K974486 · Kuraray America, Inc. · KLE · Jan 29, 1998 · Dental
Device Facts
Record ID
K974486
Device Name
CLEARFIL LINER BOND 2V
Applicant
Kuraray America, Inc.
Product Code
KLE · Dental
Decision Date
Jan 29, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3200
Device Class
Class 2
Indications for Use
1) Direct filling restorations using light-cure or chemical-cure composite resin 2) Bonded amalgam restorations 3) Treatment of hypersensitive and/or exposed root surfaces 4) Cavity sealing as a pretreatment for indirect restorations 5) Intraoral repairs of facing crowns using light-cure composite resin 6) Cementing laminate veneers, inlays and onlays made of porcelain (or composite resin) using composite resin cement
Device Story
CLEARFIL LINER BOND 2V is a resin-based dental adhesive system composed of dimethacrylate monomers; intended for application to prepared tooth cavities to improve restorative material retention. System utilizes BOND Liquid A and BOND Liquid B to enable self-cure or dual-cure techniques; includes etching agent and porcelain primer. Used by dentists in clinical settings to bond restorative materials (composite resin, amalgam) to tooth structure or repair existing crowns. Benefits include improved retention of restorations and treatment of hypersensitive root surfaces. Clinician applies adhesive to cavity; provides interface for subsequent restorative material placement; enhances clinical outcomes through improved bonding performance.
Clinical Evidence
No clinical data. Evidence consists of bench testing and biocompatibility studies. Biocompatibility evaluation of DMABB included acute oral toxicity (LD50 > 4,000 mg/kg in mice), Ames mutagenicity test (negative), and maximization sensitization test (negative). Leachable analysis using HPLC showed DMABB levels below the identification limit (0.04 ppm) after 24-hour immersion in distilled water at 37°C and 50°C.
Technological Characteristics
Resin-based dental adhesive system; components include dimethacrylate monomers, BOND Liquid A, BOND Liquid B, etching agent, and porcelain primer. Chemical composition includes DMABB. Designed for self-cure or dual-cure polymerization techniques. Materials are consistent with previously cleared Kuraray dental products.
Indications for Use
Indicated for dental patients requiring restorative procedures including direct composite resin fillings, bonded amalgam restorations, treatment of hypersensitive or exposed root surfaces, cavity sealing for indirect restorations, intraoral repair of facing crowns, and cementation of porcelain or composite resin veneers, inlays, and onlays.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{0}
KURARAY
KURARAY CO., LTD.
[CLEARFIL LINER BOND 2V]
12-39, 1-Chome, Umeda, Kita-ku, Osaka 530, JAPAN
Phone : +81-6-348-2603
Facsimile: +81-6-348-2552
K974486
JAN 29 1998
# 510(k) SUMMARY
## 1. Submitter
1) Name
Kuraray Co., Ltd.
2) Address
1-12-39, Umeda, Kita-ku, Osaka 530, Japan
3) Telephone
81(Japan)6-348-2603
4) Facsimile
81(Japan)6-348-2552
5) Contact person
Yoshinori Nagase
Dental Material Department
Medical Products Division
6) Date
December 2, 1997
## 2. Representing (Subsidiary of Kuraray Co., Ltd.)
1) Name
Kuraray America, Inc.
2) Address
30th FI. Metlife Building, 200 Park Avenue, New York, NY 10166
3) Telephone
(212)986-2230
4) Facsimile
(212)876-3543
5) Contact person
Koji Fujita
President
## 3. Name of Device
1) Proprietary Name
CLEARFIL LINER BOND 2V
2) Classification Name
Resin tooth bonding agent (21CFR 872.3200)
3) Common/Usual Name
Resin-based dental adhesive system
## 4. Predicate devices:
1) CLEARFIL LINER BOND 2 by Kuraray Co., Ltd. (K943170)
2) CLEARFIL PORCELAIN BOND by Kuraray Co., Ltd. (K871636)
3) CLEARFIL PHOTO BOND by Kuraray Co., Ltd. (K943165)
4) CLEARFIL NEW BOND by Kuraray Co., Ltd. (K943167)
5) SCOTCHBOND MULTI-PURPOSE PLUS DENTAL ADHESIVE SYSTEM by 3M COMPANY (K942493)
6) PRIME & BOND 2.1 MULTIPURPOSE DENTIN/ENAMEL BONDING AGENT WITH ACTIVATOR by DENTSPLY (K964525)
7) ALL-BOND 2 by BISCO, INC. (K910860)
8) PROBOND by LD CAULK/DENTSPLY (K934690) unknown
9) OPTIBOND by KERR MFG.CO
page 7 of 12
{1}
[CLEARFIL LINER BOND 2V]
## 5. Description for the premarket notification
CLEARFIL LINER BOND 2V is classified into the resin tooth bonding agent, CFR 21 Section 872.3200, because it is a device composed of materials such as dimethacrylate monomers intended to be painted on the interior of a prepared cavity of a tooth to improve retention of restorative materials. PROTECT LINER F is classified into tooth shade resin material, CFR 21 Section 872.3690, because it is a device composed of material such as bisphenol A glycidylmethacrylate (Bis-GMA) intended to restore carious or structural defects in teeth. CLEARFIL PORCELAIN BOND ACTIVATOR was permitted to be marketed under its 510 (k) notification submission.
This product is similar and substantially equivalent in design, composition and function to the similar products which are identified in the paragraph 4 of this summary; all of which are safe, effective and beneficial.
## 6. Statement of the intended use.
This device is used for the following indications. Each indication is same as that of similar products.
1) Direct filling restorations using light-cure or chemical-cure composite resin
a) CLEARFIL LINER BOND 2 by Kuraray Co., Ltd. (K943170)
b) CLEARFIL PHOTO BOND by Kuraray Co., Ltd. (K943165)
c) CLEARFIL NEW BOND by Kuraray Co., Ltd. (K943167)
d) SCOTCHBOND MULTI-PURPOSE PLUS DENTAL ADHESIVE SYSTEM by 3M COMPANY. (K942493)
e) PRIME & BOND 2.1 MULTIPURPOSE DENTIN/ENAMEL BONDING AGENT WITH ACTIVATOR by DENTSPLY (K964525)
f) ALL-BOND 2 by BISCO, INC. (K910860)
g) PROBOND by LD CAULK/DENTSPLY (unknown)
h) OPTIBOND by KERR MFG.CO (K934690)
2) Bonded amalgam restorations
a) CLEARFIL LINER BOND 2 by Kuraray Co., Ltd. (K943170)
b) CLEARFIL PHOTO BOND by Kuraray Co., Ltd. (K943165)
c) CLEARFIL NEW BOND by Kuraray Co., Ltd. (K943167)
d) ALL-BOND 2 by BISCO, INC. (K910860)
e) PROBOND by LD CAULK/DENTSPLY (unknown)
f) OPTIBOND by KERR MFG.CO (K934690)
3) Treatment of hypersensitive and/or exposed root surfaces
a) CLEARFIL LINER BOND 2 by Kuraray Co., Ltd. (K943170)
b) ALL-BOND 2 by BISCO, INC. (K910860)
4) Cavity sealing as a pretreatment for indirect restorations
a) ALL-BOND 2 by BISCO, INC. (K910860)
b) OPTIBOND by KERR MFG.CO (K934690)
page 8 of 12
{2}
[CLEARFIL LINER BOND 2V]
5) Intraoral repairs of facing crowns using light-cure composite resin
a) CLEARFIL PORCELAIN BOND by Kuraray Co., Ltd. (K871636)
b) SCOTCHBOND MULTI-PURPOSE PLUS DENTAL ADHESIVE (K942493) SYSTEM by 3M COMPANY.
c) PRIME & BOND 2.1 MULTIPURPOSE DENTIN/ENAMEL (K964525) BONDING AGENT WITH ACTIVATOR by DENTSPLY
d) PROBOND by LD CAULK/DENTSPLY unknown
e) ALL-BOND 2 by BISCO, INC. (K910860)
f) OPTIBOND by KERR MFG.CO (K934690)
6) Cementing laminate veneers, inlays and onlays made of porcelain (or composite resin) using composite resin cement
a) SCOTCHBOND MULTI-PURPOSE PLUS DENTAL ADHESIVE (K942493) SYSTEM by 3M COMPANY.
b) PROBOND by LD CAULK/DENTSPLY unknown
c) ALL-BOND 2 by BISCO, INC. (K910860)
d) OPTIBOND by KERR MFG.CO (K934690)
7. Statement of the technological characteristics and safety
This device is developed as a improved material of CLEARFIL LINER BOND 2 (K943170).
7-1. Components
The new component, BOND Liquid B, is used as a mixture with BOND Liquid A to enable the practitioner to use self-cure or dual-cure techniques. The etching agent and the porcelain primer are introduced to use for special intended purpose. These components are same components in the CLEARFIL PHOTO BOND (K943165) and CLEAFIL PORCELAIN BOND (K871636).
7-2. Chemical ingredients
The chemical ingredients excepting DMABB have been already used in the following devices sold in the U.S. market. All of them are manufactured by Kuraray Co., Ltd.
a) CLEARFIL LINER BOND 2 (K943170)
b) PANAVIA 21(K933030) classified into the dental cement other than zinc oxide-eugenol (21 CFR 872.3275(b))
c) CLEARFIL PHOTO BOND (K943165)
d) CLEARFIL PORCELAIN BOND (K871636)
8. Summary of toxicity study
The chemical ingredients excepting DMABB have been already used in Kuraray's dental material allowed to be sold in US market as described in the paragraph 7-2. The biocompatibility of DMABB was evaluated its acute toxicity (LD50), genotoxicity and sensitization. Additionally The amount of leaching DMABB from cured resin into distilled water is evaluated.
These results suggest that CLEARFIL LINER BOND 2V is a safe dental device.
page 9 of 12
{3}
[CLEARFIL LINER BOND 2V]
# 8-1 Biological evaluation of DMABB
## 8-1-1 Acute toxicity
1) Animal
Mouse
2) Dosing route
Oral
3) Dosing period
10 days
4) Results
LD50; not less than 4,000 mg/kg
## 8-1-2 Genotoxicity test
1) Method
Ames mutagenicity test
2) Results
Negative
| Bacterial species | Genotoxicity | |
| --- | --- | --- |
| | with S-9 mix | without S-9 mix |
| Salmonella typhimurium TA100 | Negative | Negative |
| Salmonella typhimurium TA1535 | Negative | Negative |
| Escherichia coli WP2 uvrA | Negative | Negative |
| Salmonella typhimurium TA98 | Negative | Negative |
| Salmonella typhimurium TA1537 | Negative | Negative |
This test was based on ISO 10993-3 (1992-12-15), biological evaluation of medical devices-part 3.
## 8-1-3 sensitization (maximization test)
1) Method
Maximization sensitization test
2) Results
Negative
This test was based on ISO 10993-10 (1995-3-15), biological evaluation on medical devices.
## 8-2 Leachables
### 8-2-1 Amount of DMABB leached from cured resin
1) Sample
Cured resin 10mm φ × 5 mm.
2) Extraction media
Distilled water
3) Method
Two pieces of cured resin were immersed into 20ml of distilled water and stored at 37°C or 50°C for 24 hours. The amount of leached DMABB in measured using high pressure liquid chromatography.
4) Result
Less than identification limit (0.04ppm) for both conditions
page 10 of 12
{4}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
JAN 29 1998
Mr. Koji Fujita
President
Kuraray America, Incorporated
200 Park Avenue
New York, New York 10166
Re: K974486
Trade Name: Clearfil Liner Bond 2V
Regulatory Class: II
Product Code: KLE
Dated: November 25, 1997
Received: November 26, 1997
Dear Mr. Fujita:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531
{5}
Page 2 - Mr. Fujita
through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Timothy A. Ulatowski
Director
Division of Dental, Infection Control and General Hospital Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{6}
[CLEARFIL LINER BOND 2V]
Page ____ of ____
510(k) Number(if known): K974486
Device Name: CLEARFIL LINER BOND 2V
Indications For Use
1) Direct filling restorations using light-cure or chemical-cure composite resin
2) Bonded amalgam restorations
3) Treatment of hypersensitive and/or exposed root surfaces
4) Cavity sealing as a pretreatment for indirect restorations
5) Intraoral repairs of facing crowns using light-cure composite resin
6) Cementing laminate veneers, inlays and onlays made of porcelain (or composite resin) using composite resin cement
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Dental, Infection Control, and General Hospital Devices
510(k) Number K974486
Prescription Use ☑ OR Over-The-Counter Use
(Per 21 CFR 801.109)
(Optional Format 1-2-96)
page 11 of 12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.