GLOBAL ADVANTAGE SHOULDER, GLOBAL ADVANTAGE HUMERAL STEM, GLOBAL ADVANTAGE ECCENTRIC HEAD
Applicant
DePuy Orthopaedics, Inc.
Product Code
KWS · Orthopedic
Decision Date
Jul 12, 1999
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3660
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Global Advantage Shoulder, humeral stem and humeral head, are intended for use in total or hemi-arthroplasty. When used for total shoulder arthroplasty, the subject components are designed to be used with the existing Global cemented glenoid components cleared in K905786 & K914000. The humeral stem is intended to be used with bone cement or in press-fit applications (cementless).
Device Story
Global Advantage Shoulder is a modular humeral prosthesis consisting of a titanium alloy (Ti-6Al-4V) humeral stem and a cobalt-chromium-molybdenum alloy humeral head. Components utilize a morse-type taper lock system for assembly. The humeral stem features four proximal fins with suture holes, a proximal collar, and a fluted distal stem. The humeral head is semi-spherical and available in offset eccentric designs. The device is intended for use in total or hemi-shoulder arthroplasty, implanted by orthopedic surgeons in a clinical/OR setting. The stem supports both cemented and press-fit (cementless) fixation, while glenoid components (used in total arthroplasty) are for cemented use only. The system aims to restore joint function and relieve pain in patients with degenerative or traumatic shoulder conditions. The design incorporates features to improve stress transfer and stability, such as congruent mating surfaces when used with the Global glenoid reamer system.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on design, material, and technological similarities to previously cleared predicate devices.
Technological Characteristics
Modular humeral prosthesis. Materials: Titanium alloy (Ti-6Al-4V) stem, cobalt-chromium-molybdenum alloy head. Fixation: Press-fit or cemented (stem), cemented (glenoid). Design: Morse-type taper lock, four proximal fins with suture holes, fluted distal stem. Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for patients with severely painful/disabled joints due to osteoarthritis, traumatic arthritis, or rheumatoid arthritis; proximal humerus fracture-dislocations with severe comminution or vascular compromise; or difficult clinical cases where arthrodesis/resection is unacceptable (e.g., revision). Hemi-shoulder replacement also indicated for ununited humeral head fractures and avascular necrosis.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
{0}------------------------------------------------
# JUL 1 2 1999
# K992065
# Summary of Safety and Effectiveness
| NAME OF SPONSOR: | DePuy Orthopaedics, Inc.<br>700 Orthopaedic Drive<br>Warsaw, Indiana 46581-0988<br>Est. Reg. No. 1818910 |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) CONTACT: | Janet G. Johnson, RAC<br>Sr. Regulatory Associate<br>Phone: (219) 372-7176<br>FAX: (219) 267-7098 |
| TRADE NAME: | Global™ Advantage Shoulder<br>● Global™ Advantage™ Humeral Stem<br>● Global™ Advantage™ Eccentric Head |
| COMMON NAME: | Shoulder prosthesis, humeral stem |
| CLASSIFICATION: | When used as a hemi-shoulder, it is a Class II device<br>per 21 CFR §888.3690<br>When used as a total shoulder, it is a Class III device<br>per 21 CFR §888.3660 |
| DEVICE PRODUCT CODE: | 87 HSD - Prosthesis, Shoulder, Hemi-, Humeral,<br>Metallic Uncemented (Class II)<br>87 KWS - Prosthesis, Shoulder, Semi-Constrained,<br>Metal/Polymer Cemented (Class III) |
| SUBSTANTIALLY<br>EQUIVALENT DEVICE: | Global Shoulder: K911686, K974044, K984541 |
#### DEVICE DESCRIPTION:
The Global Advantage Shoulder consists of a humeral stem and a humeral head. Like the predicate devices, the components are modular in that they employ a morse-type taper lock system, the modular head having the male taper and the body with the female taper. The humeral stem is manufactured titanium alloy (Ti-6Al-4V), while the humeral head is manufactured from cobalt chromium molybdenum alloy.
The Global Advantage humeral stem is similar to the current Global humeral stem in that there are four proximal fins, a proximal collar, and a fluted distal stem. The fins are perforated with suture holes to accommodate the sutures necessary to reconstruct the proximal humerus.
The Global Advantage humeral head is similar to the current Global head in that it is fitted with an identical locking taper, is semi-spherical in design, and is also available in an offset eccentric head.
{1}------------------------------------------------
### INTENDED USE:
The Global Advantage Shoulder, humeral stem and humeral head, are intended for use in total or hemi-arthroplasty. When used for total shoulder arthroplasty, the subject components are designed to be used with the existing Global cemented glenoid components cleared in K905786 & K914000. The humeral stem is intended to be used with bone cement or in press-fit applications (cementless).
#### INDICATIONS FOR USE:
Total or hemi-shoulder replacement is indicated for:
- 1. A severely painful and/or disabled joint resulting from osteoarthritis, traumatic arthritis or rheumatoid arthritis;
- 2. Fracture-dislocations of the proximal humerus where the articular surface is severely comminuted, separated from its blood supply or where the surgeon's experience indicates that alternative methods of treatment are unsatisfactory;
- 3. Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g., revision of a failed primary component).
Hemi-shoulder replacement is also indicated for:
- 1. Ununited humeral head fractures:
- 2. Avascular necrosis of the humeral head.
Only the titanium alloy humeral stem components and the cobalt-chrome alloy humeral stem components, which are marketed under either the Global Shoulder or Global Advantage Shoulder name, are intended for press-fit or cemented fixation. The glenoid components are for cemented use only.
CAUTION: The cobalt-chrome alloy humeral components, which are marketed under the HRP Shoulder name and all glenoid components are for CEMENTED USE ONLY.
{2}------------------------------------------------
## SUBSTANTIAL EQUIVALENCE:
The fundamental scientific technologies of the Global Advantage humeral stem and head have not changed from the FDA cleared (K911686, K974044 and K984541) Global Shoulder humeral stem and head. The intended use and indications for use of the Global Advantage humeral stem and humeral head have not changed from the FDA cleared (K911686) Global Shoulder, Global Eccentric Humeral Head (K974044) and Global Fx (K984541). The humeral stem is manufactured titanium alloy (Ti-6AL-4V), identical to the humeral component cleared in K911686. With the exception of minor design modifications, the Global Advantage humeral stem and humeral head are identical to the Global Shoulder devices cleared in K911686, K974044 and K984541.
Based on conformance with the design control procedures requirements as specified in 21 CFR 820.30, similarities of design, commonly used materials, sterilization processes, indications for use, and intended use, DePuy believes that the Global Advantage humeral stem and head are substantially equivalent to the FDA cleared (K911686 K974044 and K984541) Global Shoulder, Global Eccentric humeral head and Global Fx components.
{3}------------------------------------------------
## Premarket Notification
## Class III Summary and Certification
#### Global Advantage Total Shoulder
#### CERTIFICATION:
I certify, in my capacity as Senior Regulatory Associate, of DePuy Orthopaedics, Inc. that I have conducted a reasonable search of all information known or otherwise available about the types and causes of safety or effectiveness problems that have been reported for total shoulder prostheses. I further certify that I am aware of the types of problems to which total shoulder prostheses are susceptible and that, to the best of my knowledge, the following summary of the types and causes of safety or effectiveness problems about total shoulder prostheses is complete and accurate.
Sanets Johnson
Janet G. Johnson, RAC
Senior Regulatory Associ DePuy Orthopaedics, Inc.
June 30, 1999
Date
#### CLASS III SUMMARY:
The most significant complication of total shoulder arthroplasty is currently loosening of the glenoid component. Revision surgery generally follows. Loosening rates for shoulder athroplasty range from 0% to 11%. The Global Total Shoulder addresses the issue of glenoid loosening in 3 ways:
1) Most total shoulder designs rely on hand burring or sculpting to prepare the face of the glenoid bone to accept the prosthesis. This procedure is inaccurate and allows the glenoid prosthesis to seat only on the high points of the bone. With point contact, normal glenohumeral loading tends to rock loose the anchoring fin of the glenoid component. The Global Shoulder system features a glenoid reamer which machines the glenoid to the same radius as the back side of the glenoid prosthesis. These two congruent mating surfaces provide for optimal stress transfer through the range of glenohumeral motion. Loads are transferred through the prosthetic glenoid articular surface to the bone, thereby preventing high stresses in the anchoring fins or pegs.
2) Due to the difference in elastic properties between the metal backing and the UHMWPe articular surface, and insufficient locking mechanisms two-piece glenoid components have a tendency to separate. The UHMWPe flexes more than the metal backing which causes the locking mechanism between the two components to fail. The Global Shoulder glenoid component is of a single piece UHMWPe construction, therefore, the possibility of component separation is eliminated.
3) The Global Shoulder glenoid is available in two design variations, pegged and finned. The pegged glenoid component has five fixation pegs including one anterior and one posterior peg. These anterior-posterior pegs reduce the rocking type of motion that can cause glenoid loosening. The finned glenoid component has a traditional tapered fixation fin. The finned component could be used in revision cases where a trough already exists in the glenoid bone or in cases of poor glenoid bone stock.
{4}------------------------------------------------
# Class III Summary and Certification (continued)
Other types of safety and effectiveness problems associated with total shoulder arthroplasty are also associated with other total joint replacements. These include:
- · Change in position of the prosthesis.
- · Early or late infection.
- · Cardiovascular disorders (including venous thrombosis, pulmonary embolism, and myocardial infarction which may be related to the use of bone cement).
- · Hematoma and delayed wound healing.
- · Pneumonia and atelectasis.
- · Subluxation or dislocation of the implant due to size selection, positioning of components and/or muscle and fibrous tissue laxity.
- · Failure to relieve pain.
- · Failure to restore range of motion.
In order to reduce the chance of complications with a total shoulder arthroplasty, the following conditions (which tend to impose severe loading on the affected extremity and may tend to adversely affect safety or effectiveness of a total shoulder arthroplasty) should be reduced or eliminated:
- · Heavy labor.
- · Active sports participation.
- · Likelihood of falls.
- · Alcohol or drug addiction.
- · Obesity.
The following conditions tend to adversely affect the stability (safety or effectiveness) of the implants:
- · Marked osteoporosis with poor bone stock and danger of impaired abutment of implants.
- · Systemic and metabolic disorders leading to progressive deterioration of solid bone. support for the implant (e.g. cortisone therapies, immunosuppressive therapies).
- · History of general infectious disease (e.g. erysipelas) or local infectious disease.
- · Severe deformities leading to impaired anchorage or improper positioning of the implant.
- · Tumors of the supporting bone structure.
- · Allergic reactions to the implant materials.
- · Tissue reactions to corrosion or wear products.
Attached is a bibliography of the materials upon which this summary is based.
{5}------------------------------------------------
- Neer, C.S., Watson, K.C., Stanton, F.J .: Recent Experience in Total Shoulder 1. Replacement. The Journal of Bone and Joint Surgery, 64A(3):319-336, 1982.
- 2. Faludi, D.D., Weiland, A.J.: Cementless Total Shoulder Arthroplasty: Preliminary Experience with Thirteen Cases. Orthopedics, 6(4):431-437, 1983.
- 3. Cofield, R.H .: Total Shoulder Arthroplasty: Associated Disease of the Rotator Cuff, Results and Complications. Surgery of the Shoulder, 229-233.
- 4. Boyd, A.D., Thomas, W.H., Scott, R.D., Sledge, C.B., Thornhill, T.S.: Total Shoulder Arthroplasty Versus Hemiarthroplasty. The Journal of Arthroplasty, 5(4):329-336, 1990.
- న. Gristina, A.G., Romano, R.L., Kammire, G.C., Webb, L.X.: Total Shoulder Replacement. Management of Shoulder Problems, 445-453.
- Lettin, A.W., Copeland, S.A., Scales, J.T .: The Stanmore Total Shoulder Replacement. 6. The Journal of Bone and Joint Surgery, 64B(1):47-51, 1982.
- 7. Hawkins, R.J., Bell, R.H., Jallay, B .: Total Shoulder Arthroplasty. Clinical Orthopedics and Related Research, 242:188-194, 1989.
- 8. Barrett, W.P., Franklin, J.L., Jackins, S.E., Wyss, C.R., Matsen, F.A .: Total Shoulder Arthroplasty. The Journal of Bone and Joint Surgery, 69A(6):865-872, 1987.
- Neer, C.S.: Unconstrained Shoulder Arthroplasty. Surgery of the Shoulder, 240-245. 9.
- 10. Wilde, A.H., Borden, L.S., Brems, J.J .: Experience with the Neer Total Shoulder Replacement. Surgery of the Shoulder, 224-228.
- l 1 . Orthopedic Surgical Manufacturers Association (OSMA): Petition for Reclassification of Shoulder Joint Metal/Polymer Non-constrained and Semi-constrained Cemented, Press Fit Hemi and Total Shoulder Prostheses. Submitted to FDA October 25, 1989.
- 12. DePuy Package Insert for the Total Shoulder or Hemi-Shoulder Prosthesis for Cement Fixation, No. 0902-00-457, Issued 1-95.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS) in the USA. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare, with three lines forming a wing-like shape. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 1 2 1999
Janet G. Johnson, RAC Senior Regulatory Associate DePuy Orthopaedics, Inc. 700 Orthopaedic Drive Warsaw, Indiana 46581-0988
K992065 Re:
> Trade Name: Global Advantage Shoulder: Humeral Stem and Eccentric Head Regulatory Class: III Product Code: KWS and HSD Dated: June 17, 1999 Received: June 18, 1999
Dear Ms. Johnson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{7}------------------------------------------------
Page 2 - Janet G. Johnson, RAC
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
## 510(k) Number (if known): _
# Device Name: Global™ Advantage Shoulder
## Indications for Use:
Total or hemi-shoulder replacement is indicated for:
- 1. A severely painful and/or disabled joint resulting from osteoarthritis, traumatic arthritis or rheumatoid arthritis;
- 2. Fracture-dislocations of the proximal humerus where the articular surface is severely comminuted, separated from its blood supply or where the surgeon's experience indicates that alternative methods of treatment are unsatisfactory;
- 3. Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g., revision of a failed primary component).
Hemi-shoulder replacement is also indicated for:
- 1. Ununited humeral head fractures;
- 2. Avascular necrosis of the humeral head.
Only the titanium alloy humeral stem components and the cobalt-chrome alloy humeral stem components, which are marketed under either the Global Shoulder or Global Advantage Shoulder name, are intended for press-fit or cemented fixation. The glenoid components are for cemented use only.
> CAUTION: The cobalt-chrome alloy humeral components, which are marketed under the HRP Shoulder name and all glenoid components are for CEMENTED USE ONLY.
> > Concurrence of CDRH, Office of Device Evaluation
OR
A. Seefelt
(Division \$ign-Off)
Division of General Restorative Devices 12992065
510(k) Number
Prescription Use
(Per 21 CFR 801.109)
Over-The-Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.