K073203 · Ge Medical Systems Information Technologies · MHX · Nov 29, 2007 · Cardiovascular
Device Facts
Record ID
K073203
Device Name
CARESCAPE V100
Applicant
Ge Medical Systems Information Technologies
Product Code
MHX · Cardiovascular
Decision Date
Nov 29, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The CARESCAPE V100 Vital Signs Monitor is intended to monitor a single adult, pediatric or neonatal patient's vital signs at the bedside or during intra-hospital transport. Vital signs parameters include noninvasive blood pressure (systolic, diastolic, and mean arterial pressure), pulse rate, and/or oxygen saturation (pulse oximetry) and/or temperature. The portable device is designed for use in numerous clinical settings in various hospital departments such as emergency, radiology, recovery, medical/surgical, labor and delivery, endoscopy, cardiac step-down. The CARESCAPE V100 Vital Signs Monitor can also be used in satellite areas, physicians' office, or alternate care settings.
Device Story
Portable multi-parameter monitor for adult, pediatric, neonatal patients; used in hospital departments (emergency, recovery, etc.) and physician offices. Inputs: NIBP (oscillometric), pulse rate, SpO2 (via compatible sensors), and temperature. Device processes signals to provide real-time vital sign measurements; features optional integrated printer. Operated by clinicians; no patient-controlled features. Output displayed on monitor for clinical decision-making regarding patient status during bedside monitoring or intra-hospital transport.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing, including risk analysis, requirements/design reviews, subsystem verification, integration testing, final acceptance/validation testing, performance testing, safety testing, and environmental testing.
Technological Characteristics
Portable multi-parameter monitor. Measures NIBP (oscillometric), pulse rate, SpO2, and temperature. Configurations include optional integrated printer. Connectivity for various SpO2 sensor types (Ohmeda TruSignal, Nellcor, Masimo). Battery-operated. Complies with voluntary standards for safety and performance.
Indications for Use
Indicated for monitoring vital signs in adult, pediatric, and neonatal patients in hospital, clinical, or alternate care settings. Parameters include NIBP, pulse rate, SpO2, and temperature.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
DINAMAP® ProCare Series 100N-400N Monitor (K022193)
DINAMAP® Pro 1000 Monitor with SuperSTAT (K022834)
Datex-Ohmeda TruSat Pulse Oximeter and Accessories (K040831)
Submission Summary (Full Text)
{0}------------------------------------------------
K073203
## 510(k) Summary of Safety and Effectiveness
Date: November 9, 2007 Submitter: GE Medical Systems Information Technologies NOV 2 9 2007 4502 Woodland Corporation Boulevard Tampa, FL 33614 USA Contact Person: Karen Russell Requlatory Affairs Specialist GE Medical Systems Information Technologies Phone: 503-831-3176 Fax: 503-907-8591 E-mail: Karen.Russell@med.ge.com Secondary Contact Person: Sharon Untz Senior Director, Global Quality and Regulatory Affairs Systems GE Medical Systems Information Technologies Phone: (414) 362-2436 Fax: (414) 362-2585 E-mail: Sharon.Untz@med.ge.com Device Trade Name: CARESCAPE V100 Vital Signs Monitor Common /Usual Name: Physiological or Vital Signs Monitor, Patient Monitor Classification Names: Patient Physiological Monitor (without Arrhythmia Detection) System, Measurement, Blood Pressure, Noninvasive Computer, Blood Pressure Alarm, Blood Pressure Oximeter Oximeter, Ear Thermometer, Clinical Electronic Recorder, Paper Chart Predicate Devices: DINAMAP® ProCare Series 100N-400N Monitor (K022193) DINAMAP® Pro 1000 Monitor with SuperSTAT (K022834) Datex-Ohmeda TruSat Pulse Oximeter and Accessories (K040831) Device Description: The CARESCAPE V100 Vital Signs Monitor is a small, portable monitor for use in a sub-acute hospital and non-hospital settings. The CARESCAPE V100 is for use on adult, pediatric, or neonatal patients -
Page 1 OF 3
{1}------------------------------------------------
one at a time. The battery operated monitor provides noninvasive determination of systolic blood pressure, diastolic blood pressure, mean arterial pressure, pulse rate, oxygen saturation, and temperature.
The CARESCAPE V100 is a multi-parameter device that provides Sp02 monitoring for use in spot-checking or continuous monitoring. The CARESCAPE V100 also measures NIBP using the oscillometric method. There are no features of CARESCAPE V100 that can be controlled by the patient.
The CARESCAPE V100 monitor is available in four standard configurations with or without an integrated printer:
- CARESCAPE V100 with NIBP (Classic, Classic Auscultatory, or . SuperSTAT), Pulse
- CARESCAPE V100 with NIBP (Classic, Classic Auscultatory, or . SuperSTAT), Pulse, and Temperature
- . CARESCAPE V100 with NIBP (Classic, Classic Auscultatory, or SuperSTAT), Pulse, and Sp02 (Ohmeda TruSignal, Nellcor, or Masimo)
- . CARESCAPE V100 with NIBP (Classic, Classic Auscultatory, or SuperSTAT), Pulse, Temperature, and Sp02 (Ohmeda TruSignal, Nellcor, or Masimo)
- Intended Use: The CARESCAPE V100 Vital Signs Monitor is intended to monitor a single adult, pediatric or neonatal patient's vital signs at the bedside or during intra-hospital transport. Vital signs parameters include noninvasive blood pressure (systolic, diastolic, and mean arterial pressure), pulse rate, and/or oxygen saturation (pulse oximetry) and/or temperature. The portable device is designed for use in numerous clinical settings in various hospital departments such as emergency, radiology, recovery, medical/surgical, labor and delivery, endoscopy, cardiac step-down. The CARESCAPE V100 Vital Signs Monitor can also be used in satellite areas, physicians' office, or alternate care settings.
## Technology:
The CARESCAPE V100 Vital Signs Monitor employs the same functional scientific technology as its predicate device(s).
{2}------------------------------------------------
## Test Summary:
The subject of this 510(k) is a design modification for the ProCare V1 Vital Signs Monitor. The CARESCAPE V100 Vital Signs Monitor complies with the voluntary standards as detailed in Section 4.2 of this submission. The following quality assurance measures were applied to the development of the CARESCAPE V100 Vital Signs Monitor:
- Risk Analysis ●
- Requirements Reviews ●
- Design Reviews .
- Subsystern Verification .
- Integration testing (System verification) .
- Final acceptance testing (Validation) .
- . Performance testing
- Safety testing .
- Environmental testing .
## Conclusion:
The results of these measurements demonstrated that the CARESCAPE V100 Vital Signs Monitor is as safe, as effective, and performs as well as the predicate device(s),
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle with three stripes representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 2 9 2007
GE Medical Systems Information Technologies c/o Ms. Karen Russell Regulatory Affairs Specialist 4502 Woodland Corporate Blvd. Tampa, FL 33614
Re: K073203
Trade Name: CARESCAPETM V100 Vital Signs Monitor Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm Regulatory Class: Class II (two) Product Code: MHX Dated: November 9, 2007 Received: November 14, 2007
Dear Ms. Russell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Karen Russell
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Blymmimata for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use K073203
510(k) Number (if known):
Device Name:
CARESCAPE V100 Vital Signs Monitor
Indications for Use:
The CARESCAPE V100 Vital Signs Monitor is intended to monitor a single adult, pediatric or neonatal patient's vital signs at the bedside or during intra-hospital transport. Vital signs parameters include non-invasive blood pressure (systolic, diastolic, and mean arterial pressure), pulse rate, and/or oxygen saturation (pulse oximetry) and/or temperature.
The portable device is designed for use in numerous clinical settings in various hospital departments such as emergency, radiology, recovery, medical/surgical, labor and delivery, endoscopy, cardiac stepdown. The CARESCAPE V100 Vital Signs Monitor can also be used in satellite areas, physicians' office, or alternate care settings.
Over-The-Counter Use Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Blymmuma
(Div Sign-Off)
Div n of Cardiovascular Devices
510 Number K073203
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.