors. A crimping instrument is supplied to crimp the stop bead to the stainless steel core suture thus controlling the overall length of the implanted device.
#### Intended Use
The Ortheon Medical Teno Fix™ tendon repair system is indicated for the repair of severed or lacerated digital flexor tendons.
#### Summary of Technological Characteristics
Eleven (11) technological characteristics of the Teno Fix™ were compared to the predicate device and were found to be similar.
## Summary of Performance Data
The Teno Fix™ meets the requirements of the following recognized consensus standards.
- ASTM F899 95, Standard Specification for Stainless Steel Billet, Bar and Wire for Surgical Instruments
- ASTM F138 00. Standard Specification for Wrought 18 Chromium-14 Nickel-2.5 Molybdenum Stainless Steel Bar and Wire for Surgical Implants (UNS S31673)
{1}------------------------------------------------
#### 510(k). Ortheon Medical Teno Fix™ Repair System
[previously titled ASTM F138-92, Standard Specification for Stain Wire for Surgical Implants (Special Quality)]
- Sterilization of health care products-Requirements for validation and routine control-Radiation sterilization, ANSI/AAMI/ISO 11137-1994, Approved 25 May 1994 by Association for the Advancement of Medical Instrumentation, Approved 11 July 1994 by American National Standards Institute, Inc.
Comparative bench testing has established substantial equivalence to the predicate devices. Furthermore, the device has similar technological characteristics to the predicate devices.
Comparative testing in canines shows substantial equivalence to the predicate devices, demonstrating the biocompatibility of the materials, stability of the implant within the tendon, and the ability of the product to improve healing of the injury over the predicate devices by reducing the gap between severed ends of the tendon.
Results of a comparative clinical study in humans shows the Teno Fix™ Tendon Repair System performs as well or better than the predicate devices, and that it is safe and effective.
The Teno Fix™ Tendon Repair System:
- meets the requirements of the stated standards;
- embodies technological characteristics essentially identical to those of the predicate . devices:
- has been shown through clinical study that it performs as well or better than the predicate devices:
- has been shown through clinical study it is safe and effective: and .
- has been designed and developed utilizing design control methods in compliance . with 21CFR820.
The Teno Fix™ is manufactured per specifications and good manufacturing practices which ensure the device is safe and effective for its intended use.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, represented by three curved lines that suggest wings and a head.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 1 5 2003
Ortheon Medical, LLC c/o Mr. Jonathan S. Kahan Hogan & Hartson L.L.P. 555 Thirteenth Street, NW Washington, D.C. 20004
Re: K023594
Trade/Device Name: Teno Fix™ Tendon Repair System Regulation Number: 21 CFR 878.4495 Regulation Name: Stainless steel suture Regulatory Class: II Product Code: GAQ Dated: March 5, 2003 Received: March 5, 2003
Dear Mr. Kahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Jonathan S. Kahan
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
Page_1_of_1_
510(k) Number (if known): K023594
Device Name Teno Fix™ Tendon Repair System
Indications for Use:
The Ortheon Medical Teno Fix™ Tendon Repair System is intended for the repair of severed or lacerated digital flexor tendons.
Muriam C. Provost
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number K023594
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.