1) Direct filling restorations using light-curing composite or compomer 2) Cavity sealing as a pretreatment for indirect restorations 3) Treatment of hypersensitive and/or exposed root surfaces 4) Intraoral repairs of fractured facing crowns made of porcelain, hybrid ceramics or composite resin using light-curing composite 5) Surface treatment of prosthetic appliances made of porcelain, hybrid ceramics and cured composite resin
Device Story
CLEARFIL SE BOND is a resin-based dental adhesive system; composed of dimethacrylate monomers; used by dentists in clinical settings. Device functions as a bonding agent to improve retention of restorative materials to tooth structure. Applied to prepared tooth cavities or prosthetic surfaces; facilitates adhesion of light-curing composites or compomers. Benefits include improved restoration retention, reduced hypersensitivity, and enabling repair of fractured dental crowns. No complex electronic processing or software involved.
Clinical Evidence
Bench testing only. Bond strength to human enamel, dentin, and porcelain was evaluated and compared to predicate device CLEARFIL LINER BOND 2V. Results indicate performance is substantially equivalent to the predicate.
Technological Characteristics
Resin-based dental adhesive system consisting of Primer, Bonding Agent, and Etching Agent. Primary chemical components include dimethacrylate monomers. System is a simplified version of the predicate CLEARFIL LINER BOND 2V. No electronic components, software, or connectivity features.
Indications for Use
Indicated for dental patients requiring direct light-cured composite/compomer restorations, cavity sealing for indirect restorations, treatment of hypersensitive/exposed root surfaces, intraoral repair of fractured crowns (porcelain, hybrid ceramic, or composite), and surface treatment of prosthetic appliances.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{0}------------------------------------------------
FEB 400000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000 ধ
K990040
(K943170)
(K943165)
# KURAF
12-39, 1-Chome, Umeda, Kita-ku, Osaka 530-86 : +81-6-6348-2603 Phone Facsimile: +81-6-6348-2552
## 510(k) SUMMARY
#### 1. Submitter
| 1) Name | KURARAY CO., LTD. |
|-------------------|-----------------------------------------------------------------------------|
| 2) Address | 1-12-39, Umeda, Kita-ku, Osaka 530-8611, Japan |
| 3) Telephone | 81(Japan)-6-6348-2603 |
| 4) Facsimile | 81(Japan)-6-6348-2552 |
| 5) Contact person | Yoshinori Nagase<br>Dental Material Department<br>Medical Products Division |
| 6) Date | January 6, 1999 |
#### 2. Representing (Subsidiary of KURARAY CO., LTD.)
| 1) Name | KURARAY AMERICA INC. |
|-------------------|-------------------------------------------------------------------|
| 2) Address | 30th Fl. Metlife Building, 200 Park Avenue, New York,<br>NY 10166 |
| 3) Telephone | (212)-986-2230 |
| 4) Facsimile | (212)-867-3543 |
| 5) Contact person | Koji Fujita<br>President |
#### 3. Name of Device
| 1) Proprietary Name | CLEARFIL SE BOND |
|------------------------|--------------------------------------------|
| 2) Classification Name | Resin tooth bonding agent (21CFR 872.3200) |
| 3) Common/Usual Name | Resin-based dental adhesive system |
#### র্ব-Predicate devices:
| | 1. CLEARFIL LINER BOND 2V by KURARAY CO.,LTD. | (K974486) |
|--|-----------------------------------------------|-----------|
|--|-----------------------------------------------|-----------|
- CLEARFIL LINER BOND 2 by KURARAY CO.,LTD. 2.
- 3. CLEARFIL PHOTO BOND by KURARAY CO.,LTD.
- PRIME & BOND 2.1 MULTIPURPOSE DENTIN/ENAMEL (K964525) 4. BONDING AGENT WITH ACTIVATOR by DENTSPLY
-
- GLUMA ONE BOND by HERAEUS KULZER, INC. (K974390) 5. (K#unknown) ONE STEP UNIVERSAL DENTAL ADHESIVE SYSTEM by BISCO, 6.
- INC. (K910860) 7. ALL-BOND 2 by BISCO, INC.
- (K934690) OPTIBOND by KERR MFG.CO 8.
- 5. Description for the premarket notification
CLEARFIL SE BOND is classified into the resin tooth bonding agent, CFR 21 Section 872.3200, because it is a device composed of materials such as dimethacrylate monomers intended to painted on the interior of a prepared cavity of a tooth to improve retentive materials.
This product is similar and substantially equivalent in design, composition and function to the
{1}------------------------------------------------
similar products which are identified in the paragraph 4 of this summary; all of which are safe, effective and beneficial.
#### 6. Statement of the intended use
Statement of the firended aso products.
| | 1) Direct filling restorations using light-curing composite or compomer | |
|----|-----------------------------------------------------------------------------------------|-------------|
| a) | CLEARFIL LINER BOND 2V by KURARAY CO., LTD. | (K974486) |
| b) | CLEARFIL LINER BOND 2 by KURARAY CO., LTD. | (K943170) |
| c) | PRIME & BOND 2.1 MULTIPURPOSE DENTIN/ENAMEL<br>BONDING AGENT WITH ACTIVATOR by DENTSPLY | (K964525) |
| d) | GLUMA ONE BOND by HERAEUS KULZER, INC. | (K974390) |
| | | |
| | 2) Cavity sealing as a pretreatment for indirect restorations | |
| a) | CLEARFIL LINER BOND 2V by KURARAY CO., LTD. | (K974486) |
| b) | ONE STEP UNIVERSAL DENTAL ADHESIVE SYSTEM by<br>BISCO, INC. | (K#unknown) |
| c) | ALL-BOND 2 by BISCO, INC. | (K910860) |
| d) | OPTIBOND by KERR MFG.CO | (K934690) |
| | | |
| | 3) Treatment of hypersensitive and/or exposed root surfaces | |
| a) | CLEARFIL LINER BOND 2V by KURARAY CO., LTD. | (K974486) |
| b) | CLEARFIL LINER BOND 2 by KURARAY CO., LTD. | (K943170) |
- PRIME & BOND 2.1 MULTIPURPOSE DENTIN/ENAMEL (K964525) c) BONDING AGENT WITH ACTIVATOR by DENTSPLY
- ONE STEP UNIVERSAL DENTAL ADHESIVE SYSTEM by (K#unknown) d) BISCO, INC.
- Intraoral repairs of fractured facing crowns made of porcelain, hybrid ceramics or 4) composite resin using light-curing composite
| a) CLEARFIL LINER BOND 2V by KURARAY CO., LTD. | (K974486) |
|--------------------------------------------------------------------------------------------|-----------|
| b) PRIME & BOND 2.1 MULTIPURPOSE DENTIN/ENAMEL<br>BONDING AGENT WITH ACTIVATOR by DENTSPLY | (K964525) |
- 5) Surface treatment of prosthetic appliances made of porcelain, hybrid ceramics and cured composite resin
| a) CLEARFIL LINER BOND 2V by KURARAY CO., LTD. | (K974486) |
|------------------------------------------------|-----------|
| b) CLEARFIL PHOTO BOND by KURARAY CO., LTD. | (K943165) |
- 7. Statement of the technological characteristics and safety
Buttems of the wonners great as a simplified system of CLEARFIL LINER BOND 2V permitted to be marketed (K974486). CLEARFIL SE BOND is substantially equivalent to those of products sold in the U.S. market in design, components and functions.
- 7-1 Components
CLEARFIL SE BOND consists of Primer, Bonding Agent, Etching Agent and accessories. These components are similar to those of the products in the paragraph 4 of this summary.
{2}------------------------------------------------
#### 7-2 Performance
Periormance
There is no ISO standard applicable to CLEARFIL SE BOND. The bond strengths to human There Is no too standard appricable to cl and porcelain were evaluated in comparison with enamel, numan dentife, precious mocal are performances are substantially equivalent to those of CLEARFIL LINER BOND 2V.
### 7-3 Chemical ingredients and safety
The chemical ingredients and bacty The chemical ingredictive have been use an in ial y equivalent to the predicated devices.
- CLEARFIL LINER BOND 2V by KURARAY CO., LTD. (K974486) a) (K983361) PANAVIA F by KURARAY CO., LTD. b)
- (K982164) ESTENIA by KURARAY CO., LTD. c)
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB ! 4 1999
Kuraray Co., Ltd. C/O Mr. Koji Fujita President Kuraray America, Incorporated 30th FI Metlife Building 200 Park Avenue New York, New York 10166-3098
Re : K990040 CLEARFIL SE BOND Trade Name: Regulatory Class: II Product Code: KLE January 6, 1999 Dated: January 6, 1999 Received:
Dear Mr. Fujita:
We have reviewed your Section 510(k) notification of intent to we have reviewed your becoed above and we have determined the market the device refly equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Commerce prior co nay ents, or to devices that have been Medical Device imendance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, rood, Drug, and commons to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of Include requiremufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may (opecial ochorsuch additional controls. Existing major be subject to back accoundevice can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A rederal Regulation ent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic (URL) regections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP vegulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of
{4}------------------------------------------------
Page 2 - Mr. Fujita
the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA described in your tial equivalence of your device to a legally Finding of substancial equires in a classification for your marketed predicate device burned to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be Information on Jour Chilison of Small Manufacturers Assistance at its toll-free number (800) 638-20#1 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours
Timothy A. Ulatowski
Timo Directbr Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radioloqical Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the text "K990040" in a bold, sans-serif font. Below this, the text "[CLEARFIL SE BOND, Kuraray]" is printed in a smaller, serif font. The text appears to be part of a label or product identifier.
Page 1 of 1
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ K990040
CLERFIL SE BOND Device Name: _
Indications For Use
CLEARFIL SE BOND is indicated for the following applications:
Ard III OL DOND is managing light-curing composite or compomer 1)
- Cavity sealing as a pretreatment for indirect restorations 2)
- Oavity seaming as a prositive and/or exposed root surfaces 3)
- Treatment of dypersensitive andor exposed root successor
Intraoral repairs of fractured facing crowns made of porcelain, hybrid ceramics and cured 4) composite resin
- composite resin
Surface treatment of prosthetic appliances made of porcelain, hybrid ceramics and cured 5) composite resin
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) *** Over-The-Counter Use_ OR Prescription Use_6 (Part 21 CFR 801.109) (Optional Format 1-2-96) us not of the 1100 (Division Sign-Off) Division of The'. Infect නාවිතා and General Siot pital Develop remov 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.