INTERMITTENT COMPRESSOR SYSTEM

K964957 · Eischoco, Inc. · IRP · Feb 24, 1997 · Physical Medicine

Device Facts

Record IDK964957
Device NameINTERMITTENT COMPRESSOR SYSTEM
ApplicantEischoco, Inc.
Product CodeIRP · Physical Medicine
Decision DateFeb 24, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5650
Device ClassClass 2
AttributesTherapeutic

Indications for Use

This device is an inflatable garment or boot that is intended for use on patients’ lower limbs.

Device Story

Inflatable garment/boot for lower limbs; two-chamber design (upper chamber envelopes extremity, lower chamber under foot). Operation: patient applies pressure to lower chamber via foot against surface (floor, wall, bed, wheelchair); air transfers to upper chamber via internal passageways; manual inflation via hand pump to desired pressure; manual deflation via valve stem. Used by patient; home or clinical setting. Output: intermittent compression to lower limbs. Benefit: provides compression therapy without external power source.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Inflatable garment; two-chamber pneumatic system; manual hand pump; manual valve stem for deflation; non-powered; patient-actuated.

Indications for Use

Indicated for patients requiring intermittent compression therapy on lower limbs.

Regulatory Classification

Identification

A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.

Predicate Devices

Submission Summary (Full Text)

{0} K964957 EISCHCO, Inc. EISCHCO Inflatable Massager Boot Premarket Notification Summary FEB 24 1997 ## 5. Device Description: This device is an inflatable garment or boot that is intended for use on patients’ lower limbs. It consists of two chambers. The upper chamber envelopes the lower extremity and the lower chamber is located under the foot. When pressure is applied to the lower chamber by the patient’s foot, air moves from the lower chamber to the upper chamber through the air passageways shaded in blue. The boot is placed on the patient and then inflated by attaching a hand pump to the port at the top of the boot. The boot is inflated to the desired pressure. The patient repeats this process until the therapy session is completed. To deflate the boot, the patient presses inward on the valve stem with his/her finger or uses the cap that is attached to the valve. The deflated boot is then easily removed. ## 6. Substantial Equivalency Comparisons: EISCHCO’s Inflatable Massager Boot is substantially equivalent to Chattanooga Group’s, Inc. Intermittent Compression System K832920. A comparison of EISCHCO’s Inflatable Massager Boot and the predicate device in Table I shows that both devices are the same except for the method for inflating the garment. The EISCHCO boot provides intermittent pressure to the garment by manual means, i.e. by actual exercise by the wearer. Pressing against a floor, wall, footboard of a bed, or possibly the foot rest of a wheelchair causes air to pass through the lower chamber to the upper chamber. As a result, pressure increases in the upper chamber and to the patient’s limbs. The predicate device provides intermittent pressure to the garment by the presssion pump. ## 7. Conclusion: The EISCHCO Inflatable Massager Boot is substantially equivalent to Chattanooga Group’s, Inc. Intermittent Compression System K832920
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...