COBE CLOSED VENOUS RESERVOIR BAGS WITH SMARX T, MODELS VRB 1200 AND VRB 1800
K980786 · Cobe Cardiovascular, Inc. · DTN · Aug 24, 1998 · Cardiovascular
Device Facts
Record ID
K980786
Device Name
COBE CLOSED VENOUS RESERVOIR BAGS WITH SMARX T, MODELS VRB 1200 AND VRB 1800
Applicant
Cobe Cardiovascular, Inc.
Product Code
DTN · Cardiovascular
Decision Date
Aug 24, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4400
Device Class
Class 2
Indications for Use
The COBE® Closed Venous Reservoir Bags With SMAR,T™ are closed, flexible reservoir bags intended to be used as blood reservoirs during adult cardiac surgical procedures requiring extracorporeal support for periods up to six hours.
Device Story
Flexible, closed blood reservoir bags (1200 ml or 1800 ml) used during adult cardiac surgery with extracorporeal support. Inputs include systemic venous and cardiotomy return blood. Features surface-modifying material to improve blood compatibility. VRB 1200 includes rigid backing plate, three-stopcock sampling system, one-way vent valve, and level-sensing magnet for integration with COBE Air Emboli Protection System. VRB 1800 is a simplified version without the rigid plate. Operated by perfusionists/clinicians in surgical settings. Device collects and stores blood, assists in air removal, and provides volume indication. Benefits include improved blood compatibility and integrated air emboli protection.
Clinical Evidence
No clinical data. Evidence consists of bench testing including SEM analysis, mechanical integrity, air challenge, bag shutoff, collapse volume, blood cell damage, and surface-modifying material leaching/compatibility testing, plus ex-vivo blood compatibility testing.
Technological Characteristics
Flexible reservoir bags with surface-modifying material on blood-contact surfaces. VRB 1200 includes rigid backing plate, stopcock sampling system, one-way vent valve, and level-sensing magnet. Single-use, sterile, non-pyrogenic. No electronic components; level sensing is mechanical/magnetic for external system integration.
Indications for Use
Indicated for adult patients undergoing cardiac surgical procedures requiring extracorporeal support for up to six hours.
Regulatory Classification
Identification
A cardiopulmonary bypass blood reservoir is a device used in conjunction with short-term extracorporeal circulation devices to hold a reserve supply of blood in the bypass circulation.
Special Controls
*Classification.* Class II (special controls), except that a reservoir that contains a defoamer or filter is classified into the same class as the defoamer or filter. The device, when it is a cardiopulmonary bypass blood reservoir that does not contain defoamers or blood filters, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
COBE® VRB® 1200™ Closed Venous Reservoir Bag (K980786)
Submission Summary (Full Text)
{0}------------------------------------------------
510(k) Pre-Market Notification: COBE® Closed Venous Reservoir Bags With SMAR, T™
K980786
20
AUG 24 1998
## 510(k) Summary
SUBMITTER:
COBE Cardiovascular, Inc. 14401 W. 65th Way Arvada, CO 80004
Lynne Leonard
CONTACT PERSON:
(303) 467-6429 Fax:
February 24, 1998 DATE PREPARED:
COBE® Closed Venous Reservoir Bags With SMAR TTM, DEVICE TRADE NAME: Model Numbers VRB® 1200™ and VRB® 1800™
Phone: (303) 467-6586
Venous Reservoir Bag COMMON/USUAL NAME:
Cardiopulmonary Bypass Blood Reservoir CLASSIFICATION NAME:
COBE® VRB® 1200™ Closed Venous Reservoir Bag PREDICATE DEVICE:
DEVICE DESCRIPTION:
The COBE® Closed Venous Reservoir Bags With SMAR TT™, Model Numbers VRB® 1200™ and VRB® 1800™ are sterile devices with non-pyrogenic fluid pathways, for single use only, and are not to be resterilized by the user. The devices accomodate both systemic venous and cardiotomy return blood. A surface-modifying material is added to the primary blood contact surfaces of the device to improve the blood compatibility of the materials.
The VRB® 1200™ with SMAR,T consists of the bag with its basic ports (venous inlet, cardiotomy inlet, reservoir outlet, reservoir vent, venous sample ports) along with several additional user features. The reservoir bag is attached directly to a rigid backing plate that contains a three stopcock blood sampling system, a reservoir vent line with a one-way valve, and a front plate that is designed to exert pressure on the bag, squeezing it between the front and back plates to assist in air removal. There is a volume indicator tape on the back plate which provides an indication of the amount of blood in the bag. There is also a level sensing magnet on the outside of the reservoir bag and a sensor in the bracket which allows the VRB® 1200™ with SMAR,T to be used with the COBE® Air Emboli Protection System. The maximum volume of the reservoir bag is 1200 ml.
The VRB® 1800™ with SMAR,T is a simplified version of the VRB® 1200™ with SMAR,T. The VRB® 1800™ with SMAR, T consists of only the bag with its basic ports (venous inlet, cardiotomy inlet, reservoir outlet, reservoir vent, venous sample port, and auxillary ports). It does not have the additional user features listed above for the VRB® 1200™ with SMAR. T. The maximum volume of the reservoir bag is 1800 ml (the reservoir bag is identical to the VRB® 1200™ with SMART, but the maximum volume is greater in the VRB® 1800™ with SMAR, T because it is not attached to the rigid back plate).
{1}------------------------------------------------
## INDICATIONS FOR USE
The COBE® Closed Venous Reservoir Bags With SMAR,T™ are closed, flexible reservoir bags intended to be used as blood reservoirs during adult cardiac surgical procedures requiring extracorporeal support for periods up to six hours.
## STATEMENT OF TECHNICAL CHARACTERISTICS COMPARISON
The COBE® Closed Venous Reservoir Bags With SMAR T™ have the same intended use as the COBE® VRB® 1200™ Closed Venous Reservoir Bag. The only differences between the COBE® Closed Venous Reservoir Bags With SMAR TTM and the currently marketed VRB® 1200™ are 1) the COBE® Closed Venous Reservoir Bags With SMAR T™ contain the surface-modifying material; 2) Model VRB® 1800™ with SMAR,T is a simplified version of the VRB® 1200™, consisting only of the bag with its basic ports.
Biocompatibility testing, in-vitro testing, and ex-vivo testing were performed to demonstrate that the COBE® Closed Venous Reservoir Bags With SMAR, T™ are substantially equivalent to the currently marketed COBE® VRB® 1200™ Closed Venous Reservoir Bag.
In-vitro testing consisted of:
SEM Analysis Mechanical Integrity Air Challenge Bag Shutoff Collapse Volume Blood cell damage Surface-modifying material leaching Surface-modifying material blood compatibility
Ex-vivo testing consisted of:
Surface-modifying material blood compatibility
These data support that the COBE® Closed Venous Reservoir Bags With SMAR, T™ are substantially equivalent to the currently marketed COBE® VRB® 1200™ Closed Venous Reservoir Bag, and the addition of the surface-modifying material does not affect safety and effectiveness.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the seal, there is a stylized graphic of three human profiles facing to the right, one behind the other.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 2 4 1998
Ms. Lynne Leonard Manager, Requlatory Submissions COBE Cardiovascular, Inc. 14401 W. 65th Way Arvada, CO 80004-3599
K980786 Re: COBE® Closed Venous Reservoir Bags with SMAR,T™ Regulatory Class: II (Two) DTN Product Code: Dated: June 10, 1998 Received: June 11, 1998
Dear Ms. Leonard:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the marice is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices maxketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
## Page 2 - Ms. Lynne Leonard
This letter will allow you to beqin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahon
Thomas J. Cal ahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications For Use
510(k) Number (If known): _ K98 0786
COBE® Closed Venous Reservoir Bags With SMAR T™, Device Names: Model Numbers VRB® 1200™ and VRB® 1800™
Indications For Use:
The COBE® Closed Venous Reservoir Bags With SMAR,T™ (Model Numbers VRB® 1200™ and The CODE Closed Transable be used in adult surgical procedures requiring extracorporeal support for periods of up to six hours.
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ben R. Russell
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number
Prescription Use
(Per 21 CFR 801.109) _/
OR
Over-The-Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.