K992533 · Philips Medical Systems North America, Inc. · LNH · Oct 18, 1999 · Radiology
Device Facts
Record ID
K992533
Device Name
PHILIPS GYROSCAN INTERA (*)
Applicant
Philips Medical Systems North America, Inc.
Product Code
LNH · Radiology
Decision Date
Oct 18, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The Gyroscan INTERA systems are indicated for use as diagnostic devices that produce transverse, sagittal, coronal and oblique cross-sectional images, spectroscopic images and/or spectra, based upon 'H and 3'P metabolites, and that display the internal structure and/or function of the head, body or extremities. These images and/or spectra when interpreted by a trained physician, yield information that may assist in diagnosis.
Device Story
Magnetic Resonance Diagnostic Device (MRDD) producing cross-sectional images and spectroscopic data (1H, 31P) of head, body, and extremities; utilizes RF receive-only coils and magnet bore; operated by clinicians in radiology settings. Features include patient comfort enhancements (forced air, observation camera, laser positioning), ceiling-suspended console with LCD display, and Compaq Alpha processor for high-speed reconstruction (40 images/sec). Supports real-time interactive imaging, diffusion/perfusion mapping (ADC, TTP, MTT), and vector ECG gating. Physician interprets output to assist in clinical diagnosis; device provides structural and functional anatomical information.
Clinical Evidence
No clinical data provided; bench testing only. Safety and effectiveness are established through comparison to the predicate device and adherence to NEMA, UL, and IEC 601 standards.
Technological Characteristics
MRDD system (0.5T, 1.5T, 1.6T configurations). Uses RF receive-only coils (e.g., Synergy Pediatric Coil). Processing via Compaq Alpha processor. Connectivity includes ceiling-suspended console with 20-inch LCD. Safety standards: NEMA, FDA MRDD guidance, UL, IEC 601. Software-based reconstruction and analysis.
Indications for Use
Indicated for use as an MR diagnostic device to produce cross-sectional images and spectra (1H and 31P) of the head, body, or extremities to assist in diagnosis by a trained physician. No specific age or gender restrictions; contraindications not specified.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a sequence of handwritten characters. The characters appear to be 'K99 25 33'. The handwriting is somewhat stylized, with the numbers being slightly more rounded.
Image /page/0/Picture/1 description: The image shows the Philips logo, which consists of the word "PHILIPS" written above a circular emblem. The emblem features four stars arranged around two wavy lines. The logo is presented in black and white, with the word "PHILIPS" in bold, uppercase letters.
OCT 1 8 1999
Page 1 of 2
# 510(k) Summary
The following information is being submitted in accordance with the requirements of 21 CFR 807.92.
| Company Name | : | Philips Medical Systems North America Company. |
|-----------------------|---|--------------------------------------------------------------------------------------------------------------------------------|
| Address | : | 710 Bridgeport Avenue<br>Shelton, CT 06484. |
| Registration No. | : | 1217116 |
| Contact person | : | Peter Altman |
| Device (Trade) Name | : | Philips Gyroscan INTERA (*) |
| Classification Name | : | Magnetic Resonance Diagnostic Device (MRDD). |
| Classification | : | Class II. |
| Product code | : | LNH / LNI. |
| Performance standards | : | NEMA voluntary standards, FDA MRDD guidance's, UL and IEC 601<br>appropriate safety standards and/or draft standards are used. |
| Common/Usual Name | : | Philips Gyroscan INTERA (Release 7 series). |
### Predicate Device(s).
The Philips cleared MRDD Philips GYROSCAN NT Release 6 series systems with FDA ref.K980645.
## Intended use.
The Philips Gyroscan INTERA (Release 7) series have the same intended use as its predecessor and predicate device Gyroscan NT Release 6. The Gyroscan INTERA systems are indicated for use as diagnostic devices that produce transverse, sagittal, coronal and oblique cross-sectional images, spectroscopic images and/or spectra, based upon 'H and 3'P metabolites, and that display the internal structure and/or function of the head, body or extremities. These images and/or spectra when interpreted by a trained physician, yield information that may assist in diagnosis.
# Device Description and Technological Characteristics
The new Philips MRDD being the successor of Gyroscan NT (Rel.6), is the Release 7 series called the Gyroscan INTERA series. The Gyroscan INTERA is based on the same platform as its predicate device Gyroscan NT (re.K980645) with the same technological characteristics and intended use .
<sup>(*)</sup>Gyroscan INTERA series are available in Gyroscan INTERA (0.5T), Gyroscan INTERA 1.6T (1.5T) and the CV version Gyroscan INTERA (1.5T)
{1}------------------------------------------------
The main enhanced and new feature of the Gyroscan INTERA series are:
- The new look in its appearance to emphasis its compactness. .
- Hardware features to improve user friendliness such as the operator's console and the use of . LCD monitor displays.
- Patients comfort by forced air flow through the magnet bore and comfort zone with Patient . Observation provision. The patient can view the environment outside the bore and the built-in camera is used to observe the patient in the magnet bore.
- Save laser light cross beam is applied for reference for the patient positioning. .
- A ceiling suspended Examination Room Operator's console with tracker-ball control and 20 . inch LCD display.
- Enhanced and new functionality's : .
- * RF receive only Synergy Pediatric Coil
- * The use of the XP1000 Compaq (Alpha processor) allowing image reconstruction up to 40 images/second.
- * Real-time Interactive Imaging.
- * On-line calculation of the ADC maps (Diffusion package)
- * On-line calculation of TTP, Negative Integral, Index, and MTT maps (Perfusion package).
- * Three-points Plan scan ( enhancement Free Style Plan Scan).
- * Vector ECG signals for MR scan synchronization (triggering) and gating.
#### Safety parameters.
The safety parameters of the Gyroscan INTERA Release 7 remains the same as with its predecessor and predicate device Gyroscan NT Release 6 series ( re. K980645).
#### General Safety and effectiveness.
The safety and effectiveness of the Gyroscan INTERA a re the same as with its predicate device the GYROSCAN NT Release 6 systems (ref.K980645)
It does not induce other safety issues and warnings than already valid for its predecessor and predicate device.
#### Substantial Equivalence.
The Gyroscan INTERA (Release 7 series) is substantially equivalent to the predicate device Philips GYROSCAN NT Release 6 series systems with FDA ref.K980645.
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 8 1999
Peter Altman Director of Regulatory Affairs Philips Medical Systems North America Company 710 Bridgeport Avenue P.O. Box 860 Shelton, CT 06484-0917
Re:
K992533
Gyroscan Intera (Release 7 Series) Dated: July 28, 1999 Received: July 29, 1999 Regulatory class: II 21 CFR 892.1000/Procode: 90 LNH
Dear Mr. Altman:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If vour device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in . regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Capt. Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
| 510(k) Number (if known): | Unknown K992533 |
|---------------------------|--------------------------|
| Device Name : | Philips Gyroscan INTERA. |
#### Indication For Use :
The Philips Gyroscan INTERA (Release 7) series are indicated for use as magnetic resonance diagnostic devices (MRDD's) that produce transverse, sagittal, coronal and oblique crosssectional images, spectroscopic images and/or spectra, based upon 'H and 31P metabolites, and that display the internal structure and/or function of the head, body or extremities. These images and/or spectra when interpreted by a trained physician, yield information that may assist in diagnosis.
## ( PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109) ✓
OR Over-The-Counter Use
(Optional Format 1-2-96)
David h. Segears
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological I 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.