For use with Automatic and Manual External Defibrillators and Heartstream Electrodes. For Defibrillation, Monitoring and Pacing.
Device Story
Heartstream electrode adapters serve as physical interface between Heartstream electrodes and various manual/automatic defibrillators; enable external cardiovascular pacing, monitoring, and delivery of defibrillation shocks up to 360J. Device consists of rigid thermoplastic body with gold-plated conductors; provided as reusable stand-alone accessory. Used in clinical settings where defibrillation/pacing is required. Adapters ensure secure connection between electrodes and defibrillator, maintaining electrical continuity without exposing users to accidental electrical contact or adversely affecting device functions. Benefits include compatibility across different defibrillator models, allowing clinicians to utilize Heartstream electrodes with existing equipment.
Clinical Evidence
Bench testing only. Testing per IEC 601-1, IEC 601-2-4, IEC 601-2-25, and ANSI/AAMI DF39-1993 demonstrated effective delivery of defibrillation energy up to 360J, successful external transcutaneous cardiac pacing, and mechanical/electrical compatibility with manual and automatic defibrillators.
Technological Characteristics
Rigid thermoplastic body with gold-plated conductors. Reusable accessory. Complies with IEC 601-1, IEC 601-2-4, IEC 601-2-25, and ANSI/AAMI DF39-1993. Passive electrical interface; no internal software or active circuitry.
Indications for Use
Indicated for patients requiring external cardiovascular pacing, monitoring, and delivery of defibrillation shocks up to 360J via manual or automatic external defibrillators.
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
{0}------------------------------------------------
DEC 17 1998
# 984286 510(k) Summary Heartstream Electrode Adapters
## General Information
| Trade Name | Heartstream Electrode Adapter |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| FDA Panel and Classification | HLW<br>Cardiovascular, 74 MKJ, Class III |
| Contact Person | William D. Jordan<br>Regulatory Affairs Specialist |
| Address | Heartstream, Inc.<br>A wholly-owned subsidiary of<br>Hewlett Packard Company<br><br>2401 4th Ave.<br>Suite 500<br>Seattle, WA 98121 |
## Substantially Equivalent Devices
Manufacturer
Product
Heartstream, Inc. A wholly-owned subsidiary of Hewlett Packard Company
Heartstream Electrode Adapter
### Description of Device & Intended Use
Heartstream electrode adapters serve as an interface to allow Heartstream electrodes to be connected to various manual and automatic defibrillators for external cardiovascular pacing as well as monitoring and delivery of defibrillation shocks up to 360J.
Heartstream electrode adapters are made of rigid thermoplastic and conductive material (or equivalent), with varying configurations depending upon the defibrillator for which they are designed. The adapters will be provided as reusable stand-alone accessories to be used in conjunction with standard Heartstream electrodes.
# Technological Characteristics
All Heartstream electrode adapters are designed to comply with applicable portions of relevant standards, including:
- . IEC 601-1, Medical electrical equipment, Part 1: General requirements for safety, 1993
- IEC 601-2-4, Medical electrical equipment, Part 2: Particular Requirements for the Safety of . Cardiac Defibrillators and Cardiac Defibrillator/Monitors, 1993
{1}------------------------------------------------
- IEC 601-2-25, Medical electrical equipment, Part 2: Particular requirements for the safety of . electrocardiographs
- ANSI/AAMI DF39-1993, Automatic external defibrillators and remote-control defibrillators. . September 16, 1993
Testing per these standards demonstrated that the Heartstream electrodes effectively deliver defibrillation energy up to 360J, allow for external transcutaneous cardiac pacing, demonstrate mechanical and electrical compatibility with manual and automatic defibrillators, and minimize safety risks to the user and patient.
#### Summary of Substantial Equivalence
The Heartstream Electrode Adapter is very similar in form, fit and function to the predicate device. All adapter applications are tested using the actual model defibrillators specified on the packaging label. The adapter engineering specifications for the predicate and new device are the same - full compatibility with the specified defibrillator. The technology is the same - thermoplastic (or equivalent) adapter body with gold plated conductors. The applicable voluntary standards and quidances for both the predicate and new device are the same.
As with the predicate, the adapters are designed so that both connectors are firmly seated when in use, and so they do not expose a user to a risk of accidental electrical contact. The adapters are also designed to ensure that all pacing, monitoring and/or defibrillation functions are not adversely affected by their placement.
Heartstream, Inc. concludes that the new adapter does not raise any new safety or effectiveness concerns and that the predicate and new devices are substantially equivalent.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of human profiles.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 17 1998
Mr. William D. Jordan Hewlett-Packard Company Heartstream Operation 2401 Fourth Avenue, Suite 300 Seattle, WA 98121-1436
K984286 Re: Heartstream Electrode Adapter Regulatory Class: III (three) Product Code: MLN November 30, 1998 Dated: Received: December 1, 1998
Dear Mr. Jordan:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
#### Page 2 - Mr. William D. Jordan
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial en your of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.qov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1 Submitted: November 18, 1998
510(k) Number: Device Names: Heartstream Electrode Adapter Heartstream Electrode Adapter: For use with Automatic and Indications for Use: Manual External Defibrillators and Heartstream Electrodes. For Defibrillation, Monitoring and Pacing.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) �����������������������������������������������������������������������������������������������������������������������������������������������������������������������������
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <i>Mark Kramer</i> |
|-------------------------------------------------------------------|---------------------|
| | (Division Sign-Off) |
| Division of Cardiovascular, Respiratory, and Neurological Devices | |
| 510(k) Number | K984286 |
| Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-The-Counter Use ______ |
|------------------------------------------|----|-----------------------------|
|------------------------------------------|----|-----------------------------|
(Optional Format 1-2-96)
----------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.