K982814 · Medrad, Inc. · DXT · Oct 9, 1998 · Cardiovascular
Device Facts
Record ID
K982814
Device Name
MEDRAD VISTRON CT INJECTOR SYSTEM
Applicant
Medrad, Inc.
Product Code
DXT · Cardiovascular
Decision Date
Oct 9, 1998
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Vistron CT Injection System is a syringe-based fluid delivery system indicated for delivery of contrast media during computed tomography applications. It is intended to be used for the specific purpose of injecting intravenous contrast medium into the human vascular system for diagnostic studies in computed tomography.
Device Story
Programmable, syringe-based fluid delivery system for CT contrast media. Components: Injector Head, System Power Unit, Remote Monitor. Inputs: user-programmed injection parameters (volume, flow rate, pressure limit, scan delay) via keyboard/control panel. Operation: mechanical motor/plunger piston delivers fluid from 125ml or 200ml sterile disposable syringes. Features: syringe sensing, heater, autofill, multi-phase programming, remote start/hold. Used in clinical CT settings by healthcare professionals. Output: controlled contrast injection into patient vascular system. Safety: secondary microprocessor monitors electrical signals for over-volume, over-flow, and pressure/stall; hardware comparator for over-pressure; automatic stall termination. Benefits: precise, automated contrast delivery for diagnostic imaging.
Clinical Evidence
Bench testing only. No clinical data presented. Verification and validation testing performed to confirm compliance with design controls and safety standards including IEC 601-1, IEC 601-1-2, IEC 1000-4-6, IEC 1000-4-8, IEC 1000-4-11, IEC 1000-3-2, IEC 1000-3-3, IEC 529, UL 2601, and CSA C22.2 No. 601.1-M90.
Technological Characteristics
Syringe-based fluid delivery system. Components: Injector Head, System Power Unit, Remote Monitor. Power: 110/220 VAC. Sensing: position encoder, syringe size/presence sensors. Heating: syringe heater (87.8°F to 105.8°F). Safety: secondary microprocessor monitoring, hardware pressure comparator, electrical stop. Connectivity: wired remote monitor (up to 100ft). Standards: IEC 601-1, IEC 601-1-2, IEC 1000 series, IEC 529, UL 2601, CSA C22.2 No. 601.1-M90.
Indications for Use
Indicated for delivery of intravenous contrast media into the human vascular system for diagnostic computed tomography (CT) studies. Prescription use only.
Regulatory Classification
Special Controls
*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
MEDRAD VISTRON CT INJECTION SYSTEM 510(K) SUMMAR
August 7, 1998
K982814
| Official Contact: | Jan Dobscha, Senior Regulatory Affairs Coordinator<br>Medrad, Inc.<br>One Medrad Drive<br>Indianola, PA 15051<br>(412) 767-2400 ext. 3280 |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Classification: | Injector with Syringe, Angiographic |
| Common/Usual Name: | Powered Injector with Syringe |
| Proprietary Name: | Medrad Vistron CT Injection System |
| Predicate Device: | Medrad EnVision CT Injection System (K934086) |
Device Description:
#### Technical Description: Medrad, Inc.
· Medrad Drive
15051-0780
:12) 767-2400
The Medrad Vistron CT Injection System is a programmable, syringe-based fluid delivery system for delivery of contrast media during computed tomography procedures. The Vistron CT Injector is a modified version of the Medrad EnVision CT Injector. It has the same indications for use and the same intended use as the predicate device. No changes to the fundamental scientific technology were made nor are any new risks introduced by this modification.
The Vistron CT Injector is comprised of three main components - Injector Head, System Power Unit (SPU), and Remote Monitor (a.k.a. Intelligent Hand Unit). The Medrad Vistron CT Injector System, like the EnVision CT Injector System, uses a 125ml and 200ml Front Load Sterile Disposable Syringe.
- SYSTEM POWER UNIT (SPU) The System Power Unit houses the injector's main power . supply. It receives 110/220 VAC input and distributes power to the Injector Head via the Injector Head cable and to the Remote Monitor.
- INJECTOR HEAD The Injector Head is the main user interface. It is the main point of user data entry and display feedback data. It contains the mechanical subsystem (motor and plunger piston) that delivers fluid through an attached syringe. It also consists of the syringe interface, position encoder, syringe position/size sensors, user display/control interface, electronic controls, and safety circuits. The syringe heat maintainer and the hand held startswitch connect externally to the head.
The Injector Head processes keyboard inputs from the hard keys on the control panel, It displays injection parameters and operating Remote Monitor, or hand held startswitch. information on the control panel's LEDs, and processes handswitch input. It also provides a standby switch which places the system in a standby mode; that is, it disables power to functional system components but keeps power supplied to all system microprocessors.
- REMOTE MONITOR (a.k.a. Intelligent Hand Unit) The purpose of the Remote Monitor is to allow the operator to interact with the Injector Head without being near the Head. It permits the operator to view two phases of the injector's programmed parameters and to control the injector's Start/Hold and Disarm functions. The Remote Monitor provides LED readouts to display Scan Delay, Injected Volume, Injection, Phase, Flow Rate, and Volume information. It connects to the System Power Unit by means of an integral cord set. It can be placed up to 100 feet from the System Power Unit.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "MEDRAD" in a stylized, bold, sans-serif font. The letters are all capitalized and appear to be slightly slanted to the right. A small circled "R" symbol, indicating a registered trademark, is present to the upper right of the letter "D".
- MANUAL STARTSWITCH A hand-held push button handswitch, connected to the Injector . Head via a cable, allows the operator to initiate, suspend, or resume a programmed injection.
#### Functional Features:
The volume of contrast medium delivered is settable for each phase Programmed Volume of the iniection profile. The range is 10 ml to 125 ml or 10 ml to 200 ml depending on the size of syringe used.
The flow rate is the ratio of fluid volume delivered to the injection Programmed Flow Rate duration (ml/sec) and is settable, in 0.1 ml/sec increments, for each phase of the injection profile. The range is 0.1 ml/sec to 9.9 ml/sec.
The system has a settable pressure limit. The range is 50 psi to 300 Pressure Limiting psi with increments of 50 psi. In the event of a pressure limit being reached, the injection will proceed at a reduced flow rate until the programmed volume has been delivered, unless a stall condition occurs: A stall condition will automatically terminate the injection. The user is alerted via audible beeps and an indicator light on the Injector Head control panel.
The user can program the injector with fluid volume, flow rate, Program Parameters injection duration, pressure limit value, program phase/level, and scan delay parameters.
Injection Profile The user can program a sequence of one or more program phases, up to a maximum of 4, in which each phase delivers fluid according to the program parameters defined for that phase.
- "Hold" allows the user to temporarily interrupt an injection. When the user Hold enables the "Hold" function, the system remains armed, but the injection stops until the user cancels the hold function, whereupon the injection resumes from the point at which it was interrupted. The injector will remain in the "Hold" condition for a maximum of ten minutes, after which the unit disarms, an audible beep is emitted and a message is displayed on the Iniector Head and Remote Monitor.
- Scan Delay "Scan Delay" is a function that the user can program into the injection profile to indicate to the user that thev should begin the scan. The "Scan Delay" interval begins simultaneously with the first phase of the injection profile. The interval duration can be set from 0 seconds to 99 seconds. A visible "Scan Delay" countdown clock, which shows the time remaining in 1 second increments, is located on the Injector Head control panel and Remote Monitor. When the countdown clock reaches 0 an audible beep emitted from the Injector Head and Remote Monitor alerts the user.
- "Autofill" allows the user to load a specified amount of fluid into a syringe by Autofill actuating a control on the Iniector Head. Once the "Autofill" control is activated the head will load the specified amount of fluid into the syringe without further operator interaction. The user may select volumes of 25, 50, 75, 100, 125, 150, 175, or 200ml depending on the syringe size used. The fill rate is fixed at 2ml/sec.
- Single/Multi Arm The user can select one of two arming modes. "Single Arm" allows the execution of a single injection profile after pressing the start switch. After the programmed injection has been completed, the unit disarms. "Multiple Arm" allows the programmed injection to be repeated without re-arming the injector. After completing its programmed injection, the injector remains in the armed state. The user can press the Start switch again to repeat the same programmed injection. Repeated activation of the Start switch will
Image /page/1/Picture/12 description: The image shows a logo with the letter 'Q' at the top. Below the 'Q', the word 'Quality' is written in a smaller font size. The logo appears to be for a company or organization that emphasizes quality.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "medran" in a stylized, bold, sans-serif font. The letters are black and have a slightly distressed or textured appearance, giving them a rugged look. A small registered trademark symbol is visible to the right of the letter 'n'.
deliver the programmed injection until there is insufficient contrast media in the syringe to complete a full injection. At this point, the injector will inform the user, by flashing the Volume Remaining display, that there is insufficient volume for another injection. The operator is permitted to complete the injection with less than the programmed volume available in the syringe.
- "Syringe Sensing" is a system on the injector head that determines which of Syringe Sensing the two compatible Medrad syringe sizes (125ml or 200ml) is installed on the injector head, and also senses whether or not the syringe has been installed properly. After determining the size of the syringe that has been installed, the iniector adjusts its pressure, volume and flow characteristics accordingly. If the syringe has not been properly installed, the injector issues an error message on the Injector Head control panel alerting the user that the syringe is not correctly installed.
- Svringe Heater The Syringe heater maintains the temperature of preheated contrast medium within the syringe while the syringe is attached to the Injector. The heater operating temperature range is 87.8° to 105.8°F.
- Safety Stop All critical system functions are continuously monitored during an injection to insure that the actual injection values do not deviate from the programmed values beyond a specified limit. A secondary microprocessor in the Injector Head constantly monitors the electrical signals of the injector system. It monitors for over-volume, over flow rate, and pressure/stall; a hardware comparitor circuit monitors for over-pressure. When the backup monitoring system detects one of these conditions, it will open a relay to cut power to the piston motor and the injector system is placed in a maintenance state.
- Retract Control A control provided to return the piston to its fully retracted station. A single touch of a button is required to activate and complete this operation. This function is only operational when the syringe is detached from the injector system. The Syringe Sensing system detects whether or not a syringe is in place.
# Intended Use:
The Vistron CT Injection System is a svringe-based fluid delivery system indicated for delivery of contrast media during computed tomography applications. It is intended to be used for the specific purpose of injecting intravenous contrast medium into the human vascular system for diagnostic studies in computed tomography.
Image /page/2/Picture/8 description: The image shows a logo with a stylized letter 'Q' that forms a circular arrow, suggesting a cycle or continuous process. Below the 'Q', the words 'Quality' and 'of Life' are stacked vertically. The logo appears to represent a concept related to improving or maintaining the quality of life through a cyclical or ongoing effort.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a word in a bold, sans-serif font. The word appears to be "MELROSE", with each letter clearly defined and spaced. The font is uniform in size and thickness, contributing to the word's legibility.
Technological Characteristics:
| Feature | Medrad Vistron CT Injector<br>(New Device) | Medrad EnVision CT Injector<br>(K934086) |
|--------------------------------|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| Information Display | Large Colored LED's | Monochrome LCD |
| Programming Keys | Dedicated Function Keys | Non-Dedicated Keys - Software<br>Determined |
| Multi-Phase | 1 - 4 Phases Per Injection | 1 - 10 Phases Per Injection |
| Arming Modes | Single/Multi Arm | Single/Multi Arm |
| Protocol Storage<br>Capability | Stores last protocol used | 50 Protocols |
| Hold Capability | 0 - 600 seconds | 0 - 600 seconds |
| Scan Delay | 0-99 seconds | 0 - 99 seconds |
| Safety Stop Mechanism | Electrical Stop with a Software<br>Backup System | Electrical Stop with a Software Backup<br>System |
| Syringe System | 200ml, 125ml | 200ml, 125ml |
| Programmed Volume | 10 to 125 ml or 10 to 200 ml<br>Depending on Syringe Size | 1 to 125 ml or 1 to 200 ml<br>Depending on Syringe Size |
| Volume Remaining<br>Readout | LED | LED |
| Fill Rate | 2 ml/sec | 2 ml/sec to 9.9 ml/sec |
| Flow Rate | Settable .1 ml/sec to 9.9 ml/sec<br>(Settable for each injection during<br>programming only) | Variable .1 ml/sec to 9.9 ml/sec<br>(Settable for each injection during<br>programming and during injection.) |
| Pressure Limit | Settable from 50 to 300 PSI | Settable from 25 to 300 PSI |
| Pause | No | Programmable - 1 sec to 10 min |
| KVO | No | Continuous or Intermittent - Avg<br>minimum flow rate of .6 ml/sec |
| Autofill | Fill rate fixed at 2 ml/sec | Fill rate 2 ml/sec to 9 ml/sec in 1 ml/sec<br>increments |
| Retract Control | Yes | Yes |
| Remote Start Switch | Yes | Yes |
# Supporting Data:
Medrad has established, as part of its Quality System, design controls in compliance with FDA's Quality System Requirements. These design controls are applied to all Medrad product development processes and product design changes. These design controls were applied to the development of the Vistron CT Injector and meet the requirements of the FDA's QSRs.
As part of the design control a risk analysis was performed, and design verification and validation testing was conducted to support the conclusion drawn by the risk analysis.
# Conclusion:
The results of the risk analysis concluded that the modifications to the EnVision CT Injector did not introduce any new or additional risks relative to the use of this device. Testing also demonstrated that the Vistron CT Injection System met the applicable requirements of the following standards: IEC 601-1, IEC 601-1-2, IEC 1000-4-6, IEC 1000-4-8, IEC 1000-4-11, IEC 1000-3-2, IEC 1000-3-3, IEC 529, UL 2601, and CSA C22.2 No. 601.1-M90. Therefore, it has been determined that the Vistron CT Injection System is safe and effective for its intended use when used as prescribed in the User Operation Manual.
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows a black and white graphic. The graphic appears to be a stylized logo or emblem, possibly associated with a department or organization. The design features a series of curved lines or strokes that converge and overlap, creating an abstract shape. The word "DEPARTMENT" is partially visible along the left edge of the graphic, oriented vertically.
OCT - 9 1998
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Jan E. Dobscha Regulatory Affairs Coordinator Medrad, Inc. One Madrad Drive Indianola, PA 15051
Re: K982814 Trade Name: Medrad Vistron CT Injection System Requlatory Class: II Product Code: DXT Dated: September 11, 1998 Received: September 14, 1998
Dear Ms. Dobscha:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to leqally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, qood manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act
{5}------------------------------------------------
Paqe 2 - Ms. Jan E. Dobscha
for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation wo (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4586. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.qov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahon
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory -Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "merad" in a stylized font. The letters are blocky and have a slightly pixelated appearance. There is a circled "R" symbol to the right of the word, indicating that it is a registered trademark.
### INDICATION FOR USE
K982814 510(k) Number:
Device Name: Medrad Vistron CT Injection System
Indications for Use/Intended Use:
The Vistron CT Injection System is a syringe-based fluid delivery system indicated for delivery of contrast media during computed tomography …… applications. It is intended to be used for the specific purpose of injecting intravenous contrast medium into the human vascular system for diagnostic studies in computed tomography.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
M. Ringer
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number _
Prescription Use
(Per 21 CFR 801.109) ✓
OR
Over-The-CounterUse_
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.