DINAMAP PROCARE SERIES MONITOR, MODELS 100N, 200N, 300N, 400N
K022193 · Ge Medical Systems Information Technologies · DQA · Aug 1, 2002 · Cardiovascular
Device Facts
Record ID
K022193
Device Name
DINAMAP PROCARE SERIES MONITOR, MODELS 100N, 200N, 300N, 400N
Applicant
Ge Medical Systems Information Technologies
Product Code
DQA · Cardiovascular
Decision Date
Aug 1, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The DINAMAP® ProCare Series 100N-400N Monitor is intended to monitor a single adult, pediatric or neonatal patient's vital signs at the bedside or during intrahospital transport. Vital signs parameters include non-invasive blood pressure (systolic, diastolic, and mean arterial pressure), pulse rate, and/or oxygen saturation (pulse oximetry) and/or temperature. The portable device is designed for use in numerous clinical settings in various hospital departments such as emergency, radiology, recovery, medical/ surgical, labor and delivery, endoscopy, cardiac stepdown. It can also be used in satellite areas, physicians' offices, or alternate care settings.
Device Story
Portable vital signs monitor; provides NIBP, pulse rate, SpO2, and temperature monitoring. Inputs: NIBP cuffs, pulse oximetry sensors, and temperature probes. Operation: utilizes established NIBP algorithms, Alaris IVAC Turbo thermometry, and Nellcor N-595 SpO2 technology. Output: real-time vital signs displayed on screen and via integrated printer. Used in hospital departments (emergency, recovery, etc.) and physician offices by healthcare professionals. Facilitates patient assessment during bedside monitoring or intrahospital transport; supports clinical decision-making through continuous or spot-check vital sign data.
Clinical Evidence
Bench testing only. Studies included electromagnetic compatibility, electrical safety, and mechanical/environmental testing to demonstrate safety and effectiveness.
Technological Characteristics
Portable monitor with integrated printer; AC mains or internal lead-acid rechargeable battery. Parameters: NIBP (proprietary algorithm), Alaris IVAC Turbo thermometry, Nellcor N-595 SpO2. No new technologies compared to predicate. Standalone device.
Indications for Use
Indicated for monitoring vital signs (NIBP, pulse rate, SpO2, temperature) in adult, pediatric, and neonatal patients in clinical settings including bedside, intrahospital transport, emergency, radiology, recovery, medical/surgical, labor and delivery, endoscopy, cardiac stepdown, and physician offices.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circle. There are decorative swirls or flourishes around the letters, adding a touch of elegance to the design. The logo is in black and white.
K022193
GE Medical Systems Information Technologies
General Electric Company 4502 Woodland Corporate Blvd., Tampa, FL 33614 813 887-2000
#### SUMMARY OF SAFETY AND EFFECTIVENESS July 3, 2002
# DINAMAP® ProCare Series 100N-400N Monitor
#### A. Submitter
GE Medical Systems Information Technologies 4502 Woodland Corporate Boulevard Tampa, FL 33614
#### B. Company Contact
Primary: Melissa Robinson Regulatory Affairs Specialist Phone: 813-887-2133 Fax: 813-887-2552
#### Secondary:
Tom English Global QA/RA Phone: 813-887-2107 Fax: 813-887-2413
#### C. Common Name
Physiological or Vital Signs Monitor, Patient Monitor
| Classification Name | Product Code | 21 CFR |
|--------------------------------------------------|--------------|----------|
| System, Measurement, Blood Pressure, Noninvasive | DXN | 870.1130 |
| Computer, Blood Pressure | DSK | 870.1110 |
| Alarm, Blood Pressure | DSJ | 870.1100 |
| Oximeter | DQA | 870.2700 |
| Oximeter, Ear | DPZ | 870.2710 |
| Thermometer, Clinical Electronic | FLL | 880.2910 |
| Recorder, Paper Chart | DSF | 870.2810 |
#### D. Predicate/Legally Marketed Devices
DINAMAP® ProCare Series Monitor -K014255 GE Medical Systems Information Technologies
{1}------------------------------------------------
# E. Device Description
The DINAMAP ProCare Series 100N-400N Monitor is a prescription device intended for use only by health care professionals. Four configurations of the monitor-all with integrated printer-will offer the following vital signs parameters:
- DPC 100N: Non-Invasive Blood Pressure and Pulse Rate .
- DPC 200N: Non-Invasive Blood Pressure and Pulse Rate, Temperature .
- DPC 300N: Non-invasive Blood Pressure and Pulse Rate, Pulse Oximetry .
- DPC 400N: Non-Invasive Blood Pressure and Pulse Rate. Pulse Oximetry . and Temperature.
This portable device includes an integrated printer and is capable of operation from an external AC mains power source or an internal lead-acid rechargeable battery.
## F. Intended Use
The DINAMAP® ProCare Series 100N-400N Monitor is intended to monitor a single adult, pediatric or neonatal patient's vital signs at the bedside or during intrahospital transport. Vital signs parameters include non-invasive blood pressure (systolic, diastolic, and mean arterial pressure), pulse rate, and/or oxygen saturation (pulse oximetry) and/or temperature. The portable device is designed for use in numerous clinical settings in various hospital departments such as emergency, radiology, recovery, medical/ surgical, labor and delivery, endoscopy, cardiac stepdown. It can also be used in satellite areas, physicians' offices, or alternate care settings.
## G. Technological Characteristics
The DINAMAP® ProCare Series 100N-400N Monitor has the same technological characteristics as the predicate device, the DINAMAP® ProCare Series Monitor. There are no new technologies used on the DINAMAP® ProCare Series 100N-400N Monitor.
## H. Parameter Technology
The DINAMAP® ProCare Seies 100N-400N Monitor has the following parameter technologies:
- ·NIBP algorithm as implemented on the ProCare Series Monitor
- ·Alaris IVAC Turbo thermometry technology as implemented on the ProCare Series Monitor
- · Wholly implemented Nellcor N-595 SpO2 technology
# I. Testing
Several bench studies were conducted which demonstrate safety and effectiveness of the DINAMAP® ProCare Series 100N-400N Monitor:
- · Electromagnetic Compatibility
- · Electrical Safety
- Mechanical and Environmental
{2}------------------------------------------------
# K. Substantial Equivalence
| ProCare Series 100N-<br>400N | Predicate Device & Model | 510(k)<br>Numbers |
|------------------------------|------------------------------|------------------------|
| | Monitor | DINAMAP ProCare Series |
| Pulse Oximetry | Nellcor N-595 Pulse Oximeter | K012891 |
| Temperature | Alaris Medical System | K955846 |
| NIBP | DINAMAP ProCare Series | K014255 |
Page 3 of 3
00037
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
Public Health Service
AUG 01 2002
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Melissa Robinson Regulatory Affairs Specialist GE Medical Systems Information Technologies 4502 Woodland Corporation Boulevard Tampa, Florida 33614
Re: K022193
Trade/Device Name: DINAMAP® ProCare Series 100-400N Monitor Regulation Number: 870.2700 and 870.2300 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA and MWI Dated: July 3, 2002 Received: July 5, 2002
Dear Ms. Robinson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
Page 2 - Ms. Robinson
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html
Sincerely yours,
Timoth A. Ulatowski Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE
KO22193 510(k) Number (if known):
Device Name: DINAMAP® ProCare Series 100N-400N Monitor
Indications for Use:
The DINAMAP® ProCare Series 100N-400N Monitor is intended to monitor a single adult, pediatric or neonatal patient's vital signs at the bedside or during intra-hospital transport. Vital signs parameters include non-invasive blood pressure (systolic, diastolic, and mean arterial pressure), pulse rate and/or temperature and/or oxygen saturation (pulse oximetry). The portable device is designed for use in numerous clinical settings in various hospital departments such as emergency, radiology, recovery, medical/surgical, labor and delivery, endoscopy, cardiac step-down. It can also be used in satellite areas, physicians' offices, or alternate care settings.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (per 21 CFR 801.109
-------------------------
OR
Over-The Counter Use (Optional Format 1-2-96)
A.W. Anderson
(Sivision Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 510(k) Number_KOzz193
4.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.