INTERMITTENT COMPRESSION BOOT FOR OTC

K991921 · Eischoco, Inc. · IRP · Aug 13, 1999 · Physical Medicine

Device Facts

Record IDK991921
Device NameINTERMITTENT COMPRESSION BOOT FOR OTC
ApplicantEischoco, Inc.
Product CodeIRP · Physical Medicine
Decision DateAug 13, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5650
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The EISCHO Intermittent compression Boot is intended for medical purposes, such as to relieve minor aches and pains and to increase circulation in the patient's lower limbs. It simulates kneading and stroking of tissues by using an inflatable pressure cuff. The boot is not life sustaining is not implanted and does not use software. It is constructed of plastic that is designed for use over the lower extremity. It appears as an inflated boot. There are no toxicology implications. It is not sterile and is intended for the use of one person. The device is for home, hospital or clinical use. There are no drug or biological components. See also exhibit 6, product labels, for indications for use.

Device Story

Inflatable garment/boot for lower limbs; two-chamber design (upper chamber for limb, lower chamber for foot). Operation: patient-actuated via foot pressure; weight-bearing transfers air between chambers to provide intermittent compression; manual hand pump used for initial inflation; valve stem/cap used for deflation. Used in home, hospital, or clinical settings by patients. Provides kneading/stroking tissue simulation to increase circulation and relieve minor aches/pains. No software, electronics, or power source; purely mechanical pneumatic operation.

Clinical Evidence

Bench testing only; no clinical data provided.

Technological Characteristics

Plastic inflatable garment; two-chamber pneumatic system; manual hand pump inflation; mechanical valve deflation; non-sterile; single-patient use; no software; no electrical energy source.

Indications for Use

Indicated for first aid treatment of minor sports injuries, relief of minor aches and pains, and increasing circulation in lower limbs for patients requiring compression therapy.

Regulatory Classification

Identification

A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 991921/ ### 1. Submitter: Clem Eischen Owner EISCHO, Inc. 1232 S.E. 282 Ave. Grisham, OR 97080 Tel: (503) 2232 Fax: (503) 2232 ### Device Name 2. - Classification: 2.1 - 2.2 Common/Usual Name: - Proprietary Name: 2.3 Panel 74, Class II, 890.5650 Power Inflatable Tube Messager Intermittent Compression Boot ### Predicate Device: 3. - 3.1 EISCO, Inc.'s Intermittent Compression Boot for professional use, K964957 ### 4. Intended Use: The EISCHO Intermittent compression Boot is intended for medical purposes, such as to relieve minor aches and pains and to increase circulation in the patient's lower limbs. It simulates kneading and stroking of tissues by using an inflatable pressure cuff. The boot is not life sustaining is not implanted and does not use software. It is constructed of plastic that is designed for use over the lower extremity. It appears as an inflated boot. There are no toxicology implications. It is not sterile and is intended for the use of one person. The device is for home, hospital or clinical use. There are no drug or biological components. See also exhibit 6, product labels, for indications for use. ### స్ Device Description: This device is an inflatable garment or boot that is intended for use on patients' lower limbs. It consists of two chambers. The upper chamber envelops the lower extremity and the lower chamber is located under the foot. The interior panel has been sealed onto the exterior panel. When pressure is applied to the lower chamber by the patient's foot, air moves from the lower chamber to the upper chamber through the air passageways shaded in blue. The boot is placed on the patient and then inflated by attaching a hand pump to the port at the top of the boot. The boot is then inflated by means of the patient placing pressure on the foot.. At this point, the pressure in the upper chamber increases which causes a compression of the limb. When the patient lifts his/her foot, the air passes from the upper chamber to inflate the {1}------------------------------------------------ # EISCHO, Inc. Intermittent Compression Boot Premarket Notification Summary Of Data lower chamber, which causes the overall boot pressure to return to nominal pressure. The pressure may be applied while walking, standing, sitting or lying. The patient repeats this process until the therapy session is completed. To deflate the boot, the patient presses inward on the valve stem with his/her finger or uses the cap that is attached to the valve. The deflated boot is then easily removed. ### 6. Substantial Equivalency Comparisons: The comparison of EISCHO's Intermittent Compression Boot for over the counter use and EISCHO's Intermittent Compression Boot for professional use K964957 are the same The intermittent compression boot for over the counter use is indicated for first aid use on minor sports injuries and the intermittent compression boot for professional use is indicated for post phlebitis syndrome; swelling of foot, ankle and lower leg due to total knee replacement; bruising, swollen knee and ankle; and gravitational edema after cast removal. #### 7. Conclusions: Comparison of EISCHO's Intermittent Compression Boot for over the counter use and EISCHO's Intermittent Compression Boot for professional use indicates that they are substantially equivalent. The intermittent compression boot for over the counter use is use for first aid minor injuries. This poses no new safety of functional issues. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. The text is in all caps and appears to be in a sans-serif font. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG 13 1999 Mr. Clem Eischen President/Sales Eischco, Incorporated 1232 S.E. 282nd Avenue Gresham, Oregon 97080 Re: K991921 > Trade Name: Intermittent Compression Boot for OTC Regulatory Class: II Product Code: IRP Dated: May 28, 1999 Received: June 7, 1999 Dear Mr. Eischen: We have reviewed your Section 510(k) notification of intent to market the device referenced above, and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general control provisions of the Act. The general control provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General Regulation (21 CFR Part 820), and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 – Mr. Clem Eischen This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Colin M. Witten, Ph.D., M.D. Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use Statement K991921 Unknown 510(k) Number (if known): Intermittent Compression Boot For Over The Counter Use Device Name: Indications For Use: The EISCHO Intermittent compression Boot is intended for first aid treatment of minor sportrs injuries. It is intended to relieve minor aches and pains and to increase circulation in the patient's lower limbs. It simulates kneading and stroking of tissues by using an inflatable pressure cuff. . . . ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter USE X tcaief Division Sign-Off) ivision of General Restorative Devic 510/k) Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...