GLOBAL FX HUMERAL STEM, GLOBAL ADVANTAGE HUMERAL HEAD
K984541 · DePuy Orthopaedics, Inc. · KWT · Jan 14, 1999 · Orthopedic
Device Facts
Record ID
K984541
Device Name
GLOBAL FX HUMERAL STEM, GLOBAL ADVANTAGE HUMERAL HEAD
Applicant
DePuy Orthopaedics, Inc.
Product Code
KWT · Orthopedic
Decision Date
Jan 14, 1999
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The subject humeral stem and humeral head are intended for use in total or hemiarthroplasty. When used for total shoulder arthroplasty, the subject components are designed to be used with the existing Global cemented glenoid components cleared in K905786 & K914000. The humeral stem is intended to be used with bone cement or in press-fit applications (cementless).
Device Story
Global Shoulder system (humeral stem and head) provides prosthetic replacement for shoulder joint. Components feature modular morse-type taper lock; humeral head (male) mates with humeral body (female). Designed for total or hemiarthroplasty; humeral stem supports cemented or press-fit fixation. Modifications include reduced A/P and M/L dimensions for smaller metaphyses/fracture cases; lateral fin profile reduced to avoid biceps tendon contact/varus tilting; medial fin includes suture hole for tuberosity repositioning; collar/head interface modified with recess to reduce gap and increase articular surface area. Used by orthopedic surgeons in clinical settings to restore joint biomechanics and alleviate pain in patients with arthritis, fractures, or necrosis.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on design control procedures, material analysis, and comparison of technological characteristics to predicate devices.
Technological Characteristics
Modular humeral prosthesis; Co-Cr-Mo alloy (F-75). Features: proximal fins, proximal collar, fluted distal stem, morse-type taper lock. Dimensions modified for smaller metaphyses. Fixation: press-fit or cemented. Sterilization: standard methods implied. No software or electronic components.
Indications for Use
Indicated for patients with severely painful/disabled joints due to osteoarthritis, traumatic arthritis, or rheumatoid arthritis; proximal humerus fracture-dislocations with severe comminution or vascular compromise; or other conditions where arthrodesis/resection are unacceptable (e.g., revision). Hemi-shoulder replacement also indicated for ununited humeral head fractures and avascular necrosis.
Regulatory Classification
Identification
A shoulder joint metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-Aluminum 4-Vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-Chromium-Molybdenum Casting Alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-Chromium-Molybdenum Alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal Coatings,”
(vii) F 1378-97 “Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for DePuy, a Johnson & Johnson company. The logo features a solid black circle to the left of the name "DePuy" in a bold, sans-serif font. Below the name is the text "a Johnson-Johnson company" in a smaller font. The logo is simple and professional, reflecting the company's focus on medical devices and healthcare.
K98454/
JAN 1 4 1999
### Summary of Safety and Effectiveness
| NAME OF SPONSOR: | DePuy Orthopaedics, Inc.<br>700 Orthopaedic Drive<br>Warsaw, Indiana 46581-0988<br>Est. Reg. No. 1818910 |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) CONTACT: | Arlene C. Saull, RAC<br>Sr. Regulatory Associate<br>Phone: (219) 372-7176<br>FAX: (219) 267-7098<br>E-Mail: Arlene_Saull@ccgate.depuy.com |
| TRADE NAME: | Global™ Shoulder<br>● Global™ Fx Humeral Stem<br>● Global™ Advantage® Humeral Head |
| COMMON NAME: | Shoulder prosthesis, humeral stem & head components |
| CLASSIFICATION: | When used as a hemi-shoulder, it is a Class II device per 21 CFR §888.3690<br>When used as a total shoulder, it is a Class III device per 21 CFR §888.3660 |
| DEVICE PRODUCT CODE: | 87 HSD Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented (Class II)<br>87 KWS Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented (Class III) |
| SUBSTANTIALLY EQUIVALENT DEVICE: | Global Shoulder: K911686, K914695 |
# DEVICE DESCRIPTION:
The Global Shoulder components, which are the subjects of this submission, consist of a humeral body (stem) and a humeral head. Like the predicate devices, the components are modular in that they employ a morse-type taper lock system, the modular head having the male taper and the body with the female taper. Both components (head and body) will be made from Co-Cr-Mo alloy.
Continued on next page . . . . .
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the DePuy logo. The logo consists of a black circle on the left, followed by the word "DePuy" in a bold, sans-serif font. Below the word "DePuy" is the phrase "a Johnson & Johnson company" in a smaller font.
### Summary of Safety and Effectiveness (Continued)
The subject Global humeral stem is similar to the current Global humeral stem in that there are four proximal fins, a proximal collar, and a fluted distal stem. The fins are perforated to accommodate the sutures necessary to reconstruct the proximal humerus.
The Global humeral head is similar to the current Global head in that it is fitted with an identical locking taper, is semi-spherical in design, and it has similar head sizes and heights. The size range will allow options for joint tensioning and restoration of normal joint biomechanics.
Modifications and Reason: Although the currently marketed design already includes the indication for use in fracture cases, the need for a smaller proximal humeral stem has been identified for those patients with smaller metaphyses and for use in fracture cases when the predicate Global humeral component is too large for the particular patient. DePuy believes the subject humeral components, containing the following described modifications, fill this need.
- The proximal humeral body has been reduced in the A/P and M/L dimensions to ease ■ repositioning of the fractured humeral tuberosities.
- The lateral fin has a reduced profile to prevent contact with the biceps tendon and to avoid tilting the prosthesis in a varus orientation.
- 트 The medial fin has a suture hole added for passing of sutures to reposition fractured tuberosities.
- . The lateral fin has one centrally located suture hole versus the three suture holes in the current design.
- 해 The collar diameter has been reduced and contains a flat profile to mate with humeral heads.
- 미 The humeral head has been modified to contain a recess in the undersurface in order to mate with the collar of the humeral stem component. The collar/head design reduces collar/head gap and increases the effective articular surface area.
### INTENDED USE:
The subject humeral stem and humeral head are intended for use in total or hemiarthroplasty. When used for total shoulder arthroplasty, the subject components are designed to be used with the existing Global cemented glenoid components cleared in K905786 & K914000. The humeral stem is intended to be used with bone cement or in press-fit applications (cementless).
Continued on next page . . . . .
000011
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for DePuy, a Johnson & Johnson company. The logo features a black circle to the left of the name "DePuy" in a stylized font. Below the name is a horizontal line, and beneath that is the text "a Johnson & Johnson company" in a smaller font. The logo is simple and professional, reflecting the company's association with healthcare and medical devices.
### Summary of Safety and Effectiveness (Continued)
### INDICATIONS FOR USE:
Total or hemi-shoulder replacement is indicated for:
- 1. A severely painful and/or disabled joint resulting from osteoarthritis, traumatic arthritis or rheumatoid arthritis;
- 2. Fracture-dislocations of the proximal humerus where the articular surface is severely comminuted, separated from its blood supply or where the surgeon's experience indicates that alternative methods of treatment are unsatisfactory;
- 3. Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g., revision of a failed primary component).
Hemi-shoulder replacement is also indicated for:
- 1. Ununited humeral head fractures;
- 2. Avascular necrosis of the humeral head.
Only the titanium alloy humeral stem components and the cobalt-chrome alloy humeral stem components, which are marketed under the Global Shoulder name, are intended for press-fit or cemented fixation. The glenoid components are for cemented use only.
#### CAUTION:
The cobalt-chrome alloy humeral components, which are marketed under the HRP Shoulder name and all glenoid components are for CEMENTED USE ONLY.
Continued on next page . . . . .
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for DePuy, a Johnson & Johnson company. The logo features a solid black circle to the left of the name "DePuy" in a bold, sans-serif font. Below the name is the text "a Johnson & Johnson company" in a smaller, serif font.
### Summary of Safety and Effectiveness (Continued)
### BASIS OF SUBSTANTIAL EQUIVALENCE:
The fundamental scientific technology of the humeral stem and head has not changed from the FDA cleared (K911686) Global Shoulder humeral components. The intended use and indications for use of the subject Global humeral body and head, as described in its labeling has not changed from the FDA cleared (K911686) Global Shoulder devices. The addition of the Global Fx Stem to the Global Shoulder line does not require an application for a new indication because the fracture indication already exists for the currently marketed design. Like the predicate devices in K911686, the humeral head will be manufactured from cobalt chrome alloy. The humeral stem will be manufactured from F-75 cobalt chrome alloy, identical to the humeral component cleared in K914695. With the exception of the minor design modifications previously described, the subject humeral stem and humeral head are identical to the Global Shoulder devices cleared in K911686 and K914695.
Based on conformance with the design control procedures requirements as specified in 21 CFR 820.30, similarities of design, commonly used materials, sterilization processes, indications for use, and intended use, DePuy believes that the subject humeral stem and humeral head components are substantially equivalent to the FDA cleared (K911686 and K914695) Global Shoulder components.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/4/Picture/2 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or a person.
JAN 1 4 1999
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Arlene C. Saull, RAC Senior Regulatory Associate DePuy Orthopaedics, Inc. P.O. Box 988 700 Orthopaedic Drive Warsaw, Indiana 46581-0988
K984541 Re: Global™ Fx Humeral Stem, Global™ Advantage® Humeral Head Requlatory Class: III Product Codes: KWT and HSD Dated: December 18, 1998 December 21, 1998 Received:
Dear Ms. Saull:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Page 2 - Arlene C. Saull, RAC
This letter will allow you to begin marketing your device as described in your 510 k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for DePuy Synthes. The logo consists of a black circle on the left, followed by the word "DePuy" in a bold, sans-serif font. The "®" symbol is located at the bottom right of the word "DePuy". The logo is simple and modern, and it is likely used to represent the company's brand.
#### K984541 510(k) Number (if known)
Device Name: Global™ Shoulder
## Indications for Use:
Total or hemi-shoulder replacement is indicated for:
- 1. A severely painful and/or disabled joint resulting from osteoarthritis, traumatic arthritis or rheumatoid arthritis;
- 2. Fracture-dislocations of the proximal humerus where the articular surface is severely comminuted, separated from its blood supply or where the surgeon's experience indicates that alternative methods of treatment are unsatisfactory;
- 3. Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g., revision of a failed primary component).
Hemi-shoulder replacement is also indicated for:
- 1. Ununited humeral head fractures;
- 2. Avascular necrosis of the humeral head.
Only the titanium alloy humeral stem components and the cobalt-chrome alloy humeral stem components, which are marketed under the Global Shoulder name, are intended for press-fit or cemented fixation. The glenoid components are for cemented use only.
CAUTION: The cobalt-chrome alloy humeral components, which are marketed under the HRP Shoulder name and all glenoid components are for CEMENTED USE ONLY.
Concurrence of CDRH, Office of Device Evaluation
P. Cottofo
ral Restorative Devices
Prescription Use
(Per 21 CFR 801.109) X
OR
Over-The-Counter Use
000009
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.