The intended use of the "Digi-Snap ™" device is to screen patients for apnea and snoring and to provide quantitative and qualitative analysis of apnea and snoring. The Digi-Snap testing system is only intended for short term monitoring such as to record the oximetry and snoring sounds continuously during the night. The system is not intended as a substitute for full polysomnography when additional parameters such as sleep stages, limb movements or EEG activity are required. The target population is patients who are suspected of apnea and/or complain about snoring. The majority of the screenings are going to take place at the patient's home, although some may take place in a sleep laboratory. Both pediatric and adult patients may be tested. CAUTION: US Federal law restricts this device to sale by or on the order of a physician, Use of this device must be under the direct supervision of a qualified adult (parent or guardian) or health care practitioner trained in the use of the "Digi-Snap™" device.
Device Story
Digi-Snap is a screening system for apnea and snoring; records oximetry, snoring sounds, and heart rate. Inputs: microphone/cannula for snoring sounds, finger probe for oximetry/heart rate. Device digitizes snoring sounds at bedside; stores data on 100 MB Zip drive. Used primarily in home settings or sleep labs; operated by patient or trained practitioner. After overnight recording, patient returns disk to service center for analysis. Output: quantitative/qualitative analysis of apnea/snoring events. Healthcare providers use output to screen for sleep-disordered breathing; aids in clinical decision-making regarding further diagnostic needs. Benefits patients by providing accessible, short-term monitoring.
Clinical Evidence
Bench testing and user testing performed. No clinical trial data provided. Testing focused on device setup, recording functionality, and comparison of digital recording performance against the predicate DAT-based system.
Technological Characteristics
System includes 100 MB Zip drive, microphone/cannula, and Nonin OEM II/Xpod oximeter. Energy source: 90-240V internal medical-grade power supply. Channels: snoring sounds, oximetry, heart rate. Compliance: FCC Class B, CISPR 11 Class B, UL-2601 electrical safety.
Indications for Use
Indicated for pediatric and adult patients suspected of apnea or complaining of snoring. Used for short-term, overnight screening and quantitative/qualitative analysis of apnea and snoring. Not for use when sleep stages, limb movements, or EEG are required.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
Nonin Model 8500 (OEM II Module or Xpod Module) (K893221)
Submission Summary (Full Text)
{0}------------------------------------------------
K984169
## EXHIBIT 2
Snap Laboratories, LLC 3633 West Lake Avenue Suite 406 Glenview. IL 60025 USA 847-657-8100 Fax: 847-657-8105 Contact Name: Gil Raviv, President February 21 1999
## 510(k) Summary of Safety and Effectiveness
- 1. Identification of the Device Proprietary-Trade Name: "Digi-Snap TM" Classification Name: 73 MNR Apnea/Snoring Recording and analysis Device and 74DQA, Oximeter. Common/Usual Name: Snoring and Apnea Recording and Analysis Device
- 2. Equivalent legally marketed devices This product is similar in design and function to the "Oxi-Snap TM" Testing Device," K971184, and incorporates a different Pulse Oximeter, Nonin Model 8500 (OEM II Module or the Xpod Module), K893221.
- 3. The intended use of the "Digi-Snap ™" device is to screen patients for apnea and snoring and to provide quantitative and qualitative analysis of apnea and snoring. The Digi-Snap testing system is only intended for short term monitoring such as to record the oximetry and snoring sounds continuously during the night. The system is not intended as a substitute for full polysomnography when additional parameters such as sleep stages, limb movements or EEG activity are required. The target population is patients who are suspected of apnea and/or complain about snoring. The majority of the screenings are going to take place at the patient's home, although some may take place in a sleep laboratory. Both pediatric and adult patients may be tested. CAUTION: US Federal law restricts this device to sale by or on the order of a physician, Use of this device must be under the direct supervision of a qualified adult (parent or guardian) or health care practitioner trained in the use of the "Digi-Snap™" device.
- 4. Description of the Device: This notification is for a modification to the existing device, the "Oxi-Snap™" Testing Device. The modified device is called the "Digi-Snap. TM" The modification involves the use of a 100 MB "Zip" drive in place of the DAT tape recorder. This eliminates one A/D-D/A conversion because the snoring sounds are digitized at the first opportunity, in the unit at the patient's bedside.
- 5. Safety and Effectiveness, comparison to predicate device. The results of bench and user testing indicates that the modified device is as safe and effective as the predicate
{1}------------------------------------------------
device. The modified device is easy for the user to set up at home or in the sleep laboratory. The modification involves the use of a fully digital recording technique. The oximeter is connected to the unit along with the usual microphone/cannula apparatus and the oximeter sensor is slipped over the patient's finger in the usual manner. The patient turns on the unit, then goes to sleep. Apnea and snoring events are then recorded. After awakening, the patient returns the disk and the equipment to the analysis service center, where the disk is analyzed.
- 6. Substantial Equivalence Chart
| Characteristic | Predicate device: Oxi-<br>SNAPTM testing device<br>K971184 | Modified device:<br>"Digi-Snap TM" |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| Labeling: | (Original submission) | The User's Manual has been<br>updated to show the new<br>configuration, with new operation<br>instructions. |
| Intended Use: | Recording and analysis of snoring<br>and apnea | Same |
| Physical characteristics: | | |
| Recording device: | Sony TDC-D7 DAT digital audio<br>tape recorder | 100 MB "Zip" Drive and<br>proprietary interface. |
| Channels acquired: | Two: snoring sounds on one<br>channel, Oximetry level on the<br>other channel. | Three: snoring sounds, Oximetry<br>level, and Heart Rate |
| User equipment: | DAT recorder, tape, cannula,<br>microphone, and Oximeter, Palco<br>Model 305 | Zip unit, Zip Disk, cannula,<br>microphone, and Oximeter, Nonin<br>OEM II or Xpod. |
| Energy Source: | 120 V 60~wall mount AC-DC<br>converter 12W (recommended) or<br>batteries Plus NiCad rechargeable<br>batteries for the oximeter (12 hour<br>life per charge) Charger is UL<br>listed | 90-240 V, 50/60~ internal Medical<br>grade power supply |
| Anatomical sites: | Upper lip and finger probe | Same |
| Performance testing: | Summarized above | Same |
| Safety characteristics: | | |
| Electrical safety: | Per applicable sections of UL-2601 | Same |
| EMI: | Per FCC part 15 Class A | FCC class B, CISPR 11 Class B |
| Oximetry | Included | Same |
| Intended population | Adults and pediatrics | Same |
| Home use | Yes | Same |
## 7. Conclusion
After analyzing both bench and user testing data, it is the conclusion of Snap Laboratories that the "Digi-Snap ™" testing snoring and apnea testing device is as safe and effective as the predicate device and has no new indications for use, thus rendering it substantially equivalent to the predicate Snap Testing Device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the logo is an abstract symbol that resembles a stylized caduceus or a representation of human figures.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 2 6 1999
Gil Raviv, Ph.D. Snap Laboratories 3633 West Lake Avenue, Suite 406 Glenview, IL 60025
Re: K984169 "Digi-Snap™" Snoring and Apnea Recording and Analysis Device Regulatory Class: II (two) Product Code: 73 MNR February 23, 1999 Dated: February 26, 1999 Received:
Dear Dr. Raviv:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Gil Raviv, Ph.D.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. C allahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## i) Indications for Use
K984169 510(k) Number
Device Name: "Digi-Snap ™" Snoring and Apnea recording and analysis system
Indications for Use:
The intended use of the "Digi-Snap ™" device is to screen patients for apnea and snoring and to provide quantitative and qualitative analysis of apnea and snoring.
The Digi-Snap testing system is only intended for short term monitoring such as to record the oximetry and snoring sounds continuously during the night. The system is not intended as a substitute for full polysomnography when additional parameters such as sleep stages, limb movements or EEG activity are required.
The target population is patients who are suspected of apnea and/or complain about snoring. The majority of the screenings are going to take place at the patient's home, although some may take place in a sleep laboratory. Both pediatric and adult patients may be tested.
CAUTION: US Federal law restricts this device to sale by or on the order of a physician. Use of this device must be under the direct supervision of a qualified adult (parent or guardian) or health care practitioner trained in the use of the "Digi-Snap TM" device.
Concurrence of CDRH, Office of Device Evaluation (ODE)
R.H.A. Garlurchi.
Prescription Use \/ ***_*** OR Over the Counter Use ***_***
(Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.